Cellvizio 100 Series System with Confocal Miniprobes
K191144 · Mauna Kea Technologies · OWN · Jan 24, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K191144
Device Name
Cellvizio 100 Series System with Confocal Miniprobes
Applicant
Mauna Kea Technologies
Product Code
OWN · Gastroenterology, Urology
Decision Date
Jan 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K191144 · Jan 24, 2020
Cellvizio 100 Series System with Confocal Miniprobes
Mauna Kea Technologies
Independent clinical findings from clinical researchers; International independent Health Technology Assessment organizations
The sponsor utilized RWE and independent clinical findings to support the extension of the device's Indications for Use to include the visualization of blood flow using Fluorescein Sodium as a contrast agent.
Blood flow visualization; Fluorescein Sodium; Clinical literature; Health Technology Assessment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical demonstration based on literature review; Literature review of independent clinical findings
Patients undergoing procedures requiring visualization of blood flow and tissue microstructure
Not applicable for this study
Visualization of blood flow and tissue microstructures using Fluorescein Sodium
Indications for Use
Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells, vessels and their organization or architecture. The Cellvizio® 100 Series System F400-v2 is indicated for imaging blood flow in vascular areas, including microvasculature and capillaries. The GastroFlex™ UHD and ColoFlex™ UHD Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., gastrointestinal systems, accessed by an endoscope or endoscopic accessories. The AlveoFlex™ Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e. respiratory systems, accessed by an endoscope or endoscopic accessories. The CholangioFlex™ series of Confocal Miniprobe™ is intended to allow imaging of the upper gastrointestinal tract including biliary and pancreatic ducts, accessed by an endoscope or endoscopic accessories. The AQ-Flex™ 19 Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA needles). The CystoFlex™ (F, UHD-R) and UroFlex™ B of Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories. The CelioFlex™ UHD 5 of Confocal Miniprobe™ is intended to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures. The CranioFlex™ Confocal Miniprobe™ is indicated to provide visualization within central nervous system during cranial diagnostic and therapeutic procedures such as tumor biopsy and resection.
Device Story
Cellvizio 100 Series is a confocal laser endomicroscopy system using fiber optic probes to provide real-time, high-magnification in-vivo imaging of tissue microstructure. The system utilizes a 488 nm blue laser to excite tissue autofluorescence or Fluorescein Sodium contrast agent. The fiber scanner probe is placed in direct contact with target tissue; the system reconstructs the reflected/emitted light into images displayed on a monitor. Used in OR or clinical settings by physicians (endoscopists, surgeons, neurosurgeons) during diagnostic and therapeutic procedures (e.g., biopsy, resection). Output provides visual information on cellular organization and blood flow, assisting clinicians in real-time decision-making during procedures. Benefits include enhanced visualization of microvasculature and tissue architecture, potentially improving diagnostic accuracy and surgical guidance.
Clinical Evidence
No new clinical trials conducted. Substantial equivalence supported by literature review, Real World Evidence (RWE), and independent clinical findings from researchers and Health Technology Assessment organizations. Previous bench testing from cited predicates (K111047, K122042, K123676, K133466, K150831, K151593, K160416, K171345, K172844, K180270, K183640) remains applicable.
Technological Characteristics
Confocal laser scanning system; 488 nm continuous blue laser source (Class 2M). Fiber optic probe sensing/actuation. Real-time imaging (9-12 Hz). Connectivity via DICOM PACS, shared drive, and USB. Software controls laser emission, scanning, image reconstruction, and display. No changes to materials or design from predicates.
Indications for Use
Indicated for imaging internal tissue microstructure, cells, and vessels in gastrointestinal, respiratory, urinary, and central nervous system tracts, as well as body cavities and organs during endoscopic, laparoscopic, or robot-assisted procedures. Also indicated for imaging blood flow in microvasculature and capillaries using Fluorescein Sodium contrast agent.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Cellvizio® 100 Series Confocal laser imaging system with Confocal Miniprobe™ (K172844, K180270, K183640)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Mauna Kea Technologies % Michael Daniel President Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460
January 24, 2020
Re: K191144
Trade/Device Name: Cellvizio 100 Series System with Confocal Miniprobes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OWN, GCJ, GWG Dated: December 19, 2019 Received: December 20, 2019
Dear Michael Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT5: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191144
Device Name
Cellvizio® 100 Series system with Confocal Miniprobes™
### Indications for Use (Describe)
Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells, vessels and their organization or architecture. The Cellvizio® 100 Series System F400-v2 is indicated for imaging blood flow in vascular areas, including microvasculature and capillaries.
The GastroFlex™ UHD and ColoFlex™ UHD Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., gastrointestinal systems, accessed by an endoscope or endoscopic accessories.
The AlveoFlex™ Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e. respiratory systems, accessed by an endoscope or endoscopic accessories.
The CholangioFlex™ series of Confocal Miniprobe™ is intended to allow imaging of the upper gastrointestinal tract including biliary and pancreatic ducts, accessed by an endoscope or endoscopic accessories.
The AQ-Flex™ 19 Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA needles).
The CystoFlex™ (F, UHD-R) and UroFlex™ B of Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories.
The CelioFlex™ UHD 5 of Confocal Miniprobe™ is intended to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures.
The CranioFlex™ Confocal Miniprobe™ is indicated to provide visualization within central nervous system during cranial diagnostic and therapeutic procedures such as tumor biopsy and resection.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <div><span style="font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K191144
# 2 Premarket Notification 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## 510(k) Number: K191144
## Applicant Information:
Date Prepared: January 24, 2020
| Manufacturer Contact Person: | Aline Criton |
|------------------------------|-------------------------------------------------------------|
| Name: | Mauna Kea Technologies |
| Address: | 9 rue d'Enghien<br>F-75010 Paris, France |
| Phone: | +33 1 48 24 03 45 |
| Fax: | +33 1 48 24 12 18 |
| Contact Person: | Michael A Daniel, Consultant<br>madaniel@clinregconsult.com |
| Phone Number: | (415) 407-0223 |
| Office: | (775) 392-2970 |
| Facsimile Number: | (610) 545-0799 |
## Device Information:
| Device Trade Name: | Cellvizio® 100 Series Confocal laser imaging<br>systems with Confocal Miniprobes™ |
|-------------------------------------------|-----------------------------------------------------------------------------------|
| Common Name: | Endoscope and Accessories Neurological Endoscope |
| Classification Name(s): | Confocal Optical Imaging |
| Primary regulation and Product Code: | OWN / GCJ 21 CFR 876.1500 |
| Secondary regulation and Product<br>Code: | GWG 21 CFR 882.1480 |
| Classification: | Class II |
# Classification: Predicate Devices:
- Previously cleared versions of Cellvizio® 100 Series Confocal laser imaging system with ● Confocal Miniprobe™ (K172844, K180270 and K183640).
## Reference Devices:
- . The Zeiss CONVIVO (K181116).
{5}------------------------------------------------
## Indications for Use:
Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells, vessels and their organization or architecture. The Cellvizio® 100 Series System F400-v2 is indicated for imaging blood flow in vascular areas, including microvasculature and capillaries.
The GastroFlex™ UHD and ColoFlex™ UHD Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., gastrointestinal systems, accessed by an endoscope or endoscopic accessories.
The AlveoFlex™ Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., respiratory systems, accessed by an endoscope or endoscopic accessories.
The CholangioFlex™ series of Confocal Miniprobe™ is intended to allow imaging of the upper gastrointestinal tract including biliary and pancreatic ducts, accessed by an endoscope or endoscopic accessories.
The AQ-Flex™ 19 Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA needles).
The CystoFlex™ (F, UHD-R) and UroFlex™ B of Confocal Miniprobe™ are intended to allow imaging of anatomical tracts, i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories.
The CelioFlex™ UHD 5 of Confocal Miniprobe™ is intended to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures.
The CranioFlex™ Confocal Miniprobe™ is indicated to provide visualization within central nervous system during cranial diagnostic and therapeutic procedures such as tumor biopsy and resection.
## Device Description:
Confocal Miniprobes™ are used with Cellvizio® 100 Series System (F400-v2), which is a confocal imaging system with fiber optic probes which allows visualization of internal microstructure of tissues and blood flow including, but not limited to, the identification of cells, vessels and their organization or architecture, during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures and during neurosurgical procedures.
Fluorescein Sodium is used as a fluorescence contrast agent to allow imaging of microvasculature and visualization of blood flow. Fluorescein Sodium can be used as a contrast agent with Cellvizio® 100 Series system with Confocal Miniprobes™ without change of formulation, mode of action, approved dose or route of administration; it is delivered independent of Cellvizio® 100 Series system in accordance with Fluorescein Sodium instruction for use.
{6}------------------------------------------------
Materials, design and intended use of the aforementioned Cellvizio® 100 Series Confocal laser imaging systems and their Confocal Miniprobes™ remain exactly the same as what were previously cleared in K172844, K180270 and K183640.
## Comparison to Predicate Device:
The table below details the difference in Indications for Use between the subject device and the previously cleared predicate device (See Table 7-1).
{7}------------------------------------------------
| Aspect | Subject Device<br>Cellvizio® 100 Series System F400-V2<br>with Confocal Miniprobes™ with<br>Fluorescein | Predicate Device<br>Cellvizio® 100 Series System F400-V2<br>with Confocal Miniprobes™ | Comparison<br>to predicate |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Cellvizio® 100 Series System with<br>Confocal Miniprobes™ | Cellvizio® 100 Series System with<br>Confocal Miniprobes™ | Unchanged |
| Manufacturer | Mauna Kea Technologies<br>9 Rue d'Enghien<br>F-75010 Paris, France | Mauna Kea Technologies<br>9 Rue d'Enghien<br>F-75010 Paris, France | Unchanged |
| Model | Cellvizio® 100 Series F400-V2 with<br>Confocal Miniprobes™ | Cellvizio® 100 Series F400-V2 with<br>Confocal Miniprobes™ | Unchanged |
| 510 (k)# | Not Assigned | K172844, K180270, and K183640 | / |
| Regulation<br>Number | 21 CFR 876.1500<br>21 CFR 882.1480 | 21 CFR 876.1500<br>21 CFR 882.1480 | Unchanged |
| Class | II | II | Unchanged |
| Classification<br>Advisory<br>Committee | General & Plastic Surgery and Neurological<br>Devices Panel | General & Plastic Surgery and Neurological<br>Devices Panel | Unchanged |
| Device Class /<br>Name | Confocal Optical Imaging | Confocal Optical Imaging | Unchanged |
| Product Code | OWN/GCJ/GWG | OWN/GCJ/GWG | Unchanged |
| Combination<br>Device | Yes | No | Extension of<br>the<br>Indications<br>for Use with<br>Fluorescein<br>Sodium dye |
| Indication For<br>Use | The Cellvizio® 100 Series system with<br>Confocal Miniprobes™ is a confocal laser<br>system with fiber optic probes that is<br>intended to allow imaging of the internal<br>microstructure of tissues including, but not<br>limited to, the identification of cells, vessels<br>and their organization or architecture. The<br>Cellvizio® 100 series system F400-V2 is<br>indicated for imaging blood flow in vascular<br>areas, including microvasculature and<br>capillaries.<br>The CranioFlex™ Confocal Miniprobe™ is<br>indicated to provide visualization within<br>central nervous system during cranial<br>diagnostic and therapeutic procedures such<br>as tumor biopsy and resection.<br>The GastroFlex™ UHD and ColoFlex™<br>UHD Confocal Miniprobes™ are intended<br>to allow imaging of anatomical tracts, i.e.,<br>gastrointestinal systems, accessed by an<br>endoscope or endoscopic accessories.<br>The AlveoFlex™ Confocal Miniprobe™ is<br>intended to allow imaging of anatomical<br>tracts, i.e., respiratory systems, accessed by<br>an endoscope or endoscopic accessories.<br>The CholangioFlex™ (or GastroFlex™ M)<br>series of Confocal Miniprobes™ is intended<br>to allow imaging of the upper<br>gastrointestinal tract including biliary and<br>pancreatic ducts, accessed by an endoscope<br>or endoscopic accessories.<br>The AQ-Flex™ 19 Confocal Miniprobe™ is<br>intended to allow imaging of anatomical<br>tracts, i.e., gastrointestinal tracts and<br>respiratory tracts accessed by an endoscope<br>or endoscopic accessories, including<br>through endoscopic needles.<br>The CystoFlex™ (F, UHD R) and<br>UroFlex™ B Confocal Miniprobes™ are<br>intended to allow imaging of anatomical<br>tracts, i.e., urinary, including, but not limited<br>to, urethra, bladder, and ureter, accessed<br>through an endoscope or endoscopic<br>accessories.<br>The CelioFlex™ UHD 5 Confocal<br>Miniprobe™ is intended to provide<br>visualization of body cavities, organs, and<br>canals during endoscopic and laparoscopic<br>surgical procedures, including robot-assisted<br>procedures. | Cellvizio® 100 Series System with<br>Confocal Miniprobes™ is a confocal laser<br>system with fiber optic probes that are<br>intended to allow imaging of the internal<br>microstructure of tissues, including, but not<br>limited to, the identification of cells and<br>vessels and their organization or<br>architecture.<br>The CranioFlex™ Confocal Miniprobe™ is<br>indicated to provide visualization within<br>central nervous system during cranial<br>diagnostic and therapeutic procedures such<br>as tumor biopsy and resection.<br>The GastroFlex™ UHD and ColoFlex™<br>UHD Confocal Miniprobes™ are intended<br>to allow imaging of anatomical tracts, i.e.,<br>gastrointestinal systems, accessed by an<br>endoscope or endoscopic accessories.<br>The AlveoFlex™ Confocal Miniprobe™ is<br>intended to allow imaging of anatomical<br>tracts, i.e., respiratory systems, accessed by<br>an endoscope or endoscopic accessories.<br>The CholangioFlex™ (or GastroFlex™ M)<br>series of Confocal Miniprobes™ is intended<br>to allow imaging of the upper<br>gastrointestinal tract including biliary and<br>pancreatic ducts, accessed by an endoscope<br>or endoscopic accessories.<br>The AQ-Flex™ 19 Confocal Miniprobe™ is<br>intended to allow imaging of anatomical<br>tracts, i.e., gastrointestinal tracts and<br>respiratory tracts accessed by an endoscope<br>or endoscopic accessories, including<br>through endoscopic needles.<br>The CystoFlex™ (F, UHD R) and<br>UroFlex™ B Confocal Miniprobe™ are<br>intended to allow imaging of anatomical<br>tracts, i.e., urinary, including, but not limited<br>to, urethra, bladder, and ureter, accessed<br>through an endoscope or endoscopic<br>accessories.<br>The CelioFlex™ UHD 5 Confocal<br>Miniprobe™ is intended to provide<br>visualization of body cavities, organs, and<br>canals during endoscopic and laparoscopic<br>surgical procedures, including robot-assisted<br>procedures. | There is an<br>extension of<br>the Indication<br>For Use of the<br>Cellvizio®<br>100 series<br>system F400-V2 to<br>visualize<br>blood flow<br>with<br>Fluorescein<br>Sodium<br>contrast<br>agent.<br>The<br>indications<br>for use of the<br>different<br>Confocal<br>Miniprobes<br>™ do not<br>change. |
{8}------------------------------------------------
Table 7-1: Comparison of Indication for Use between the Subject Device and Previously Cleared Predicate Device
The table below further demonstrates the substantial equivalence of the subject device to the predicate device.
{9}------------------------------------------------
| Aspect | Comparison of Predicate Device to Cellvizio® 100 Series system with Confocal<br>Miniprobes™, with extended Indications for Use with Fluorescein Sodium contrast<br>agent |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technical equivalence | No change of design and performances<br>Same technical design<br>Same operating principle<br>Same safety profile and risk level<br>Identical technological characteristics<br>Same reprocessing methods<br>Same device packaging |
| Biological equivalence | Same biocompatible patient-contacting materials and same biocompatibility test<br>requirements<br>Same target tissues to image |
| Application and usage<br>equivalence | Same intended use of the device<br>Same environment of use and same intended user<br>Same instructions for use and operating procedure<br>Same precautions and safety instructions |
Table 7-2: Comparison of Predicate Device to Cellvizio® 100 Series system with Confocal Miniprobes™, with new Indications for Use.
# Comparison to Reference Device:
The table below provides the comparison between the subject device and the reference device (Zeiss CONVIVO Confocal Endomicroscopy system K181116).
| Attribute | Subject Device<br>Cellvizio® 100 Series System F400-V2<br>with Confocal Miniprobes™ with<br>Fluorescein | Reference Device<br>CONVIVO K181116 | Comparison to<br>Reference |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Device name | Cellvizio® 100 Series System with<br>Confocal Miniprobes™ | CONVIVO | / |
| Manufacturer | Mauna Kea Technologies<br>9 Rue d'Enghien<br>F-75010 Paris, France | Carl Zeiss Meditec AG Goeschwitzer<br>Strasse 51-52 D-07745 Jena, Germany | / |
| 510(k) | Not Assigned | K181116 | / |
| Classification<br>Product Code | GWG (primary product code) OWN<br>(secondary product code) | GWG (primary product code) OWN<br>(secondary product code) | Same |
| Regulation<br>No. | 21 CFR 882.1480<br>21 CFR 876.1500 | 21 CFR 882.1480<br>21 CFR 876.1500 | Same |
| Class | II | II | Same |
| Classification<br>Adv.<br>Committee | General & Plastic Surgery and<br>Neurological Devices Panel | General & Plastic Surgery and<br>Neurological Devices Panel | Same |
| Device Class /<br>Name | Confocal Optical Imaging | Confocal Optical Imaging | Same |
| Product Code | OWN/GCJ/GWG | OWN/GCJ/GWG | Same |
| Combination<br>Device | Yes | Yes | Both devices<br>can be used<br>with<br>Fluorescein<br>Sodium as a<br>contrast<br>agent to<br>image blood<br>flow. |
| Indications for<br>use | The Cellvizio® 100 Series system with<br>Confocal Miniprobes™ is a confocal laser<br>system with fiber optic probes that is | The ZEISS CONVIVO is a surgical<br>endomicroscope intended for viewing intra-<br>operative blood flow in the cerebral | Similar<br>Indication for<br>Use for blood |
{10}------------------------------------------------
| | intended to allow imaging of the internal<br>microstructure of tissues including, but not<br>limited to, the identification of cells,<br>vessels and their organization or<br>architecture. Cellvizio® 100 series F400-<br>V2 is indicated for imaging blood flow in<br>vascular areas, including microvasculature<br>and capillaries.<br><br>The CranioFlex™ Confocal Miniprobe™ is<br>indicated to provide visualization within<br>central nervous system during cranial<br>diagnostic and therapeutic procedures such<br>as tumor biopsy and resection.<br>The GastroFlex™ UHD and ColoFlex™<br>UHD Confocal Miniprobes™ are intended<br>to allow imaging of anatomical tracts, i.e.,<br>gastrointestinal systems, accessed by an<br>endoscope or endoscopic accessories.<br>The AlveoFlex™ Confocal Miniprobe™ is<br>intended to allow imaging of anatomical<br>tracts, i.e., respiratory systems, accessed by<br>an endoscope or endoscopic accessories.<br>The CholangioFlex™ (or GastroFlex™ M)<br>series of Confocal Miniprobes™ is<br>intended to allow imaging of the upper<br>gastrointestinal tract including biliary and<br>pancreatic ducts, accessed by an endoscope<br>or endoscopic accessories.<br>The AQ-Flex™ 19 Confocal Miniprobe™<br>is intended to allow imaging of anatomical<br>tracts, i.e., gastrointestinal tracts and<br>respiratory tracts accessed by an endoscope<br>or endoscopic accessories, including<br>through endoscopic needles.<br>The CystoFlex™ (F, UHD R) and<br>UroFlex™ B Confocal Min…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.