← Product Code [MBI](/productcode/MBI) · K191319

# Dunamis Fixation Button System (K191319)

_Dunamis, LLC · MBI · Aug 29, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191319

## Device Facts

- **Applicant:** Dunamis, LLC
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Aug 29, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Dunamis Fixation Button System is to be used for fixation of bone or soft tissue to bone. The components are intended to serve as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, the Dunamis Fixation Button System is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon repair/ reconstruction, including providing fixation during the healing process following acromioclavicular separations, as an adjunct to fracture repair in syndesmotic trauma, and ACL and PCL repair.

## Device Story

The Dunamis Fixation Button System is a mechanical orthopedic implant used for bone-to-bone or soft tissue-to-bone fixation. The system consists of titanium fixation buttons, non-absorbable suture loops, optional passing sutures, and optional button supports. It functions as a fixation post or distribution bridge to manage suture tension during ligament or tendon repair. The device is used by surgeons in an operating room setting during orthopedic procedures such as ACL/PCL reconstruction, acromioclavicular repair, or syndesmotic fracture stabilization. By providing stable fixation during the healing process, the device supports tissue approximation and structural integrity. The surgeon implants the device to secure soft tissue or bone fragments, facilitating natural healing.

## Clinical Evidence

Bench testing only. Testing included static and dynamic mechanical testing, sterilization validation, sterile barrier packaging performance, and shelf-life aging. The device complies with LAL testing requirements for orthopedic implants per AAMI ST-72.

## Technological Characteristics

System components include titanium fixation buttons and non-absorbable suture loops. The device functions as a mechanical fixation fastener. It is provided sterile. No software or electronic components are involved.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex TightRope ([K043248](/device/K043248.md), [K052776](/device/K052776.md), [K112990](/device/K112990.md))
- RiverPoint OrthoButton CL ([K160655](/device/K160655.md))
- RiverPoint OrthoButton AL ([K171060](/device/K171060.md))
- Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device ([K082095](/device/K082095.md))

## Reference Devices

- Dunamis Force DFX Suture ([K150327](/device/K150327.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

Dunamis LLC % Hollace Saas Rhodes Vice President, Orthopedic Regulatory Affairs MCRA. LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001

Re: K191319

Trade/Device Name: Dunamis Fixation Button System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HTN Dated: July 24, 2019 Received: July 24, 2019

Dear Ms. Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191319

Device Name Dunamis Fixation Button System

#### Indications for Use (Describe)

The Dunamis Fixation Button System is to be used for fixation of bone or soft tissue to bone. The components are intended to serve as fixation posts, a distributing suture tension over areas of ligament or tendon repair.

Specifically, the Dunamis Fixation Button System is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon repair/ reconstruction fixation during the healing process following acromioclavicular separations, as an adjunct to fracture repair in syndesmotic trauma, and ACL and PCL repair.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Manufacturer:       | Dunamis Medical , LLC<br>509 E. Commerce Street, Suite 3<br>Greenville, AL 36037<br>Phone: 731.217.2533                                                              |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:            | Dr. PrithviRaj Chavan<br>President                                                                                                                                   |
| Prepared By:        | MCRA, LLC<br>1050 K Street, NW, Suite 1000<br>Washington, DC 20001<br>Phone: 202.552.5800                                                                            |
| Date Prepared:      | August 28, 2019                                                                                                                                                      |
| Device Trade Name:  | Dunamis Fixation Button System                                                                                                                                       |
| Device Common Name: | Fastener, fixation, nondegradable, soft tissue                                                                                                                       |
| Classification:     | 21 CFR 888.3040 – Fastener, fixation, nondegradable, soft tissue<br>21 CFR 888.3030 – Single/multiple component metallic bone<br>fixation appliances and accessories |
|                     | Class II                                                                                                                                                             |
|                     | Product Codes:                                                                                                                                                       |

## Indications for Use:

The Dunamis Fixation Button System is to be used for fixation of bone or soft tissue to bone. The components are intended to serve as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair.

Specifically, the Dunamis Fixation Button System is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon repair/ reconstruction, including providing fixation during the healing process following acromioclavicular separations, as an adjunct to fracture repair in syndesmotic trauma, and ACL and PCL repair.

## Device Description:

The Dunamis Fixation Button System is to be used for fixation of bone or soft tissue to bone, and is intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. The Dunamis Fixation System consists of titanium fixation buttons, non-absorbable suture loops, optional passing sutures, and optional button supports.

{4}------------------------------------------------

## Predicate Devices:

The Dunamis Fixation Button System is substantially equivalent to the following predicates with respect to indications, design, material and function: Arthrex TightRope (K043248, K052776, K112990), RiverPoint OrthoButton CL (K160655), RiverPoint OrthoButton AL (K171060) and Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device (K082095). The sutures proposed for use with the subject system are those used with the reference device, Dunamis Force DFX Suture (K150327), with respect to material, design, and function.

## Substantial Equivalence:

Non-clinical testing performed included static testing, dynamic testing, sterilization, sterile barrier packaging performance and aging of sterile medical device. The non-clinical performance testing conducted demonstrates that the insertion and fixation properties of the Dunamis Fixation Button System (with Continuous Loop or Adjustable Loop) are substantially equivalent to the predicate device, Arthrex TightRope (K112990). Additionally, the Dunamis Fixation Button System is in compliance with LAL testing requirements for orthopedic implants per AAMI ST-72 testing.

---

**Source:** [https://fda.innolitics.com/device/K191319](https://fda.innolitics.com/device/K191319)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
