K191103 · Parkell, Inc. · EJK · Nov 22, 2019 · Dental
Device Facts
Record ID
K191103
Device Name
Parkell Desensitizer Gel
Applicant
Parkell, Inc.
Product Code
EJK · Dental
Decision Date
Nov 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Parkell® Desensitizer Gel is indicated for reduction of tooth hypersensitivity by the following treatments: - treatment of dentin exposed by toothbrush abrasion, periodontal disease, and/or acid erosion; - treatment of dentin after mechanical tooth cleaning, scaling, and/or root planning; - treatment of tooth hypersensitivity associated with bleaching: and - treatment of prepared dentin for fillings and/or prosthetic restorations.
Device Story
Parkell Desensitizer Gel treats dental hypersensitivity; applied to dentin via standard applicator brush. Device releases calcium and phosphate ions; stimulates hydroxyapatite formation on dentin surface and within dentinal tubules. Hardens post-application; forms mineral hydroxyapatite layer and plugs to occlude tubules. Used in dental clinic setting by dental professionals. Output is physical occlusion of tubules; reduces patient sensitivity to thermal or mechanical stimuli. Packaged in single-chamber syringe with snap-in tips.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993 series) and bioactivity analysis (ISO 23317:2014). Microshear bond strength testing demonstrated performance ≥15MPa, consistent with predicate.
Technological Characteristics
Composition: water, calcium phosphate-based fillers. Form: gel. Delivery: single-chamber syringe with applicator tips. Principle: chemical-induced hydroxyapatite formation and tubule occlusion. Biocompatibility: ISO 10993 compliant. Bioactivity: ISO 23317 compliant.
Indications for Use
Indicated for reduction of tooth hypersensitivity in patients with dentin exposure due to toothbrush abrasion, periodontal disease, acid erosion, or post-mechanical cleaning/scaling/root planing; hypersensitivity associated with bleaching; and prepared dentin for restorations.
Regulatory Classification
Identification
A calcium hydroxide cavity liner is a device material intended to be applied to the interior of a prepared cavity before insertion of restorative material, such as amalgam, to protect the pulp of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2019
Parkell, Inc. David Mott Vice President - Legal, R&D, and Regulatory 300 Executive Drive Edgewood, New York 11717
Re: K191103
Trade/Device Name: Parkell Desensitizer Gel Regulation Number: 21 CFR 872.3250 Regulation Name: Calcium Hydroxide Cavity Liner Regulatory Class: Class II Product Code: EJK, LBH Dated: October 24, 2019 Received: October 25, 2019
Dear David Mott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191103
Device Name Parkell® Desensitizer Gel
Indications for Use (Describe)
Parkell® Desensitizer Gel is indicated for reduction of tooth hypersensitivity by the following treatments:
- · treatment of dentin exposed by toothbrush abrasion, periodontal disease, and/or acid erosion;
- · treatment of dentin after mechanical tooth cleaning, scaling, and/or root planning;
- · treatment of tooth hypersensitivity associated with bleaching: and
- · treatment of prepared dentin for fillings and/or prosthetic restorations.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | <span> <svg height="15" width="15"> <path d="M2 2 L13 13 M13 2 L2 13" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <svg height="15" width="15"> <rect fill="white" height="15" stroke="black" stroke-width="1" width="15"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Parkell Inc. The logo is in black and white and features the word "parkell" in a bold, sans-serif font. To the right of "parkell" and slightly below is the word "inc." in a smaller, italicized font. The overall design is simple and professional.
# 510(k) Summary – K191103
#### 1. Submitter
Parkell, Inc. 300 Executive Drive Edgewood, NY 11717 Phone: 631-389-1545 Fax: 631-389-1546 dmott@parkell.com
Contact Person: David Mott, Vice President - Legal & Regulatory Date of Initial Submission: April 24, 2019 Date of Submission of Final Revision: October 24, 2019
#### 2. Device
| Device Proprietary Name: | Parkell® desensitizer |
|--------------------------|--------------------------------|
| Common or Usual Name: | Tooth desensitizer |
| Classification Name: | Calcium hydroxide cavity liner |
| Regulation Number: | 21 CFR 872.3250 |
| Product Code: | EJK |
| Device Classification | II |
#### 3. Primary Predicate
Teethmate Desensitizer, K131068, Kuraray Noritake Dental Inc.
#### 4. Device Description
Parkell® Desensitizer Gel alleviates dental hypersensitivity at treatment sites. It is a gel which requires no mixing prior to use and which is applied to treatment sites using a standard applicator brush. Parkell® Desensitizer releases calcium and phosphate ions at the treatment site and stimulates hydroxyapatite formation on the surface of treated dentin and in dentinal tubules. In this manner, the Device hardens following application and forms a layer of mineral hydroxyapatite as well as hydroxyapatite plugs which occlude dentinal tubules, which results in the treatment of dental hypersensitivity.
Parkell® Desensitizer Gel is packaged in any standard single-chamber syringe. The syringe is accompanied by standard snap-in tips and standard applicator brushes.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Parkell Inc. The logo is in black and white and features the word "parkell" in a bold, sans-serif font. To the right of "parkell" and slightly below is the word "inc." in a smaller, italicized font. The logo is simple and professional.
#### 5. Indications for Use
Parkell® Desensitizer Gel is indicated for reduction of tooth hypersensitivity by the following treatments:
- treatment of dentin exposed by toothbrush abrasion, gingival recession, periodontal disease, and/or acid erosion;
- · treatment of dentin after mechanical tooth cleaning, scaling, and/or root planning;
- treatment of tooth hypersensitivity associated with bleaching; and
- treatment of prepared dentin for fillings and/or prosthetic restorations.
## 6. Comparison of Technological Characteristics
Information provided in this 510(k) submission demonstrates that Parkell® Desensitizer Gel is substantially equivalent to the Primary Predicate (Teethmate Desensitizer, K131068), in terms of the intended use and technological characteristics.
Any differences between the Device and the Primary Predicate, for purposes of this 510(k) submission, are of a minor nature as is demonstrated in Table 5-A and throughout this submission. Thus, this submission demonstrates the substantial equivalence between Parkell® desensitizer and the Primary Predicate. A brief comparison of Parkell® Desensitizer Gel to the Primary Predicate is provided below:
| Property | Parkell® Desensitizer Gel<br>(Parkell, Inc.) | Primary Predicate<br>Teethmate Desensitizer (K131068, Kuraray Noritake Dental Inc.) |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended uses | Parkell® Desensitizer Gel is indicated for reduction of tooth hypersensitivity by the following treatments: treatment of dentin exposed by toothbrush abrasion, gingival recession, periodontal disease, and/or acid erosion; treatment of dentin after mechanical tooth cleaning, scaling, and/or root planning; treatment of tooth hypersensitivity associated with bleaching; and treatment of prepared dentin for fillings and/or prosthetic restorations. | Teethmate Desensitizer is indicated for reduction of tooth hypersensitivity by the following treatments: treatment of dentin exposed by toothbrush abrasion, gingival recession, periodontal disease, and/or acid erosion; treatment of dentin after mechanical tooth cleaning, scaling, and/or root planning; treatment of tooth surface before and/or after bleaching; and treatment of prepared dentin for fillings and/or prosthetic restorations. |
| Classification Product Code | EJK | EJK |
| Regulation Number | 21 CFR 872.3250 | 21 CFR 872.3250 |
| Principle of operation | tooth desensitizer | tooth desensitizer |
| Material form | gel | powder/liquid |
## Table 5-A:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "parkell inc.". The word "parkell" is written in a bold, sans-serif font, with the letters closely spaced together. To the right of "parkell", the word "inc." is written in a smaller, italicized font. The logo is in black and white.
| | | Parkell® Desensitizer<br>Gel<br>(Parkell, Inc.) | Primary Predicate<br>Teethmate Desensitizer<br>(K131068, Kuraray) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------------------------|
| Chemical composition | | water; calcium<br>phosphate-<br>based filler(s) | water; calcium<br>phosphate-<br>based filler(s) |
| Surface coverage of treated dentin | | yes | yes |
| Occludes dentinal tubules | | yes | yes |
| Microshear bond<br>strength to dentin of<br>resin-based bonding<br>agents following<br>pretreatment with<br>Device or Predicate<br>Device<br>(Pass/fail criteria of at<br>least about 15MPa) | Commercially-available<br>bonding agent #1<br>[self-etch mode] | ≥15MPa | ≥15MPa |
| | Commercially-available<br>bonding agent #1<br>[total-etch mode] | ≥15MPa | ≥15MPa |
| | Commercially-available<br>bonding agent #2<br>[self-etch mode] | ≥15MPa | ≥15MPa |
#### Table 5-A (continued):
As seen above, the differences between the subject and primary predicate devices are limited to materials of construction and material form. These technological differences do not raise issues with respect to substantial equivalence and are addressed by comparative performance data provided within this submission.
#### 7. Biocompatibility and Bench Data
An evaluation of biocompatibility was conducted on Parkell® desensitizer in accordance, e.g., with ISO10993-1:2009, ISO10993-5:2009, ISO10993-10:2010, ISO10993-11:2017, ISO10993-12: 2012, and the FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (June 16, 2016). The conclusion of the evaluation is that Parkell® Desensitizer Gel is biocompatible. In addition, an evaluation of the bioactivity of Parkell® desensitizer was performed through analysis of treated dental tissue and via tests performed in accordance with ISO23317:2014.
#### 8. Clinical Performance Data
There was no clinical testing required to support the Device as the intended uses are equivalent to the Predicate Device. These types of devices have been on the market for many years with no reported adverse events. The non-clinical testing detailed in this submission supports the substantial equivalence of the Device.
#### 9. Statement of Substantial Equivalence:
The information provided above supports that Parkell® Desensitizer Gel is substantially equivalent to the Primary Predicate. Although minor differences in design and technology exist between the subject and primary predicate devices, the testing supports that these differences do not raise questions as to the substantial equivalence between the devices. Therefore, it is concluded that Parkell® Desensitizer Gel is substantially equivalent to the Primary Predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.