Leksell Vantage Stereotactic System

K190887 · Elekta Instrument AB · HAW · Aug 1, 2019 · Neurology

Device Facts

Record IDK190887
Device NameLeksell Vantage Stereotactic System
ApplicantElekta Instrument AB
Product CodeHAW · Neurology
Decision DateAug 1, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.

Device Story

Leksell Vantage Stereotactic System provides coordinate referencing and skull fixation for neurosurgical procedures. System includes head frame, fixation pins, stereotactic arc, and accessories. Operates on established center-of-arc principle (190 mm radius); uses Cartesian (x,y,z) coordinates to triangulate brain targets. Input includes patient anatomical images (CT/MR) with fiducial markers. Used in operating rooms by neurosurgeons to guide interventional instruments (biopsy needles, cannulas, electrodes) to specific intracranial targets. Output is physical guidance of instruments at precise trajectory angles. Benefits include accurate, minimally invasive access to brain targets for stimulation, recording, lesioning, or drug delivery.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated through bench testing, including system/subsystem verification, accuracy testing (total system, mechanical, and handling), MR safety compliance (ASTM F2052, F2119, F2182, F2213, F2503), and design/usability validation by qualified personnel.

Technological Characteristics

Materials: PEEK with Aluminum tip (invasive components), Glass fiber reinforced Epoxy (head frame), Aluminum (arc). Sensing/Actuation: Mechanical coordinate referencing and instrument guidance. Energy: None (mechanical). Connectivity: None. Sterilization: EtO (ISO 11135:2014, SAL 10^-6). Form factor: Head frame with semi-circular arc system.

Indications for Use

Indicated for patients requiring stereotactic neurosurgical procedures, including deep brain stimulation, lesioning, biopsies, targeted injections, aspirations, and minimally invasive tumor treatments, where target localization and head fixation are necessary.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 1, 2019 Elekta Instrument AB Alf Laurell Regulatory Affairs Engineer Kungstensgatan 18 Stockholm, 10393 SE Re: K190887 Trade/Device Name: Leksell Vantage Stereotactic System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: July 1, 2019 Received: July 3, 2019 Dear Alf Laurell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K190887 ### Device Name Leksell® Vantage™ Stereotactic System ### Indications for Use (Describe) The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. Type of Use (Select one or both, as applicable) | <span></span> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> <path d="M2 2 L13 13 M13 2 L2 13" stroke="#000000" stroke-width="1"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span></span> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Dokumentnamn/Name of document Traditional 510(k) | Issuer | Ref nr/Dok nr/Ref no/Doc no | Edition | |--------------------------------------------------------|-----------------------------|---------| | Alf Laurell | -- | -- | | Regarding | Directory | | | CT accessories for Leksell Vantage Stereotactic System | -- | | # 510(k) Summary As Required by 21 CFR 807.92(c) 510 (k) Summary #### 1. Subscribers Name & Address Elekta Instrument AB Kungstensgatan 18, P.O. Box 7593 SE-103 93 Stockholm, Sweden Tel: (011) 46 8 587 254 00 Fax: (011) 46 8 587 255 00 Official Correspondent: Mats Premfors, QA Manager Date summary prepared: 2019-07-27 #### 2. Trade Name Leksell® Vantage™ Stereotactic System #### 3. Common Name Stereotaxic instrument #### Device Classification 4. | Trade Name | Product Code | Class | Regulation Number | |---------------------------------------|--------------|-------|-------------------| | Leksell® Vantage™ Stereotactic System | HAW | II | 21 CFR 882.4560 | #### Classification Panel 5. Neurology #### Predicate Device Identification 6. | Legally marketed devices to which equivalence is being claimed | 510(k) # | |----------------------------------------------------------------|----------| | Leksell Stereotactic System | K080355 | | Leksell® Vantage™ Stereotactic System (Primary) | K171123 | #### 7. Other relevant submissions N/A, no other relevant submissions #### 8. Device Description The Leksell® Vantage™ Stereotactic System is a device used for minimally invasive neurosurgical procedures. It enables coordinate referencing and fixation of the patient's skull {4}------------------------------------------------ | Dokumentnamn/Name of document | Traditional 510(k) | |-------------------------------|--------------------| |-------------------------------|--------------------| Traditional 510(k) | Issuer | Alf Laurell | |-----------------------------|--------------------------------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Edition | -- | | Regarding | CT accessories for Leksell Vantage Stereotactic System | | Directory | -- | and brain during image acquisition and treatment. The coordinate referencing enables target localization and accurate stereotactic treatment of brain targets; for example for aspirations and biopsies, electrode placements, injections of cells & drugs, heamatoma evacuations as well as lesioning using a number of methodologies. The system consists of a head frame that is fixated to the patient skull by minimally invasive disposable fixation pins and a number of accessories for frame application and imaging as well as a stereotactic arc and operating room accessories for the sterile surgical procedure. The system is based on the established Leksell center-of-arc principle, which allows guidance of compatible interventional instruments at any trajectory angle to the brain target. Compatible interventional instruments are based on the Leksell principle of an active working length of 190 mm (as the arc radius) and use the correct mechanical interfacing of the Arc instrument stop and guide holders. Various instruments, such as biopsy needles, cannulas, and electrodes can be used with the system to perform biopsies, hematoma evacuation and injections of radioactive nuclides and cytostatic agents, puncture of cysts, positioning electrodes for stimulation/ recording or lesioning and stereotactic endoscope guidance. #### 9. Summary of clinical testinq Clinical testing was deemed not required to demonstrate substantial equivalence with the predicate devices. #### 10. Summary of Non Clinical and Performance testing The verification activities have covered requirements at system and subsystem level to show that the design output meet the design input. Design and usability validation of the system has been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs. The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety (FRS) was effective. Results from verification and validation testing demonstrates that conformance to applicable technical requirement specifications and user needs have been met showing that the system is safe to use and is substantially equivalent to the predicate devices. Testing is summarized in the table below. {5}------------------------------------------------ Dokumentnamn/Name of document Traditional 510(k) | Issuer | Alf Laurell | |-----------------------------|--------------------------------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Edition | -- | | Regarding | CT accessories for Leksell Vantage Stereotactic System | | Directory | -- | | Test | Test Method Summary | Results | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sterilization | Sterility validation (EtO) according to<br>ISO 11135:2014 was performed to<br>demonstrate that the device have a<br>sterility assumed level (SAL) of 10-6.<br><br>The tests were performed on the<br>predicate device Leksell® Vantage™<br>Stereotactic System (K171123). | Results met pre-<br>established acceptance<br>criteria, demonstrating<br>equivalence to the<br>predicate device in terms<br>of sterility. | | MR testing | The device was subjected to<br>compliance testing to applicable<br>consensus safety standards, such as:<br>ASTM F2052; ASTM F2119; ASTM<br>F2182, ASTM F2213 and ASTM F2503,<br>and FDA guidance regarding safety and<br>compatibility in the MR environment. | The result demonstrates<br>that the device is<br>compliant to the FDA<br>guidance and applicable<br>consensus standards. This<br>compliance demonstrates<br>substantial equivalence. | | Accuracy | Accuracy was tested on different<br>levels of the subject device:<br>- Total system accuracy from<br>imaging to treatment.<br>- Total mechanical accuracy of<br>the system<br>- Mechanical accuracy of each<br>component<br>- Handling accuracy of various<br>steps like repeated mounting<br>of parts to each other and<br>scale setting accuracy.<br>- Displacement from load of<br>head when fixed in the frame. | All test cases were passed,<br>demonstrating that the<br>overall system accuracy of<br>the Vantage system when<br>used with the CT<br>accessories is substantial<br>equivalent to the predicate<br>device. | | Design and Usability<br>validation | Design and usability validation of the<br>system has been performed by trained<br>competent and professionally qualified<br>personnel to ensure that the product<br>fulfils the intended use and user<br>needs. | No unacceptable human<br>factor risks were identified.<br>Residual risks have<br>been weighted against the<br>benefit of the product and<br>are deemed acceptable<br>demonstrating that the<br>Leksell® Vantage™<br>Stereotactic System used | | | | with the CT accessories is<br>substantial equivalent to<br>the predicate device. | {6}------------------------------------------------ Dokumentnamn/Name of document Traditional 510/lz) Traditional 510(k) | Issuer | Alf Laurell | |-----------------------------|--------------------------------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Edition | -- | | Regarding | CT accessories for Leksell Vantage Stereotactic System | | Directory | -- | #### 11. Intended Use The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. ## 12. Technological Characteristics The CT accessories of Leksell® Vantage™ Stereotactic System has equivalent technological characteristics as its predicate device (K080355), for further details see the table below. #### Substantial Equivalence 13. The intended use, the basic functionality and the fundamental technological characteristics of the device Leksell® Vantage™ Stereotactic System remain unchanged with the introduction of the CT accessories and the Leksell Vantage Head Frame Holder. Leksell® Vantage™ Stereotactic as compared to the predicate devices has: - Same mechanical accuracy - . Same coordinate system - Same Center of Arc Principle - Intended for localizing the same intracranial anatomical targets - Used for the same clinical interventions and/or treatments - Intended for use by the same health care professionals In addition, the risk evaluation regarding endotoxin testing in routine production for product FirmFix™ 1 -5 has been updated. The conclusion from verification and validation is that the Leksell® Vantage™ Stereotactic System is substantial equivalent to its predicate devices. The table below shows a comparison between the Leksell® Vantage™ Stereotactic System and the predicate devices. {7}------------------------------------------------ Dokumentnamn/Name of document Traditional 510(k) | Issuer | Alf Laurell | |-----------------------------|--------------------------------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Edition | -- | | Regarding | CT accessories for Leksell Vantage Stereotactic System | | Directory | -- | | | Subject Device | Predicate Device (K080355) | Predicate Device (K171123) | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Leksell® Vantage™ Stereotactic System | Leksell Stereotactic System® (K080355) | Leksell® Vantage™ Stereotactic System (K171123) | | Regulation Number | 21 CFR 882.4560 | 21 CFR 882.4560 | 21 CFR 882.4560 | | Product Code | HAW | HAW | HAW | | Product Class | II | II | II | | Description | Head Frame and semi- circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | Head Frame and semi- circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | Head Frame and semi- circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | | Technical Features | Coordinate System: Leksell Coordinate System using Cartesian coordinates | Coordinate System: Leksell Coordinate System using Cartesian coordinates | Coordinate System: Leksell Coordinate System using Cartesian coordinates | | | Fixation: Head Frame applied to patient skull with four fixation pins. | Fixation: Head Frame applied to patient skull with four fixation pins. | Fixation: Head Frame applied to patient skull with four fixation pins. | | | Fixation Pins: Single use Fixation Pins and Inserts that secures the Vantage head frame to the patient skull using Leksell Vantage Keys | Fixation Pins: Titanium Fixation Pins, Reusable Fixation Pins and Disposable Inserts that secures the Leksell Coordinate Frame G to the patient skull using Instrument Screw Drivers | Fixation Pins: Single use Fixation Pins and Inserts that secures the Vantage head frame to the patient skull using Leksell Vantage Keys | | Imaging modality | CT and MR | CT and MR | MR | | Technical characteristics | CT accessories: The Leksell Vantage CT Fiducial Box is used for referencing of the stereotactic coordinate system on the patient's anatomical CT images. | CT accessories: The Open CT Indicator Box is used for referencing of the stereotactic coordinate system on the patient's anatomical CT images. | CT accessories: N/A | | | Leksell® Vantage™ Head Frame interface: | Leksell® Vantage™ Head Frame interface: | Leksell® Vantage™ Head Frame interface: | | | Subject Device | Predicate Device<br>(K080355) | Predicate Device<br>(K171123) | | | Leksell® Vantage™ Frame<br>Holder is a modified OR<br>adapter. In addition to<br>holding the Leksell®<br>Vantage™ Head Frame in<br>the operating room, the<br>Frame Holder is also<br>interfacing the CT Fiducial<br>box via one hole on each<br>side. | N/A | Leksell® Vantage™ OR<br>adapter | | Principles of Use | Center of the arc<br>principle using an arc<br>radius of 190 mm.<br>Cartesian coordinates<br>applied to the brain of<br>the patient by the use of<br>a head frame fixated to<br>the skull.<br>• comprises a headframe<br>and arc;<br>• is target centered<br>• uses Cartesian (x,y,z)<br>coordinates to<br>triangulate target<br>position in frame<br>• utilizes fiducial markers<br>that can be registered<br>to a stereotactic brain<br>atlas and/or<br>intracranial image;<br>• provides support and<br>accessories for surgical<br>instruments to the<br>target; | Center of the arc<br>principle using an arc<br>radius of 190 mm.<br>Cartesian coordinates<br>applied to the brain of<br>the patient by the use<br>of a head frame fixated<br>to the skull.<br>• comprises a<br>headframe and arc;<br>• is target centered<br>• uses Cartesian (x,y,z)<br>coordinates to<br>triangulate target<br>position in frame<br>• utilizes fiducial<br>markers that can be<br>registered to a<br>stereotactic brain<br>atlas and/or<br>intracranial image;<br>• provides support and<br>accessories for<br>surgical instruments<br>to the target; | Center of the arc<br>principle using an arc<br>radius of 190 mm.<br>Cartesian coordinates<br>applied to the brain of<br>the patient by the use<br>of a head frame fixated<br>to the skull.<br>• comprises a<br>headframe and arc;<br>• is target centered<br>• uses Cartesian (x,y,z)<br>coordinates to<br>triangulate target<br>position in frame<br>• utilizes fiducial<br>markers that can be<br>registered to a<br>stereotactic brain<br>atlas and/or<br>intracranial image;<br>• provides support and<br>accessories for<br>surgical instruments<br>to the target; | | Material | Invasive components:<br>PEEK with Aluminum tip. | Invasive components:<br>Aluminum or Titanium. | Invasive components:<br>PEEK with Aluminum tip. | | | Head Frame:<br>Glass fiber reinforced<br>Epoxy | Head Frame:<br>Aluminum (with glued<br>joints for insulation in<br>MRI) | Head Frame:<br>Glass fiber reinforced<br>Epoxy | | | Subject Device | Predicate Device<br>(K080355) | Predicate Device<br>(K171123) | | | Arc:<br>Aluminum | Arc:<br>Aluminum | Arc:<br>Aluminum | | Accuracy | Mechanical accuracy 0,9 mm | Mechanical accuracy 0,9 mm | Mechanical accuracy 0,9 mm | {8}------------------------------------------------ Dokumentnamn/Name of document Traditional 510(k) | Issuer | Alf Laurell | |-----------------------------|--------------------------------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Edition | -- | | Regarding | CT accessories for Leksell Vantage Stereotactic System | | Directory | -- | {9}------------------------------------------------ Dokumentnamn/Name of document Traditional 510/k) Traditional 510(k) | <i>Issuer</i> | Alf Laurell | |------------------------------------|--------------------------------------------------------| | <i>Ref nr/Dok nr/Ref no/Doc no</i> | -- | | <i>Edition</i> | -- | | <i>Regarding</i> | CT accessories for Leksell Vantage Stereotactic System | | <i>Directory</i> | -- | ### Conclusion: The conclusion is that the subject device is substantial equivalent to its predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%