Scope ProTech

K191011 · Meditech Endoscopy, Ltd. · OCU · May 7, 2019 · Gastroenterology, Urology

Device Facts

Record IDK191011
Device NameScope ProTech
ApplicantMeditech Endoscopy, Ltd.
Product CodeOCU · Gastroenterology, Urology
Decision DateMay 7, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Scope ProTech is a single-use, sterile endoscopic tip protector that is intended to be used during the transport and storage of endoscopes for the protection of these delicate instruments. The Scope ProTech is intended for the protection of the distal tip and the bending rubber of endoscopes with a diameter of 2.7mm-8.0mm and 8.7mm-14.7mm. The Scope ProTech will aid in the protection of distal end, the lens and other delicate components from damage. It is not intended for use during sterilization.

Device Story

Single-use, sterile endoscopic tip protector; protects distal tip and bending rubber of endoscopes during transport and storage. Device slides over distal end of endoscope; open design allows aeration to prevent biofilm formation. Manufactured via injection molding using soft polypropylene. Used by healthcare professionals/technicians in clinical settings. Provides impact protection for delicate optics and components; includes tabs to identify device status and prevent re-use. Benefits include maintaining cleanliness of reprocessed endoscopes and preventing mechanical damage during handling.

Clinical Evidence

No clinical data. Bench testing only: operating performance, scope compatibility/dimension, retention performance, and sterility/shelf-life validation.

Technological Characteristics

Single-use, sterile (EtO) endoscopic tip protector. Material: injection-molded soft polypropylene. Form factor: two sizes covering 2.7mm-14.7mm diameter endoscopes. Mechanical protection principle via physical barrier. No electronic components, software, or energy sources.

Indications for Use

Indicated for protection of the distal tip and bending rubber of endoscopes (diameters 2.7mm-14.7mm) during transport and storage in healthcare facilities. Not for use during sterilization.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it. May 7, 2019 Meditech Endoscopy Ltd % Thomas Schorre Responsible Third-Party Official Accelerated Device Approval Services, LLC 6800 S.W. 40th Street. Ste. 444 Ludlum, FL 33155-3708 K191011 Re: Trade/Device Name: Scope ProTech Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCU Dated: April 15, 2019 Received: April 16, 2019 Dear Thomas Schorre: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191011 Device Name Scope ProTech #### Indications for Use (Describe) The Scope ProTech is a single-use, sterile endoscopic tip protector that is intended to be used during the transport and storage of endoscopes for the protection of these delicate instruments. The Scope ProTech is intended for the protection of the distal tip and the bending rubber of endoscopes with a diameter of 2.7mm-8.0mm and 8.7mm. The Scope ProTech will aid in the protection of distal end, the delicate components from damage. It is not intended for use during sterilization. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------|------------------------------------------------------------------------------| | <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for Meditech Endoscopy. The logo features a blue circle with a stylized white "E" inside it. To the right of the circle are the words "MEDITECH" in a darker blue font, with the word "ENDOSCOPY" underneath in a lighter blue font. # 510(k) Summary – Summary of Safety and Effectiveness Scope ProTech # 510(k) number: K191011 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. | Submitter's Identification: | Meditech Endoscopy Ltd | |----------------------------------|--------------------------------------------------------------------------------| | Address | Bowden House<br>Luckyn Lane<br>Basildon<br>Essex<br>SS14 3AX<br>United Kingdom | | Registered Establishment Number: | n/a | | Owner/Operator Number: | n/a | | Contact: | Peter Ramsey<br>Managing Director | | Telephone: | +44 1268 534035 | | Email: | peter@i-medical.net | | Date of Summary: | 29th March 2019 | | Device/Proprietary Name: | Scope ProTech | | Common Name: | Endoscopic Storage Cover | | Classification: | Endoscope and accessories. | | Type of Submission | Traditional 510(k) – New Device | | Product Code: | OCU | | Product Class: | II | | Regulation number: | 21 CFR § 876.1500 | | Panel | Gastroenterology/Urology | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logo for Meditech Endoscopy. The logo consists of a blue circle with a white stylized letter "E" inside it. To the right of the circle, the word "MEDITECH" is written in blue, and below it, the word "ENDOSCOPY" is written in a lighter shade of blue. The logo is clean and modern, with a focus on the company's name and area of expertise. Ref: K191011 | Substantial Equivalence: | | | | |------------------------------------------|------------|----------------------|---------------| | Manufacturer | Trade Name | Regulation<br>Number | 510(k) Number | | UNITED STATES<br>ENDOSCOPY GROUP,<br>INC | ENDOBOOT | 876.1500 | K951104 | ### Description The Scope ProTech is a single-use, sterile endoscopic tip protector which provides a simple, safe, and highly effective method of protecting the delicate optics of an endoscope, while allowing the tip to 'breath' and thus not allowing biofilm to form. The Scope ProTech is available in two models to fit a range of scopes: | Product Code | Product Description | Size of scope | |--------------|---------------------|---------------| | INOV8-028S | Scope ProTech Small | 2.7 to 8mm | | INOV8-028L | Scope ProTech Large | 8.7 to 14.7mm | Table 1: Scope ProTech™ sizes ## Indications for Use The Scope ProTech is a single-use, sterile endoscopic tip protector that is intended to be used during the transport and storage of endoscopes for the protection of these delicate instruments. The Scope ProTech is intended for the protection of the distal tip and the bending rubber of endoscopes with a diameter of 2.7mm-8.0mm and 8.7mm-14.7mm. The Scope ProTech will aid in the protection of distal end, the lens and other delicate components from damage. It is not intended for use during sterilization. ## Summary of the Determination of Substantial Equivalence & Performance Data The subject device is substantially equivalent to the predicate, K951104, the Endo-Boot and a comparison of the key characteristics is summarised in Table 2. | Characteristic | Meditech Endoscopy<br>Scope ProTech | US Endoscopy<br>Endo-Boot (K951104) | Equivalence | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Device Name | Scope ProTech | Endo-Boot™ | n/a | | Characteristic | Meditech Endoscopy<br>Scope ProTech | US Endoscopy<br>Endo-Boot (K951104) | Equivalence | | Indications for Use | The Scope ProTech is a<br>single-use, sterile<br>endoscopic tip protector<br>that is intended to be used<br>during the transport and<br>storage of endoscopes for<br>the protection of these<br>delicate instruments. The<br>Scope ProTech is intended<br>for the protection of the<br>distal tip and the bending<br>rubber of endoscopes with<br>a diameter of<br>2.7mm-8.0mm and<br>8.7mm-14.7mm. The Scope<br>ProTech will aid in the<br>protection of distal end, the<br>lens and other delicate<br>components from damage.<br>It is not intended for use<br>during sterilization. | The Endo-Boot™ is used as a<br>protective cover for the tip of<br>the endoscope or other<br>surgical instruments (i.e. rigid<br>laparoscope, laparoscopic<br>instruments) during storage<br>and transport. The Endo-<br>Boot™ will aid in protecting<br>the lens and other delicate<br>components from damage. | Substantially<br>Equivalent | | Intended User/Location | Healthcare professionals/<br>technicians and non-patient<br>contacting<br>Healthcare facility | Healthcare professionals/<br>technicians and non-patient<br>contacting<br>Healthcare facility | Equivalent | | Storage and transport | The Scope ProTech is<br>intended for use immediately<br>after processing in an AER<br>and is used during the<br>storage and transport to and<br>from the location of the<br>procedure | The Endo-Boot™ is intended<br>for use immediately after<br>processing in an AER and is<br>used during the storage and<br>transport to and from the<br>location of the procedure | Equivalent | | Sterile | Yes, EtO sterilised | Non-sterile | Substantially<br>Equivalent | | Re-use | The Scope ProTech is a sterile<br>single-use device and this is<br>stated on the packaging | The Endo-Boot™ is a single<br>use device and this is stated<br>on the packaging | Equivalent | | Scope Sizes | The Scope ProTech is<br>supplied in two sizes<br>intended to fit a range of<br>scopes for each size (mm):<br>Scope ProTech Small 2.7 - 8<br>Scope ProTech Large 8.8 - 14.7 | The Endo-Boot™ is supplied<br>in two sizes intended to fit a<br>range of scopes for each size:<br>Endo-Boot™ small 3.0<br>– 8.8 mm<br>Endo-Boot™ large 8.8<br>– 15 mm | Substantially<br>Equivalent | | Characteristic | Meditech Endoscopy<br>Scope ProTech | US Endoscopy<br>Endo-Boot (K951104) | Equivalence | | Materials | The Scope ProTech is<br>manufactured using injection<br>moulding using a soft<br>polypropylene | The Endo-Boot™ is<br>manufactured from a sponge | Substantially<br>Equivalent | | Instructions for use | Instructions for use included<br>that includes graphical<br>instructions, text and<br>relevant warnings and<br>cautions | Instructions for use included<br>that includes graphical<br>instructions, text and<br>relevant warnings and<br>cautions | Equivalent | | Use on an endoscope | Slides over the distal tip of an<br>endoscope | Slides over the distal tip of an<br>endoscope | Equivalent | | Distal tip protection | The Scope ProTech is<br>intended to protect the<br>entire distal end of the<br>endoscope, covering the tip<br>and delicate components | The Endo-Boot™ is intended<br>to be used on the distal end<br>however the placement of<br>the device should ensure the<br>distal tip of the endoscope is<br>uncovered | Substantially<br>Equivalent | | | | | | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the logo for Meditech Endoscopy. The logo consists of a blue circular icon with a stylized letter "E" inside. To the right of the icon, the words "MEDITECH" and "ENDOSCOPY" are written in a sans-serif font, with "MEDITECH" in a darker blue and "ENDOSCOPY" in a lighter shade. Ref: K191011 {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Meditech Endoscopy. The logo consists of a blue circle with a white stylized letter "E" inside it. To the right of the circle are the words "MEDITECH" in blue and "ENDOSCOPY" in a lighter shade of blue. The text is aligned to the right of the circle. Ref: K191011 Table 2: Comparison of characteristics between Scope ProTech™ and Endo-Boot™ ### Discussion The Scope ProTech and the predicate both have equivalent intended use, the user, location of use, how they are used (slide onto the distal tip and during storage/ transport of endoscopes), the scope sizes covered and their non-re-use. Both devices include a set of instructions for use that includes graphical and text information with warnings/cautions. The indications for use are substantially equivalent as the intended use is equivalent however the clarification that the Scope ProTech is sterile, single-use and should not be used during sterilisation are important additions but do not impact the safety or effectiveness. The main areas of difference are in the sterility aspect, with the Scope ProTech provided sterile to reduce bioburden and maintain cleanliness of the endoscope after reprocessing whereas the predicate is not provided sterile; the materials used with the Scope ProTech using an injection moulded soft polypropylene that provides impact protection whereas the predicate is manufacture from the thin sponge foam; the sizes of the scopes covered is lower for the Scope ProTech (2.75mm compared to 3mm) however the Endo-Boot™ can stretch further (max 15mm compared to 14.75mm) however both the subject and predicate device cover the range of endoscopes used in clinical practice and the size coverage is identified on the product and this does not impact safety or effectiveness; the distal tip protection with the open design of the Scope ProTech allowing for complete coverage of the distal tip (allowing it to aerate) whereas the predicate has to leave the distal tip uncovered; the clean/dirty identification as the Scope ProTech includes tabs that identify the status of the device and prevents re-use of a used device on a clean scope whereas the predicate does not have this identification. Issue 2 {7}------------------------------------------------ Image /page/7/Picture/1 description: The image contains the logo for Meditech Endoscopy. The logo consists of a blue stylized letter "E" on the left, followed by the words "MEDITECH" in a darker blue and "ENDOSCOPY" in a lighter blue on the right. The "E" is formed by three lines that create the shape of the letter within a circle. These differences ensure the endoscope cleanliness is not affected by the use of a Scope ProTech, provide additional protection of the reprocessed scope over the distal tip and the delicate components, prevent re-use and therefore the Scope ProTech is at least as safe and effective as the predicate. ### Non-Clinical Testing The Scope ProTech has been tested for mechanical performance across the range of scope sizes and for sterility and shelf-life requirements. The Scope ProTech™ is provided individually packaged and is provided sterile (ethylene oxide). The device does not have any patient contacting parts and has an equivalent intended use to the predicate and are used in a similar way. The following bench tests were performed on the device: - Operating performance - . Scope compatibility and dimension - Retention performance ● The testing performed demonstrated that the proposed device and predicate device are equivalent. ## Clinical Testing No clinical data was used in the submission. ### Summary The Scope ProTech is accompanied by instructions for use which demonstrate how the device is used safely and the labelling and instructions for use have been compared with the predicate and found to be similar for their purpose. The Scope ProTech has been compared to similar devices and the is substantially equivalent to the predicate for the parameters and the testing has demonstrated that the device does not introduce any additional risks when used in accordance with the instructions for use. ### Conclusion Based on the information presented in this submission, it is concluded that the Scope ProTech are equivalent to the Endo-Boot (K951104) with respect to intended use, design and technological characteristics. The Scope ProTech is as safe and effective as the predicate devices for the protection of the delicate parts of the endoscope during transport and storage. There are no new indications for use, therefore by following the FDA 510(k) "Substantial Equivalence" decision making process, https://www.fda.gov/downloads/MedicalDevices/ DeviceRegulationandGuidance/Guidanc eDocuments/UCM284443.pdf, it is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...