K191194 · Carl Zeiss Meditec, Inc. · QER · Jun 25, 2019 · Ophthalmic
Device Facts
Record ID
K191194
Device Name
CLARUS
Applicant
Carl Zeiss Meditec, Inc.
Product Code
QER · Ophthalmic
Decision Date
Jun 25, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K191194 · Jun 25, 2019
CLARUS
Carl Zeiss Meditec, Inc.
Clinical study of patients with vitreoretinal and choroidal diseases
The clinical study was used to demonstrate the ability of the CLARUS 700 to capture clinically useful Fluorescein Angiography images in a range of retinal and choroidal diseases, supporting the new imaging modality.
Non-significant risk clinical study; Clinical evaluation of imaging utility
Patients with a range of vitreoretinal and choroidal diseases; Sample Size: 20 eyes from 13 subjects
Optos California Fa (K142897)
Visibility of area/lesion of interest, clinical utility of image, and absence of interfering artifacts
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Optic Nerve Head detection
Deep learning algorithm
—
—
—
—
—
—
Fluorescein Angiography image capture
—
—
85% (17/20) for area and lesion visibility; 95% (19/20) for clinically useful image; 95% (19/20) for artifact interference
—
—
Clinical study: 20 eyes from 13 subjects
—
Indications for Use
The CLARUS 700 ophthalmic camera is indicated to capture, display, annotate and store images to aid in the diagnosis and monitoring of diseases and disorders occurring in the retina, ocular surface and visible adnexa. It provides true color and autofluorescence imaging modes for stereo, widefield, and montage fields of view. The CLARUS 700 angiography is indicated as an aid in the visualization of vascular structures of the retina and the choroid.
Device Story
CLARUS 700 is a slit-scanning ophthalmic camera (Broad Line Fundus Imaging) for in-vivo eye imaging. Input: light reflected from retina/ocular surface; captured via 12-megapixel monochrome sensor. Operation: line of illumination scans retina; red/green/blue LEDs provide true color; green/blue LEDs provide autofluorescence; green LED supports fluorescein angiography. Used in clinics by ophthalmologists, optometrists, and technicians. Output: high-definition widefield (90°) and ultra-widefield (135°) images displayed on external monitor. Software includes deep learning for optic nerve head (ONH) detection to assist image montaging. Clinicians review images, use annotation/caliper tools for analysis, and compare exams to monitor disease progression. Benefits: non-mydriatic/mydriatic imaging, improved patient comfort via stable head/chin rest, and widefield visualization of retinal/choroidal vascular structures.
Clinical Evidence
Clinical study of 20 eyes from 13 subjects (mean age 62) evaluated FA imaging utility. Primary endpoints: visibility of lesion (85% pass), clinical utility/consistency with disease/transit phase (95% pass), and lack of artifact interference (95% pass). Results demonstrate ability to capture clinically useful FA images of retina and choroid across various vitreoretinal/choroidal diseases.
Technological Characteristics
Slit-scanning ophthalmic camera (BLFI). Materials: ABS/PC alloy (Bayer Bay Blend #FR3010) for patient contact. Illumination: Red (585-640nm), Green (500-585nm), Blue (435-500nm) LEDs; 785nm NIR laser for alignment. Sensors: 12MP monochrome CMOS. Connectivity: DICOM, FORUM/EMR. Software: Version 1.1, Windows 10 IoT. Standards: ISO 10940:2009, ISO 10993-1:2009, IEC 60601-1, IEC 60601-1-2, ANSI Z80.36-2016.
Indications for Use
Indicated for patients requiring retinal, ocular surface, and visible adnexa imaging to aid in diagnosis and monitoring of ocular diseases and disorders. Includes visualization of retinal and choroidal vascular structures via angiography.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 25, 2019
Carl Zeiss Meditec, Inc Saurabh Jamkhindikar Senior Regulatory Affairs Specialist 5160 Hacienda Drive Dublin, CA 94568
Re: K191194
Trade/Device Name: Clarus Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: QER, Dated: May 1, 2019 Received: May 3, 2019
Dear Saurabh Jamkhindikar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Cunningham Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191194
Device Name CLARUS Model 700
#### Indications for Use (Describe)
The CLARUS 700 ophthalmic camera is indicated to capture, display, annotate and store images to aid in the diagnosis and monitoring of diseases and disorders occurring in the retina, ocular surface and visible adnexa. It provides true color and autofluorescence imaging modes for stereo, widefield, and montage fields of view. The CLARUS 700 angiography is indicated as an aid in the visualization of vascular structures of the retina and the choroid.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92), the 510(k) Summary for the CLARUS 700 with Software Version 1.1 is provided below.
# GENERAL INFORMATION
| Manufacturer: | Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, California 94568<br>(925) 557-4100 (phone)<br>(925) 557-4259 (fax)<br>Est. Reg. No. 2918630 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Saurabh Jamkhindikar<br>Senior Regulatory Affairs Specialist<br>Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, California 94568<br>saurabh.jamkhindikar@zeiss.com<br>(925) 557-4696 (phone)<br>(925) 557-4259 (fax) |
| Date Summary Prepared: | June 21, 2019 |
| Classification Name: | Camera, Ophthalmic, Slit-scanning |
| Regulation Description: | Ophthalmic Camera (acc. 21 CFR 886.1120) |
| Classification: | Class II (acc. 21 CFR 886.1120) |
| Product Code: | QER |
| Common Name: | Ophthalmic Camera |
| Trade/Proprietary Name: | CLARUS |
| Model(s): | 700 |
# PREDICATE DEVICE
| Company: | Carl Zeiss Meditec, Inc. |
|----------|--------------------------|
| Device: | CLARUS 500 (K181444) |
{5}------------------------------------------------
### REFERENCE DEVICE
| Company: | OPTOS PLC. |
|----------|-----------------------------------------|
| Device: | P200DTx - Optos California Fa (K142897) |
### INTENDED USE / INDICATIONS FOR USE
The CLARUS 700 ophthalmic camera is indicated to capture, display, annotate and store images to aid in the diagnosis and monitoring of diseases and disorders occurring in the retina, ocular surface and visible adnexa. It provides true color and autofluorescence imaging modes for stereo, widefield, ultra-widefield, and montage fields of view.
The CLARUS 700 angiography is indicated as an aid in the visualization of vascular structures of the retina and the choroid.
### DEVICE DESCRIPTION
The CLARUS™ model 700 is a new addition to the CLARUS product family consisting of existing model 500 (K181444). The CLARUS 700 is an active, software controlled, highresolution ophthalmic imaging device for In-vivo imaging of the human eye. Imaging modes include True color, Fundus Auto-fluorescence with green excitation. Fundus Auto-fluorescence with blue excitation. Fluorescein Angiography, Stereo and External eve. All true color images can be separated into red, green and blue channel images to help enhance visual contrast of details in certain layers of the retina. The CLARUS 700 angiography imaging aids in the visualization of the vascular structures of the retina and the choroid. With a single capture, CLARUS 700 produces a 90° high definition widefield image. Widefield images are automatically merged to achieve a 135° ultra-widefield of view. The CLARUS 700 makes use of a deep learning algorithm for Optic Nerve Head (ONH) detection. The ultra-widefield montage on CLARUS 700 is no longer dependent just on the patient accurately fixating their gaze on the internal fixation. With the ONH detection, the software will find the optic nerve and determine based on the image(s) captured where the patient was gazing at the point of capture. The CLARUS 700 device allows clinicians to easily review and compare high-quality images captured during a single exam while providing annotation and caliper measurement tools that allow in-depth analysis of eye health. CLARUS 700 is designed to optimize each patient's experience by providing a simple head and chin rest that allows the patient to maintain a stable, neutral position while the operator brings the optics to the patient. facilitating a more comfortable imaging experience. The ability to swivel the device between the right and left eye helps technicians capture an image without realigning the patient. Live IR Preview allows the technician to confirm image quality and screen for lid and lash obstructions, prior to imaging, ensuring fewer image recaptures.
The CLARUS 700 device's principle of operation is Slit Scanning Ophthalmic Camera also referred to as Broad Line Fundus Imaging (BLFI), same as the predicate CLARUS 500 (K181444). During image capture, a line of illumination passes through the slit and scans across the retina. A 2D monochromatic camera captures the returned light to image the retina. A single sweep of the
{6}------------------------------------------------
illumination is used to illuminate the retina for image capture. Repeated sweeps of near infrared light are used for a live retina view for alignment. Red, green and blue LEDs sequentially illuminate to generate true color images. Blue and green LED illumination enables Fundus Autofluorescence (FAF) imaging. Fluorescein Angiography images are captured with green LED illumination at a wavelength that stimulates fluorescence of the injected sodium fluorescein dye.
The CLARUS 700 system is mainly comprised of an acquisition device, all-in-one PC, keyboard, mouse, instrument lift table and external power supply.
The CLARUS 700 hardware is based off the predicate CLARUS 500 (K181444) hardware. New FA imaging mode on the CLARUS 700 require the below hardware changes:
- Added filters to support FA imaging mode .
- Updated slim turret and motor with new positions for reliability, angiography filters and ● FPGA code
- Updated calibration tool for new turret positions and differentiation
- Change to lightbox board for reliability and support higher duty cycle in support of FA imaging
- Updated Onyx All-in-one Computer for 32GB RAM and 2TB HDD storage space
- Updated belt driven slit for reliability and to support FA imaging mode ●
- . Updated camera to support FA imaging mode
The CLARUS software provides the user the capability to align, capture, review and annotate images. The software has two installation configurations: Software installed on the Instrument (Acquisition & Review) as well as Software installed on a separate 'Review Station' (Laptop or Computer) (only Review).
The CLARUS software version 1.1 is based off the predicate CLARUS software version 1.0 (K181444). Added image capture modality includes Fluorescein Angiography. Other changes implemented in the software version 1.1 include:
- Automated Optic Nerve Head (ONH) detection for montaging ●
- Smart (Region of Interest) Focus ●
- Auto brightness for FA image series
- Calibration software update for DEVICE hardware changes ●
- . FORUM/ Other EMR connectivity updates for new FA imaging mode
The CLARUS 700 device meets the requirements of ISO 10940:2009 standard. The device technical specifications are identical to the predicate device. The performance specifications relevant to the user are summarized in the Table 1 below.
{7}------------------------------------------------
| Feature | Specification |
|----------------------------------|------------------------------------------------------------------------------------|
| FoV – Widefield (single capture) | • 90° |
| FoV - Ultra-widefield (montage) | • 135° |
| Image Resolution | • 60 lp/mm at central field (0°)<br>• 40 lp/mm at 23° FOV<br>• 25 lp/mm at 45° FOV |
| Sensors | • 12-megapixel monochrome |
| Sensor Resolution | • 3000 x 3000 pixels |
| Focusing Range | • +20 D to -24D |
| Pixel Pitch on the Fundus | • 7.3 µm/pixel |
Table 1 - Specifications
# RISK MANAGEMENT AND GENERAL SAFETY AND EFFECTIVENESS
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements and validation of the clinical workflow to ensure that the product meets its intended uses. To minimize electrical, mechanical and radiation hazards, ZEISS adheres to recognized and established industry practice and relevant international standards.
# BIOCOMPATIBILITY
The CLARUS 700 has two patient-contact components, i.e. the patient chin rest and the forehead rest, which are surface contacting and have transient contact. The device uses the same patient contacting materials as used in the predicate device. The materials have been evaluated for Biocompatibility and comply with requirements of ISO 10993-1:2009 standard.
# PERFORMANCE DATA & SUMMARY OF VERIFICATION AND VALIDATION ACTIVITY
The following performance testing for the CLARUS 700 is provided to support the substantial equivalence of the subject device:
### Design Verification Testing
The design verification testing results demonstrate that the system complies with the established system requirements.
{8}------------------------------------------------
### Design Validation Testing
The purpose of the design validation testing is to measure customer acceptance of the intended use, features and the workflow. The validation results demonstrate that the device meets the requirements set out by the Product Requirements Specifications and all aspects of the user experience met the acceptance criteria.
### Testing to Consensus Standards
The device was tested (as needed) to meet the requirements for conformity (where applicable) to multiple industry standards. The R&D evaluation of the relevant testing to consensus sis documented.
#### Electrical Safety Testing
The electrical safety for the device has been evaluated and was found to be in compliance with the ANSI AAMI 60601-1:2005/(R) 2012 and A1:2012 (Ed 3.1) standard.
#### Electromagnetic Compatibility Testing
The electromagnetic compatibility (EMC) for the device has been evaluated and was found to be in compliance with the IEC 60601-1-2:2014 Ed 4.0 standard.
#### Optical Safety
The same optical pathway is used in both the CLARUS 700 device and the predicate. The device includes additional filters in the optical pathway for Fluorescein Angiography imaging modality.
The optical safety of the device for Fluorescein Angiography imaging mode with recognized consensus standard ANSI Z80.36-2016 and ISO 15004-2:2007 has been demonstrated by performing hazard analysis assessment. The device is determined to be a Group 1 instrument.
There are no changes to the design or use of the near Infrared laser for CLARUS 700 from the predicate device. The classification and safety of the device's laser components in accordance with the recognized consensus standard IEC 60825-1:2007 has been established. The device is determined to be a Class 1 laser system.
#### Environmental Conditions
There are no significant hardware or requirement changes for the CLARUS 700 device from the predicate device requiring re-testing for environmental conditions. The device complies with the requirements for environmental conditions for use, storage and transport as specified in the ISO 15004-1:2009 standard.
#### Software Verification:
Software verification testing for the device was conducted and documentation is provided as recommended by the FDA's Guidance for Industry and FDA staff, "Guidance for the Content of premarket Submissions for Software Contained in Medical Devices."
{9}------------------------------------------------
### DICOM Conformity Assessment
The device complies with the NEMA PS 3.1-3.20 (2016) standard and a DICOM Conformance Statement is provided.
# CLINICAL TESTING
A non-significant risk clinical study was completed with the CLARUS 700 to support indications for use statement for the product. The objective of this study was to demonstrate the ability of the CLARUS 700 device in capturing fluorescein angiography (FA) images of the retina and choroid in a range of vitreoretinal and choroidal diseases. The study also included comparison data obtained on a reference device Optos California Fa (K142897) during the same angiogram procedure in some cases.
As part of this study, 20 eyes from 13 subjects (11 male, 2 female) were enrolled and imaged. The average age of the recruited subjects was 62 years old. Overall, Fluorescein Angiography images acquired on the CLARUS 700 device showed good clinical utility. The passing rate was as below.
| | Criteria# Criteria Description | Passing Rate |
|---|------------------------------------------------------------------------------------------------------|--------------|
| | Area and lesion of interest is visible on the angiogram | 17/20 (85%) |
| | Clinically useful image. Image appearance is consistent with<br>the disease and transit phase of dye | 19/20 (95%) |
| ಗ | Artifacts, if any, do not interfere with ability to interpret image | 19/20 (95%) |
The clinical study results demonstrate the ability of CLARUS 700 to capture clinically useful Fluorescein Angiography images of the retina and choroid in a range of vitreoretinal and choroidal diseases. In addition, the wide field of view allows more retinal area to be imaged at once, making it possible to capture the entire posterior pole during the early phase of dye transit, and reducing the number of peripheral sweeps needed during the later phases to cover the retina. This may be useful in documenting diseases that affect large parts of the retina, or diseases that affect the retina outside the posterior pole, such as diabetic retinopathy, retinal vein and artery occlusions.
# SUBSTANTIAL EQUIVALENCE DISCUSSION
The predicate device is Carl Zeiss Meditec, Inc's CLARUS model 500 (K181444). The intended use of both devices is the same. They are both used to capture, display, store, annotate and review images of the human retina and the surrounding parts of the eye under mydriatic and nonmydriatic conditions. Both devices support the diagnosis and monitoring of eye diseases.
The differences in the Indications for Use statement for the devices do not affect their intended use, anatomical site of application or target population.
The technological characteristics of both the devices are identical. Both devices use the Slit
{10}------------------------------------------------
Scanning Ophthalmoscope technique (also referred to as Broad Line Fundus Imaging (BLFI)). Both devices have the same basic functions for capturing, displaying, storing and reviewing images of the human retina and surrounding parts of the eye. Both devices provide a wide field of view (FoV) of 90° with a single capture. Both devices have an identical patient-operator interface and controls for operating the device. With both devices, an external monitor is used to view images. For both devices, the software provides image review functionality.
Both devices use the same light sources to illuminate the retinal area and digital image sensors for recording images. Both devices provide retinal imaging modes under non-mydriatic and mydriatic conditions.
The CLARUS 700 provides Fluorescein Angiography imaging mode not available in the predicate. Fluorescein Angiography fundus imaging is a well-established imaging modality and has been widely used by eye care professionals in documenting various ocular diseases and conditions. Design Verification testing was completed to demonstrate that the specified Fluorescein Angiography imaging mode is provided and confirmed. A clinical study was completed to demonstrate the ability of the CLARUS 700 device to capture clinically useful Fluorescein Angiography images of the retina and choroid in a range of vitreoretinal and choroidal diseases. The study also includes comparison data obtained on a reference device during the same angiogram procedure for some of the cases.
A comparison of the subject device to the predicate device is provided in Table 2.
Table 2: Comparison Table of Proposed Device CLARUS 700 with SW Version 1.1 and Predicate CLARUS 500 with SW Version 1.0 (K181444)
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) – Proposed Device | CLARUS Model 500 with SW version 1.0<br>(K181444) - Predicate Device |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, CA 94568, USA | Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, CA 94568, USA |
| 510(k) | • K191194 | • K181444 |
| Proprietary Name | • CLARUS model 700 | • CLARUS model 500 |
| Device<br>Classification<br>Name | • Camera, Ophthalmic, Slit-Scanning | • Camera, Ophthalmic, Slit-Scanning |
| Generic/ Common<br>Name | • Ophthalmic camera | • Ophthalmic camera |
| Product Code | • QER | • QER |
| Regulation<br>Number | • 886.1120 | • 886.1120 |
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) - Proposed Device | CLARUS Model 500 with SW version 1.0<br>(K181444) - Predicate Device |
| Class | II<br>● | II<br>● |
| Review Panel | Ophthalmic<br>● | Ophthalmic<br>● |
| Intended Use/<br>Indications for<br>Use | The CLARUS 700 ophthalmic camera is<br>indicated to capture, display, annotate and<br>store images to aid in the diagnosis and<br>monitoring of diseases and disorders<br>occurring in the retina, ocular surface and<br>visible adnexa. It provides true color and<br>autofluorescence imaging modes for<br>stereo, widefield, ultra-widefield, and<br>montage fields of view.<br>The CLARUS 700 angiography is<br>indicated as an aid in the visualization of<br>vascular structures of the retina and the<br>choroid. | The CLARUS 500 ophthalmic camera is<br>indicated to capture, display, annotate and<br>store images to aid in the diagnosis and<br>monitoring of diseases and disorders<br>occurring in the retina, ocular surface and<br>visible adnexa. It provides true color and<br>autofluorescence imaging modes for stereo,<br>widefield, ultra-widefield, and montage<br>fields of view. |
| Target Population | Opticians<br>●<br>Ophthalmic Photographers<br>●<br>Optometrists<br>●<br>Ophthalmologists<br>●<br>Medical Assistants<br>●<br>Clinical Researchers<br>● | Opticians<br>●<br>Ophthalmic Photographers<br>●<br>Optometrists<br>●<br>Ophthalmologists<br>●<br>Medical Assistants<br>●<br>Clinical Researchers<br>● |
| Anatomical Site | Retina, ocular surface and visible<br>●<br>adnexa | Retina, ocular surface and visible<br>●<br>adnexa |
| Device Type | Fundus Camera<br>● | Fundus Camera<br>● |
| Methodology | Slit Scanning Ophthalmoscope, also<br>●<br>referred to as Broad Line Fundus<br>Imaging (BLFI) | Slit Scanning Ophthalmoscope, also<br>●<br>referred to as Broad Line Fundus<br>Imaging (BLFI) |
| Principle of<br>Fundus Image<br>Capturing | Broad Line Fundus Imaging (BLFI)<br>●<br>using LEDs and laser illumination<br>Near-IR laser illumination for live<br>●<br>retina preview<br>LEDs (red, green, blue) used for true<br>●<br>color, fundus autofluorescence and<br>fluorescein angiography imaging<br>modes<br>Live preview and image-capture with<br>●<br>12-megapixel monochrome sensors | Broad Line Fundus Imaging (BLFI)<br>●<br>using LEDs and laser illumination<br>Near-IR laser illumination for live<br>●<br>retina preview<br>LEDs (red, green, blue) used for true<br>●<br>color and fundus autofluorescence<br>imaging modes<br>Live preview and image-capture with<br>●<br>12-megapixel monochrome sensors |
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) - Proposed Device | CLARUS Model 500 with SW version 1.0<br>(K181444) - Predicate Device |
| Image Capture<br>Modes | True color (with red, green and blue<br>●<br>channel separation in review mode)<br>Fundus autofluorescence with green<br>●<br>excitation<br>Fundus autofluorescence with blue<br>●<br>excitation<br>Fluorescein Angiography<br>●<br>Stereo<br>●<br>External<br>● | True color (with red, green and blue<br>●<br>channel separation in review mode)<br>Fundus autofluorescence with green<br>●<br>excitation<br>Fundus autofluorescence with blue<br>●<br>excitation<br>Stereo<br>●<br>External<br>● |
| Illumination<br>Source for Image<br>Acquisition | Red LED: 585-640 nm<br>●<br>Green LED: 500-585 nm<br>●<br>Blue LED: 435-500 nm<br>●<br>Near-infrared laser diode: 785 nm<br>● | Red LED: 585-640 nm<br>●<br>Green LED: 500-585 nm<br>●<br>Blue LED: 435-500 nm<br>●<br>Near-infrared laser diode: 785 nm<br>● |
| Laser Class<br>(Based on IEC<br>60825-1:2007) | Class 1<br>● | Class 1<br>● |
| Device Group<br>(Optical safety) | Group 1 (ANSI Z80.36-2016)<br>● | Group 1 (ANSI Z80.36-2016)<br>● |
| Illumination<br>Source for Live<br>Preview | Near-infrared laser diode: 785 nm<br>● | Near-infrared laser diode: 785 nm<br>● |
| Detector Type | Internal CMOS sensors - 12<br>●<br>megapixels | Internal CMOS sensors - 12 megapixels<br>● |
| Filters | Barrier and Excitation Filters for<br>●<br>Fundus Autofluorescence Blue<br>Barrier and Excitation Filters for<br>●<br>Fundus Autofluorescence Green<br>Barrier and Excitation Filters for<br>●<br>Fluorescein Angiography | Barrier and Excitation Filters for<br>●<br>Fundus Autofluorescence Blue<br>Barrier and Excitation Filters for<br>Fundus Autofluorescence Green |
| Field of View | 90° (widefield - single shot image)<br>●<br>135° (ultra-widefield - two shot auto<br>●<br>montage) | 90° (widefield - single shot image)<br>●<br>135° (ultra-widefield - two shot auto<br>●<br>montage) |
| Refractive Error<br>Compensation | +24 D -20 D, continuous<br>● | +24 D -20 D, continuous<br>● |
| Minimum pupil<br>size | 2.5 mm<br>● | 2.5 mm<br>● |
| Working Distance | 25 mm (patient's eye - front lens)<br>● | 25 mm (patient's eye - front lens)<br>● |
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) – Proposed Device | CLARUS Model 500 with SW version 1.0<br>(K181444) – Predicate Device |
| Fixation Target | • Internal and External | • Internal and External |
| Configuration | • Acquisition unit consists of the acquisition head (includes image acquisition optics), cross-table and the patient support<br>• Peripheral components include the All-in-One PC, keyboard, mouse and the instrument lift table | • Acquisition unit consists of the acquisition head (includes image acquisition optics), cross-table and the patient support<br>• Peripheral components include the All-in-One PC, keyboard, mouse and the instrument lift table |
| System Control<br>Input Devices | • Joystick for operating image acquisition head<br>• Keyboard/Touchpad/Mouse for operating internal computer | • Joystick for operating image acquisition head<br>• Keyboard/Touchpad/Mouse for operating internal computer |
| Patient Contacting<br>Materials | • Chin rest and forehead rest | • Chin rest and forehead rest |
| Materials – Patient<br>Contact | • Chin rest and forehead rest - ABS/PC Alloy, Bayer Bay Blend #FR3010, Color – Medium Gray RAL-U713 | • Chin rest and forehead rest - ABS/PC Alloy, Bayer Bay Blend #FR3010, Color – Medium Gray RAL-U713 |
| Operating System | • Microsoft Windows 10 IoT Enterprise Version 64 bit | • Microsoft Windows 10 IoT Enterprise Version 64 bit |
| Display Types | • 22" Full HD MVA LCD with LED Backlight (external) | • 22" Full HD MVA LCD with LED Backlight (external) |
| Data Storage | • 2 TB (At-Instrument Computer) | • 1 TB (At-Instrument Computer) |
| Export | • Via FORUM/PACS, DICOM, network drive and USB thumb drives | • Via FORUM, DICOM, network drive and USB thumb drives |
| Software Features | • Review and Analysis of Fundus Images | • Review and Analysis of Fundus Images |
| Electrical<br>Requirements | • Input Voltage: 100-240 V~, 50/60 Hz, 4A max<br>• Power consumption: Max. 1315 VA for DEVICE and lift table configuration (standalone DEVICE not available) | • Input Voltage: 100-240 V~, 50/60 Hz, 4A max<br>• Power consumption: Max. 1315 VA for DEVICE and lift table configuration (standalone DEVICE not available) |
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) – Proposed Device | CLARUS Model 500 with SW version 1.0<br>(K181444) - Predicate Device |
| Electrical Safety<br>parameters | Protection Class: 1 Device Type (IEC 60601-1): B Enclosure Protection: IPX0 (No degree of protection against ingress of water or particulate matter) | Protection Class: 1 Device Type (IEC 60601-1): B Enclosure Protection: IPX0 (No degree of protection against ingress of water or particulate matter) |
| Environmental<br>Conditions:<br>Operation | Temperature: +10° to +35° C Relative Humidity: 30% to 90% (excluding condensation) Atmospheric Pressure: 800 hPa to 1060 hPa | Temperature: +10° to +35° C Relative Humidity: 30% to 90% (excluding condensation) Atmospheric Pressure: 800 hPa to 1060 hPa |
| Environmental<br>Conditions:<br>Storage | Temp. -10° to +55° C Relative Humidity 10% to 95% (excluding condensation) Atmospheric Pressure 700 hPa to 1060 hPa | Temp. -10° to +55° C Relative Humidity 10% to 95% (excluding condensation) Atmospheric Pressure 700 hPa to 1060 hPa |
| Environmental<br>Conditions:<br>Transport | Temp. -40° to +70° C Relative Humidity 10% to 95% (excluding condensation) Atmospheric Pressure 500 hPa to 1060 hPa | Temp. -40° to +70° C Relative Humidity 10% to 95% (excluding condensation) Atmospheric Pressure 500 hPa to 1060 hPa |
{11}------------------------------------------------
{12}------------------------------------------------
# K191194 – 510(k) Summary
{13}------------------------------------------------
{14}------------------------------------------------
A comparison of the subject device to the reference device Optos California fa is provided in Table 3. The reference device Optos California fa is used to claim equivalence for Fluorescein Angiography imaging mode in CLARUS 700.
# Table 3: Comparison Table of Proposed Device CLARUS 700 with SW Version 1.0 and Reference device Optos California fa (K142897)
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) – Proposed Device | Optos California P200DTx (K142897)<br>- Reference Device |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, CA 94568, USA | OPTOS PLC.<br>QUEENSFERRY HOUSE, CARNEGIE<br>BUSINESS CAMPUS<br>Dunfermline, GB Ky11 8gr |
| 510(k) | • K191194 | • K142897 |
| Proprietary Name | • CLARUS model 700 | • Optos California fa |
| Device<br>Characteristics | CLARUS Model 700 with SW version<br>1.1 (K191194) – Proposed Device | Optos California P200DTx (K142897) - Reference Device |
| Device<br>Classification<br>Name | Camera, Ophthalmic, Slit-Scanning | Ophthalmoscope, Laser, Scanning |
| Generic/ Common<br>Name | Ophthalmic camera | Ophthalmoscope |
| Product Code | QER | MYC |
| Regulation<br>Number | 886.1120…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.