K191020 · CooperSurgical, Inc. · PUB · Jan 14, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K191020
Device Name
G210 InviCell Plus with SignipHy pH monitoring
Applicant
CooperSurgical, Inc.
Product Code
PUB · Obstetrics/Gynecology
Decision Date
Jan 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6120
Device Class
Class 2
Indications for Use
The G210 InviCell Plus with SignipHy™ pH Monitoring System is a bench-top incubator that is intended to provide a controlled environment at or near body temperature and gas levels (CO2, O2, and N2) for the development of gametes and/or embryos during In Vitro Fertilization (IVF) / Assisted Reproductive Technology (ART) treatments. The G210 InviCell Plus with SignipHy™ pH Monitoring System includes an accessory for pH monitoring of surrogate samples of bicarbonate-based culture media used for ART procedures.
Device Story
Bench-top incubator for IVF/ART; provides controlled temperature and gas environment (CO2, O2, N2) for gametes/embryos. Includes SignipHy™ pH Monitoring System: TrakPod (LED-based optical measurement instrument), sv2 Sensor (single-use polystyrene vessel with pH-sensitive dye membrane), TrakStation (tablet with proprietary software), and qc² Alignment Tool. TrakPod sends green light (518 nm) through fiber optic to sensor; dye membrane reflects light (600 nm) proportional to pH. TrakStation displays/stores pH data. Used in clinical labs by embryologists/technicians. pH monitoring provides environmental data to assist in maintaining culture conditions. Benefits include real-time pH tracking of surrogate media to ensure optimal development environment.
Clinical Evidence
No clinical data. Bench testing only. Includes sterilization validation (ISO 11135, AAMI TIR28), shelf-life/package integrity (ASTM F1886, F88, F1929), electrical safety (IEC 61010-2-010/101), EMC (IEC 61326-1), and software/cybersecurity validation. pH monitoring validation confirmed linearity, stability, precision, and accuracy. Light safety testing confirmed no toxicity. Mouse Embryo Assay (MEA) met specification (≥90% blastocysts at 96h) under worst-case operation.
Technological Characteristics
Bench-top incubator with 10 incubation chambers and 1 preparation chamber. Resistive heating foil. Gas supply: internally regulated CO2, O2, N2. pH monitoring: LED-based optical measurement (518 nm excitation, 600 nm emission). Materials: Polystyrene (sv2 sensor). Connectivity: USB (TrakPod to TrakStation). Sterilization: Ethylene oxide (sv2 sensor). Software: Minor level of concern.
Indications for Use
Indicated for use in IVF/ART treatments to provide a controlled environment (temperature, CO2, O2, N2) for gamete and embryo development, and to monitor pH of surrogate bicarbonate-based culture media samples.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 14, 2020
CooperSurgical Inc. Kyle Hooper Regulatory Affairs Associate 95 Corporate Drive Trumbull, CT 06611
Re: K191020
Trade/Device Name: G210 InviCell Plus with SignipHyTM pH Monitoring System Regulation Number: 21 CFR 884.6120 Regulation Name: Assisted Reproduction Accessories Regulatory Class: II Product Code: PUB Dated: December 12, 2019 Received: December 13, 2019
Dear Kyle Hooper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K191020
Device Name
G210 InviCell Plus with SignipHyTM pH Monitoring System
Indications for Use (Describe)
The G210 InviCell Plus with SignipHy™ pH Monitoring System is a bench-top incubator that is intended to provide a controlled environment at or near body temperature and gas levels (CO2, O2, and N2) for the development of gametes and/or embryos during In Vitro Fertilization (IVF) / Assisted Reproductive Technology (ART) treatments.
The G210 InviCell Plus with SignipHy™ pH Monitoring System includes an accessory for pH monitoring of surrogate samples of bicarbonate-based culture media used for ART procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> ☑ </span> </div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> ☐ </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cooper Surgical. The logo is in blue and consists of the words "Cooper Surgical" in a stylized font. Below the logo is the tagline "Healthy women, babies, and families" in a smaller font, also in blue. The logo is simple and professional, and the tagline conveys the company's mission.
K191020 Page 1 of 7
# 510(k) Summary
K191020
### Submitter Information
| Company Name: | CooperSurgical Inc. |
|------------------|-------------------------------------------------------------------|
| Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611 USA |
| Contact Person: | Kyle Hooper<br>Telephone: +1-203-601-5200<br>Fax: +1-203-601-9870 |
January 13, 2020
## Device Information
Date Prepared:
| Trade Name: | G210 InviCell Plus with SignipHy™ pH Monitoring System |
|--------------------|--------------------------------------------------------|
| Common Name: | Embryo Incubator with pH monitoring |
| Regulation Number: | 21 CFR 884.6120 |
| Regulation Name: | Assisted Reproduction Accessories |
| Product Code: | PUB (Accessory, Assisted Reproduction, Exempt) |
| Regulatory Class: | II |
## Predicate Device Information
Trade Name: BT37 Incubator (510(k) number: K121566) Common Name: IVF Incubator Classification Name: Accessory, Assisted Reproduction (21 CFR 884.6120) Product Code: MQG
The predicate device has not been subject to a design-related recall.
## Device Description
The G210 InviCell Plus with SignipHy™ pH Monitoring System is an assisted reproduction incubator that includes an accessory pH monitoring feature. The incubator component of this device (G210 Invicell) is regulated under 21 CFR 884.6120, product code PBH (exempt). The incubator includes 10 incubation chambers and one larger preparation chamber that is used for the equilibration of plates/oil before use. The tri-gas incubator provides a controlled environment (temperature, CO2/O2/N2) for gametes and embryos in the incubation chambers.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for CooperSurgical. The logo is in blue and consists of the words "CooperSurgical" in a stylized font. Below the logo is the tagline "Healthy women, babies, and families™" also in blue.
The pH monitoring feature of the G210 InviCell Plus with SignipHy™ pH Monitoring System is controlled separately and does not impact the operation or incubator functions of the device. The SignipHy™ pH Monitoring System consists of the following components:
- . SignipHy™ TrakPod – An LED-based, optical pH measurement instrument. The SignipHy TrakPod is physically supported within the chassis of the incubator and allows monitoring of one incubator chamber. It is a USB connected fluorescent measurement device with a fiber optic cable and fixture that connects to the SignipHy sv2 Sensor inside an incubator chamber. The SignipHy TrakPod and SignipHy sv² Sensor together detect the pH of a liquid sample.
- . SignipHy™ sv2 Sensor – The SignipHy sv2 Sensor is a single-use, polystyrene, ethylene oxide sterilized vessel that is loaded with a sample of bicarbonate-buffered assisted reproduction technology (ART) media with a pH between 6.8 and 7.2 for pH tracking. The bottom of the sensor includes a membrane that is impregnated with a dye that is affected by the pH of the media sample. The sensor fits into the SignipHy TrakPod fiber optic fixture located in an incubation chamber. SignipHy sv2 Sensors can be used for measuring pH at one-minute intervals for three days or 30-minute intervals for seven days.
- . SignipHy™TrakStation – A tablet computer running proprietary software that initiates pH readings, displays results, and stores pH measurement data over time. Up to eight SignipHy TrakPods can be connected to a single SignipHy TrakStation.
- SignipHy™ qc² Alignment Tool - A fluorescent reference device for use in system alignment and quality control.
To make a pH measurement, the SignipHy TrakPod sends green light flashes (peak 518 nm) through the fiber optic fixture. The dye in the membrane reacts by sending back flashes of light at a different wavelength (peak 600 nm). The SignipHy TrakPod reads this result and calculates the pH of the media sample that is then displayed on and stored in the SignipHy™TrakStation.
# Indications for Use
The G210 InviCell Plus with SignipHy™ pH Monitoring System is a bench-top incubator that is intended to provide a controlled environment at or near body temperature and gas levels (CO2, O2, and N2) for the development of gametes and/or embryos during In Vitro Fertilization (IVF) / Assisted Reproductive Technology (ART) treatments.
The G210 InviCell Plus with SignipHy™ pH Monitoring System includes an accessory for pH monitoring of surrogate samples of bicarbonate-based culture media used for ART procedures.
{5}------------------------------------------------
Table 1 provides a comparison of the indications for use and technological characteristics of the subject device and predicate device.
| Attribute | Subject<br>G210 InviCell Plus with SignipHy™ pH Monitoring System | Predicate<br>BT37 Incubator | Discussion |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K191020 | K121566 | Not Applicable |
| Manufacturer | CooperSurgical, Inc. | Planer Plc | Not Applicable |
| Indications for<br>Use | The G210 InviCell Plus with<br>SignipHy™ pH Monitoring System<br>is a bench-top incubator that is<br>intended to provide a controlled<br>environment at or near body<br>temperature and gas levels (CO2,<br>O2, and N2) for the development<br>of gametes and/or embryos<br>during In Vitro Fertilization (IVF) /<br>Assisted Reproductive<br>Technology (ART) treatments.<br>The G210 InviCell Plus with<br>SignipHy™ pH Monitoring System<br>includes an accessory for pH<br>monitoring of surrogate samples<br>of bicarbonate-based culture<br>media used for ART procedures. | The Planer BT37 Incubator is<br>intended to be used to provide<br>an environment with controlled<br>temperature at or near body<br>temperature, CO2, O2 and N2<br>gases, and elevated humidity for<br>the development of gametes and<br>embryos during in vitro<br>fertilization (IVF) / assisted<br>reproductive technology (ART)<br>treatments | Different:<br>The subject and predicate devices<br>are both indicated for<br>maintaining temperature and gas<br>concentrations sufficient to<br>support gametes and embryos.<br>The predicate device indication<br>also lists maintenance of<br>humidity within the chamber.<br>Lack of humidity control in the<br>subject device does impact the<br>overall intended use of the<br>device (maintaining environment<br>for ART procedures). FDA has<br>cleared humidified and non-<br>humidified ART incubators.<br>The subject device indications for<br>use also lists the ability of the<br>device to measure pH of<br>bicarbonate-based media within<br>the incubator. The pH<br>information is given to laboratory<br>personnel to provide additional<br>information on the<br>environmental conditions within<br>the incubator, as pH changes are<br>influenced by CO2 concentrations.<br>Although the predicate device<br>does not include a pH<br>measurement function, the<br>overall intended use of the<br>subject and predicate devices is<br>the same (i.e., providing a<br>controlled environment for<br>gamete and embryo<br>development in vitro). |
| Attribute | Subject<br>G210 InviCell Plus with SignipHy™<br>pH Monitoring System | Predicate<br>BT37 Incubator | Discussion |
| General Design | Bench-top incubator with 10<br>incubation chambers and one<br>preparation chamber | Bench-top incubator with 2<br>incubation chambers | Different: The subject device has<br>more incubation chambers and<br>an additional chamber to<br>equilibrate plates/oil before use.<br>This difference does not raise<br>different questions of safety and<br>effectiveness (S&E). |
| Gas Supply<br>Type | Internally regulated mixture of<br>CO2, O2 and N2 | Premixed blend of 6% CO2, 5% O2,<br>89% N2 | Different: The subject and<br>predicate devices use different<br>mechanisms to control gas<br>concentrations in incubator<br>chambers. This difference does<br>not raise different S&E questions. |
| Gas Supply<br>Pressure | 0.5-1.0 bar<br>(50 – 100 kPa) | 150 kPa +/- 15 kPa | Different: The subject and<br>predicate devices have different<br>gas supply pressures based on<br>the different gas supply/control<br>methods. This difference does<br>not raise different S&E questions. |
| Gas flow rate<br>capability | 1 l/h CO2 & 5 l/h N2 | 0 mL/min to 450 ml/min. Normal<br>bleed set to 15 ml/min and purge<br>at 180 ml/min for 3 minutes.<br>Flows fully adjustable in 1 ml/min<br>increments | Different: The subject and<br>predicate devices have different<br>gas flow rates based on the<br>different gas supply/control<br>methods. This difference does<br>not raise different S&E questions. |
| Gas flow rate<br>accuracy | CO2: 1.5 l/h<br>N2: 7 l/h | ±10% or ± 3% ml/min | Different: The flow rate accuracy<br>is different for the subject and<br>predicate devices that use<br>different gas supply types.<br>Differences in flow rate accuracy<br>do not raise different questions<br>of S&E. |
| Gas flow<br>pattern | Constant | Pulsed or non-pulsed (constant)<br>flow pattern. | Different: The subject device<br>does not include a pulsed flow<br>mode. This difference does not<br>raise different S&E questions. |
| Incubation<br>Chamber<br>Heating | Resistive heating foil - bottom,<br>sides and lid of each chamber | Not known | Different: The heating<br>mechanism used in the predicate<br>device is not known. However,<br>differences in heating methods<br>do not raise different S&E<br>questions. |
| Chamber<br>temperature<br>capability | Ambient to 42.9 °C | 5 °C above ambient to 40 °C | Similar |
| Chamber<br>temperature<br>accuracy | ± 0.2 °C calibration point | ± 0.2 °C calibration point | Same |
| Attribute | Subject<br>G210 InviCell Plus with SignipHy™<br>pH Monitoring System | Predicate<br>BT37 Incubator | Discussion |
| Humidification<br>system | No | Yes | Different: The subject device<br>does not include a humidification<br>feature. This difference does not<br>raise different S&E questions.<br>FDA has cleared both humidified<br>and non-humidified ART<br>incubators. |
| pH<br>Measurement<br>Function | Device includes a pH<br>measurement system that<br>includes a non-contact fiber optic<br>pH probe and sensor that holds a<br>representative media within one<br>incubation chamber, while a<br>separate display showing pH<br>information is on the exterior of<br>the incubator. | No | Different: The predicate device<br>does not include a pH<br>measurement system. The<br>inclusion of the pH system in one<br>incubation chamber provides<br>additional information on the<br>incubation environment. This<br>difference does not raise<br>different questions of S&E as<br>compared to the predicate (e.g.,<br>the ability to aid in maintaining<br>an appropriate culture<br>environment, materials do not<br>raise potential embryotoxicity<br>concerns, etc.). |
Table 1: Subject and Predicate Device Comparison
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for CooperSurgical. The logo is written in a blue sans-serif font. The word "Cooper" is connected by a line underneath the word.
Healthy women, babies, and families™
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for CooperSurgical. The logo is written in a blue sans-serif font. The word "Cooper" is connected to the word "Surgical".
Healthy women h
As shown in the table above, the subject and predicate devices have different indications for use statements; however, the intended uses of the predicate and subject devices are the same (i.e., providing a controlled environment for gamete and embryo development in vitro during assisted reproduction technology procedures).
In regards to technological characteristics, the subject and predicate devices have similarities (e.g., more than one incubation chamber, chamber temperature accuracy, etc.); however, as noted in the table above, technological differences were identified between the subject and predicate devices (e.g., pH monitoring system, gas supply, gas management, total number of incubation chambers, humidification of chambers, etc.). As stated in the table, the technological differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
#### Non-Clinical Performance Testing
The following performance data were provided in support of the substantial equivalence determination:
{8}------------------------------------------------
Healthy women, babies, and families™
- Sterilization Validation of SignipHy™ sv2 Sensors: ●
- o ISO 11135:2014 Sterilization of health-care products -- Ethylene oxide --Requirements for the development, validation and routine control of a sterilization process for medical devices
- AAMI TIR28:2016 Product Adoption and Process Equivalence for Ethylene Oxide o Sterilization
- Residual testing was not conducted, as the sensor is non-patient and nono gamete/embryo contacting.
- Shelf-Life of SignipHy™ sv2 Sensors: ●
- o Package integrity testing at the end of shelf-life in real-time aged samples after shipping/distribution:
- Visual Inspection per ASTM F1886/F1886M-16 .
- . Seal Strength per ASTM F88/F88M-15
- I Dye penetration per ASTM F1929-15
- o SignipHy™ sv2 Sensor Performance at the end of shelf-life in real-time aged samples after shipping/distribution:
- Comparison of subject device pH measurements to gold standard (blood gas analyzer) at a single time point.
- I Assessment of pH under worst-case conditions (samples collected at one minute intervals for three days) as compared to gold standard (blood gas analyzer).
- Reprocessing: Cleaning and disinfection validation testing conducted in accordance with ● the FDA guidance document, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
- Electrical Safety, Electromagnetic Compatibility (EMC), and Wireless Technology:
- o Electrical Safety Testing:
- IEC 61010 2-010: 2014 (Third Edition)
- . IEC 61010-2-101: 2015 (Second Edition)
- o EMC Testing:
- = IEC 61326-1:2012
- Wireless Technology: Information provided in accordance with the FDA o guidance document, "Radio Frequency Wireless Technology in Medical Devices.
- Software and Cybersecurity: ●
- o Software documentation in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" to support device software with a minor level of concern.
- Cybersecurity information in accordance with FDA guidance document "Content o of Premarket Submissions for Management of Cybersecurity in Medical Devices."
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for CooperSurgical. The logo is in blue and features the company name in a modern, sans-serif font. Below the company name is the tagline "Healthy women, babies, and families" also in blue.
- pH Monitoring Validation Testing: ●
- o The submission included testing to support the ability of the device to measure pH in bicarbonate-based ART media. Testing showed that the device met all specifications for each parameter assessed below:
- I Linearity
- l Stability
- Precision
- Accuracy
- Interfering substances
- Light Safety: Light energy exposure testing was performed to assess the potential for light toxicity associated with pH measurements taken under worst-case exposure conditions (samples taken every minute – total light exposure of 8.8 seconds/day). The results showed that the light wavelengths and worst-case light energy exposure duration did not raise safety concerns as compared to conventional imaging using standard laboratory ART imaging procedures.
- Mouse Embryo Assay (MEA): Testing was performed to assess the impact of the device ● (SignipHy TrakPod and SignipHy sv2 Sensor) when in operation under worst-case conditions (samples taken every minute for 96h). The results of testing met the acceptance specification of "1-Cell MEA: ≥90% blastocysts at 96h."
- Incubator Performance Testing: The incubator component of this device is 510(k) exempt and remains unchanged from that currently marketed. The additional pH monitoring system is separate from the incubator and does not impact its function. Therefore, performance testing on the incubator was not requested.
#### Conclusion
The results of the testing described above demonstrate that the subject G210 InviCell Plus with SignipHy™ pH Monitoring System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.