K191166 · Dignitana, Inc. · PMC · Jun 26, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191166
Device Name
DigniCap Delta
Applicant
Dignitana, Inc.
Product Code
PMC · General, Plastic Surgery
Decision Date
Jun 26, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DigniCap Delta® Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.
Device Story
DigniCap Delta is a scalp cooling system designed to reduce chemotherapy-induced alopecia. The device cools fluid to a set temperature and circulates it through a cooling cap worn by the patient. It features a solid-state thermoelectric cooling unit with an integrated control system operated via a touchscreen monitor. The system monitors fluid temperature via sensors in the fluid lines to ensure consistent cooling. It is intended for use in a clinical setting. By maintaining scalp hypothermia during chemotherapy, the device aims to reduce hair loss. Healthcare providers use the touchscreen interface to manage treatment parameters. The device is a single-cap system, an evolution from the predicate's dual-cap design.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), accessory performance, stable fluid temperature verification, alarm system testing, electrical safety (IEC 60601-1:2005(R)2012, IEC 62133-2:2017), electromagnetic compatibility (IEC 60601-1-2:2014), and software verification and validation.
Technological Characteristics
Solid-state thermoelectric cooling unit; touchscreen interface; single-cap configuration; cooling fluid (isopropyl alcohol/water mixture); temperature sensors in fluid lines; cooling range -7 to 1.5°C. Biocompatible materials per ISO 10993-1. Electrical safety per IEC 60601-1:2005(R)2012 and IEC 62133-2:2017. EMC per IEC 60601-1-2:2014.
Indications for Use
Indicated for cancer patients with solid tumors undergoing chemotherapy to reduce the likelihood of chemotherapy-induced alopecia.
Regulatory Classification
Identification
A scalp cooling system to reduce the likelihood of chemotherapy-induced alopecia is a prescription device intended to reduce the frequency and severity of alopecia during chemotherapy in which alopecia-inducing chemotherapeutic agents are used.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use. This information must include testing to demonstrate accuracy of the temperature control mechanism.
(2) Performance testing must demonstrate the electromagnetic compatibility and electrical safety of the device.
(3) Software verification, validation, and hazard analysis must be performed.
(4) The patient contacting components of the device must be demonstrated to be biocompatible. Material names must be provided.
(5) Labeling must include the following:
(i) A statement describing the potential risk of developing scalp metastasis.
(ii) Information on the patient population and chemotherapeutic agents/regimen for which the device has been demonstrated to be effective.
(iii) A summary of the non-clinical and/or clinical testing pertinent to use of the device.
(iv) A summary of the device technical parameters, including temperature cooling range and duration of cooling.
(v) A summary of the device- and procedure-related adverse events pertinent to use of the device.
(vi) Information on how the device operates and the typical course of treatment.
(6) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, and adverse effects/complications.
(ii) Information on how the device operates and the typical course of treatment.
(iii) Information on the patient population for which there is clinical evidence of effectiveness.
(iv) The potential risks and benefits associated with use of the device.
(v) Postoperative care instructions.
(vi) A statement describing the potential risk of developing scalp metastasis.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Dignitana Inc. % Valerie Defiesta-Ng Vice President, Regulatory Affairs Experien Group 224 Airport Parkway, Suite 250 San Jose, California 95110
June 26, 2019
#### Re: K191166
Trade/Device Name: DigniCap Delta Regulation Number: 21 CFR 878.4360 Regulation Name: Scalp Cooling System to Reduce the Likelihood of Chemotherapy-Induced Alopecia Regulatory Class: Class II Product Code: PMC Dated: April 30, 2019 Received: May 1, 2019
Dear Valerie Defiesta-Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden, MS Assistant Director THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191166
Device Name DigniCap Delta® Scalp Cooling System
Indications for Use (Describe)
The DigniCap Delta® Scalp Cooling System is indicated toreduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, induding suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Notification K191166
#### GENERAL INFORMATION
#### Applicant:
Dignitana, Inc. 10925 Estate Lane, Suite W-185 Dallas, Texas 75238 Phone: 469-917-5555 info@dignitana.com
#### Contact Person:
Valerie Defiesta-Ng Vice President, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA
Date Prepared: April 30, 2019
DEVICE INFORMATION
Trade/Proprietary Name: DigniCap Delta®
### Generic/Common Name:
Scalp Cooling System
# Classification:
21 CFR §878.4360
#### Product Code: PMC
Establishment Registration: 3012146457- Dignitana, Inc. Dallas, Texas 2019, Complaint File Establishment
# Manufacturing Facility: Dignitana, Inc.
10925 Estate Lane, Suite W-185 Dallas, Texas 75238
{5}------------------------------------------------
#### PREDICATE DEVICE(S)
DigniCap Delta is substantially equivalent to the predicate device, the Dignitana DigniCap Scalp Cooling System, cleared under K170871. DigniCap Delta is intended for use in the same population, with the same mechanism of action and Indications for Use. Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness.
#### DEVICE DESCRIPTION
The DigniCap Delta Scalp Cooling System ("DigniCap Delta")is a device that is intended to function as a cooling system to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The proposed therapy of the new device is comparable to the therapy of the predicate device, DigniCap Scalp Cooling System (K170871).
DigniCap Delta cools fluid to a prescribed set temperature and circulates that cooled fluid through a cooling wrap and then back to the device. This mode of operation is equivalent to other scalp cooling devices. User operation of the DigniCap Delta system is carried out on the illuminated graphics display with the integrated touch control, located on the front of the unit. The navigation and selection buttons on the display are used to interface with the device.
#### INDICATIONS FOR USE
The DigniCap Delta® Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
DigniCap Delta is substantially equivalent to the predicate device, the Dignitana DigniCap® Scalp Cooling System, cleared under K170871. Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. DigniCap Delta has the same Indications for Use, population and mechanism of action as the predicate device. Thus, DigniCap Delta is substantially equivalent to the predicate device.
The tables below compare the regulatory information (Table 5.1) and technical characteristics (Table 5.2) for the proposed and predicate devices.
{6}------------------------------------------------
# 510(k) SUMMARY
| | DigniCap Delta® Scalp Cooling System<br>(Subject Device) | DigniCap® Scalp Cooling System<br>(Predicate Device) | Analysis of<br>Differences |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | | | |
| 510(k)<br>Number | To Be Assigned | K170871 | N/A |
| Classification | II | II | Same |
| Product<br>Code | PMC | PMC | Same |
| Regulatory<br>Number | \$878.4360 | \$878.4360 | Same |
| Regulation<br>Name | Scalp Cooling System | Scalp Cooling System | Same |
| Indications<br>for Use | The DigniCap Delta® Scalp Cooling System is indicated to<br>reduce the likelihood of chemotherapy-induced alopecia in<br>cancer patients with solid tumors. | The DigniCap® Scalp Cooling System is indicated to<br>reduce the likelihood of chemotherapy-induced alopecia in<br>cancer patients with solid tumors. | Same |
| Mechanism<br>of Action | The DigniCap Delta Scalp Cooling System transports<br>temperature controlled cooled fluid from the device to a<br>cooling cap to cool the patient's scalp thus reducing<br>chemotherapy-induced alopecia | The DigniCap Scalp Cooling System transports<br>temperature controlled cooled fluid from the device to a<br>cooling cap to cool the patient's scalp thus reducing<br>chemotherapy-induced alopecia. | Same |
| Population | Chemotherapy Patients with Solid Tumors | Chemotherapy Patients with Solid Tumors | Same |
| | DigniCap Delta® Scalp Cooling System<br>(Subject Device) | DigniCap® Scalp Cooling System<br>(Predicate Device) | Analysis of Differences |
| Feature | | | |
| Technological<br>Features | The DigniCap Delta Scalp Cooling System<br>consists of a solid-state thermoelectric<br>cooling unit with integral control system<br>operated via a touchscreen monitor, capable<br>of precisely controlling the temperature of<br>one cooling cap. | The DigniCap Scalp Cooling System<br>consists of a refrigerator unit with integral<br>control system operated via a touch screen<br>monitor and capable of controlling two<br>separate cooling caps. | Any differences in the technological<br>characteristics between the devices do not<br>raise different questions of safety or<br>effectiveness. |
| User Interface | Touchscreen | Touchscreen | Same<br>Updated for user ease. |
| | Cooling Cap Features | | |
| Inner Cooling<br>Cap | Yes, includes adjustable tabs | Yes | The differences in the technological<br>characteristics between the devices are<br>minor and do not raise different questions<br>of safety or effectiveness. |
| Detachable<br>Coolant Lines | Yes | Yes | Same |
| Outer Cover/<br>Material | Yes/Neoprene | Yes/Neoprene | Same |
| Cooling Fluid<br>Required | Yes | Yes | Cooling fluid is similar; the DigniCap Delta<br>uses an isopropyl alcohol/water mixture,<br>while the predicate uses MPG<br>(monopropylene glycol). The cooling fluids<br>are used for the same purpose. This<br>difference does not raise different questions<br>of safety or effectiveness as demonstrated<br>by performance testing. |
| Fluid<br>Temperature | Fluid temperatures are regulated through<br>temperature sensors in the fluid lines. | Fluid temperatures are regulated through<br>temperature sensors in the cooling cap. | The sensors are located in a different region<br>but are used for the same purpose and have<br>been tested to prove that they work as<br>effectively as the predicate device. This<br>difference does not raise different questions<br>of safety or effectiveness as demonstrated<br>by performance testing. |
| Feature | DigniCap Delta® Scalp Cooling System<br>(Subject Device) | DigniCap® Scalp Cooling System<br>(Predicate Device) | Analysis of Differences |
| Pre /Post<br>Cooling Phase | Yes | Yes | Same |
| Coolant<br>Temperature<br>Range | -7 to 1.5°C | -15°C to 5°C | This range is within the range originally<br>cleared with the predicate device but is<br>tighter to provide more consistent cooling<br>to the patient's scalp. This difference does<br>not raise different questions of safety or<br>effectiveness. |
| Outer Cooling<br>Cap | Single Patient Use | Multiple Patient Use | This difference does not raise different<br>questions of safety or effectiveness. |
| Inner Cooling<br>Cap - Single<br>Use | Single Patient Use | Multiple Patient Use | This difference does not raise different<br>questions of safety or effectiveness. |
| Device Testing | | | |
| Biocompatibility<br>Testing | Yes. Cap materials are biocompatible in<br>accordance with ISO 10993-1. | Yes. Cap materials are biocompatible in<br>accordance with ISO 10993-1. | Same |
| Electrical Safety<br>Testing | Yes.<br>IEC 60601-1:2005/(R)2012<br>IEC 62133-2:2017 | Yes.<br>IEC 60601-1:2007 | Electrical Safety Testing was conducted to<br>the current standard. |
| Electromagnetic<br>Compatibility<br>Testing | Yes.<br>IEC60601-1-2:2014 | Yes.<br>IEC 60601-1-2:2007 | Electrical Safety Testing was conducted to<br>the current standard. |
# Table 5.1: Substantial Equivalence Table – Regulatory Information
{7}------------------------------------------------
# 510(k) SUMMARY
# Table 5.2: Substantial Equivalence Table – Technological Information
{8}------------------------------------------------
# 510(k) SUMMARY
# Table 5.2: Substantial Equivalence Table – Technological Information (cont.)
{9}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE
DigniCap Delta has the same Indications for Use, patient population and mechanism of action as the predicate device (K170871). Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. Thus, DigniCap Delta is substantially equivalent to the predicate device.
#### PERFORMANCE DATA
All necessary non-clinical performance testing was conducted on DigniCap Delta to support a determination of substantial equivalence to the predicate device.
#### Nonclinical Testing Summary:
DigniCap Delta was tested to confirm the device met all performance requirements and is considered safe and effective for its intended use. Testing was completed per relevant FDA guidance documents. Specifically, the device was evaluated for the following through nonclinical, bench testing:
- . Biocompatibility Testing of Cooling Cap (Cap materials are biocompatible in accordance with ISO 10993-1)
- Accessory Performance Testing ●
- Stable Fluid Temperature Verification
- Notifications and Alarms System Testing
- Electrical Safety in accordance with IEC 60601-1:2005(R)2012 and IEC 62133-2:2017 ● and Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014
- Software verification and validation testing
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of DigniCap Delta meet the established specifications necessary for consistent performance during its intended use. The bench testing has demonstrated that the DigniCap Delta can maintain and adjust cooling fluid temperature as needed to meet the predetermined treatment specifications. The collective bench testing demonstrates that DigniCap Delta does not raise new questions of safety or effectiveness for scalp cooling when compared to the predicate devices.
#### Clinical Testing Summary:
Not applicable. Clinical testing was not performed to support this 510(k) submission.
#### CONCLUSIONS
The collective results of the performance testing demonstrate that DigniCap Delta meets the established specifications necessary for consistent performance during its intended use. In addition, the collective performance testing demonstrate that DigniCap Delta does not raise different questions of safety or effectiveness when compared to the predicate device.
Based on the non-clinical performance testing results, the differences in the technological characteristics between the subject and predicate devices do not raise any new issues of safety or
{10}------------------------------------------------
effectiveness. DigniCap Delta is intended for use in the same population and with the same Indications for Use as its predicate device. Thus, DigniCap Delta is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.