ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White-to-white (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
Device Story
ARGOS is a non-invasive, non-contact optical biometer using swept-source optical coherence tomography (SS-OCT) to measure ocular parameters; device captures axial length, corneal thickness, anterior chamber depth, lens thickness, K-values, astigmatism, white-to-white diameter, and pupil size. It also provides reference image capture for preoperative/postoperative documentation. Operated by ophthalmologists or eye-care professionals in clinical settings; device performs automated calculations for IOL power and type selection. Output includes ocular measurements and reference images, which can be transferred to external image-guided surgical systems to assist in preoperative planning. Device benefits patients by providing precise biometric data necessary for accurate IOL selection during cataract or refractive surgery.
Clinical Evidence
Bench testing only. Performance evaluated against standards including ISO 15004-1, ANSI Z80-36, IEC 60825, and ISO 10343. Hardware validation confirmed accuracy for axial/lateral distance measurements, signal-to-noise ratio, and depth attenuation. Usability engineering performed per IEC 62366. Software verification and validation conducted per FDA guidance. No clinical data presented.
Technological Characteristics
SS-OCT biometer; 1060nm swept laser (3kHz rate); 16-LED video keratometry (850nm); white LED for reference imaging; 630nm fixation LED. Class 1 laser product. Complies with ANSI/AAMI ES 60601-1, IEC 60529 (IP20), ANSI Z80-36, IEC 62471, and ISO 10343. Connectivity to external surgical planning systems (Verion, ORA).
Indications for Use
Indicated for ophthalmologists, physicians, and eye-care professionals to perform non-invasive ocular measurements (axial length, corneal thickness, anterior chamber depth, lens thickness, K-values, astigmatism, white-to-white, pupil size) and IOL power calculations for patients undergoing intraocular lens placement.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 16, 2019
Santec Corporation % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, NJ 07059
Re: K191051
Trade/Device Name: Argos Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: MXK Dated: April 17, 2019 Received: April 19, 2019
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K191051
Device Name ARGOS (ARGOS ver1.5)
#### Indications for Use (Describe)
ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Clamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
#### Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(k) SUMMARY
The information contained in this premarket notification 510(k) summary is submitted as required by 21 CFR 807.92:
### 1. Type of Submission
510k Traditional submission for New Devices
#### 2. Applicant Company:
SANTEC CORPORATION 5823 Ohkusa-Nenjozaka, Komaki, Aichi 485-0802, JAPAN Phone: 81 (0568) 79 3535 www.santec.com
#### 3. Applicant (Contact Person) Name:
Changho Chong
# 4. Date Summary Prepared:
May 15, 2019
#### 5. Device Trade / Proprietary Name (Model name):
ARGOS (ARGOS ver1.5)
## 6. Common Name:
Optical Biometer
## 7. Regulation
Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Bio-Microscope Regulatory Class: Class II Product Code: MXK
{4}------------------------------------------------
### 8. Indications for Use:
ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K - values (Radii of flattest and steepest meridians). Astigmatism. White-to-white (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
The indications for use are similar between the proposed and predicate ARGOS devices. The proposed ARGOS device has additional reference image functionality as minor difference.
#### 9. Device Descriptive
The ARGOS is substantially equivalent to the predicate device identified previously:
- the ARGOS (Santec Corporation) that was cleared by the FDA on October 2nd, 2015(K150754) as the primary predicate as it has the most similar intended use and characteristics
- the IOLMaster 700 (Carl Zeiss Meditec) that was cleared by the FDA on June 29, ● 2015 (K143275, K170171) as an additional (secondary) predicate to support a new feature of reference image capture on the version of ARGOS in this submission.
The predicate devices are Class 2 devices to premarket notification, as defined per regulation number 21 CFR 886.1850. In addition, the predicate devices have product codes of MXK(ARGOS), and HJO(IOLMaster700).
The version of ARGOS in this submission is a modified version of the Argos cleared under K150754 which is substantially equivalent with regard to intended use, operating principle, function, materials, and energy source. The differences from the predicate ARGOS (K150754) that are subject of this 510(k) submission are:
- An additional feature of reference image capture function ●
- Labeling change including change in the intended use adding the feature of reference . image capture.
- . This image can be transferred to image guided devices in order to support the execution of preoperative plan.
{5}------------------------------------------------
The changes described in this submission do not affect how the hardware is used to acquire measurements as a biometer, nor do these changes affect the principle of operation of the device.
| Compliant standard | Description | Classification |
|----------------------|------------------------------------------------------------------------|---------------------|
| ANSI/AAMI ES 60601-1 | Electrical safety | Class 1 Type B |
| IEC 60601-1-2 | Electrical safety | Refer to Appendix 4 |
| IEC 60529 | Housing protection | IP20 |
| ANSI.Z80-36 | Ophthalmics - Light Hazard<br>Protection For Ophthalmic<br>Instruments | Group 2 |
| IEC 62471 | LED light hazard | Exempt group |
| IEC 60825-1 | Laser product safety | Laser Class 1 |
| ISO 10343 | Ophthalmometer | |
#### 10. Guidance on the Recognition and Use of Consensus Standards:
ARGOS complies with the requirements of listed FDA Recognized Consensus Standards
{6}------------------------------------------------
# 11. Substantial equivalence to the predicate device
## 11.1 Comparisons of technological characteristics with Predicate ARGOS devices
| | | Table of Comparison between the ARGOS and the Predicate Device | | | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--|
| | New Device (Subjective device) | Primary Predicate Device | Secondary Predicate device | Comparison | |
| Manufacturer | santec corporation | santec corporation | Zeiss Meditech | - | |
| Model | ARGOS ver1.5 | ARGOS | IOLMaster 700 | - | |
| 510(k) number | | K150754 | K143275, K170171 | - | |
| Class | II | II | II | Same | |
| Classification<br>panel | Ophthalmic device panel | Ophthalmic device panel | Ophthalmic device panel | Same | |
| Product code | MXK | MXK | HJO | Same | |
| Classification<br>number | AC-Powered Slitlamp Bio-<br>Microscope | Device, Analysis, Anterior<br>Segment | AC-Powered Slitlamp Bio-<br>Microscope | Same | |
| Type | Optical biometer | Optical biometer | Biometer | Same | |
| Intended Use | ARGOS is a non-invasive, non-contact<br>biometer based on swept-source<br>optical coherence tomography (SS-<br>OCT). The device is intended to<br>acquire ocular measurements as well<br>as perform calculations to determine<br>the appropriate intraocular lens (IOL)<br>power and type for implantation during<br>intraocular lens placement. ARGOS<br>measures the following 9 parameters:<br>Axial Length, Corneal Thickness,<br>Anterior Chamber Depth, Lens<br>Thickness, K-values (Radii of flattest<br>and steepest meridians), Astigmatism,<br>White-to-white (corneal diameter) and<br>Pupil Size. The Reference Image<br>functionality is intended for use as a<br>preoperative and postoperative image<br>capture tool.<br>It is intended for use by<br>ophthalmologists, physicians, and<br>other eye-care professionals and may<br>only be used under the supervision of<br>a physician. | ARGOS is a non-invasive, non-<br>contact biometer based on swept-<br>source optical coherence<br>tomography (SS-OCT). The device<br>is intended to acquire ocular<br>measurements as well as perform<br>calculations to determine the<br>appropriate intraocular lens (IOL)<br>power and type for implantation<br>during intraocular lens placement.<br>ARGOS measures the following 9<br>parameters: Axial Length, Corneal<br>Thickness, Anterior Chamber<br>Depth, Lens Thickness, K-values<br>(Radii of flattest and steepest<br>meridians), Astigmatism, White-to-<br>white (corneal diameter) and Pupil<br>Size.<br>lt is intended for use by<br>ophthalmologists, physicians, and<br>other eye-care professionals and may<br>only be used under the<br>supervision of a physician. | The indications for use are identical<br>between the subject and predicate<br>IOLMaster 700 devices.<br>The IOLMaster 700 is intended for<br>biometric measurements and<br>visualization of ocular structures. The<br>measurements and visualization assist<br>in the determination of the appropriate<br>power and type of intraocular lens.<br>The IOLMaster 700 measures:<br>• Lens thickness<br>• Corneal curvature and thickness<br>• Axial length<br>• Anterior chamber depth<br>• Pupil diameter<br>• White-to-white distance (WTW)<br>For visualization, the IOLMaster 700<br>employs optical coherence<br>tomography (OCT) to obtain two-<br>dimensional images of ocular<br>structures of the anterior and posterior<br>segments of the eye.<br>The Reference Image functionality is<br>intended for use as a preoperative and<br>postoperative image capture tool. | Similar to Primary,<br>additional feature<br>identical to<br>secondary<br>predicate | |
| Measured<br>parameters | Axial Length, Corneal Thickness,<br>Anterior Chamber Depth, Lens<br>Thickness, K-values (Radii of<br>flattest and steepest meridians),<br>Astigmatism (Toric Angle), White-<br>to-white (corneal diameter), and<br>Pupil Size. | Axial Length, Corneal<br>Thickness, Anterior Chamber<br>Depth, Lens Thickness, K-<br>values (Radii of flattest and<br>steepest meridians),<br>Astigmatism (Toric Angle),<br>White-to-white (corneal<br>diameter), and Pupil Size. | Lens thickness, Corneal<br>curvature and thickness, Axial<br>length, Anterior chamber<br>depth, Pupil diameter, White-to-<br>white distance (WTW) | Same parameters | |
| In-vivo<br>repeatability | Axial length: range 14-38mm, SD<br>0.01mm | Axial length: range 14-38mm,<br>SD 0.01mm | Axial length: range 14-38mm,<br>SD 0.005mm | Same between<br>subjective and<br>primary, Similar<br>between subjective<br>and secondary<br>predicate | |
| | Corneal thickness: range 200-<br>120um, SD 10um | Corneal thickness: range 200-<br>120um, SD 10um | Corneal thickness: range 200-<br>120um, SD 2.5um | | |
| | Anterior Chamber Depth: range<br>0.7-8.0mm SD 0.01mm | Anterior Chamber Depth:<br>range 0.7-8.0mm SD 0.01mm | Anterior Chamber Depth: range<br>0.7-8mm SD 0.007mm | | |
| | Lens Thickness: range 0.5-<br>10.0mm SD 0.02mm | Lens Thickness: range 0.5-<br>10.0mm SD 0.02mm | Lens Thickness: range 1-10.0mm<br>SD 0.006mm | | |
| | Keratometry: range 5.5-10.0mm<br>SD 0.02mm | Keratometry: range 5.5-<br>10.0mm SD 0.02mm | Keratometry: range 5-11.0mm SD<br>0.09D(~0.02mm) | | |
| | Astigmatism: range 0-180deg, SD<br>5deg (Cylinder>1D) | Astigmatism: range 0-180deg,<br>SD 5deg (Cylinder>1D) | Astigmatism: range 0-180deg, SD<br>3.8deg (Cylinder>0.75D) | | |
| | Pupil size: range 2-13mm SD<br>0.09mm | Pupil size: range 2-13mm SD<br>0.09mm | Pupil size: (no specification<br>available) | | |
| | White-to-White: range 7-15mm<br>SD 0.06mm | White-to-White: range 7-15mm<br>SD 0.06mm | White-to-White: range 8-16mm<br>SD 0.044mm | | |
| Method for<br>biometry | Interferometry (SS-OCT) | Interferometry (SS-OCT) | Interferometry (SS-OCT) | Same | |
| Method for<br>keratometry | Video keratometry (16 LED) | Video keratometry (16 LED) | Video keratometry (IR LED) | Same | |
| Optical<br>radiation | | | | | |
| Type of light<br>source | Wavelength swept laser (3kHz<br>swept rate)<br>Wavelength 1060nm<br>Power <0.72mW (<0.1mW<br>average at 1mm aperture during<br>B-scan) | Wavelength swept laser (3kHz<br>swept rate)<br>Wavelength 1060nm<br>Power <0.72mW (<0.1mW<br>average at 1mm aperture<br>during B-scan) | Wavelength swept laser (2kHz<br>swept rate)<br>Wavelength 1055nm<br>unavailable | Same | |
| Laser class | Class 1 (IEC 60825)<br>ANSI-Z80.36-2016 Group2 | Class 1 (IEC 60825)<br>ISO15004-2 Group2 | Class 1 (IEC 60825)<br>ISO15004-2 Group2 | Same<br>(ANSI.Z80-36 is<br>equivalent to<br>ISO15004-2) | |
| Keratometry | L…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.