← Product Code [NQW](/productcode/NQW) · K191169

# NuVasive® Camber Laminoplasty System (K191169)

_Nu Vasive, Incorporated · NQW · Jul 3, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191169

## Device Facts

- **Applicant:** Nu Vasive, Incorporated
- **Product Code:** [NQW](/productcode/NQW.md)
- **Decision Date:** Jul 3, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The NuVasive® Camber Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Caminoplasty System is used to hold the allograft or autograft material in place in order to prevent the allograft or autograft material from expulsion, or impinging the spinal cord.

## Device Story

The NuVasive® Camber Laminoplasty System consists of titanium alloy plates and screws used to reinforce the spinal canal during laminoplasty procedures. The device is implanted by a surgeon to secure allograft or autograft material, preventing graft expulsion or spinal cord impingement. It is used in the lower cervical and upper thoracic spine (C3 to T3). The system provides mechanical stabilization of the posterior elements after canal expansion. Clinical benefit is derived from maintaining graft position and spinal canal patency.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence was demonstrated through non-clinical bench testing, including axial pullout (ASTM F2193) and static/dynamic 3-point bend testing.

## Technological Characteristics

Constructed from titanium alloy (ASTM F136). System comprises plates and screws of various sizes. Mechanical fixation device; no energy source or software components. Sterilization method not specified.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- NuVasive® Laminoplasty Fixation System & LeVerage LFS System (K1091623)
- Medronic Centerpiece Plate Fixation System ([K050082](/device/K050082.md))

## Reference Devices

- NuVasive® VuePoint® II OCT System ([K180198](/device/K180198.md))
- NuVasive Affix II Spinous Process Plate System ([K132411](/device/K132411.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 3, 2019

NuVasive, Incorporated Thao Huynh Specialist. Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121

Re: K191169

Trade/Device Name: NuVasive® Camber Laminoplasty System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: NQW Dated: April 30, 2019 Received: May 1, 2019

Dear Thao Huynh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K191169

Device Name NuVasive® Camber Laminoplasty System

### Indications for Use (Describe)

The NuVasive® Camber Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Caminoplasty System is used to hold the allograft or autograft material in place in order to prevent the allograft or autograft material from expulsion, or impinging the spinal cord.

| Type of Use (Select one or both, as applicable)  |                                             |
|--------------------------------------------------|---------------------------------------------|
| ×   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NuVasive. The logo consists of a stylized leaf shape in purple and gray on the left, followed by the word "NUVASIVE" in gray. The logo is simple and modern, and the colors are muted.

# 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations. Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

#### Submitted By A.

Thao Huynh Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 320-5256

Date Prepared: July 2, 2019

#### B. Device Name

| Trade Name:          | NuVasive® Camber Laminoplasty System        |
|----------------------|---------------------------------------------|
| Common Name:         | Spinal Interlaminal Fixation Orthosis       |
| Classification Name: | Orthosis, Spine, Plate, Laminoplasty, Metal |
| Regulation Number:   | 21 CFR § 888.3050                           |
| Product Code:        | NQW                                         |

#### C. Predicate Devices

The subject NuVasive® Camber Laminoplasty System is substantially equivalent to the primary predicate device, NuVasive® Laminoplasty Fixation System & LeVerage LFS System (K1091623), and additional predicate Medronic Centerpiece Plate Fixation System (K050082). NuVasive® VuePoint® II OCT System (K180198) and NuVasive Affix II Spinous Process Plate System (K132411) are also used as reference predicates.

#### D. Device Description

The NuVasive® Camber Laminoplasty System consists of plates and screws of various sizes made from titanium alloy (ASTM F136) to provide reinforcement while expanding the spinal canal and preserving the posterior elements. Instruments required to implant the device are also available.

#### E. Indications for Use

The NuVasive® Camber Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Camber Laminoplasty System is used to hold the allograft or autograft material in place in order to prevent the allograft or autograft material from expulsion, or impinging the spinal cord.

#### F. Technological Characteristics

As was established in this submission, the subject Camber Laminoplasty System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape in purple and gray on the left, followed by the word "NUVASIVE" in gray. The leaf shape is partially purple and partially gray, with the purple portion being on top. The word "NUVASIVE" is in a sans-serif font and is all uppercase.

The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate devices through comparison in areas including design, labeling/intended use, material composition, and function.

#### G. Performance Data

Non-clinical testing was performed to demonstrate that the subject Camber Laminoplasty System is substantially equivalent to other predicate devices. Engineering rationale and the following testing was performed:

- . Axial Pullout (per ASTM F2193)
- Static 3-Point Bend
- Dynamic 3-Point Bend

The results demonstrate that the subject Camber Laminoplasty System meets the same criteria as the predicate devices, and the subject device was therefore found to be substantially equivalent to the predicate. No clinical studies were conducted.

#### H. Conclusions

The subject NuVasive® Camber Laminoplasty System has been shown to be substantially equivalent to legally marketed predicate devices for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K191169](https://fda.innolitics.com/device/K191169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
