Magic Clear Aligners

K190948 · Voodoo Manufacturing, Inc. · NXC · May 9, 2019 · Dental

Device Facts

Record IDK190948
Device NameMagic Clear Aligners
ApplicantVoodoo Manufacturing, Inc.
Product CodeNXC · Dental
Decision DateMay 9, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Intended Use

Magic Clear Aligners are indicated for the alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.

Device Story

Magic Clear Aligners are custom-made, removable thermoplastic orthodontic appliances. Input consists of digital or traditional dental impressions provided by a dental professional. Specialized orthodontic CAD/CAM software (3Shape Ortho System) processes these inputs to design intermediate tooth states and treatment plans. The prescribing doctor reviews and approves the plan. The software generates 3D files used to produce physical models; Voodoo Manufacturing thermoforms plastic sheets over these models to create aligner trays. Trays are delivered to the dental professional, who manages the patient's treatment by providing trays in sequence. The patient wears the aligners to apply continuous gentle forces to reposition teeth. The device is used in a clinical setting under the supervision of a dental health professional. The aligners are removable by the patient. Treatment benefits include gradual tooth alignment to correct malocclusions.

Clinical Evidence

No clinical data included. Bench testing only. Manufacturing validation assessed dimensional accuracy of digital models, 3D printed molds, and thermoformed aligners using 3rd party inspection software. All translational measurements were within 0.15 mm of target values. Biocompatibility testing performed per ISO 10993-1 (Cytotoxicity, Sensitization, Irritation, Subacute/SubChronic Toxicity, Genotoxicity).

Technological Characteristics

Thermoplastic material; non-sterile; custom-fabricated via thermoforming over 3D-printed models. Software-driven treatment planning (3Shape Ortho System). Biocompatibility per ISO 10993-1. Mechanism of action: continuous gentle force application via sequential tray wear.

Indications for Use

Indicated for the alignment of teeth during orthodontic treatment of malocclusions in patients requiring tooth repositioning via continuous gentle forces.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size. May 9, 2019 Voodoo Manufacturing, Inc. % Mark Job Official Correspondent Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K190948 Trade/Device Name: Magic Clear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: April 9, 2019 Received: April 11, 2019 Dear Mark Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Statement 4. | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | | Form Approved: OMB No. 0910-012<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | |------------------------------------------------------------------------------------------------|---------|--------------------------------------------------------------------------------------------| | 510(k) Number (if known) | K190948 | | | Device Name | | | Indications for Use (Describe) Magic Clear Aligners are indicated for the alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|-----------------------------------------------| | <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6746 EF {3}------------------------------------------------ K190948 # 510(k) Summary Voodoo Manufacturing, Inc Magic Clear Aligners 4/1/2019 #### 5.1 ADMINISTRATIVE INFORMATION | Manufacturer Name | Voodoo Manufacturing, Inc | |-------------------|---------------------------| | | 361 Stagg Street #408 | | | Brooklyn, NY 112062 | | | USA | | Telephone: | +1 (646) 854-4173 | | Fax: | n/a | | Official Contact | Max Friefeld, CEO | |------------------|-------------------| | Email: | max@voodoomfg.com | #### 5.2 DEVICE NAME AND CLASSIFICATION | Trade/Proprietary Name: | Magic Clear Aligners | |-------------------------|----------------------| | Common Name: | Aligners, sequential | | Classification Name: | Orthodontic Plastic Bracket | |-----------------------------|-----------------------------| | Classification Regulations: | 21 CFR 872.5470 | | Device Class: | Class II | | Product Code: | NXC | | Classification Panel: | Dental Products Panel | | Reviewing Branch: | Dental Devices Branch | #### 5.3 PREDICATE DEVICE INFORMATION The devices within this submission are substantially equivalent in indications, intended use and design principles to the following predicate and reference devices: | 510(k) | Predicate Device Name | Company Name | |---------|-----------------------|-----------------| | K180241 | Twin Aligner | Ortho Caps GmbH | | 510(k) | Reference Device Name | Company Name | ClearCorrect System #### 5.4 DEVICE DESCRIPTION K113618 Magic Clear Aligners consist of a series of custom-made removable plastic orthodontic appliances which sequentially reposition teeth by way of continuous gentle force. ClearCorrect LLC A digital or traditional mold impression of the patient's teeth is provided by a dental health professional (e.g. orthodontist or dentist). From the digital data created of the patient's teeth, specialized orthodontic CAD/CAM software is used to develop treatment plans. Using the software, dental technicians design a series of intermediate models corresponding to each stage of treatment, gradually realigning the patient's teeth according to the dental health professional's prescription. {4}------------------------------------------------ The prescribing doctor reviews and approves the model scheme and treatment plan before the molds/models are produced. Once approved, the specialized orthodontic software is used to generate standard format 3D files which are used to physically produce each model/mold in the treatment plan for aligner fabrication. Voodoo Manufacturing produces the aligner trays by thermal forming a plastic sheet over each model in the treatment plan. The aligner trays are sent to the dental health professional who then delivers them to the patient, confirming fit and design. Over a period of time, additional trays are provided sequentially to the patient by the dental health professional to gradually move the teeth to the desired position. The dental health professional monitors treatment from the first aligner is delivered to when the final aligner is delivered and treatment complete. The aligners are held in place by pressure and can be removed by the patient at any time. The specialized orthodontic treatment planning software has a 510k clearance for the intended use under FDA Classification Product Code PNN, regulation 872.5470. The manufacturing process has been internally validated by Voodoo Manufacturing, Inc. The technology is identical to that used by the Predicate device, Ortho Caps Twin Aligner (K180241) and a number of other sequential aligner systems currently being legally marketed. #### 5.5 INDICATIONS FOR USE Magic Clear Aligners are indicated for the alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces. #### EQUIVALENCE TO MARKETED DEVICE 5.6 The Subject device is substantially equivalent to the predicate device with respect to Indications for Use and technological principles. The Comparison table below compares parameters and characteristics of the subject device and predicate/reference devices. | Parameter | Subject Device | Predicate Device | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Magic Clear Aligner | Ortho Caps<br>Twin Aligner<br>K180241 | | Regulation # | 21 CFR 872.5470 | 21 CFR 872.5470 | | Device Classification<br>Name | Orthodontic Plastic<br>Bracket | Orthodontic Plastic<br>Bracket | | Product Code | NXC | NXC | | Classification | Class II | Class II | | Indications for Use | Magic Clear Aligners are indicated for the<br>alignment of teeth during orthodontic treatment of<br>malocclusions by way of continuous gentle forces. | The Orthocaps TwinAligner® System is indicated for<br>the alignment of teeth during orthodontic treatment<br>of malocclusions by way of continuous gentle forces. | | Mode of action | Orthodontic movement occurs through continuous<br>gentle forces applied to the dentition as each tooth<br>follows the programmed displacement based on a<br>doctor's prescription. | Orthodontic movement occurs through continuous<br>gentle forces applied to the dentition as each tooth<br>follows the programmed displacement based on a<br>doctor's prescription. | | Method of use | Each preformed plastic tray is worn in sequence by<br>the patient as prescribed by the dental practitioner. | Each preformed plastic tray is worn in sequence by<br>the patient as prescribed by the dental practitioner. | | Function of the<br>software | Standard dental software for tooth alignment uses<br>digital scan (untreated state) to generate the image<br>of a final, provisional treated state and then<br>interprets a series of images that represent<br>intermediate teeth states. The dental practitioner<br>then reviews these images and has the option to<br>reject or request modifications to the set-up prior<br>to approving it for aligner fabrication. Once the | Standard dental software for tooth alignment uses<br>digital scan (untreated state) to generate the image<br>of a final, provisional treated state and then<br>interprets a series of images that represent<br>intermediate teeth states. The dental practitioner<br>then reviews these images and has the option to<br>reject or request modifications to the set-up prior to<br>approving it for aligner fabrication. Once the dental | # Predicate Device Comparison Table {5}------------------------------------------------ | Parameter | Subject Device | Predicate Device | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | | Magic Clear Aligner | Ortho Caps | | | | Twin Aligner | | | | K180241 | | | dental practitioner approves the treatment plan,<br>the software converts the files to produce the | practitioner approves the treatment plan, the<br>software converts the files to produce the series of | | | series of 3D models used to produce<br>thermoformed aligners. | 3D models used to produce thermoformed aligners. | | Material | Thermoplastic | Thermoplastic | | Material Properties | Demonstrates sufficient tensile strength, elasticity,<br>ductility, chemical resistance, and clarity for use as<br>a clear tray aligner. | Demonstrates sufficient tensile strength, elasticity,<br>ductility, chemical resistance, and clarity for use as a<br>clear tray aligner. | | Design | Image: Clear aligner | Image: Clear aligner | | Biocompatible | Yes | Yes | | OTC or Rx | Rx | Rx | | Sterile | Non-sterile | Non-sterile | ## TECHNOLOGICAL CHARACTERISTICS 5.7 Orthodontic tooth movement occurs through forces applied to the appliance as each tooth follows the predetermined displacement based on a dental health professional's prescription. The Subject device mechanism of action is identical to the Predicate device and supports a determination of substantial equivalence. The Subject device use of Software is identical to the Predicate device and supports a determination of substantial equivalence. The Subject and Predicate devices are both fabricated of a non-sterile, biocompatible thermoplastic material which supports a determination of substantial equivalence. #### 5.8 PERFORMANCE DATA Due to the difficulty in evaluating this type of dental device in a laboratory environment, no direct performance bench testing of the aligners was performed. The use of thermoplastic materials for sequential aligners intended to treat malocclusions have been well documented in scientific literature regarding incremental tooth moving forces. Software used for treatment planning and creation of models/casts for Magic Clear Aligners is manufactured by 3Shape A/S under the name Ortho System. It has a 510(k) clearance (K180941) under the PNN product code and is indicated for the intended use. A manufacturing validation was performed to demonstrate the dimensional accuracy of the manufacturing process for Magic Clear Aligners. Three critical aspects of the manufacturing process were assessed for accuracy: digital dentition models from treatment planning, 3D printed molds, and the final thermoformed aligners. Independent 3rd party reverse engineering/inspection software was used to perform point-to-point and critical displacement measurements. All translational measurements were within 0.15 mm of the target input value, the predefined tolerance of the manufacturing process. There were no statistical differences in the intended and measured values observed from any of the groups. This test has met the pre-established acceptance criteria to demonstrate dimensional accuracy. {6}------------------------------------------------ ## 5.9 CLINICAL TESTING The performance of sequential aligners in the clinical environment has been well established since the first such devices were cleared by the FDA in 1998 under product code NXC. There no clinical testing is required to support Magic Clear Aligners, as the Indications for Use is equivalent to the Predicate device, which also was not subject to clinical testing. No clinical data is included in this submission. #### 5.10 BIOLOGICAL TESTING Biocompatibility testing for the aligner material, the only patient contacting material, was conducted in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The results of the tests satisfy the requirements of the study protocols and comply with ISO 10993-1 for the intended use. The following biological tests were performed: | Biological Endpoint | Relevant Standard | |------------------------------|-------------------| | Cytotoxicity | ISO 10993-5:2009 | | Sensitization | ISO 10993-10:2010 | | Irritation | ISO 10993-10:2010 | | Subacute/SubChronic Toxicity | ISO 10993-11:2006 | | Genotoxicity | ISO 10993-3:2014 | #### 5.11 CONCLUSION Indications for Use statement for the Subject and Predicate devices is identical. The Technological Characteristics, Materials, Prescription Use and Non-sterile status of the Subject device is identical to the Predicate device. The use of Software to produce the Subject and Predicate devices is identical. Overall, the Magic Clear Aligners are substantially equivalent to the Predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%