← Product Code [MAI](/productcode/MAI) · K191242

# HEALIX ADVANCE SP BIOCOMPOSITE Anchor (K191242)

_Medos International SARL · MAI · Aug 22, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191242

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [MAI](/productcode/MAI.md)
- **Decision Date:** Aug 22, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The HEALIX ADVANCE SP BIOCOMPOSITE Anchor is indicated for use in the following procedures for reattachment of soft tissue to bone: Shoulder: Rotator cuff repair, biceps tenodesis, deltoid repair Knee: Posterior oblique repair, medial collateral ligament (MCL) reconstruction, lateral ligament reconstruction, iliotibial (IT) band tenodesis Foot/Ankle: Achilles tendon repair

## Device Story

Two-piece self-punching suture anchor system; consists of cannulated threaded anchor and dilator; preloaded on disposable inserter shaft with handle. Anchor molded from absorbable Biocryl biocomposite (PLA and TCP); dilator made of non-absorbable PEEK. Held by UHMWPE and PET stay-suture. Used in orthopedic surgery for soft tissue reattachment to bone. Implanted permanently. Provides mechanical fixation of soft tissue to bone; facilitates healing. Sterile, single-use device.

## Clinical Evidence

Bench testing only. Performance evaluation included in-vitro fixation strength at zero, six, and twelve weeks; insertion torque; torque to failure; and cyclic migration at time zero. Sterilization validated per ANSI/AAMI/ISO 11135:2014; EO residuals per AAMI/ANSI/ISO 10993-7:2008; non-pyrogenicity per ANSI/AAMI ST-72:2011.

## Technological Characteristics

Two-piece cannulated threaded anchor (Biocryl PLA/TCP) and dilator (PEEK). UHMWPE/PET stay-suture. Sterile (Ethylene Oxide per ANSI/AAMI/ISO 11135:2014). Sizes: 4.9mm and 5.5mm. Mechanical fixation principle.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex PushLock Anchors ([K101679](/device/K101679.md))
- Arthrex PushLock Anchors ([K082810](/device/K082810.md))
- Arthrex PushLock Anchors ([K071177](/device/K071177.md))
- Arthrex PushLock Anchors ([K061863](/device/K061863.md))
- Arthrex PushLock Anchors ([K051219](/device/K051219.md))

## Reference Devices

- HEALIX ADVANCE™ SP PEEK Anchor ([K182941](/device/K182941.md))
- Bio-INTRAFIX Tibial Tapered Screws and Sheaths ([K032167](/device/K032167.md))
- SPEEDTRAP™ Graft Preparation System ([K161638](/device/K161638.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

Medos International SARL % Elizabeth Messana Regulatory Affairs Specialist II DePuy Synthes Mitek, a Johnson and Johnson Company 325 Paramount Drive Raynham, Massachusetts 02767

Re: K191242

Trade/Device Name: HEALIX ADVANCE™ SP BIOCOMPOSITE Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: July 11, 2019 Received: July 15, 2019

Dear Ms. Messana:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191242

Device Name HEALIX ADVANCE™ SP BIOCOMPOSITE Anchor

Indications for Use (Describe)

The HEALIX ADVANCE SP BIOCOMPOSITE Anchor is indicated for use in the following procedures for reattachment of soft tissue to bone:

· Shoulder: Rotator cuff repair, biceps tenodesis, deltoid repair

· Knee: Posterior oblique repair, medial collateral ligament (MCL) reconstruction, lateral ligament reconstruction, iliotibial (IT) band tenodesis

· Foot/Ankle: Achilles tendon repair

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY HEALIX ADVANCE™ SP BIOCOMPOSITE Anchor Date Prepared: 05/7/2019

| Submitter's<br>Name and<br>Address | DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>On behalf of:<br>Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                        |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Contact Person                     | Elizabeth Messana<br>Regulatory Affairs Specialist II<br><br>DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Telephone: 508-828-3150<br>Email: emessan1@its.jnj.com |
| Name of Medical<br>Device          | Proprietary Name: HEALIX ADVANCE™ SP BIOCOMPOSITE<br>Anchor<br>Classification Name: Single/multiple component metallic bone fixation<br>appliances and accessories<br>Product Code: MAI<br>Common Name: Suture Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                        |
| Substantial<br>Equivalence         | The HEALIX ADVANCE SP BIOCOMPOSITE Anchor is<br>substantially equivalent to:<br>K101679, K082810, K071177, K061863, K051219 Arthrex PushLock Anchors              Reference Devices:  K182941, HEALIX ADVANCE™ SP PEEK Anchor K032167, Bio-INTRAFIX Tibial Tapered Screws and Sheaths K161638, SPEEDTRAP™ Graft Preparation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                        |
| Device<br>Classification           | HEALIX ADVANCE SP BIOCOMPOSITE Anchor is classified as:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                        |
|                                    | Single/multiple component metallic bone fixation appliances and<br>accessories, classified as Class II, product code MAI, regulated under<br>21 CFR 888.3030.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                        |
| Device Panel                       | Orthopedic Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                        |
| Device<br>Description              | The HEALIX ADVANCE SP BIOCOMPOSITE Anchor is a two-piece<br>"self-punching" design consisting of a cannulated, threaded anchor and<br>dilator which are permanently implanted into the body. The anchor-<br>dilator are preloaded on a disposable inserter shaft with handle, held in<br>place by a non-absorbable stay suture comprised of Ultra-High<br>Molecular Weight Polyethylene (UHMWPE) and Green poly (ethylene<br>terephthalate) (PET).  The proposed HEALIX ADVANCE SP<br>BIOCOMPOSITE Anchor is offered in two sizes, 4.9 mm and 5.5 mm<br>and is molded from Absorbable Biocryl Biocomposite ((Polylactic Acid<br>(PLA) and Tricalcium Phosphate (TCP)) material.  The dilator is<br>comprised of Non-absorbable PEEK (Polyetheretherketone).  The<br>HEALIX ADVANCE SP BIOCOMPOSITE Anchor is provided sterile<br>via Ethylene Oxide (EO) sterilization and is for single use only. |                                                        |
| Technological<br>Characteristics   | The suture anchor design, anchor materials, principal of operation and<br>indications for use are similar when compared to the predicate Arthrex<br>PushLock Anchors (K101679, K082810, K071177, K061863,<br>K051219).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                        |
|                                    | The suture anchor design, principal of operation and indications for use<br>are identical when compared to the reference device- HEALIX<br>ADVANCE™ SP PEEK Anchor (K182941)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                        |
|                                    | The Biocryl (PLA/ ß-TCP) Biocomposite material of the anchor is<br>identical to the Biocryl in the reference device- Bio-INTRAFIX Tibial<br>Tapered Screws and Sheaths (K032167).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                        |
|                                    | The PEEK (Polyetheretherketone) design and material of the dilator<br>component is identical to the PEEK dilator in the reference device-<br>HEALIX ADVANCE™ SP PEEK Anchor (K182941).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                        |
|                                    | The stay-suture material (UHMWPE/PET) is identical to the winding<br>suture present in the reference device- SPEEDTRAPTM Graft<br>Preparation System (K161638).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                        |
| Indications for Use                | The HEALIX ADVANCE SP BIOCOMPOSITE Anchor is indicated<br>for use in the following procedures for reattachment of soft tissue to<br>bone:<br><br>• Shoulder: Rotator cuff repair, biceps tenodesis, deltoid repair<br>• Knee: Posterior oblique repair, medial collateral ligament (MCL)<br>reconstruction, lateral collateral ligament reconstruction, iliotibial<br>(IT) band tenodesis<br>• Foot/Ankle: Achilles tendon repair                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                        |
| Non-clinical Testing               | Verification activities were performed on the proposed device and / or<br>its predicates. Performance testing included evaluation of <i>in-vitro</i><br>fixation strength at zero, six and twelve weeks, insertion torque, torque<br>to failure and cyclic migration at time zero.<br><br>Ethylene Oxide Sterilization was validated according to<br>ANSI/AAMI/ISO 11135: 2014 to a SAL of 1 x 10-6.<br><br>EO residuals were tested per AAMI/ANSI/ISO 10993-7:2008<br><br>The proposed device has been determined to be non-pyrogenic per the<br>requirements set forth in ANSI/AAMI ST-72:2011, United States<br>Pharmacopeia (USP), and European Pharmacopeia (EP) using the<br>bacterial endotoxin testing (BET) method.                                                                                                                                                                       |                                                        |
| Safety and Performance             | Results of performance testing have demonstrated that the proposed<br>devices are suitable for their intended use.<br><br>Based on similarities in the indications for use, technological<br>characteristics, and performance in comparison to the predicate devices,<br>the proposed HEALIX ADVANCE SP BIOCOMPOSITE Anchor has<br>shown to be substantially equivalent to the predicate devices under the<br>Federal Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                        |

{4}------------------------------------------------

{5}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K191242](https://fda.innolitics.com/device/K191242)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
