BonAlive® granules is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. BonAlive® granules is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product contains a bone void filler that resorbs and is replaced with bone during the healing process. When used in the extremities and pelvis, BonAlive® granules is intended to be used alone.
Device Story
BonAlive® Granules are synthetic, osteoconductive bone void fillers composed of bioactive glass (SiO2, Na2O, CaO, P2O5). Used by surgeons to fill non-structural bony voids or gaps in the extremities and pelvis. The material is packed into the defect; in the presence of body fluids, it leaches ions to form a silica-gel layer, which facilitates calcium phosphate precipitation and hydroxyapatite crystallization. This process enables chemical bonding to surrounding bone and promotes new bone formation. The granules are radiodense, allowing for postoperative radiologic evaluation, and resorb over time as they are replaced by natural bone. The device is provided sterile for single-use.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including apatite formation, heavy metal composition, crystallinity, particle size distribution, surface area, and biocompatibility (cytotoxicity, sensitization, systemic toxicity, muscle implantation).
Technological Characteristics
Synthetic bioactive glass (SiO2, Na2O, CaO, P2O5). Amorphous, non-porous, random-shaped particles. Granule sizes: 0.5-0.8 mm and 1.0-2.0 mm. Radiodense. Sterilized via gamma irradiation (SAL 10^-6). Biocompatible per ISO 10993.
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (extremities and pelvis) not intrinsic to structural stability. Defects may be surgically created or traumatic. Contraindicated for use in posterolateral spine applications.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
BonAlive Biomaterials Ltd. % Elisa Maldonado-Holmertz RA/QA Consultant Obelix Biotech Solutions 12416 Fairfax Ridge Place Austin, Texas 78738
Re: K191274
Trade/Device Name: BonAlive® Granules Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: May 10, 2019 Received: May 15, 2019
Dear Ms. Maldonado-Holmertz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191274
Device Name
BonAlive® Granules
#### Indications for Use (Describe)
BonAlive® granules is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. BonAlive® granules is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. The product contains a bone void filler that resorbs and is replaced with bone during the healing process. When used in the extremities and pelvis, BonAlive® granules is intended to be used alone.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 005 - 510(k) Summary
### 1. Submission Sponsor
Jimmy Lucchesi Chief Technology Officer BonAlive Biomaterials Ltd. Biolinja 12 20750 Turku Finland Email: Jimmy.Lucchesi@bonalive.com Tel number: +358 50 552 4671
### 2. Submission Correspondent
Obelix Consulting, LLC 12416 Fairfax Ridge Place Austin, TX 78738 USA Elisa Maldonado-Holmertz RA/QA Consultant Email: elisamh@obelixconsult.com Tel number: 512.431.6069
#### 3. Date Prepared
10 May 2019
## 4. Device Identification
| Type of 510(k) Submission: | Traditional |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Trade or Proprietary Name: | BonAlive Granules |
| Common or Usual Name: | Filler, Bone Void, Calcium Compound |
| Regulation Description: | Resorbable calcium salt bone void filler device |
| Regulation Classification: | 888.3045 |
| Product Code: | MQV |
| Class of Device: | Class II |
| Review Panel: | Office of Device Evaluation (ODE)<br>Division of Orthopedic Devices (DOD)<br>Restorative and Repair Devices Branch (RRDB) |
| Reason for Submission: | New device |
| Prior Related Submissions: | Submission K110024 received NSE due to<br>osteostimulation and antibacterial claims. This<br>current submission has removed such claims. |
| Multiple Devices: | None |
{5}------------------------------------------------
## 5. Legally Marketed Predicate Device(s)
K071199 (Rx Only) BonAlive Granules by Vivoxid Ltd. (now called BonAlive Biomaterials, Ltd.)
## 6. Device Description
BonAlive® granules is composed of osteostimulative calcium-phosphorous-sodium-silicate (glass 553P4) granules (size 0.5-0.8 mm or 1.0-2.0 mm) and is a sterile medical device. This synthetic, osteoconductive material is comprised of SiO2, Na2O, CaO and P2O5. Bioactive glass is characterized by its ability to attach firmly to living tissue. Other properties include its ability to facilitate bone tissue growth, bond chemically with surrounding bone, and promote new bone formation in the implanted area.
In aqueous solution (e.g. body fluids), bioactive glass works by leaching out ions and developing a silica-gel layer which acts as a template for a calcium phosphate (CaP) precipitation. The CaP crystallizes to hydroxyapatite, which resembles the mineral phase of natural bone in its chemical composition and structure, thus enabling bonding of the bioactive glass to the surrounding bone.
The BonAlive® granules resorb and are replaced with bone slowly over a period of years. The bioactive glass in the BonAlive® granules is radiodense thus enabling postoperative radiologic evaluation.
Best results are obtained by ensuring close contact of the device with surrounding bone tissue and by carefully following the Instructions for Use.
BonAlive® granules is sterilized by irradiation and is available in different granule and unit sizes.
## 7. Indication for Use Statement
BonAlive® granules is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. BonAlive® granules is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product contains a bone void filler that resorbs and is replaced with bone during the healing process. When used in the extremities and pelvis, BonAlive® granules is intended to be used alone.
RX Only
{6}------------------------------------------------
## 8. Substantial Equivalence Discussion
The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| | Comparison of Characteristics | | |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Manufacturer | SUBJECT DEVICE<br>BonAlive Biomaterials, Ltd | PREDICATE DEVICE<br>Vivoxid Ltd (now called<br>BonAlive Biomaterials, Ltd.) | SIGNIFICANT<br>DIFFERENCES |
| Trade Name | BonAlive Granules | BonAlive Granules | None |
| 510(k) Number | - | K071199 | |
| Product Code | MQV | MQV | None |
| Regulation Number | 888.3045 | 888.3045 | None |
| Regulation Name | Resorbable calcium salt<br>bone void filler device | Resorbable calcium salt<br>bone void filler device | None |
| Indications for Use | BonAlive® granules is<br>indicated only for bony<br>voids or gaps that are not<br>intrinsic to the stability of<br>the bony structure.<br>BonAlive® granules is<br>indicated to be gently<br>packed into bony voids or<br>gaps of the skeletal system<br>(i.e. the extremities and<br>pelvis). These defects may<br>be surgically created<br>osseous defects or osseous<br>defects created from<br>traumatic injury to the<br>bone. The product<br>contains a bone void filler<br>that resorbs and is<br>replaced with bone during<br>the healing process. When<br>used in the extremities and<br>pelvis, BonAlive® granules<br>is intended to be used<br>alone. | BonAlive® granules is<br>indicated only for bony<br>voids or gaps that are not<br>intrinsic to the stability of<br>the bony structure.<br>BonAlive® granules is<br>indicated to be gently<br>packed into bony voids or<br>gaps of the skeletal system<br>(i.e. the extremities and<br>pelvis). These defects may<br>be surgically created<br>osseous defects or osseous<br>defects created from<br>traumatic injury to the<br>bone. The product provides<br>a bone void filler that<br>resorbs and is replaced<br>with bone during the<br>healing process. When<br>used in the extremities and<br>pelvis, BonAlive® granules<br>is intended to be used<br>alone. The device is not<br>intended for use in<br>posterolateral spine<br>applications. | None |
Comparison of Characteristics
{7}------------------------------------------------
| Manufacturer | New Product | PREDICATE | SIGNIFICANT |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Manufacturer | BonAlive Biomaterials, Ltd | Vivoxid Ltd (now called<br>BonAlive Biomaterials, Ltd.) | DIFFERENCES |
| Trade Name | BonAlive Granules | BonAlive Granules | None |
| | | indicated for use in load-<br>bearing applications,<br>therefore, standard<br>internal or external<br>stabilization techniques<br>must be followed to obtain<br>rigid stabilization. | |
| Rx or OTC | Rx | Rx | None |
| Physical Form | Amorphous, non-porous<br>random-shaped particles | Amorphous, non-porous<br>random-shaped particles | None |
| Color | Brown | Slightly pink | Color is different |
| Materials<br>Composition | SiO2, Na2O, CaO and P2O5 | SiO2, Na2O, CaO and P₂O5 | None |
| | Granule sizes:<br>0.5-0.8 mm<br>1.0-2.0 mm | Granules sizes:<br>0.5-0.8 mm<br>1.0-2.0<br>2.0-3.15 mm | Largest granule size<br>2.0-3.15 mm is<br>obsolete |
| | Product volumes:<br>1 cc<br>2.5 cc<br>5 cc<br>10 cc | Product volumes:<br>1 cc<br>2 cc<br>4 cc<br>16 cc | Volumes are<br>different |
| Biocompatibility | Biocompatible<br>ISO 10993 | Biocompatible<br>ISO 10993 | None |
| Sterilization, SAL | Gamma<br>Sterile, SAL 10-6 | Dry Heat<br>Sterile, SAL 10-6 | None |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | None |
| Single Use/<br>Reuse | Single use only | Single use only | None |
| Mode of action | Works by leaching out ions<br>that react with the body<br>fluids transforming the<br>glass surface chemically<br>into one that by its<br>chemical composition and<br>structure resembles the<br>mineral phase found in<br>natural bone. | Works by leaching out ions<br>that react with the body<br>fluids transforming the<br>glass surface chemically<br>into one that by its<br>chemical composition and<br>structure resembles the<br>mineral phase found in<br>natural bone. | None |
| Properties | Synthetic | Synthetic | None |
{8}------------------------------------------------
| Manufacturer | New Product<br>BonAlive Biomaterials, Ltd | PREDICATE<br>Vivoxid Ltd (now called<br>BonAlive Biomaterials, Ltd.) | SIGNIFICANT<br>DIFFERENCES |
|--------------|-------------------------------------------|----------------------------------------------------------------------|----------------------------|
| Trade Name | BonAlive Granules | BonAlive Granules | None |
| | Osteoconductive | Osteoconductive | |
#### 9. Non-Clinical Performance Data
The following testing has been performed to support substantial equivalence to the predicate:
- Apatite
- Composition Heavy Metals ●
- Crystallinity
- Particle Size Distribution
- Surface Area
- Manufacturing & Specifications Validation
As part of demonstrating safety and effectiveness of BonAlive Granules, and in showing substantial equivalence to the predicate device, BonAlive completed a number of tests. The BonAlive Granules meets all the requirements for overall design and confirms that the output meets the design inputs and specifications. The BonAlive Granules passed all testing stated above as shown by the acceptable results obtained.
The BonAlive Granules complies with the applicable voluntary, national and international standards and guidance.
## 10. Biocompatibility
Biocompatibility testing was conducted in accordance with ISO-10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" The following biocompatibility studies were successfully performed with the BonAlive® granules: cytotoxicity, sensitization, systemic toxicity, and muscle implantation.
The granules predicate device was depyrogenated at 250 degrees Celsius for one hour prior to dry heat sterilization. Bacterial endotoxins were investigated in a LAL gel clot test. The analysis was performed according to Ph. Eur. 2.6.14 method A. The log reduction in endotoxin levels in the dry heat depyrogenation cycles was 4.6 logs. The results met USP requirement for bacterial endotoxins.
Similarly, as for the predicate device, a depyrogenation step at 250 degrees Celsius is performed for the granules subject device prior the gamma irradiation. The granules material and the depyrogenation method are the same for the predicate and the subject devices, thus the log reduction in the endotoxin levels for the predicate and the subject devices are equivalent. The testing performed on the predicate device supports the pyrogen status of the subject device. The predicate and subject devices are substantially equivalent in their nonpyrogenic nature.
{9}------------------------------------------------
## 11. Sterility and Shelf Life
The BonAlive Granules and Applicator (Granules) are sterilized with radiation. The sterility of the Granules was assured by using a validated sterilization method qualified in accordance with EN ISO 11137. After the sterilization cycle was completed, the Granules were placed on real-time shelf-life testing for a length of time to determine any degradation or loss of sterility over the anticipated shelf-life period of five (5) years. The validations and shelf life testing were completed successfully in accordance with ISO 11607.
### 12. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
### 13. Statement of Substantial Equivalence
A device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device. BonAlive Granules is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.