K191216 · Hermes Medical Solutions AB · LLZ · Oct 17, 2019 · Radiology
Device Facts
Record ID
K191216
Device Name
Voxel Dosimetry™ v1.0
Applicant
Hermes Medical Solutions AB
Product Code
LLZ · Radiology
Decision Date
Oct 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of Voxel Dosimetry™ is to provide estimates (deterministic) of absorbed radiation dose at voxel level as a result of administering one of the supported radionuclides and to provide a dose map. This is dependent on input data regarding bio distribution being supplied to the application. Voxel Dosimetry™ only allows voxel-based dose calculations for patients who have been administered with radioisotopes. Warning! The Voxel Dosimetry™ is only intended for calculating dose for FDA approved radiopharmaceuticals for any clinical purpose, and calculation of unapproved drugs can only be used for research purpose.
Device Story
Voxel Dosimetry™ is a software tool for calculating voxel-level absorbed radiation dose following radiotracer administration. Inputs include SPECT/CT or PET/CT DICOM data from PACS or data manager. Workflow: registration of multi-timepoint studies to a common reference; generation of voxel-level time-activity curves; Semi-Monte Carlo (SMC) dose calculation for photons and electrons using patient-specific CT scans. Output: DICOM-formatted voxel-level absorbed dose map. Used in clinical settings by healthcare providers to perform dose-volume histograms and curve analysis. Enables patient-specific dosimetry for organs and lesions, improving upon organ-level average estimates. Benefits include accurate, localized dose assessment for internal radionuclide therapy planning.
Clinical Evidence
Validation included phantom testing and clinical comparison. Phantom study used XCAT phantoms for various isotopes (Ga68, I123, I131, In111, Lu177, Tc99m, Y90) comparing cumulated activities and doses against analytical models and PENELOPE Monte Carlo code. Clinical validation compared kidney doses in 6 patients (12 treatment cycles) undergoing Lu177-DOTATATE therapy against OLINDA/EXM® v2.0. Results showed high correlation (Pearson's r=0.97-0.98) and average relative difference of -2%.
Technological Characteristics
Software-based dosimetry tool; utilizes Semi-Monte Carlo (SMC) method for dose calculation; inputs patient-specific CT scans for photon/electron transport; outputs DICOM dose maps; supports isotopes including Lu-177, I-131, Ga-68, I-123, In-111, Tc-99m, Y-90; integrates with PACS/data managers; standalone software application.
Indications for Use
Indicated for patients administered with radioisotopes requiring voxel-based absorbed radiation dose calculation and dose mapping for clinical purposes using FDA-approved radiopharmaceuticals.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
October 17, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hermes Medical Solutions AB % Joakim Arwidson VP Quality and Regulatory Affairs Strandbergsgatan 16 112 51 Stockholm SWEDEN
Re: K191216
Trade/Device Name: Voxel Dosimetry™ v1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: September 4, 2019 Received: September 18, 2019
Dear Joakim Arwidson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191216
Device Name Voxel Dosimetry v1.0
The intended use of Voxel Dosimetry™ is to provide estimates (deterministic) of absorbed radiation dose at voxel as a result of administering one of the supported radionuclides and to provide a dose map. This is dependent on input data regarding bio distribution being supplied to the application.
Voxel Dosimetry™ only allows voxel-based dose calculations for patients who have been administered with radioisotopes.
Warning! The Voxel Dosimetry™ is only intended for calculating dose for FDA approved radiopharmaceuticals for any clinical purpose, and calculation of unapproved drugs can only be used for research purpose.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) SUMMARY
#### A. Submitted by:
- Submitters name and address: Hermes Medical Solutions AB Strandbergsgatan 16 112 51 Stockholm Sweden
#### . Submitters telephone number
| Phone: | +46 8 19 03 25 |
|---------|-----------------------------------|
| Cell: | +46 708 19 03 08 |
| E-mail: | ioakim.arwidson@hermesmedical.com |
#### ● Contact person
Joakim Arwidson VP Quality & Regulatory Hermes Medical Solutions AB Strandbergsgatan 16 112 51 Stockholm Sweden
- Registration number . 9710645
#### B. Preparation date:
2019-03-12
#### C. Proprietary/Trade name, Common name, Classification name:
- Proprietary/Trade name ● Voxel Dosimetry™ v1.0
- . Common name System, Image Processing, Radiological
- . Classification name Picture archiving and communications system, Class II, 21CFR892.2050.
#### D. Legally marketed device (predicate device):
Following legally marketed device has been used for comparison.
- MIM-MRT Dosimetry (K182624) ●
- . OLINDA/EXM® v2.0 (K163687)
#### E. Description of the device that is subject of this premarket notification:
Voxel Dosimetry™ is a tool for voxel level absorbed dose calculation resulting from radiotracer injection. Voxel Dosimetry™ workflow consists of the following steps:
- SPECT/CT or PET/CT DICOM data loading from the data manager GOLD or ● PACS
- . Registration of different time-points to a common reference study
- . Generation and integration of voxel-level time-activity curves
{5}------------------------------------------------
- . Voxel-level absorbed dose calculation using a Monte Carlo-method
- . Saving of the absorbed dose-map back to GOLD or PACS in DICOM format
#### F. Intended use:
The intended use of Voxel Dosimetry™ is to provide estimates (deterministic) of absorbed radiation dose at voxel level as a result of administering one of the supported radionuclides and to provide a dose map. This is dependent on input data regarding bio distribution being supplied to the application.
Voxel Dosimetry™ only allows voxel-based dose calculations for patients who have been administered with radioisotopes.
Warning! The Voxel Dosimetry™ is only intended for calculating dose for FDA approved radiopharmaceuticals for any clinical purpose, and calculation of unapproved drugs can only be used for research purpose.
### G. Technological characteristics:
MIM – MRT Dosimetry is based on voxel S-values (VSV), which is a voxel-based schema published in MIRD Pamphlet No. 17 (MIRD pamphlet No. 17: The dosimetry of nonuniform activity distributions-radionuclide S values at the voxel level. MIRD, J Nucl Med. 1999 Jan;40(1):115-36S), in which sources and targets are defined at the voxel level and voxel S-values calculated in a 3D matrix. The VSV approach is limited to lesions located in a homogeneous medium. As mentioned below, the Semi-Monte Carlo (SMC) method used in Voxel Dosimetry™ operates on a voxel level and performs dose calculation for photons and electrons based on patient specific CT scans. Voxel Dosimetry refers to the same patient population as MIM - MRT Dosimetry, support of the same isotopes (Lu-177, I-131) and has equivalent intended use.
The output from Voxel Dosimetry and MIM – MRT Dosimetry are dose maps containing voxel level doses to be used for dose volume histograms and curve analysis.
OLINDA/EXM® v2.0 is based on the use of S-factors, which are calculated on patient-like phantoms using a Monte Carlo method. The S-factors are equal to the average absorbed dose to a target organ generated by a unit of activity in a source organ. OLINDA/EXM® v2.0 dose calculations can thus be performed by multiplying the source organ time-activity curve integral by the S-factor. The SMC method used in Voxel Dosimetry™, on the other hand, operates on a voxel level and performs dose calculations for photons and electrons based on patient specific CT scans. Therefore, Voxel Dosimetry™ is patient-specific and produces voxel-level dose-maps instead of average organ-level dose estimates as OLINDA/EXM® v2.0 provides. Voxel Dosimetry ™ can also perform accurate lesion dosimetry because doses are calculated on a voxel-level and the same method can be used for lesions as for organs. This is not possible with OLINDA/EXM® v2.0, with which only a rough estimate of lesion doses is possible.
The output from Voxel Dosimetry v1.0 is a dose map containing voxel level doses in comparison to the predicate device OLINDA/EXM® v2.0, where the output is a CSV file including estimated dose for the organs.
{6}------------------------------------------------
#### H. Testing:
The tests for verification and validation followed Hermes Medical Solutions AB design-controlled procedures. The Risk analysis was completed, and risk control implemented to mitigate identified hazards.
Voxel Dosimetry and OLINDA/EXM® v2.0 Lu177-DOTATATE kidney doses were compared, because this is one of the most common situations in internal radionuclide dosimetry today. The average of the relative differences between SMC and OLINDA/EXM® v2.0 was found to be 2% (Hippeläinen ET, Tenhunen MJ, Mäenpää HO, Heikkonen JJ, Sohlberg AO. Dosimetry software Hermes Internal Radiation Dosimetry: from quantitative image reconstruction to voxel-level absorbed dose distribution. Nuclear Medicine Communications 2017;38:5:357-365.), which is less than the uncertainty in Lu177 kidney dosimetry (Gustafsson J, Brolin G, Cox M, Ljungberg M, Johansson L, Gleisner KS. Uncertainty propagation for SPECT/CT-based renal dosimetry in Lu-177 peptide receptor radionuclide therapy. Phys Med Biol. 2015;60(21):8329–8346.) and thus the difference will not affect safety or effectiveness.
The validation (see TAB 5 - CL92.01 P55V1.0 Clinical Validation in the original 510k application), was performed by generating an XCAT phantom for each isotope, with four time points dependent on the isotope. The XCAT phantom code also generated cumulated activities for each voxel using a mono-exponential model with the effective half-lives of the isotopes. The phantom data was processed in Voxel Dosimetry™ which generated cumulated activities for each voxel based on the trapezoidal integration. The Voxel Dosimetry cumulated activities were compared to the true cumulated activities calculated with the mono-exponential model and the mentioned effective half-lives. The true cumulated activity was obtained by analytically integrating the monoexponential model. The true cumulated activity phantom data was also used to compare the Voxel Dosimetry™ dose calculations to the dose calculations made with the Monte Carlo code PENELOPE.
| Isotope | Kidney | Tumor | Spleen |
|---------|--------|-------|--------|
| Ga68 | 6 | 6 | 7 |
| I123 | 3 | 1 | 2 |
| I131 | 7 | 2 | 3 |
| In111 | 11 | 7 | 7 |
| Lu177 | 7 | 3 | 3 |
| Tc99m | 8 | 7 | 6 |
| Y90 | 12 | 8 | 8 |
Difference (100% x (Voxel Dosimetry -true)/true) between Voxel Dosimetry™ cumulated activity and the true cumulated activity.
Difference (100% x (Voxel Dosimetry -PENELOPE) between Voxel Dosimetry™ dose and PENELOPE dose
| | Difference [%] | | |
|---------|----------------|-------|--------|
| Isotope | Kidney | Tumor | Spleen |
| I123 | 2 | 3 | 3 |
| I131 | 3 | 3 | 3 |
| Ga68 | 12 | 12 | 12 |
| In111 | 2 | 2 | 3 |
| Lu177 | 1 | 1 | 1 |
| Tc99m | 2 | 3 | 3 |
| Y90 | 5 | 6 | 4 |
{7}------------------------------------------------
The cumulated activities and doses obtained with Voxel Dosimetry™ match the reference values well. The differences in cumulated activities are due to differences in time-activity curve shapes. The piecewise linear model used in Voxel Dosimetry™ did not always perfectly match the simulated mono-exponential kinetics. Differences in doses were small. The only larger difference is seen with the Ga68 isotope, which has high positron energy.
In addition to phantom testing, Voxel Dosimetry™ was also validated against the predicate device OLINDA/EXM® v2.0. Twelve treatment cycles of six patients who underwent Lu177-DOTATE treatments were compared. The patients were scanned with a Siemens Symbia T2 SPECT/CTscanner 1, 24, 72 and 168 hours after the treatment. The SPECT/CT scans were reconstructed and left and right kidney doses obtained with Voxel Dosimetry™ and OLINDA/EXM® were compared. Voxel Dosimetry and OLINDA/EXM® v2.0 left and right kidney doses were found to be highly correlated (Pearson's rleft=0.97 and rleft=0.98). The average of the relative difference was -2% when compared to OLINDA/EXM® v2.0. These results are presented in more detail in the publication (Hippeläinen E, Tenhunen M, Mäenpää H, Heikkonen J, Sohlberg A. Dosimetry software Hermes Internal Radiation Dosimetry: from quantitative image reconstruction to voxellevel absorbed dose distribution. Nucl Med Commun 2017; 38:357-365).
The testing results support that all the software specifications have met the acceptance criteria.
#### I. Substantially Equivalent/Conclusions:
MIM – MRT Dosimetry is based on voxel S-values (VSV), which is a voxel-based schema published in MIRD Pamphlet No. 17 (MIRD pamphlet No. 17: The dosimetry of nonuniform activity distributions--radionuclide S values at the voxel level. MIRD, J Nucl Med. 1999 Jan;40(1):115-36S), in which sources and targets are defined at the voxel level and voxel S-values calculated in a 3D matrix. The VSV approach is limited to lesions located in a homogeneous medium. As mentioned below, the Semi-Monte Carlo (SMC) method used in Voxel Dosimetry™ operates on a voxel level and performs dose calculation for photons and electrons based on patient specific CT scans. Voxel Dosimetry refers to the same patient population as MIM – MRT Dosimetry, support of the same isotopes (Lu-177, I-131) and has equivalent intended use.
OLINDA/EXM® v2.0 is based on the use of S-factors, which are calculated on patient-like phantoms using a Monte Carlo method. The S-factors are equal to the average absorbed dose to a target organ generated by a unit of activity in a source organ. OLINDA/EXM® v2.0 dose calculations can thus be performed by multiplying the source organ time-activity curve integral by the S-factor. The SMC method used in Voxel Dosimetry™, on the other hand, operates on a voxel level and performs dose calculations for photons and electrons based on patient specific CT scans. Therefore, Voxel Dosimetry™ is patient-specific and produces voxel-level dose-maps instead of average organ-level dose estimates as OLINDA/EXM® v2.0 provides. Voxel Dosimetry ™ can also perform accurate lesion dosimetry because doses are calculated on a voxel-level and the same method can be used for lesions as for organs. This is not possible with OLINDA/EXM® v2.0, with which only a rough estimate of lesion doses is possible.
{8}------------------------------------------------
In summary, the Voxel Dosimetry™ v1.0, described in this submission has equivalent intended use and is, in our opinion, substantially equivalent to a combination of the predicate devices MIM-MRT Dosimetry (K182624) and OLINDA/EXM® v2.0 (K163687) and supports its clinical effectiveness, safety and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.