The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.
Device Story
Cybird WPT is a handheld LED curing light used by dental professionals to polymerize light-cured dental materials. The device features an industrial-grade aluminum body for heat dissipation and uses blue/violet LEDs to emit light in the 400-480 nm range. It incorporates black fiber optic glass light guides to reduce high-intensity light exposure to patients and operators. The device offers multiple curing modes (High Power, Plasma/Ortho) with adjustable exposure times. It is powered by a 3.7V Li-ion battery, rechargeable via an inductive wireless charger. The clinician selects the mode and time via buttons on the wand handle. By delivering specific light intensity over controlled durations, the device ensures proper polymerization of restorative composites and bonding agents, facilitating dental procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including irradiance intensity, spectral output, depth of cure (per ADA Specification No. 48), biocompatibility (ISO 10993-5, ISO 10993-10), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Industrial-grade aluminum body; blue/violet LED light source (dual peak ~405nm, 460nm); 3.7V 2600mAh Li-ion battery; inductive wireless charging; black fiber optic glass light guides; chemical disinfection (isopropyl alcohol, ethyl alcohol-based cleaners). Complies with AAMI/ANSI/ISO 10993, AAMI ANSI ES 60601-1, and IEC 60601-1-2.
Indications for Use
Indicated for polymerization of resinous dental materials, restorative composites, orthodontic brackets, and bonding/sealing materials in the 400-480 nm visible light waveband. For use by dental professionals.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
Predicate Devices
Cybird LED Curing Light (Model: Gold / XD) (K173876)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
January 10, 2020
DXM Co., Ltd % Dave Kim President Mtech Group 8310 Buffalo Speedway Houston, Texas 77025
Re: K191221
Trade/Device Name: Cybird LED Curing Light Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator for Polymerization Regulatory Class: Class II Product Code: EBZ Dated: November 29, 2019 Received: December 11, 2019
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191221
Device Name Cybird LED Curing Light Model: Cybird WPT
### Indications for Use (Describe)
The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the DXM logo. The logo consists of a green curved shape on the left and the letters "dxm" in blue on the right. The green shape is abstract and appears to be a stylized leaf or wave. The letters "dxm" are in lowercase and have a slightly blurred or faded appearance.
# 510(k) Summarv
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
## Date 510k summary prepared:
December 6, 2019
+713-467-2607
#### I. SUBMITTER
Submitter's Name Submitter's Address
Submitter's Telephone
Submitter's Fax
Contact person
DXM Co., Ltd. Baekseok-dong, Ilsan technotown 1003, 138, Ilsan-ro, Ilsandong-gu Goyang-si, Gyeonggi-do, Republic of Korea 411-360 +82 (31) 909-8275 +82 (31) 909-8276 Seung Hyun Kim / RA Team Manager
Dave Kim (davekim(@mtech-inc.net)
7707 Fannin St. Ste 200 V111, TX 77054
Official Correspondent Address Telephone
#### II. DEVICE
Trade/proprietary Name Model: Common or Usual Name Regulation Name Regulation Number Product Code Regulatory Class Review panel
Cybird LED Curing Light Cybird WPT Dental Curing Light Device Ultraviolet Activator for Polymerization 21 CFR 872.6070 EBZ Class II Dental
#### III. PREDICATE DEVICE
Primary Manufacturer Trade/proprietary Name Model: 510K Number Common or Usual Name Regulation Name Regulation Number Product Code Regulatory Class Review panel
DXM Co., Ltd Cybird LED Curing Light Cybird GOLD / Cybird XD K173876 Dental Curing Light Device Ultraviolet Activator for Polymerization 21 CFR 872.6070 EBZ Class II Dental
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for DXM. The logo consists of a green curved shape on the left, resembling a stylized checkmark or swoosh. To the right of the green shape, the letters "dxm" are written in a light blue sans-serif font. The overall design is simple and modern.
#### IV. DEVICE DESCRIPTION
Cybird is a LED curing light intended for polymerization of light-cured materials by dental professionals. This product is effective on various light-cured materials. Cybird's body is made from industrial-grade aluminum which is a durable material intended to ensure heat dissipation. The Cybird features multiple curing modes to ensure functionality.
#### V. INDICATIONS FOR USE:
The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photopolymerized in the 400-480 nm waveband of visible light.
| Descriptive<br>Information | Device: Cybird LED Curing Light<br>Model: Cybird WPT | Device: Cybird LED Curing Light<br>Model: Gold, XD, (K173876) | SE |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Intended Use | The Cybird LED Curing Light is intended to polymerize resinous dental<br>materials, restorative composite<br>materials, and orthodontic brackets,<br>bonding and sealing materials that are<br>photo-polymerized in the 400-480 nm<br>waveband of visible light. | The Cybird LED Curing Light is intended to polymerize resinous dental<br>materials, restorative composite<br>materials, and orthodontic brackets,<br>bonding and sealing materials that are<br>photo-polymerized in the 400-480 nm<br>waveband of visible light. | Same |
| Device<br>Design | Batteries: Li-ion with a working voltage<br>of: 3.7V, 2600mAh. Their safety<br>ratings: CE, RoHS, WEEE<br><br>Charger: Inductive Wireless Charger<br><br>Power On Button: Located on the<br>handle of the wand, front and back side | Batteries: Li-ion with a working voltage<br>of: 3.7V, 2600mAh. Their safety<br>ratings: CE, RoHS, WEEE<br><br>Charger: 4.2VDC Lithium Ion<br>battery charger<br><br>Power On Button: Located on the<br>handle of the wand, front and back side | SE1 |
| Operational<br>modes | High Power Mode: (5, 10, 15, 20 sec)<br>Plasma / Ortho Mode:<br>(3, 6, 9, sec, Ortho mode 55sec)<br>Device indicates illumination<br>time selection | High Power Mode: (5, 10, 15, 20 sec)<br>Plasma / Ortho Mode:<br>(2, 3, 4, sec. Ortho mode 55sec)<br>Device indicates illumination<br>time selection | SE2 |
| Light source | LED light, blue and violet<br>Wavelengths<br>8 light head diameter | LED light, blue and violet<br>Wavelengths<br>8 light head diameter | Same |
| Accessories | | | |
| Composition<br>of patient<br>contacting<br>Materials | Aluminum, anodized white<br>Barrier Sleeves. Light Shield. | Aluminum, anodized white<br>Barrier Sleeves. Light Shield. | Same |
| Light Guide | Fiber optic glass Black (11x8, 11x11) | Fiber optic glass Clear (11x8, 11x11) | SE3 |
| Light<br>Intensity | Plasma E. Mode / Ortho Mode:<br>11x8 Black Long(2034mW/Cm2)<br>11x11 Black Long(1294mW/Cm2)<br>High Power Mode:<br>11x8 Black Long(1104mW/Cm2)<br>11x11 Black Long (675mW/Cm2) | Plasma E. Mode / Ortho Mode:<br>11x8 Clear Shor(2581mW/Cm2)<br>11x11 Clear Short(1800mW/Cm2)<br>High Power Mode:<br>11x8 Clear Shor(1521mW/Cm2)<br>11x11 Clear Short(10430mW/Cm2)W | SE4 |
| Peak<br>wavelength | Dual peak: Approximately 405nm,<br>460nm | Dual peak: Approximately 405nm,<br>460nm (GOLD), 460nm (XD) | Same |
| Depth of<br>cure | Plasma mode (3, 6, 9 sec)<br>Spident Escom 100 (K110428):<br>- 11x8 Black: 3.08~4.47 mm<br>- 11x11Black: 2.24 ~ 2.88 mm<br>High Power mode (5,10, 15, 20 sec)<br>Spident Escom 100 (K110428):<br>- 11x8 Black: 3.5~4.99mm<br>- 11x11Black: 3.05~3.95 mm | Plasma mode (2, 3, 4 sec)<br>Spident Escom 100 (K110428):<br>- 11x8 Clear: 3.21~4.12 mm<br>- 11x11Clear: 2.89 ~ 3.45 mm<br>High Power mode (5, 10, 15, 20 sec)<br>Spident Escom 100 (K110428):<br>- 11x8 Clear: 3.78~5.23 mm<br>- 11x11Clear: 3.41~4.45 mm | SE5 |
| Parameters<br>of<br>Disinfection | Chemical disinfection with approved<br>cleaning/sanitizing agents:<br>Cavicide products<br>Isopropyl alcohol (75%)<br>Ethyl alcohol based cleaners Lysol<br>disinfectant (alcohol-based only) | Chemical disinfection with approved<br>cleaning/sanitizing agents:<br>Cavicide products<br>Isopropyl alcohol (75%)<br>Ethyl alcohol based cleaners Lysol<br>disinfectant (alcohol-based only) | Same |
| Usability/Erg<br>onomics | 2 buttons – 1 power, 1 mode select | 2 buttons – 1 power, 1 mode select | Same |
| Power<br>source | 3.7V Li-ion Battery with inductive<br>wireless charging | 3.7V Li-ion Battery | Same |
#### VI. PREDICATE COMPARISON
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for DXM. The logo consists of a green curved shape on the left and the letters "dxm" in blue on the right. The green shape is abstract and appears to be a stylized representation of a wave or curve. The letters "dxm" are in lowercase and have a sans-serif font.
K191221
Discussion of the substantial equivalence decision;
Cybird WPT LED Curing Light has identical characteristics and intended uses as the 510(k) cleared light curing units, the Cybird LED curing Light (Model: Gold & XD, K173876), manufactured with identical materials and using the same energy source for the photopolymerization of dental materials, restorative composite materials and polymerization of bonding and sealing materials.
The intended use of the modified device has not changed.
The differences between the subject and the predicate device are listed below:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the DXM logo. The logo consists of a green curved shape on the left and the letters "dxm" in blue on the right. The curved shape is abstract and appears to be a stylized leaf or wave. The letters "dxm" are in lowercase and have a gradient effect, with the color fading from dark blue to light blue.
| SE1 | Device Design | Wireless Charger: The subject device is equipped with an Inductive Wireless Charger whereas the predicate device has 4.2VDC Lithium Ion battery charger. | The new wireless charger for the subject device delivers the same user convenience of a cordless charger for the predicate device. the spectral irradiance for both devices are identical. |
|-----|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SE2 | Operational modes | The exposure time in Plasma mode for Cybird WPT has been increased to 3, 6, 9 sec from 2, 3, 4 sec of the predicate device. | A longer light curing time for Plasma mode to achieve equivalent polymerization of RBC of black light guides in comparison with clear light guides, |
| SE3 | Light Guide | Fiber optic glass Black | Less light penetration through opaquer shade of a black light guide compared to a clear light guide. |
| SE4 | Light intensity with Light Guide. | The irradiance intensity of black light guides for Cybird WPT is about 20~30% lower than the Cybird Gold with clear light guides. | The irradiance intensity of Cybird WPT, the subject device, with a black light guide is about 20~30% lower than Cybird Gold with a clear light guide. |
| SE5 | Depth of cure | Many factors affect the degree of polymerization including the radiant intensity, exposure time and filler type | The depth of cure performances for the subject and predicate device are similar using Spident ESCOM – K110428. Other benefits of lower light intensity with longer exposure time during the resin polymerization are discussed. |
The battery charging method is an inductive wireless charger for the subject device and 4.2VDC Lithium Ion battery charger for the predicate device. The exposure time setting for the Plasma mode of the subject device with black guides is also increased to obtain to achieve equivalent polymerization of RBC in comparison with the predicate device with clear light guides.
Cybird WPT is equipped with black light guides, in spite of lower radiant intensity compared to Cybird Gold with a clear light guide, for the following reasons:
The use of black light guides can reduce high intensity light exposure to the eyes of patients and operators through filtering.
Moreover, relatively low radiant intensity over a longer exposure time may improve the uniform distribution and the degree of convergence of dental composites during the resin polymerization process.
These differences do not represent significant issues to clinical evaluation to fulfil the indication for use. In conclusion, the Cybird WPT LED Curing Light is substantially equivalent to the identified predicate device in terms of intended use, materials of construction, performance attributes and technological characteristics.
#### VII. SUMMARY OF NON-CLINICAL TESTS
{7}------------------------------------------------
Cybird LED Curing Light has been designed and tested according to the FDA guidance for Dental Curing Lights-Premarket Notification. Verification activities included curing hardness, depth of cure per ADA Specification No. 48.
Performance testing
Irradiance intensity test and spectral irradiance output were tested for both devices for comparison. The performance test results demonstrated that the Cybird WPT LED Curing light performed similar or better compared with the predicate device.
In addition, Cybird LED Curing Light complies with voluntary standards for biocompatibility (in vitro cytotoxicity, irritation and sensitization testing), electrical safety and EMC testing. The following data were provided to support the substantial equivalence determination:
## Biocompatibility:
Cybird WPT LED Curing Light is manufactured with identical materials as the 510(k) cleared light curing units, the Cybird LED curing Light (Model: Gold & XD, K173876). All materials to be connected to the patient's skin are in conformity with AAMI / ANSI / ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical Devices-- Part 5: Tests for In Vitro Cytotoxicity (L929 Assay) and AAMI / ANSI / ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
Electrical Safety and EMC essential performance testing was conducted in accordance with - AAMI ANSI ES 60601-1:2005/(R) 2012 and A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD). - IEC 60601-1-2 Ed.3: 2007-03 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
## Software:
Software verification and validation testing as recommended in IEC62304:2006 Medical device software - Software life cycle processes and FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005).
#### VIII. SUMMARY OF CLINICAL TESTS
Clinical testing was not required to demonstrate the substantial equivalence of Cybird WPT LED Curing Light to its predicate device.
#### IX. CONCLUSIONS
Validation and verification tests have been conducted to FDA guidance document Dental Curing Lights -Premarket Notification [510(k)]. Based on these comparison data, the sponsor believes that Cybird WPT LED Curing Light is substantially equivalent to Cybird Gold / XD LED Curing Light, the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.