Megadyne Foot Switch

K191328 · Megadyne Medical Products, Inc. · FYD · Jul 26, 2019 · General, Plastic Surgery

Device Facts

Record IDK191328
Device NameMegadyne Foot Switch
ApplicantMegadyne Medical Products, Inc.
Product CodeFYD · General, Plastic Surgery
Decision DateJul 26, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5070
Device ClassClass 2

Indications for Use

The Megadyne Foot Switch is an accessory that is intended to interface with compatible smoke evacuation systems to control on/off inputs for manual operation.

Device Story

Megadyne Foot Switch is a wired accessory designed to interface with compatible smoke evacuation systems. It functions as a manual input control, allowing clinicians to toggle the on/off state of the smoke evacuation system during surgical procedures. The device is operated by the user's foot in the surgical operating room environment. By providing a hands-free method to activate or deactivate smoke evacuation, it assists in maintaining a clear surgical field and managing surgical smoke, potentially benefiting the patient by reducing exposure to hazardous smoke byproducts. The device is non-sterile and intended for use with specific compatible equipment.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), IPX8 ingress protection (IEC 60529), and mechanical durability testing.

Technological Characteristics

Wired foot switch; non-sterile; IPX8 rated per IEC 60529; compliant with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC). Non-patient contact. No wireless capability.

Indications for Use

Indicated for use as an accessory to interface with compatible smoke evacuation systems to control on/off inputs for manual operation. No specific patient population, age, or gender restrictions; no contraindications.

Regulatory Classification

Identification

Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.

Special Controls

*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 26, 2019 Megadyne Medical Products, Inc. % Ryoji Sakai Associate Director, Regulatory Affairs Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, Ohio 45242 Re: K191328 Trade/Device Name: Megadyne Foot Switch Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus For A Surgical Operating Room Regulatory Class: Class II Product Code: FYD Dated: May 15, 2019 Received: May 16, 2019 Dear Ryoji Sakai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191328 Device Name Megadyne Foot Switch Indications for Use (Describe) The Megadyne Foot Switch is an accessory that is interface with compatible smoke evacuation systems to control on/off inputs for manual operation. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary Company Megadyne Medical Products, Inc. 11506 South State St. Draper, UT 84020 - Contact Ryoji Sakai Associate Director, Regulatory Affairs Ethicon Endo-Surgery, Inc. Telephone: (513) 337-8586 Email:rsakai3@its.jnj.com # Date Prepared July 17, 2019 # Device Name Trade Name: Megadyne Foot Switch Common Name: Foot Switch # Classification Name Air-handling apparatus for surgical operating room (21 CFR 878.5070, Product Code FYD) # Regulatory Class Class II # Predicate Devices IR Wireless Foot Switch cleared under K053510 on March 20, 2006 # Device Description The Megadyne Foot Switch is used with a compatible smoke evaluation system to control on/off inputs for manual operation. # Indications for Use The Megadyne Foot Switch is an accessory that is intended to interface with compatible smoke evacuation systems to control on/off inputs for manual operation. {5}------------------------------------------------ # Comparison with The Predicate Device | Indications for Use | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Megadyne Foot Switch<br>(Subject device) | IR Wireless Foot Switch<br>(Predicate device) | | The Megadyne Foot Switch is an accessory that<br>is intended to interface with compatible smoke<br>evacuation systems to control on/off inputs for<br>manual operation. | The Linemaster wireless foot switch will be an<br>accessory to and provide foot switch input<br>control to any medical device that uses a<br>switch closure (on/off), to activate said device. | | Contraindications | | | None | None | #### Device Comparison Table - Technology, and Performance Specifications | Characteristic/Specification | Megadyne Foot Switch<br>(Subject device) | IR Wireless Foot Switch<br>(Predicate device) | |------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------| | Sterility Method | Non-Sterile | Same | | Connection method | Wired connection | Wireless connection | | Compatible generator | Smoke evacuation system<br>(Classification Name:<br>Air-handling apparatus for<br>surgical operating room) | Any class II medical device<br>that uses a foot switch | | Electrical Safety<br>/ Mechanical Safety | Tested and compliant with IEC<br>60601-1 and IEC 60601-1-2 | Same | | Biocompatibility Safety | Non-patient contact | Same | | Radiation Safety | Non-radioactive | Same | The subject Megadyne Foot Switch and predicate IR Wireless Foot Switch control the same on/off inputs, however the Megadyne Foot Switch does not have the wireless capability. Electromagnetic Compatibility and Electrical Safety testing were performed to ensure that the different technological characteristics do not raise different questions of safety and effectiveness. # Summary of Non-Clinical Testing # Electrical safety and electromagnetic compatibility (EMC ) Electrical safety and EMC testing were conducted on the Megadyne Foot Switch to comply with IEC 60601-1-2 for electromagnetic compatibility and IEC 60601-1 for electrical safety. # Durability of Marking Test This footswitch meets the requirements for an IPX8 rating per IEC 60529 Ed. 2.2 and Legibility of the Label and Durability of the marking per IEC 60601-1 Ed. 3.1. # Mechanical Test Report The footswitch meets the IEC 60601-1 Ed. 3.1 standard for Shipping, Storage and ME Test. # Clinical Studies {6}------------------------------------------------ # Conclusion The conclusions drawn from the nonclinical tests demonstrate that the Megadyne Foot Switch is as safe, as effective, and performs as well as or better than the legally marketed device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...