SeeFactorCT3

K190856 · Epica International, Inc. · OAS · Sep 20, 2019 · Radiology

Device Facts

Record IDK190856
Device NameSeeFactorCT3
ApplicantEpica International, Inc.
Product CodeOAS · Radiology
Decision DateSep 20, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2

Indications for Use

SeeFactorCT3 is a cone beam computed tomography x-ray imaging system that acquires sequences of the head including the ear, nose and throat (ENT), dento-maxillofacial complex, teeth, mandible and jaw, temporo-mandibular joint (TMJ), other areas of human skull and neck with sections of upper cervical spine, and upper and lower extremities for use in diagnostic support. The device displays two and three dimensional images for each examined anatomical region. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.

Device Story

SeeFactorCT3 is a cone beam computed tomography (CBCT) system; utilizes High Definition Computed Tomography (HDCT) for high-resolution 3D volume acquisition; features 2D high-resolution scouting via dynamic scouting and single pulse modality. Operated by physicians, dentists, and x-ray technologists in clinical settings. Device captures X-ray projections of anatomical regions; transforms inputs into 2D and 3D diagnostic images. Output allows healthcare providers to visualize internal structures for diagnostic support; aids in clinical decision-making for ENT, dental, and orthopedic conditions. Benefits include high-resolution imaging, portability, and ability to be covered during surgical procedures.

Clinical Evidence

No clinical testing performed. Bench testing included imaging of living and deceased animals to demonstrate diagnostic image quality.

Technological Characteristics

CBCT X-ray imaging system; utilizes HDCT for 3D volume and 2D scouting. Features include dynamic scouting and single pulse modality. Portable form factor. No biocompatibility testing performed due to limited patient contact duration.

Indications for Use

Indicated for diagnostic imaging of the head (ENT, dento-maxillofacial, teeth, mandible, jaw, TMJ, skull, neck), upper cervical spine, and upper/lower extremities in patients requiring diagnostic support.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION". September 20, 2019 Epica International Inc. % Frank Pokrop Consultant (510K) Frank Pokrop 3577 Sand Court CARLSBAD, CA 92010 Re: K190856 Trade/Device Name: SeeFactorCT3 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: April 10, 2019 Received: April 15, 2019 Dear Frank Pokrop: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190856 Device Name SeeFactorCT3 #### Indications for Use (Describe) SeeFactorCT3 is a cone beam computed tomography x-ray imaging system that acquires of the head including the ear, nose and throat (ENT), dento-maxillofacial complex, teeth, mandible and jaw, temporo-mandibular joint (TMI), other areas of human skull and neck with sections of upper and lower extremities for use in diagnostic support. The device displays two and three dimensional images for each examined anatomical region. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) 7 Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # 5.1 Applicant/Submitter | Company Name | : | Epica International Inc. | |-----------------------------------|---|------------------------------------| | Establishment Registration Number | : | | | Phone Number | : | 1-425-941-7365 | | Company Street Address | : | 2753 Camino Capistrano. Suite A-10 | | Fax Number | : | (949) 238-6322 | | City | : | San Clemente | | State | : | CA | | Country | : | USA | | Zip Code | : | 92672 | ## 5.2 Contact Person | Full Name | : Frank Pokrop | |-----------|-------------------------| | Job Title | : Consultant (510(k)) | | Phone | : 1-442-273-4827 | | Email | : pokropf53@outlook.com | # 5.3 Date of Preparation Date of Preparation : 03/29/2019 ### 5.4 Device Information ### Table - Device Information | Trade Name | SeeFactorCT3 | |----------------------|-----------------------------------| | Common or Usual Name | Computed tomography x-ray system. | | Classfication Name | 21 CFR 892.1750 | | Regulatory Class | 2 | | Product Code | OAS | {4}------------------------------------------------ ### 5.5 Predicate Device(s) #### Table - Predicate Device(s) | Predicate Type | 510(k) Number | Device Name | Manufacturer | |------------------|---------------|---------------------------------|----------------------| | Primary Device | K142222 | NewTom 5G | THEMA S.r.l. (Italy) | | Reference Device | K130442 | NewvTom VG i and]<br>NewTomn 5G | THEMA S.r.l. (Italy) | #### 5.6 Device Description See Factor CT3 is a Diagnostic CBCT (Cone Beam Computed Tomography) device that integrates High Definition Computed Tomography (HDCT) for obtaining High Resolution 3D Volume with 2D High Resolution Scouting, through Dynamic Scouting and Single Pulse Modality. ### 5.7 Intended Use/Indications for Use SeeFactorCT3 is a cone beam computed tomography x-ray imaging system that acquires sequences of the head including the ear, nose and throat (ENT), dento-maxillofacial complex, teeth, mandible and jaw, temporo-mandibular joint (TMJ), other areas of human skull and neck with sections of upper cervical spine, and upper and lower extremities for use in diagnostic support. The device displays two and three dimensional images for each examined anatomical region. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals. ### 5.8 Comparison of Technological Characteristics with Predicate See Appendix #1 for a detailed comparsion of the proposed device and the predicate device. #### Method of Preformance and Operation : Both the proposed device and the predicate device utilize the same type of energy and the same type of methodology for obtaining diagnostics imaging: cone-beam computed tomography x-ray imaging. #### 5.9 Performace Data #### Non Clinical : Living and deceased animals were imaged during the development of this device. A substantial number of diagnostic-quality animal images are shown in the appendices. See Appendices 2, 3 and 4. #### 5.10 Biocompatibility Testing {5}------------------------------------------------ Biocompatibility testing for patient contact materials wsa not performed due to the limited period of patient exposure which typically occurs at lengthy intervals. The factory statement is attached here. ### 5.11 Clinical Testing No clinical testing was performed in the development of this product. A substantial number of diagnostic-quality animal images are shown the appendices. #### 5.12 Conclusion The proposed device operates the same as the predicate device using the same source of power, the same method of image acquisition, the same intended use, the same indications for use and both are targeted towards the same group of professional healthcare workers. The proposed device while identical in many ways and similar in a some others, does have superior imaging resolution, it weighs less, it can be moved and it can be covered during surgical procedures should there be a need.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...