← Product Code [JAK](/productcode/JAK) · K191332

# XACT Robotic System (K191332)

_Xact Robotics, Ltd. · JAK · Oct 3, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191332

## Device Facts

- **Applicant:** Xact Robotics, Ltd.
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Oct 3, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.

## Device Story

XACT Robotic System is a user-controlled robotic positioning system for CT-guided percutaneous procedures. Input: CT images for trajectory planning. Operation: System registers coordinate system to CT images; physician defines trajectory; robot positioning unit (placed on patient) and insertion module hold and steer instrument. Physician controls advancement via foot pedal; system monitors respiration-related motion and makes real-time adjustments to maintain trajectory. Output: Automated instrument positioning and advancement to target. Used in clinical settings by physicians. Benefits: Enhanced accuracy in needle placement during interventional procedures.

## Clinical Evidence

Clinical data from 2 centers involving 31 subjects undergoing abdominal CT-guided procedures. The device successfully reached the target in 27/31 cases; 4 cases were completed manually. No serious adverse events reported. Preclinical animal testing (porcine model, 40 procedures) showed 100% target reach with average accuracy of 1.41mm ± 0.94mm. Bench testing included mechanical accuracy (ASTM F2554-10), registration precision, CT phantom accuracy, rigidity, axes mapping, needle properties, and respiration synchronization.

## Technological Characteristics

System comprises a robot positioning unit/insertion module, control unit, workstation, and cart. Materials biocompatible per ISO 10993. Sensing/actuation: Robotic positioning with respiration synchronization. Connectivity: Workstation for trajectory planning. Electrical safety: IEC 60601-1; EMC: IEC 60601-12. Mechanical accuracy: ASTM F2554-10.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- iSYS Medizintechnik GmbH iSYS 1 ([K131433](/device/K131433.md))
- Perfint Healthcare Pvt. Ltd. Maxio ([K132108](/device/K132108.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

October 3, 2019

XACT Robotics Ltd. % Mr. Jonathan S. Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004

Re: K191332

Trade/Device Name: XACT Robotic System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: September 4, 2019 Received: September 4, 2019

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K191332

Device Name

#### XACT Robotic System

Indications for Use (Describe)

The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

x Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY XACT Robotic System (K191332)

#### Submitter

- XACT Robotics Ltd. 8 Hatochen Street POB 3097 Caesarea Industrial Park North Caesarea 3079861 Israel Phone: +972 (0)4 770 0153 Facsimile: +972 (0)4 770 0161
Contact Person: Chen Levin, CEO

## Date Prepared:

September 4, 2019

### Name of Device:

XACT Robotic System

### Common or Usual Name:

CT Stereotactic Accessory

### Classification Name:

21 CFR 892.1750; Computed tomography X-ray system

### Regulatory Class:

Class II

### Product Code:

JAK

### Predicate Devices

iSYS Medizintechnik GmbH iSYS 1 (K131433) Perfint Healthcare Pvt. Ltd. Maxio (K132108)

### Device Description

The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of instruments during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning based on CT images and is intended to assist the physician in positioning of an instrument, such as a needle, and reviewing instrument position during advancement to the target. The system quides (i.e., positions and

{4}------------------------------------------------

steers) the instrument according to a predefined trajectory. The physician controls advancement of the instrument along the trajectory using a foot pedal. The system also allows for monitoring of motion associated with respiration during the procedure.

The XACT Robotic System comprises the following main components:

- . XACT Robot Positioning Unit & Insertion Module Assembly which is placed on the patient
- XACT Control Unit (Central Computer) .
- . XACT Computer Work Station which includes the user interface for trajectory planning and review of instrument position
- XACT System Cart ●

# Intended Use / Indications for Use

The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.

## Summary of Technological Characteristics

The XACT Robotic System allows for planning of percutaneous CT-guided procedures and tracking and positioning of the instrument during the procedure.

Both the XACT System and the cleared iSYS 1 predicate device are designed to provide positioning of instruments for percutaneous intervention under imaging guidance from CT scanners. The systems position the instrument according to a predefined trajectory following a registration process between the device's coordinate system and real-time CT images. achieving this, both the XACT Robotic System and the iSYS 1 System comprise similar components, namely a control unit, a bed/table, floor or patient mount, a robotic positioning unit and tool guide adapter and insertion module.

The XACT Robotic System robot positioning unit is placed directly on the patient in the region of interest for the percutaneous procedure, whereas the iSYS 1 System is placed on the bed/table using various adapters.

While the iSYS 1 System is only used to initially position the instrument prior to insertion into the patient and the insertion of the instrument is completed manually by the physician holding the instrument, the XACT Robotic System Insertion Module Assembly includes a jig which holds the instrument. Advancement of the instrument is controlled by the physician depressing a foot pedal to move the needle forward with the XACT Robotic System with the robot positioning unit making adjustments to the positioning of the instrument to ensure that the instrument remains on the physician defined trajectory path.

The XACT Robotic System also includes software for planning the instrument trajectory overlaid on CT images. The iSYS 1 System is used with separately cleared planning software to perform this function, though the cleared Perfint Maxio device includes its own planning software.

{5}------------------------------------------------

### Performance Data

The following performance/safety tests were conducted with the XACT Robotics System:

- . Mechanical Accuracy in accordance with the ASTM F2554-10 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems
- . Registration Precision in a simulated clinical environment under CT quidance
- CT Phantom Accuracy to evaluate device performance and system accuracy under simulated clinical conditions under CT quidance.
- . Rigidity of the system under different weight loads
- . Axes Mapping to evaluate the performance specifications of each of the axes of movement when individually operated
- . Needle Properties Testing to evaluate the mechanical strength of compatible needles under forces which may be applied by the XACT Robot during insertion
- Body Contour & Movement/Respiration Synchronization to evaluate compatibility with different body contours and that the movement sensor is synchronized with patient breathing motions
- System Lifetime Testing
- Electrical safety in accordance with IEC 60601-1
- . Electromagnetic compatibility in accordance with IEC 60601-1-2
- . Biocompatibility in acordance with ISO 10993 for all patient-contacing componnets of the system

Pre-clinical animal testing was conducted on the XACT Robotic System to demonstrate the safety and effectiveness of the device. The study in a porcine model evaluated the accuracy of the XACT system as determined by the measurement of the distance from the target and reporting on adverse events. A total of 40 interventional procedures were performed in different body organs while animals were under general anesthesia. The XACT system reached the target in all of the 40 interventional procedures. The average accuracy of the system was 1.41mm ± 0.94mm.

The results of the testing demonstrate that the XACT System meets its specifications and that CT-quided interventional procedures can be performed safely and successfully consistent with the target accuracy specification.

Preliminary clinical data, to supplement the bench and preclinical animal testing and confirm the safety and performance of the system was gathered in 2 clinical centers, where the XACT Robotic System was used to plan an instrument trajectory and advance the instrument to the target in subjects undergoing CT-guided procedures in the abdominal area. Out of the 31 cases from an interim set of subjects where an instrument was inserted using the XACT Robotic System, the device successfully reached the target in 27 cases. In 4 out of the 31 subjects, the procedure was completed successfully with manual insertion according to routine clinical procedure. The system met its predefined specifications for use to guide and advance an instrument during CT-guided interventional procedures. There were no serious adverse events reported for any of the subjects.

# Conclusions

The XACT Robotic System has the same intended uses and similar indications for use,

{6}------------------------------------------------

technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the XACT Robotic System and its predicate devices do not raise different issues of safety or effectiveness. Preclinical performance data demonstrate that the XACT Robotic System meets its defined specifications and that it is as safe and effective as the predicate devices. Preclinical and clinical performance with the device further supports its substantial equivalence to the predicates, providing evidence that the XACT Robotic System can be used to assist in the planning and advancement of an instrument during CT-guided percutaneous procedures with no serious adverse events specific to use of the system. Thus, the XACT Robotic System is substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K191332](https://fda.innolitics.com/device/K191332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
