K191347 · Shenzhen Wisonic Medical Technology Co. , Ltd. · IYN · Feb 7, 2020 · Radiology
Device Facts
Record ID
K191347
Device Name
Paragon XHD Diagnostic Ultrasound System
Applicant
Shenzhen Wisonic Medical Technology Co. , Ltd.
Product Code
IYN · Radiology
Decision Date
Feb 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Diagnostic Ultrasound System is applicable for adults and pediatric patients. It is intended for use in pediatric, small organ (breast, thyroid, testicles, prostate), musculoskeletal (conventional), musculoskeletal (superficial), peripheral vessel and dermatological exams.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires and displays ultrasound data in B-Mode, Color Flow Doppler, Power/Dirpower mode, or combinations. System comprises main unit, ultrasound probe, power adapter, and cables. Operated by healthcare professionals in clinical settings to visualize anatomical structures. Provides software analysis packages for measurements (area, angle, volume, velocity). Output displayed to clinician for diagnostic decision-making. Benefits include non-invasive imaging for assessment of small organs, musculoskeletal, vascular, and dermatological conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1, -5, -10), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing (IEC 60601-2-37) to evaluate measurement accuracy and system sensitivity.
Technological Characteristics
Mobile ultrasound system; B-Mode, Color Flow Doppler, Power/Dirpower modes. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-37 (ultrasonic performance). Probes and materials biocompatible per ISO 10993. Software-controlled with measurement/calculation packages.
Indications for Use
Indicated for adult and pediatric patients undergoing diagnostic ultrasound exams of small organs (breast, thyroid, testicles, prostate), musculoskeletal (conventional and superficial), peripheral vessels, and dermatological structures.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
FUJIFILM SonoSite Vevo MD Imaging System (K160674)
Clover 50/Clover60/Clover70 Diagnostic Ultrasound System (K163712)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Shenzhen Wisonic Medical Technology Co., Ltd. % Jiang Xiaosan Regulatory Engineer 1st, 2nd & 5th Floor, No. 6 Building, Pingshan Tech. Park, Taoyuan Street, Nanshan Shenzhen, Guangdong 518055 CHINA
Re: K191347
Trade/Device Name: Paragon XHD Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 4, 2020 Received: January 9, 2020
### Dear Jiang Xiaosan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K191347
Device Name
Paragon XHD Diagnostic Ultrasound System
Type of Use (Select one or both, as applicable)
Indications for Use (Describe)
The Diagnostic Ultrasound System is applicable for adults and pediatric patients. It is intended for use in pediatric, small organ (breast, thyroid, testicles, prostate), musculoskeletal (conventional), musculoskeletal (superficial), peripheral vessel and dermatological exams.
| | Over-The-Counter Use (21 CFR 801 Subpart C)<br> X Prescription Use (Part 21 CFR 801 Subpart D) | | | | |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | |
| | Department of Health and Human Services<br>Food and Druq Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | | | | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of | | | | |
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K191347
In accordance with 21 CFR 807.92 the following summary of information is provided:
### 1. Identification of Sponsor
Shenzhen Wisonic Medical Technology Co., LTD.
1st, 2nd & 5th Floor, NO.6 Building, Pingshan Technology Park, Address: Taoyuan Street, Nanshan, Shenzhen. Guangdong, 518055, P.R. CHINA
Contact Person: Jiang Xiaosan
Tel: (+86) 755 86007788
Fax: (+86)755 86007799
Jan 30, 2019 Date prepared
### 2. Identification of Proposed Device
| Device Name: | Paragon XHD Diagnostic Ultrasound System |
|----------------------|---------------------------------------------|
| Common/Usual Name: | Diagnostic Ultrasound System |
| Regulation number: | 21 CFR 892.1550 |
| Regulation Name: | Ultrasonic pulsed doppler imaging system |
| Regulation Class: | Class II |
| Product Code: | IYN, ITX, IYO |
| Classification Name: | System, Imaging, Pulsed Doppler, Ultrasonic |
| Model: | Paragon XHD |
### 3. Identification of Predicate Device
Primary Predicate Device:
K160674, FUJIFILM SonoSite Vevo MD Imaging System
Secondary Predicate Device:
K163712, Clover 50/Clover60/Clover70 Diagnostic Ultrasound System
### 4 Device Description
The Paragon XHD Diagnostic Ultrasound System is a mobile software controlled ultrasonic system. Its function is to acquire and display ultrasound data in B-Mode, Color Flow Doppler, Power/Dirpower mode or the combination of these modes. The system can also measure anatomical structures and offer software analysis
{4}------------------------------------------------
packages performance to provide information based on which the competent health care professionals can make the diagnosis.
The Paragon XHD Diagnostic Ultrasound System consists of the main unit, an ultrasound probe, power adapter, connecting cable.
### 5. Intended Use
The Diagnostic Ultrasound System is applicable for adults and pediatric patients. It is intended for use in pediatric, small organ (breast, thyroid, testicles, prostate), musculoskeletal (conventional), musculoskeletal (superficial), peripheral vessel and dermatological exams.
### 6. Substantial Equipment
The Paragon XHD Diagnostic Ultrasound System is comparable with and substantially equivalent to the following predicate devices:
| Predicate<br>Device | Manufacturer | Model | 510(k) Control<br>Number |
|------------------------|----------------------|-----------------------------|--------------------------|
| Primary<br>Predicate | FUJIFILM<br>SonoSite | Vevo MD | K160674 |
| Secondary<br>Predicate | Wisonic | Clover 50/Clover60/Clover70 | K163712 |
### Compared of indications:
Indications of the predicate device (FUJIFILM SonoSite /Vevo MD/K160674) and the subject device (WISONIC/Paragon XHD) are the same.
### Compared of Technological Characteristics:
The proposed device has the similar type probes and similar technical characteristics, including design, operation controls, display modes, operation modes, measurement items and image parameters as the predicate device Vevo MD(K160674).
The proposed device has the same connector port as the predicate device Clover 50/Clover 60/Clover 70(K163712).
The proposed device has the same capability in term of measurements and calculation functions as predicate device, except for adding area (ellipse, circle), angle and volume measurements in B-mode, C-mode, PW-mode and combined mode, and velocity in PW-mode. These measurements are substantially equivalent with the predicate device Clover 50/Clover 60/Clover 70(K163712).
The acoustic power level of proposed device Paragon XHD is below the limit of FDA, which is the same as the predicated device Clover 60/Clover 60/Clover 70(K163712).
All the above differences in technological characteristics do not rise different questions of safety and effectiveness as compared to the predicate device.
{5}------------------------------------------------
# wisor
### 7. Non-clinical data
The following non-clinical data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the Paragon XHD Diagnostic Ultrasound System was conducted in accordance with the International Standard ISO 10993-1:2009/ (R)2013, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:
- Cytotoxicity
- Sensitization
- Skin Irritation
The ultrasonic probes and glue of the Diagnostic Ultrasound System are considered to contact directly with human body for a duration of less than 24 hours. The test results of cytotoxicity, sensitization and skin irritation complied with ISO 10993-5:2009. Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. It demonstrates substantial equivalences to the predicate device.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Diagnostic Ultrasound System. The device complies with the IEC 60601-1:2005/A1:2012, standard for safety and the IEC 60601-1-2:2014, standard for EMC. It demonstrates substantial equivalences to the predicate device.
### Performance testing
The Diagnostic Ultrasound System, was tested according to IEC 60601-2-37:2015 Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, and met the same standard, Performance testing was conducted on the Paragon XHD Diagnostic Ultrasound System, to evaluate the clinic measurement accuracy and system sensitivity, and all of the tested parameters met the predefined acceptance criteria. The comparisons of performance of Paragon XHD Diagnostic Ultrasound System to predicate devices are listed in the table below.
The performance of the device was compared with the predicate devices, and it is concluded that the proposed device is substantially equivalent to the predicate device.
### Software Verification and Validation Testing
{6}------------------------------------------------
# wison
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider. It demonstrates substantial equivalences to the predicate device.
# Animal Study
The subject of this premarket submission, the Paragon XHD Diagnostic Ultrasound System, does not require animal studies to support substantial equivalence.
# 8. Clinical data
The subject of this premarket submission, the Paragon XHD Diagnostic Ultrasound System, did not require clinical studies to support substantial equivalence.
# 9. Conclusion
The non-clinical data support the safety of the device and the performance testing report demonstrate that the Paragon XHD Diagnostic Ultrasound System should perform as intended in the specified use conditions and conform to applicable medical device safety standards. In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification. Shenzhen Wisonic concludes that the Paragon XHD Diagnostic Ultrasound System is as safe and as effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.