K191014 · Brainlab AG · LLZ · Jan 23, 2020 · Radiology
Device Facts
Record ID
K191014
Device Name
Elements Viewer
Applicant
Brainlab AG
Product Code
LLZ · Radiology
Decision Date
Jan 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Viewer is a software device for display of medical images and other healthcare data. It includes functions for image review, image manipulation, basic measurements and 3D visualization (Multiplanar reconstructions and 3D volume rendering). It is not intended for primary image diagnosis or the review of mammographic images.
Device Story
Viewer is a DICOM image viewing software for desktop, touch screen, and mobile devices, with added support for Magic Leap Mixed Reality glasses. It inputs DICOM data; transforms inputs via image processing for display (axial, coronal, sagittal, 3D volume rendering); and outputs visual representations for clinician review. Users interact via mouse, keyboard, touch, or pointing devices. The device enables image manipulation (scroll, zoom, pan, windowing), basic measurements (distance, angle), and annotation. In Mixed Reality, 2D views are displayed on a virtual pane, and 3D views are rendered as virtual objects in the real world. It assists clinicians in reviewing medical data; it does not perform primary diagnosis. Benefits include enhanced visualization and interaction with medical imaging data.
Clinical Evidence
Bench testing only. Validation included usability testing to ensure safe and effective user interface operation. Verification included interactive testing of the user interface, DICOM compatibility, and integration testing across platforms. Automated unit tests (Cucumber) were used for new functionalities. All tests passed acceptance criteria.
Technological Characteristics
Software-based DICOM viewer. Supports desktop, touch screen, and Mixed Reality (Magic Leap) platforms. Features include MPR, 3D volume rendering, distance/angle measurements, and annotation. Connectivity via standard DICOM protocols. No patient registration functionality. Software-based image processing.
Indications for Use
Indicated for the display of medical images and other healthcare data for review, manipulation, measurement, and 3D visualization. Not for primary diagnosis or mammographic image review.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
January 23, 2020
Brainlab AG % Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str. 9 81829 Munich, De GERMANY
Re: K191014
Trade/Device Name: Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: December 16, 2019 Received: December 26, 2019
Dear Chiara Cunico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191014
Device Name Viewer
Indications for Use (Describe)
Viewer is a software device for display of medical images and other healthcare data. It includes functions for image review, image manipulation, basic measurements and 3D visualization (Multiplanar reconstructions and 3D volume rendering). It is not intended for primary image diagnosis or the review of mammographic images.
Type of Use (Select one or both, as applicable)
| <div><span style="font-size:100%;"></span>Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|----------------------------------------------------------------------------------------------|
| <div>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (K) SUMMARY VIEWER
# IN ACCORDANCE WITH REQUIREMENTS OF 21 CFR PART 807.92
| General Information | | | |
|----------------------|---------------------------------------------------------------|--|--|
| Manufacturer | Brainlab AG, Olof-Palme-Straße 9, 81829 Munich, Germany | | |
| Establishment Reg. # | 8043933 | | |
| Device name: | Viewer | | |
| Trade name: | Elements Viewer | | |
| Common name | DICOM Viewer | | |
| Classification name: | Picture archiving and communications system (21 CFR 892.2050) | | |
| Review Panel | Radiology | | |
| Product Code: | LLZ | | |
| Device Class: | Class II | | |
| Predicate Device: | K153653 DICOM Viewer | | |
| Reference Device | K172418 OpenSight | | |
| Date of preparation | 23 September 2019 | | |
| Contact Information | |
|-----------------------------------|----------------------------------------|
| Primary contact person | Alternate contact person |
| Chiara Cunico | Regulatory Affairs Brainlab |
| Manager RA | phone: +49 89 99 15 68 0 |
| phone: +49 89 99 15 68 1738 | fax: +49 89 99 15 68 5033 |
| Email: Chiara.cunico@brainlab.com | Email: regulatory.affairs@brainlab.com |
#### Intended use:
Viewer is a software device for display of medical images and other healthcare data. It includes functions for image review, image manipulation, basic measurements and 3D visualization (Multiplanar reconstructions and 3D volume rendering). It is not intended for primary image diagnosis or the review of mammographic images.
{4}------------------------------------------------
#### Device description
Viewer is a software for viewing of DICOM data. The device provides basic measurement functionality for distances and anqles.
These are the operating principles:
- On desktop PCs the interaction with the software is mainly performed with mouse and/or keyboard. -
- On touch screen PCs and on mobile devices the software is mainly used with a touch screen interface. -
- -On Mixed Reality qlasses the interaction is performed with a dedicated pointing device.
The subject device provides or integrates the following frequently used functions:
- -Select medical images and other healthcare data to be displayed
- -Select views (e.g. axial, coronal & sagittal reconstruction views and 3D volume rendering views)
- Change view layout (e.g. maximize / minimize views, close / open / reorder views) -
- Manipulate views (e.g. scroll, zoom, pan, change windowing) -
- Perform measurements (e.g. distance or angle measurements) -
- -Place annotations at points of interests
#### Changes to Predicate Device
Convenience function have been added for view and object selection.
Magic Leap Mixed reality glasses have been added to the Subject Device as additional means of display. The Mixed Reality views are started from within the desktop version of Viewer. The view selection is done on the desktop, the views are cloned into the virtual space of Mixed Reality. The 2D views are shown on a 2D view pane in Mixed Reality. The 3D views are placed as virtual objects in to the real world.
| | Cleared device<br>feature/specification<br>DICOM Viewer (K153653) | Modified device<br>feature/specification<br>Viewer 5.0 | Substantial<br>Equivalent |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| HW Environment | Server based, for viewing via<br>HTML5 client or RDP on<br>individual client computer | Server and local installation, ,<br>for viewing via HTML5 client<br>or RDP, or local execution on<br>individual client computer | yes |
| Data inputs | DICOM | | yes |
| Viewing Data | • Skin<br>• Bone<br>• Skin/Bone<br>• Vessels<br>• Bone/Vessels<br>• Skin/Vessels<br>• Skin/Bone/Vessels<br>• Maximum Intensity Projection (MIP)<br>• Digital Reconstructed Radiograph (DRR) | | yes |
| | Cleared device<br>feature/specification<br>DICOM Viewer (K153653) | Modified device<br>feature/specification<br>Viewer 5.0 | Substantial<br>Equivalent |
| Measurement | Distance in slice sets<br>•<br>Diameter<br>•<br>Angle<br>• | Distance in slice sets<br>and X-ray images<br>•<br>Diameter<br>•<br>Angle (incl. open<br>angle) | yes |
| "Smart View" | Individual selectable visibility<br>of objects (tumor, fibers etc.) | Individual selectable visibility<br>of objects (tumor, fibers etc.)<br>Additional view: with a slider<br>it is possible to change<br>visibility of objects. | yes |
| View Layouts | Manual selection of<br>appropriate view. | Manual selection of<br>appropriate view.<br>Additional: Cranial, Spine<br>and Angio layouts load views<br>in the presence of certain<br>modalities in the Subject<br>Device. | yes |
| Probe's Eye View<br>(available in Mixed<br>Reality) | Align crop box to any desired<br>direction for the display of<br>MPR perpendicular and<br>parallel (inline) to the given<br>direction. | Align crop box to any desired<br>direction for the display of<br>MPR perpendicular and<br>parallel (inline) to the given<br>direction.<br>In Mixed Reality the user can<br>point in a desired direction and<br>view MPR perpendicular and<br>parallel (inline) to the given<br>direction. | yes |
#### Technological Characteristics of the Subject Device in comparison to the Predicate Device
{5}------------------------------------------------
{6}------------------------------------------------
Technological Characteristics of the Subject Device in comparison to the Reference Device:
| | Cleared device<br>feature/specification<br>OpenSight K172418 | Modified device<br>feature/specification<br>Viewer 5.0 | Substantial<br>Equivalent |
|------------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------|---------------------------------------------------------------------------------|
| Supported Mixed<br>Reality Device | Microsoft HoloLens | Magic Leap One | yes |
| Supported device<br>electrical medical<br>device according to<br>IEC 60601-1 | No | No | yes |
| Mixed Reality | 2D slices and MPRs<br>3D Views | 2D slices and MPRs<br>3D Views | yes |
| Registration to<br>patient | yes | no | No, this is not part of<br>the Subject Device.<br>Equivalence is not<br>claimed |
#### Verification summary:
Verification of communication and cybersecurity between Viewer and Magic Leap Mixed Reality glasses has been tested. All interactive testing of user interface have successfully passed.
## Validation summary:
The validation comprises usability tests which ensure that the user interface can be used safely and effectively. All tests were rated as successfully passed according to their acceptance criteria.
As conclusion, the nonclinical and clinical tests (discussed above) demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
| Test | Test Method Summary | Results |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| User interface | Interactive testing of user interface | All tests passed |
| DICOM compatibility | Interactive testing with companywide test data, which are<br>identical for consecutive version of the SW | All tests passed |
| Views | Interactive testing of user interface | All tests passed |
| Unit test /Automatic tests | Automated or semi-automated cucumber tests or unit<br>tests are written on the applicable level for new<br>functionalities of the Viewer in respect to previous<br>versions. Exiting tests have to pass. | All tests passed |
{7}------------------------------------------------
| Test | Test Method Summary | Results |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Integration test | Interactive testing on various platforms and combination<br>with other products following test protocols, combined<br>with explorative testing. | All tests passed |
| | The software is developed with daily builds, which are<br>explanatively tested. | |
#### Substantial equivalence
The comparison of the Viewer with the predicate device shows that the Viewer has similar functionality, intended use, technological characteristics, and typical users as the predicate device. Verification activities ensure that the design specifications are met and that the Viewer does not introduce new issues concerning safety and effectiveness. Hence the Viewer is substantial equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.