The MIS Attachments and Cutting Accessories are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE®) Console and electric and pneumatic motors. When used with these motors, the MIS Attachments and Cutting Accessories are intended to cut bone in the following manner: drilling, reaming, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications. Specific applications include Craniectomy, Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/ Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine. These devices are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices. The Footed Attachments and Cutting Accessories are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE®) Console and electric and pneumatic motors. When used with these motors, the Footed Attachments and Cutting Accessories are intended to cut bone in the following manner: drilling, reaming, dissecting, and shaping for the following medical applications: Neuro: Spine; Ear. Nose, and Throat (ENT) /Otology/Neurotology/ Otorhinolaryngology; Craniofacial and Maxillofacial; and Sternotomy. Specific applications include Craniotomy, Pterional Craniotomy, Sub Occipital/Retro Sigmoid/Posterior Fossa Craniotomy, Sphenoid Wing Dissection, Laminotomy / Laminectomy, and Orthopedic Spine. These devices are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices. It is also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices.
Device Story
Stryker MIS and Footed Attachments serve as mechanical interfaces between powered motors (CORE Console) and cutting accessories; transmit torque from motor to accessory; provide ergonomic grip for surgeon. Used in OR for neuro, spine, ENT, and craniofacial procedures. Attachments are stainless steel; feature color-coded bands at proximal end to ensure correct assembly and compatibility with cutting accessories. Modification involves color band material/colorant. No change to principle of operation, energy source, or clinical application. Benefits include standardized assembly and reliable bone cutting/shaping during surgical procedures.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing (colorfastness and durability) and biocompatibility evaluation (chemical characterization via ICP-MS and toxicological risk assessment) per ISO 10993-1.
Technological Characteristics
Stainless steel construction; mechanical torque transmission; color-coded proximal bands for assembly identification. Compatible with electric and pneumatic motors via CORE console. Biocompatibility compliant with ISO 10993-1 for limited exposure (<24 hours).
Indications for Use
Indicated for patients requiring bone cutting (drilling, reaming, shaping, dissecting, shaving, smoothing) in neurosurgical, spinal, ENT, otological, craniofacial, maxillofacial, and sternotomy procedures; also for preparation of fixation sites for screws, wires, pins, and metal hardware.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
Stryker MIS Attachments and cutting accessories (K143540)
Stryker Footed Attachments and cutting accessories (K143399)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 18, 2019
Stryker Corporation Paminder Khurmi Staff Regulatory Affairs Specialist 4100 E Milham Ave Kalamazoo, Michigan 49001
Re: K191049
Trade/Device Name: Stryker MIS and Footed Attachments Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE, ERL Dated: April 18, 2019 Received: April 19, 2019
Dear Paminder Khurmi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191049
Device Name Stryker MIS attachments
### Indications for Use (Describe)
The MIS Attachments and Cutting Accessories are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE®) Console and electric and pneumatic motors. When used with these motors, the MIS Attachments and Cutting Accessories are intended to cut bone in the following manner: drilling, reaming, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications.
Specific applications include Craniectomy, Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/ Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.
These devices are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices.
Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191049
Device Name Stryker Footed attachments
### Indications for Use (Describe)
The Footed Attachments and Cutting Accessories are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE®) Console and electric and pneumatic motors. When used with these motors, the Footed Attachments and Cutting Accessories are intended to cut bone in the following manner: drilling, reaming, dissecting, and shaping for the following medical applications: Neuro: Spine; Ear. Nose, and Throat (ENT) /Otology/Neurotology/ Otorhinolaryngology; Craniofacial and Maxillofacial; and Sternotomy.
Specific applications include Craniotomy, Pterional Craniotomy, Sub Occipital/Retro Sigmoid/Posterior Fossa Craniotomy, Sphenoid Wing Dissection, Laminotomy / Laminectomy, and Orthopedic Spine.
These devices are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices. It is also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies onlyto requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regardingthis burden estimate or any other as pect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Stryker Instruments 4100 Milham Ave. Kalamazoo, MI 49001 t: 269 389 7700 f: 269-324-5412 www.stryker.com
# 510(k)Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
#### SUBMITTER l.
| 510(k) Owner: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, Michigan 49001 USA<br>Ph: +1-269-323-7700<br>Fax: +1-269-324-5412 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Paminder Khurmi<br>Staff Regulatory Affairs Specialist<br>Phone: +1-269-389-2264<br>Paminder.Khurmi@Stryker.com |
| FDA Establishment<br>Registration No. | 1811755 |
| Date Submitted: | April 18, 2019 |
#### DEVICE II.
| Trade Name: | Stryker® MIS and Footed Attachments |
|--------------------------|--------------------------------------------------------------------------------------------------------|
| Common Name: | Surgical Drill Handpieces |
| Primary Classification | Drills, Burs, Trephines & Accessories (Simple, Powered)<br>(21 CFR 882.4310, Product code HBE) |
| Secondary Classification | Drill, Surgical ENT (Electric or Pneumatic) including Handpiece<br>(21 CFR 874.4250, Product code ERL) |
| Classification: | II |
{6}------------------------------------------------
Stryker Instruments 4100 Milham Ave. Kalamazoo, MI 49001 t: 269 389 7700 f: 269-324-5412 www.stryker.com
#### lll. PREDICATE DEVICE
| 510 (K) Number | Product code | Trade name | Manufacturer |
|----------------|--------------|-----------------------------------------------------------|---------------------|
| K143540 | HBE, ERL | Stryker® MIS<br>Attachments and<br>cutting accessories | Stryker Instruments |
| K143399 | HBE, ERL | Stryker® Footed<br>Attachments and<br>cutting accessories | Stryker Instruments |
#### IV. SUBJECT DEVICE DESCRIPTION
The Stryker MIS and Footed Attachments are used within a system consisting of a variety of devices, including a console, powered motors, and cutting accessories. The attachments connect to the motors and the cutting accessories to complete the system for physician use. The Stryker MIS and Footed Attachments are offered for prescription use only. The MIS and Footed Attachments are intended to serve as interfaces between powered motors and cutting accessories for the purposes of:
- > Cutting bone, bone cement, and teeth;
- A Placing or cutting screws, metal, wires, pins, and other fixation devices; and
- > Providing a location for the user to hold and grip the device system.
The Stryker MIS Attachments are available in straight, curved and angled styles.
The Stryker Footed attachments are available as fixed footed, rotating footed, and nonfooted attachments
The MIS and Footed Attachments are made of stainless steel (SST). All MIS and Footed Attachments have a color band on the proximal end of the outer surface to aid the user in correct system assembly. The color band serves to enhance the distinction of attachment and cutting accessory compatibility.
The Subject Devices Stryker MIS attachments and Stryker Footed attachments are cleared under two separate 510(k)s. Since the modification to the color band is identical for both attachment types (MIS and Footed), this premarket notification is being submitted as bundled submission in accordance with guidance document issued by FDA on June 22, 2007, "Bundling Multiple Devices or Multiple Indications in a Single Submission".
{7}------------------------------------------------
Stryker Instruments 4100 Milham Ave. Kalamazoo, MI 49001 t: 269 389 7700 f: 269-324-5412 www.stryker.com
**stryker**
The purpose of this submission is to gain clearance for the color band modifications for the Subject Device Attachments color band that exceeds the threshold as per FDA Guidance, "Deciding When to Submit a 510(k) for a Change to an Existing Device".
#### V. INDICATIONS FOR USE
As the subject premarket notification is a bundled submission, there are 2 proposed indications for use, and are broken in 2 tables for each attachment type for easy navigation. See Table 1 below for Indications for Use for Stryker MIS attachments and Table 2 for Indications for Use for Stryker Footed attachments.
| TABLE 1: COMPARISON OF INDICATIONS FOR USE FOR MIS ATTACHMENTS | |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device- Stryker MIS attachments |
| Indications<br>for Use | The MIS Attachments and Cutting Accessories are intended to be used with the<br>Stryker Consolidated Operating Room Equipment (CORE®) Console and electric<br>and pneumatic motors. When used with these motors, the MIS Attachments<br>and Cutting Accessories are intended to cut bone in the following manner:<br>drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for<br>the following medical applications: Neuro; Spine; Ear, Nose, and Throat<br>(ENT)/Otorhinolaryngology; and Endoscopic applications.<br>Specific applications include Craniotomy/Craniectomy,<br>Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded<br>Endonasal Approach (EEA)/Anterior Skull Base/<br>Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.<br>These devices are also usable in the preparation for the placement of screws,<br>metal, wires, pins, and other fixation devices. |
### TABLE 2: COMPARISON OF INDICATIONS FOR USE FOR FOOTED ATTACHMENTS
| | Subject Device- Stryker Footed attachments |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Footed Attachments and Cutting Accessories are intended to be used with<br>the Stryker Consolidated Operating Room Equipment (CORE®) Console and<br>electric and pneumatic motors. When used with these motors, the Footed<br>Attachments and Cutting Accessories are intended to cut bone in the following<br>manner: drilling, reaming, decorticating, shaping, dissecting, shaving and<br>smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and<br>Throat (ENT) /Otology /Neurotology/Otorhinolaryngology; Craniofacial and<br>Maxillofacial; and Sternotomy. |
{8}------------------------------------------------
Specific applications include Craniotomy/Craniectomy, Pterional Craniotomy, Sub Occipital/Retro Sigmoid/Posterior Fossa Craniotomy, Sphenoid Wing Dissection, Laminotomy / Laminectomy, and Orthopedic Spine. These devices are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices.
The Subject Device Indications for Use remains identical to the Predicate Devices. The modification described to the Subject Device in this 510(k) does not change the Indications for Use.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Subject Device is compared to the Predicate Device for substantial equivalence of technological characteristics based on the modification described in this submission.
### Technological Characteristics
The only change in the Subject and Predicate Devices is the different color band.
The Color Band location, Attachment to Cutting Accessory Locking Mechanism, and Attachment to Motor Locking Mechanism are identical when comparing the Subject and Predicate Device and, all other technological characteristics, remains the same between the Subject and Predicate Devices.
The technological characteristics that are not the same between the Subject and Predicate Devices are the color band material and colorant. However, the Subject Device color band material and colorant are similar to those cleared for the Predicate Device, and the different characteristics do not raise different questions of safety and effectiveness.
#### VII. PERFORMANCE DATA
The following testing were performed, and the data was provided in support of the substantial equivalence determination:
Verification testing was performed on the Subject Devices as dictated by the results of the Risk Analysis, and no new questions of safety and effectiveness were raised. The Subject Device met all pre-defined acceptance criteria. The results of the tests mention in the Table 3, support the substantial equivalence of the Subject Device to the Predicate Device. Testing data from the following was provided in support of the substantial equivalence determination.
{9}------------------------------------------------
### TABLE 3. LIST IF TESTS PERFORMED
| Tests performed | Results |
|-----------------------------------|---------|
| Colorfastness and Durability test | Passed |
| Colorfastness Test | Passed |
### Biocompatibility Testing
A biocompatibility evaluation was performed following the recommendations of ISO 10993-1: 2018 and FDA Guidance (Use of International Standard ISO- 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" June 2016) as appropriate for limited exposure (< 24 hours) externally communicating, tissue/bone/dentin devices. The following assessments were completed to further assess the potential risk of the subject device material and colorant modification impacting color additive release.
- Chemical Characterization: ICP-MS
- . Toxicological Risk Assessment
The Biocompatibility testing was adopted from predicate submissions (K143540, K143399)
### Animal Testing
Animal testing was not required as a basis for substantial equivalence.
### Clinical Testing
Clinical testing was not required as a basis for substantial equivalence.
### The Principle of Operation
The color band modification on the device does not change the Principle of Operation of the Subject Device compared to the Predicate Device.
The Principle of Operation remains: The Subject Device Attachments are combined with a power source, motor, and cutting accessory to achieve their function; the main function of the Subject Device Attachments is to provide a balanced location for the surgeon to hold and grip the device system, and to transmits torque from the motor to a cutting accessory.
{10}------------------------------------------------
Stryker Instruments 4100 Milham Ave. Kalamazoo, MI 49001 t: 269 389 7700 f: 269-324-5412 www.stryker.com
stryker
#### VIII. CONCLUSIONS
The subject devices, in comparison with the legally marketed predicates, have the same intended use, indications for use, operating principles, energy source, and functional outputs. Performance testing and risk analysis demonstrate that the devices are as safe and effective as the predicate devices and do not raise different significant questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.