K191032 · Nordicneurolab AS · LNH · Nov 27, 2019 · Radiology
Device Facts
Record ID
K191032
Device Name
fMRI Hardware System
Applicant
Nordicneurolab AS
Product Code
LNH · Radiology
Decision Date
Nov 27, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The fMRI Hardware System is a stimulus presentation and response collection system intended to be used by trained professionals to facilitate auditory and visual stimulation to be used in functional MR Imaging (fMRI) based on BOLD contrast. The visual system allows video signals from the stimulus presentation PC to enter the shielded scanner room and to be presented to the patient through a set of coil mounted displays (VisualSystemHD) or by an in-room LCD monitor.
Device Story
System facilitates fMRI studies by providing synchronized auditory/visual stimuli and collecting patient responses. Components: AudioSystem (headphones/mic), VisualSystem (coil-mounted displays or LCD monitor), ResponseGrips (handheld buttons), SyncBox (timing synchronization), and nordicAktiva (stimulus presentation software). Used in MRI suites by trained professionals (technicians/physicians). SyncBox connects to MR scanner to receive timing pulses, ensuring stimulus presentation aligns with image acquisition. Patient responses via grips are recorded by nordicAktiva. Fiber optics transmit signals into shielded scanner rooms. Output allows clinicians to provoke brain activity for BOLD contrast imaging, aiding in cognitive function investigation and pre/post-operative brain tumor mapping.
Clinical Evidence
Bench testing only. Functional testing performed per internal protocols. Biocompatibility testing for rubber eye guard conducted per ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2010 (skin irritation and sensitization). Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
System uses fiber optics for signal transmission. VisualSystem HD features 1920x1200 resolution, HDMI input, and integrated eye-tracking. Biocompatible rubber eye guards tested per ISO 10993. SyncBox provides hardware-level synchronization with MR scanner timing pulses. System operates via stimulus PC running nordicAktiva software.
Indications for Use
Indicated for use by trained professionals to facilitate auditory and visual stimulation during functional MR Imaging (fMRI) based on BOLD contrast for patients undergoing brain imaging, including those with brain tumors for pre-operative and post-operative assessment.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
November 27, 2019
NordicNeurolab AS Chandana Bhandari VP Quality Mollendalsveien 1 5009 BERGEN, NORWAY
Re: K191032
Trade/Device Name: fMRI Hardware System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: October 29, 2019 Received: October 31, 2019
Dear Chandana Bhandari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191032
#### Device Name
fMRI Hardware System (VisualSystem HD)
Indications for Use (Describe)
#### fMRI Hardware System
The fMRI Hardware System is a stimulus presentation and response collection system intended to be used by trained professionals to facilitate auditory and visual stimulation to be used in functional MR Imaging (fMRI) based on BOLD contrast.
VisualSystem
The VisualSystem allows video signals from the stimulus presentation PC to enter the shielded scanner room and to be presented to the patient through a set of coil-mounted displays [VisualSystem HD] or by an in-room LCD monitor.
| Type of Use (Select one or both, as applicable) | <table style="border:none;"><tr><td><span> <span style="border: 1px solid black; padding: 0px 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span style="padding-left:10px;"> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> <span style="border: 1px solid black; padding: 0px 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="padding-left:10px;"> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="border: 1px solid black; padding: 0px 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="padding-left:10px;"> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary NordicNeuroLab AS fMRI Hardware System
| 510(k) Type: | Traditional: Device Modification |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Date: | 25 November 2019 |
| Submitter: | NordicNeuroLab AS<br>Møllendalsveien 1<br>N-5009 Bergen<br>Norway<br>Phone: +47 55 70 70 95<br>E-mail: chandana@nordicneurolab.com<br>Establishment Registration Number: 3006738448 |
| Contact: | Chandana Gurung Bhandari<br>Møllendalsveien 1<br>N-5009 Bergen<br>Norway<br>Direct: +47 41 76 62 39<br>Phone: +47 55 70 70 95<br>E-mail: chandana@nordicneurolab.com |
| Legally marketed Device name and 510(k) number: | fMRI Hardware System. K092253 |
| Modified Device Name<br>Device Common Name:<br>Basis for Submission: | fMRI Hardware System<br>Accessory to MRI System, Nuclear Magnetic Resonance Imaging<br>Device Modification |
| Classification Regulation: | 21 CFR 892.1000 |
| Class:<br>Panel: | ll<br>Radiology |
Product Code: LNH
Trade/Proprietary Name: fMRI Hardware System
{4}------------------------------------------------
# Device Description
## 9.1 Intended Use
#### fMRI Hardware system
The fMRI Hardware System is a stimulus presentation and response collection system intended to be used by trained professionals to facilitate auditory and visual stimulation to be used in functional MR Imaging (fMRI) based on BOLD contrast.
### Visual system
The visual system allows video signals from the stimulus presentation PC to enter the shielded scanner room and to be presented to the patient through a set of coil mounted displays (VisualSystemHD) or by an in-room LCD monitor.
## 9.2 System Description
The system presents auditory and visual stimulus to the patient gives feedback through a pair of handheld grips. A synchronization module synchronizes the stimulus presentation software with the MR scanner. The System consists of five subsystems: AudioSystem, VisualSystem, ResponseGrip, SyncBox and nordicAktiva.
The system is used for fMRI studies. fMRI stands for functional Magnetic Resonance Imaging. This technique is useful when determining certain diseases, gaining more information about a patient's condition or investigating cognitive functions. The technique is also used for examining the area of the brain affected in patients suffering from a brain tumor in both the pre-operative and post-operative stages.
The System is used to present the stimulus necessary to provoke physiological processes in the brain. Visual [VisualSystem] and auditory [AudioSystem] stimulus and manual responses from the patient [ResponseGrips] are of primary interest. The timing of the visual and audio stimulation is critical to make sure that the correct MR image of the brains activity is linked to the stimulus presented. A synchronization unit [SyncBox], connected between the MR-scanner and the stimulus presentation
{5}------------------------------------------------
software [nordicAktiva], is included in the system to make sure that the synchronization is correct.
Image /page/5/Figure/1 description: The image shows a diagram of a system with several components, including a SyncBox, nordic Aciva, ResponseGrip Interface, and Communication Console. The SyncBox is connected to a "Trigger Pulse (MR scanner)" within a "Technical Room". The nordic Aciva is connected to the SyncBox, ResponseGrip Interface, and Communication Console, with "Audio" and "Video" connections to the console. The diagram also shows an LCD screen and a person inside a scanner, with "Video" and "Audio" connections to the Communication Console.
Figure 1 presents the complete configuration of the fMRI Hardware System. All signals entering or leaving the scanner room are received and transmitted by use of fiber optics. The system allows video and audio signals from the stimulus PC to enter the shielded scanner room and to be presented to the MR. The subject responds to the stimulus by using the handheld grips.
## 9.3 Sub-system components and description
#### 9.3.1 VisualSystem
The VisualSystem allows video signals from the stimulus presentation PC to enter the shielded scanner room and to be presented to the patient through a set of coil-mounted displays [VisualSystem HD] or by an in-room LCD monitor.
#### 9.3.2 Audio System
The AudioSystem allows auditory signals from the stimulus presentation PC to enter the scanner room and to be presented to the patient wearing a set of headphones. A communication console allows the operator to adjust the sound from the PC and to speak directly to the patient through a built-in microphone.
#### 9.3.3 ResponseGrip
The purpose of this component is to collect patient responses during an fMRI study. The ResponseGrip consists of two hand-held grips with two buttons each. By pressing the patient can respond to the presented stimulus. The ResponseGrip is connected to an optical-electrical adapter which converts
{6}------------------------------------------------
light to electrical signals. The electrical signal is fed to the Stimulus PC by using standard PC communication interfaces.
### 9.3.4SyncBox
The SyncBox is connected directly to the MRI scanner where it receives timing pulses sent out with each image series and demodulates this signal before it's forwarded to the stimulus PC. In this way one can ensure that the stimulus presentation software is synchronized with the MRI image recordings.
## 9.3.5nordicAktiva
nordicAktiva is a software that generates visual and auditory stimulus to the patient. The stimulus presentation is synchronized with the scanner through the SyncBox and presented to the patient through the VisualSystem and AudioSystem. nordicAktiva records the responses fed to the Stimulus PC from the ResponseGrip, as well as the synchronization pulses from the SyncBox.
# 9.4 Identification of Change to Unmodified Device
The VisualSystem will be updated with a new version of the coil mounted displays [VisualSystem HD] for attachment to the MR head coil. The brand name for the new version is VisualSystem HD. The same video signal can be presented on the VisualSystem HD and the in-room LCD-monitor that was cleared in K092253. The operator has the choice to use either the VisualSystem HD or the in-room LCD monitor as before.
## 9.4.1 Technological Characteristics and Substantial Equivalence
To summarize, the modified fMRI Hardware System is found substantial equivalent to the previously cleared device. The modified system only has a new device added to display video. The indications for use for the modified fMRI Hardware System have remained unchanged.
| Changes: | fMRI Hardware System,<br>K092253 [Coil Mounted<br>Displays] | Modified fMRI Hardware System<br>[VisualSystem HD] |
|--------------------|-------------------------------------------------------------|----------------------------------------------------|
| Display Resolution | 800x600 pixels | 1920x1200 pixels |
| Video source | VGA - Standard PC video output | HDMI - Standard PC video output |
| EyeTracking | Separate unit | Integrated |
The rationale for determining the substantial equivalence between the previously cleared device and the modified fMRI Hardware System is based on the defined intended use, indications for use, and the technical and operational characteristics. To verify that the modified device fulfils the defined characteristics and requirements, it has been subject to extensive in-house and lab testing. The successful completion of said tests verifies the claimed characteristics of the modified fMRI Hardware System, and thus supports the determination of substantial equivalence.
{7}------------------------------------------------
## 9.5 Summary of Testing
The fMRI Hardware System has been tested for function, biocompatibility and safety and fulfills all requirement specifications.
#### 9.5.1 Functional test
All functional testing was carried out as per internal test protocols. Testing involved performance and bench testing.
#### 9.5.2 Biocompatibility test results
The direct components of applied parts contacting with the users in the VisualSystem HD were tested for the following:
In Vitro cytotoxicity - Cytotoxicity testing was carried out in accordance to ISO 10993-5:2009 Biological Evaluation of Medical Devices – Part 5: Tests for In Vitro Cytotoxicity. Under the study, the test article (rubber eye guard) extract did not show potential toxicity.
Skin irritation - Skin irritation testing was carried out in accordance to ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10.Under the conditions of the test article (rubber eye guard) extract did not show any significant evidence of causing skin irritations.
Skin Sensitization testing was carried out with accordance to ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Under the conditions of the test article (rubber eye guard) extract did not induce signs of general toxicity. In coherence with the negligible proliferation response of the Lymph Node Cells (LNCs), the test items (rubber eye guard) was identified as a non-sensitizing agent in the LocaL Lymph Node Assay.
#### 9.5.3 Safety tests
Electrical safety and EMC safety testing was performed to, and passed, the following standards
- IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility – Requirements and tests
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.