Lawson Retrograde Nephrostomy Wire Puncture Set
K190903 · Cook Incorporated · LJE · Dec 16, 2019 · GU
Device Facts
| Record ID | K190903 |
| Device Name | Lawson Retrograde Nephrostomy Wire Puncture Set |
| Applicant | Cook Incorporated |
| Product Code | LJE · GU |
| Decision Date | Dec 16, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
Indications for Use
Used to gain precise percutaneous access to the kidney by means of controlled fine wire puncture from within the collecting system for patients aged 12 years and older.
Device Story
Device facilitates percutaneous nephrostomy access. Components include tip-deflecting wire guide, catheter, sheathed puncture wire, coaxial needle assembly, and working wire guide. Procedure involves inserting tip-deflecting wire guide into kidney collecting system to direct catheter; exchanging deflecting wire for sheathed puncture wire; advancing puncture wire from catheter through kidney to patient's flank/skin; advancing coaxial needle assembly over puncture wire into renal collecting system; removing puncture wire and inner needle; and placing working wire guide through outer needle cannula. Used by clinicians in clinical settings. Output is physical access to renal collecting system, enabling subsequent urological interventions. Benefits include precise, controlled access to kidney.
Clinical Evidence
Bench testing only. Testing included biocompatibility, dimensional analysis, tensile strength (hub-to-shaft and tip-to-shaft bonds), radiopacity, shelf-life (accelerated aging to three-year real-time equivalency), and packaging/sterility/stability testing. All predetermined acceptance criteria were met.
Technological Characteristics
Components: stainless steel tip-deflecting wire guide (0.045 in OD), braided nylon catheter (7 Fr, 82.5 cm), radiopaque TFE sheath (3 Fr, 85 cm), stainless steel puncture wire (0.43 mm OD, 145 cm), 22G/18G coaxial needle assembly, and stainless steel working wire guide (0.97 mm OD, 145 cm). Manual operation. Sterile, single-use. No software or energy source.
Indications for Use
Indicated for patients aged 12 years and older requiring percutaneous access to the kidney via controlled fine wire puncture from within the collecting system.
Predicate Devices
- Percutaneous Retrograde Nephrostomy Puncture Set (K833762)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, reflecting the organization's role in regulating food and drugs.
December 16, 2019
Cook Incorporated Minjin Choi Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K190903
Trade/Device Name: Lawson Retrograde Nephrostomy Wire Puncture Set Regulation Number: 21 CFR Regulation Name: None Regulatory Class: Unclassified Product Code: LJE Dated: November 15, 2019 Received: November 18, 2019
Dear Minjin Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Jessica Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
A-3-1
510(k) Number (if known)
K190903
Device Name
Lawson Retrograde Nephrostomy Wire Puncture Set
Indications for Use (Describe)
Used to gain precise percutaneous access to the kidney by means of controlled fine wire puncture from within the collecting system for patients aged 12 years and older.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> |
|-----------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters in the top portion of the logo. Below the word "COOK", the word "MEDICAL" is written in smaller, white, sans-serif letters.
K 190903
## 2.0 510(k) Summary
# Lawson Retrograde Nephrostomy Wire Puncture Set 21 CFR §807.92 Date Prepared: November 15, 2019
# Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|--------------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Primary Contact: | Minjin Choi |
| Secondary Contact: | Andrew Breidenbach, Ph.D., RAC |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Email: | RegSubmissions@CookMedical.com |
| Primary Contact Phone: | (812) 335-3575 x104901 |
| Secondary Contact Phone: | (812) 335-3575 x105147 |
| Contact Fax: | (812) 332-0281 |
# Device Information:
| Trade Name: | Lawson Retrograde Nephrostomy Wire Puncture Set |
|---------------------------------|-------------------------------------------------|
| Common Name: | Nephrostomy Catheter |
| Classification Name and Number: | None |
| Primary Product Code: | LJE (Catheter, Nephrostomy) |
| Regulatory Class: | Unclassified |
| Classification Panel: | Gastroenterology/Urology |
# Predicate Device:
The predicate device is the Percutaneous Retrograde Nephrostomy Puncture Set (K833762), cleared for market by the FDA on February 10, 1984.
### 1
#### Company Confidential
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif letters. Below the word "COOK" is a red banner with the word "MEDICAL" in white, sans-serif letters. The background of the logo is red.
Page 2
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
HONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
#### Device Description:
The Lawson Retrograde Nephrostomy Wire Puncture Set is used to gain percutaneous access to the kidney by means of a controlled fine wire puncture from within the collecting system.
The Lawson Retrograde Nephrostomy Wire Puncture Set is composed of the tipdeflecting wire guide, catheter, sheathed puncture wire, coaxial needle assembly (inner and outer needle cannulae) and working wire guide. The tip-deflecting wire guide is used to direct the catheter into the desired calyx in a retrograde manner using standard access techniques, followed by exchange of the deflecting wire with the sheathed puncture wire. To initiate percutaneous nephrostomy access, the puncture wire is unsheathed and advanced from the catheter through the kidney towards the patient's flank and through the skin. The coaxial needle assembly is then advanced over the puncture wire and into the renal collecting system, and then the puncture wire and inner needle cannula are removed to allow a working wire guide to be placed through the outer needle cannula.
- . The tip-deflecting wire guide has a three-ring handle which twist-locks with the catheter. The tip-deflecting wire guide is made from stainless steel and is available in an outer diameter of 0.045 in.
- . The catheter is available in an outer diameter of 7 Fr with a length of 82.5 cm and made of braided nylon tubing. The sheathed puncture wire can interlock with the proximal portion of the catheter.
- The sheath is made of radiopaque TFE tubing which extends 1 cm from the . catheter when locked together. The sheath is available in an outer diameter of 3 Fr with a length of 85 cm.
- . The puncture wire is made of stainless steel with a plastic pin vise handle. The puncture wire has an outer diameter of 0.43 mm with a length of 145 cm.
- . The needle set has a 22 gage inner needle that is 22.5 cm in length with the outer needle being 18 gage and 14 cm in length.
- . The working wire guide is made of stainless steel and is 0.97 mm in diameter and 145 cm in length.
{5}------------------------------------------------
The set will be supplied sterile and is intended for one-time use. The set is packaged in a peel-open pouch with a three-year shelf life.
### Indications for Use:
The Lawson Retrograde Nephrostomy Wire Puncture Set is used to gain precise percutaneous access to the kidney by means of controlled fine wire puncture from within the collecting system for patients aged 12 years and older.
#### Comparison to Predicate Devices:
The Lawson Retrograde Nephrostomy Wire Puncture Set and the predicate device, Percutaneous Retrograde Nephrostomy Puncture Set (K833762), are substantially equivalent in that they have the same intended use. Furthermore, the devices have nearly identical methods of operation, design, and dimensions. The modifications from the predicate device includes:
- . Indications for Use
- . Tip-Deflecting Wire (Dimension)
- 트 Catheter (Dimension and Materials)
- I Sheath (Materials)
Differences between the characteristics of the subject device and the predicate device are supported by testing.
## Performance Data:
The following testing was performed in order to demonstrate that the subject device. Lawson Retrograde Nephrostomy Wire Puncture Set, met applicable design requirements.
- Biocompatibility ●
- Compatibility ●
- Dimensional ●
- Tensile Strength
- o Hub-to-Shaft Bond of Catheter, Sheath, Stylet, and Needles
{6}------------------------------------------------
Image /page/6/Picture/14 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The word "MEDICAL" is smaller than the word "COOK".
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
- Tip-to-Shaft Bond of Catheter o
- Distal Tip of Deflecting Wire O
- Shaft of Puncture Wire O
- Radiopacity .
- Shelf Life following Accelerated Aging to Three-year Real-Time Equivalency .
- Packaging, Sterility and Stability .
# Conclusion:
All predetermined acceptance criteria of the testing were met. Therefore, the results of these tests support a conclusion that the Lawson Retrograde Nephrostomy Wire Puncture Set will perform as intended and support a determination of substantial equivalence to the predicate device.