← Product Code [GCJ](/productcode/GCJ) · K191043

# da Vinci Handheld Camera (K191043)

_Intuitive Surgical, Inc. · GCJ · Jun 6, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191043

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jun 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The da Vinci Handheld Camera is intended for endoscopic viewing of internal surgery sites during minimally invasive surgery. It is designed for use with compatible da Vinci Surgical Systems.

## Device Story

The da Vinci Handheld Camera is a lightweight 2D handheld camera designed for minimally invasive surgery. It connects to third-party 5 mm to 10 mm laparoscopes to capture surgical site images. The device integrates with the da Vinci Xi Vision Cart, utilizing its existing illuminator, video processor, and monitor to display video output. It consists of a camera head, endocoupler, cable assembly, light guide, and adapter. Used in the operating room by surgeons and clinical staff, the device provides visualization of internal anatomy to assist in surgical decision-making. It is intended to be used with compatible da Vinci Surgical Systems. The device includes a dedicated sterilization tray for protection during reprocessing.

## Clinical Evidence

No human clinical trials were conducted. Evidence consists of bench testing (physical, mechanical, electrical, user interface, and equipment interface verification), simulated use in a porcine animal model, and human factors usability evaluation. Results demonstrated that the device meets all design requirements and is safe and effective for its intended use.

## Technological Characteristics

Handheld 2D camera system; compatible with 5 mm to 10 mm third-party laparoscopes. Components include camera head, endocoupler, cable assembly, light guide, and adapter. Integrates with da Vinci Xi Vision Cart for illumination and video processing. Includes a sterilization tray for the camera head. No specific material standards or software algorithm classes are detailed.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Intuitive Surgical Stereo View Endoscopic System ([K080155](/device/K080155.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 6, 2019

Intuitive Surgical, Inc. Mr. Manjunath Bisalehalli Regulatory Affairs Engineer 1266 Kifer Road Sunnyvale, California 94086

Re: K191043

Trade/Device Name: da Vinci Handheld Camera Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: April 17, 2019 Received: April 19, 2019

Dear Mr. Bisalehalli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jennifer R. Stevenson Acting Division Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation & Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191043

Device Name da Vinci Handheld Camera

Indications for Use (Describe)

The da Vinci Handheld Camera is intended for endoscopic viewing of internal surgery sites during minimally invasive surgery. It is designed for use with compatible da Vinci Surgical Systems.

| Type of Use (Select one or both, as applicable)               |                                                              |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div>    Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div>    Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary 7

| 510(k) Owner:                         | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086                                                                                       |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                              | Manjunath Bisalehalli<br>Regulatory Affairs Engineer<br>Phone Number: 408-523-7089<br>Fax Number: 408-523-8907<br>Email: Manjun.Bisalehalli@intusurg.com |
| Date Summary Prepared:                | April 17, 2019                                                                                                                                           |
| Trade Name:                           | da Vinci Handheld Camera                                                                                                                                 |
| Common Name:                          | Endoscope and accessories                                                                                                                                |
| Classification:                       | Class II<br>21 CFR 876.1500, Endoscope and Accessories                                                                                                   |
| Product Codes:                        | GCJ (Endoscope and accessories)                                                                                                                          |
| Classification Advisory<br>Committee: |                                                                                                                                                          |
| Committee:                            | General and Plastic Surgery                                                                                                                              |
| Predicate Device:                     | Intuitive Surgical Stereo View Endoscopic System<br>(K080155)                                                                                            |

#### Device Description

The da Vinci Handheld Camera is a lightweight handheld 2D Camera which can be connected to any third party 5 mm to 10 mm laparoscope to view images on the da Vinci Xi Vision cart. The da Vinci Handheld Camera consists of the camera head, the light guide, camera connector and the light guide adaptor.

The da Vinci Handheld Camera leverages the illuminator, video processor, monitor and video outputs on the do Vinci Xi Vision Cart to provide common functions of a laparoscopic video tower. The da Vinci Xi Handheld Camera connects to the vision cart in the same way an endoscope does through the endoscope controller. The da Vinci Handheld Camera consists of the camera head, endocoupler, cable assembly, light guide, and an adapter. The da Vinci Handheld Camera Head Sterilization Tray is intended for use to encase and protect da Vinci Handheld Camera Head during sterilization.

Image /page/3/Picture/8 description: The image contains the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is in a smaller font and also in a light gray color, with the registered trademark symbol to the right of it.

{4}------------------------------------------------

### Intended Use

The da Vinci Handheld Camera is intended for endoscopic viewing of internal surgery sites during minimally invasive surgery. It is designed for use with compatible do Vinci Surgical Systems.

#### Technological Characteristics

The subject device, do Vinci Handheld Camera, is technologically very similar to the predicate device, Intuitive Surgical Stereo View Endoscopic System (cleared under K080155). The subject device has the same architecture design, as the predicate except it is a handheld camera.

#### Performance Data

Performance data (bench and animal testing) demonstrate that the subject device is substantially equivalent to the predicate device and that the design outputs meet the design requirements. The testing includes dimensional measurements, mechanical and functional verification, simulated use in animal model (porcine), and human factors evaluation.

## Bench Testing

The subject device, da Vinci Handheld Camera, was subjected to a series of bench tests to evaluate the performance and to demonstrate that the design outputs meet the input requirements. The design verification testing assessed the following:

- . Physical Specifications
- Mechanical Requirements
- Electrical Requirements
- User Interface Requirements
- Equipment Interface Requirements

#### Animal Validations

A series of tests were performed using simulated clinical models (animal) to evaluate the performance of the subject device, da Vinci Handheld Camera.

#### Human Factors Evaluation

As part of the Usability Engineering Process for the da Vinci Handheld Camera, the Usability Risk Analysis was updated to identify any new usability characteristics related to safety, as well as foreseeable hazards and hazardous situations. Human factors

Image /page/4/Picture/19 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is in a smaller font and also in a light gray color, with the registered trademark symbol to the right of it.

{5}------------------------------------------------

evaluation was conducted on the da Vinci Handheld Camera. Based on the results of those studies, the da Vinci Handheld Camera is found to be safe and effective for the intended users, uses, and use environments.

#### Summary

Based on the intended use, indications for use, technological characteristics, and performance data, the subject device, da Vinci Handheld Camera is substantially equivalent to the predicate device, Intuitive Surgical Stereo View Endoscopic System.

Image /page/5/Picture/6 description: The image contains the logo for Intuitive Surgical. The word "INTUITIVE" is in a larger, sans-serif font, and is placed above the word "SURGICAL", which is in a smaller font. There is a yellow dot above the "I" in "INTUITIVE".

---

**Source:** [https://fda.innolitics.com/device/K191043](https://fda.innolitics.com/device/K191043)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
