← Product Code [FOZ](/productcode/FOZ) · K190855

# BD Acute Central Line (K190855)

_Bard Access Systems, Inc. (Bard Has Joined Bd) · FOZ · Nov 1, 2019 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190855

## Device Facts

- **Applicant:** Bard Access Systems, Inc. (Bard Has Joined Bd)
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Nov 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

Acute central venous catheters are indicated to provide short-term access (<30 days) to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, central venous pressure monitoring, and power injection of contrast media.

## Device Story

BD Acute Central Line is a family of power-injectable, radiopaque, polyurethane central venous catheters with a soft tip. Designed for percutaneous insertion into jugular, subclavian, or femoral veins; used in clinical settings by healthcare professionals. Device facilitates short-term (<30 days) central venous access for I.V. therapy, blood sampling, pressure monitoring, and power injection of contrast media. Catheters feature multiple lumens (distal, medial, proximal) and are provided sterile. Healthcare providers use the device to establish a conduit to the central venous system; output is the delivery or withdrawal of fluids/blood. Benefits include reliable venous access for critical care and diagnostic imaging procedures.

## Clinical Evidence

Bench testing only. Performance testing included biocompatibility (ISO 10993-1), mechanical integrity (tensile, burst, leak, clamp engagement, guidewire drag), dimensional compliance, radiopacity (ASTM F640-12), luer compatibility (ISO 594), particulate matter (USP <788>), and MR safety. All tests met predetermined acceptance criteria.

## Technological Characteristics

Materials: Medical grade polyurethane. Design: Multi-lumen (triple), radiopaque, soft tip. Dimensions: 7 Fr, lengths 16, 20, 30 cm. Power injection: Up to 10 mL/sec at 325 psi. Sterilization: Ethylene Oxide (AAMI 11135:2014). Connectivity: None (mechanical device).

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- Arrow Central Venous Catheter ([K071538](/device/K071538.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.

November 1, 2019

Bard Access Systems, Inc. (Bard has joined BD) % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K190855

Trade/Device Name: BD Acute Central Line Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: October 3, 2019 Received: October 7, 2019

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K190855

Device Name BD Acute Central Line

#### Indications for Use (Describe)

Acute central venous catheters are indicated to provide short-term access (<30 days) to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, central venous pressure monitoring, and power injection of contrast media.

Type of Use (Select one or both, as applicable)

| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K190855 510(k) Summary for BD Acute Central Line

| General<br>Provisions | Submitter Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Bard Access Systems, Inc. (Bard has joined BD)                     |                              |                                               |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------------------------------|-----------------------------------------------|
|                       | Submitter Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 605 North 5600 West<br>Salt Lake City, UT 84116                    |                              |                                               |
|                       | Contact Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Sean Loring<br>Regulatory Affairs Specialist                       |                              |                                               |
|                       | Telephone Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 801.522.5634                                                       |                              |                                               |
|                       | Fax Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 801.522.5425                                                       |                              |                                               |
|                       | Date of Preparation:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 10/31/2019                                                         |                              |                                               |
| Subject<br>Device     | Trade Name(s):                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | BD Acute Central Line                                              |                              |                                               |
|                       | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Acute Central Line                                                 |                              |                                               |
|                       | Classification Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Catheter, Intravascular, Therapeutic, Short-Term Less than 30 days |                              |                                               |
|                       | Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 2                                                                  |                              |                                               |
|                       | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR 880.5200                                                    |                              |                                               |
|                       | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | FOZ                                                                |                              |                                               |
|                       | Classification Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | General Hospital                                                   |                              |                                               |
| Predicate<br>Device   | Predicate Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Arrow Central Venous Catheter                                      |                              |                                               |
|                       | Classification Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Catheter, Intravascular, Therapeutic, Short-Term Less than 30 days |                              |                                               |
|                       | Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 2                                                                  |                              |                                               |
|                       | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | FOZ                                                                |                              |                                               |
|                       | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR 880.5200                                                    |                              |                                               |
|                       | Premarket Notification #:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K071538                                                            |                              |                                               |
|                       | Manufacturer:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Arrow                                                              |                              |                                               |
|                       | Classification Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | General Hospital                                                   |                              |                                               |
| Device<br>Description | A family of power injectable central venous catheters constructed of medical grade polyurethane and is<br>designed for insertion into the central venous system. BD power injectable acute central lines are<br>radiopaque, and have a soft tip that is more pliable than the catheter body. Each catheter is provided in a<br>sterile package with applicable insertion kit accessories. The maximum pressure injector settings and<br>maximum power injection flow rate are specified in the table below: |                                                                    |                              |                                               |
|                       | Catheter Length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Lumen(s)                                                           | Power Injection<br>Flow Rate | Maximum Power<br>Injector Pressure<br>Setting |
|                       | 16 cm and 20 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Distal<br>Medial / Proximal                                        | 10 mL/sec<br>9 mL/sec        | 325 psi                                       |

{4}------------------------------------------------

|                                  |                        | Catheter Length                                                                                                                                                                                                                                                                                                                                                   | Lumen(s)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Power Injection<br>Flow Rate                                                                                  | Maximum Power<br>Injector Pressure<br>Setting |
|----------------------------------|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
|                                  |                        | 16 cm and 20 cm                                                                                                                                                                                                                                                                                                                                                   | Distal<br>Medial / Proximal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 10 mL/sec<br>9 mL/sec                                                                                         | 325 psi                                       |
|                                  |                        | 30 cm                                                                                                                                                                                                                                                                                                                                                             | Distal<br>Medial/ Proximal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 9 mL/sec<br>7 mL/sec                                                                                          |                                               |
| Technological<br>Characteristics | Attribute              | Subject Device – BD Acute Central Line                                                                                                                                                                                                                                                                                                                            | Predicate Device – Arrow Central Venous Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Discussion of Characteristics<br>between<br>subject and predicate                                             |                                               |
|                                  | Owner                  | Bard Access Systems, Inc.                                                                                                                                                                                                                                                                                                                                         | Arrow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                               |                                               |
|                                  | Classification         | Same                                                                                                                                                                                                                                                                                                                                                              | FOZ – 21 CFR 880.5200                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Classification of<br>subject device is<br>the same as the<br>predicate                                        |                                               |
|                                  | 510(k) Status          | Subject of this Premarket Notification                                                                                                                                                                                                                                                                                                                            | K071538 – Concurrence<br>date August 30, 2007                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                               |                                               |
|                                  | Indications<br>for Use | Acute central venous catheters are<br>indicated to provide short-term access<br>(< 30 days) to the central venous<br>system. They are designed for<br>administering I.V. fluids, blood<br>products, drugs and parenteral<br>nutrition solutions, as well as blood<br>withdrawal, central venous pressure<br>monitoring, and power injection of<br>contrast media. | The Arrow CVC is<br>intended to provide short-<br>term (<30 days) central<br>venous access for<br>treatment of diseases or<br>conditions requiring<br>central venous access,<br>including, but not limited<br>to the following:<br>• Lack of usable<br>peripheral IV sites<br>• Central venous<br>pressure<br>monitoring<br>• Total parenteral<br>nutrition (TPN)<br>• Infusions of fluids,<br>medications, or<br>chemotherapy<br>• Frequent blood<br>sampling or<br>receiving blood<br>transfusions/blood<br>products | Indications for<br>use between the<br>subject and<br>predicate<br>devices are<br>substantially<br>equivalent. |                                               |
|                                  | Commercial<br>Name     | BD Acute Central Line                                                                                                                                                                                                                                                                                                                                             | Arrow Central Venous<br>Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                               |                                               |

{5}------------------------------------------------

| Catheter<br>Dimensions                | 7 Fr Triple Lumen x 16 cm<br>7 Fr Triple Lumen x 20 cm<br>7 Fr Triple Lumen x 30 cm                                                                         | 7 Fr Triple Lumen x 16 cm<br>7 Fr Triple Lumen x 20 cm                                                                                                      | Additional<br>catheter length<br>of 30 cm does<br>not raise new or<br>different<br>questions of<br>safety or<br>effectiveness<br>and a risk<br>assessment did<br>not identify any<br>new or<br>significantly<br>modified risks. |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Duration of<br>Use                    | Same                                                                                                                                                        | Short term (<30 days)                                                                                                                                       | Both the subject<br>and predicate<br>devices are<br>indicated for use<br>for less than 30<br>days.                                                                                                                              |
| Means of<br>insertion                 | Same                                                                                                                                                        | Percutaneous                                                                                                                                                | Means of<br>insertion is<br>identical<br>between the<br>subject and<br>predicate<br>devices.                                                                                                                                    |
| Insertion Site                        | Same                                                                                                                                                        | Jugular, subclavian, or<br>femoral veins                                                                                                                    | Insertion sites for<br>the subject and<br>predicate<br>devices are<br>identical.                                                                                                                                                |
| Primary<br>Device<br>Materials        | Catheter Base Materials<br>Shaft Tubing:<br>Polyurethane<br>Luer Connector:<br>Polyurethane<br>Extension Legs:<br>Polyurethane<br>Junction:<br>Polyurethane | Catheter Base Materials<br>Shaft Tubing:<br>Polyurethane<br>Luer Connector:<br>Polyurethane<br>Extension Legs:<br>Polyurethane<br>Junction:<br>Polyurethane | Catheter base<br>materials for the<br>predicate device<br>are proprietary.                                                                                                                                                      |
| Catheter<br>Proximal<br>Configuration | Same                                                                                                                                                        | Side-hole skive                                                                                                                                             | Lumen<br>configurations<br>are substantially<br>equivalent<br>between the<br>subject and<br>predicate<br>devices.                                                                                                               |
| Catheter<br>Medial<br>Configuration   | Same                                                                                                                                                        | Side-hole skive                                                                                                                                             | Lumen<br>configurations<br>are substantially<br>equivalent<br>between the<br>subject and                                                                                                                                        |

{6}------------------------------------------------

|                                  |                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                         | predicate<br>devices.                                                                                                                |  |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--|
|                                  | Catheter<br>Distal<br>Configuration                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                   | Formed tip                                                                                                              | Tip<br>configurations of<br>the subject and<br>the predicate<br>device are<br>substantially<br>equivalent.                           |  |
|                                  | Number of<br>Lumens                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                   | Triple Lumen                                                                                                            | Same as<br>predicate device.                                                                                                         |  |
|                                  | Power<br>Injection<br>Maximum<br>Flow Rate                                                                                                                                                                                                                                                                             | 16 and 20 cm length:<br>• Distal (17 Ga.) – 10 mL/sec<br>• Medial (18 Ga.) – 9 mL/sec<br>• Proximal (18 Ga.) – 9 mL/sec<br>30 cm length:<br>• Distal (17 Ga.) – 9 mL/sec<br>• Medial (18 Ga.) – 7 mL/sec<br>• Proximal (18 Ga.) – 7 mL/sec |                                                                                                                                                                                                                                                                                                                                                                   | 16 and 20 cm length:<br>• Distal (16 Ga.) – 10 mL/sec<br>• Medial (18 Ga.) – 5 mL/sec<br>• Proximal (18 Ga.) - 5 mL/sec | Differences in<br>power injection<br>maximum flow<br>rates do not<br>raise additional<br>questions of<br>safety or<br>effectiveness. |  |
|                                  | Sterility                                                                                                                                                                                                                                                                                                              | Same (10-6)<br>Ethylene Oxide<br>AAMI 11135:2014                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                   | Provided Sterile                                                                                                        | The subject and<br>predicate<br>devices are both<br>provided sterile.                                                                |  |
|                                  | The technological differences listed above were evaluated using industry consensus standards, and as<br>defined in the Risk Assessment. Therefore, these differences in technological characteristics between the<br>subject and predicate devices do not raise new or different questions of safety or effectiveness. |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                         |                                                                                                                                      |  |
|                                  |                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                            | The following performance tests were conducted by or for BAS per guidance documents and standards in<br>conjunction with in-house protocols to establish the performance of the BD Acute Central Line, and in<br>determining substantial equivalence to the predicate Arrow Central Venous Catheter. All testing passed<br>the predetermined acceptance criteria. |                                                                                                                         |                                                                                                                                      |  |
|                                  |                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                            | Reference Standard: <i>ISO 10993-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and<br/>testing within a risk management process</i>                                                                                                                                                                                                      |                                                                                                                         |                                                                                                                                      |  |
|                                  |                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                            | Tests to confirm that the catheter is free from biological hazard per testing. A health-<br>based risk assessment per ISO 10993-1 was performed for determining the<br>acceptability of the material for the intended purpose.                                                                                                                                    |                                                                                                                         |                                                                                                                                      |  |
| Safety &<br>Performance<br>Tests | Biocompatibility Testing                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                            | Testing Performed includes:<br>• Cytotoxicity<br>• Sensitization<br>• Irritation or Intracutaneous Reactivity<br>• Acute Systemic Toxicity<br>• Pyrogenicity<br>• Subchronic Systemic Toxicity<br>• Genotoxicity<br>• Hemocompatibility<br>• Implantation                                                                                                         |                                                                                                                         |                                                                                                                                      |  |
|                                  |                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                            | Reference Standard: <i>ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General<br/>requirements</i>                                                                                                                                                                                                                                       |                                                                                                                         |                                                                                                                                      |  |
|                                  | Clamp Engagement                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                            | Test to confirm that the catheter assembly will not leak when the clamp is engaged.                                                                                                                                                                                                                                                                               |                                                                                                                         |                                                                                                                                      |  |
|                                  | Leak Test                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                   | Test to confirm that the catheter assembly will not leak when the distal end of the<br>catheter is occluded.            |                                                                                                                                      |  |

{7}------------------------------------------------

| Dimensional Test                                                                                                                             | Test to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification.                    |
|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implantable Length                                                                                                                           | Test to measure useful length for catheters to ensure compliance with dimensional specification.                                                                            |
| Extension Leg Length                                                                                                                         | Test to measure and confirm extension leg length compliance with dimensional specification.                                                                                 |
| Burst Test                                                                                                                                   | Burst pressure test to confirm the catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the distal end is occluded.    |
| Hydraulic Catheter<br>Burst Test                                                                                                             | Burst pressure test to confirm the catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the distal end is occluded.    |
| Power Injection<br>Conditioning                                                                                                              | Test to confirm the catheter does not leak or burst as a result of power injections at maximum indicated flow rate.                                                         |
| Gravity Flow                                                                                                                                 | Test to measure the gravity flow performance of a full-length catheter.                                                                                                     |
| Luer to Extension Leg<br>Tensile Test                                                                                                        |                                                                                                                                                                             |
| Extension Leg to<br>Trifurcation Tensile Test                                                                                                | Test to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force.                                                                       |
| Trifurcation to Shaft<br>Tensile Test                                                                                                        |                                                                                                                                                                             |
| Shaft Tensile Test                                                                                                                           |                                                                                                                                                                             |
| Reference Standard: ASTM F640-12 – Standard Test Methods for Radiopacity of Plastics<br>for Medical Use                                      |                                                                                                                                                                             |
| Radiopacity                                                                                                                                  | Test to demonstrate catheter radio-detectability                                                                                                                            |
| Reference Standard: ISO 10555-3:2013 - Intravascular catheters - Sterile and single-use catheters - Part 3:<br>Central venous catheters      |                                                                                                                                                                             |
| Tip Tensile                                                                                                                                  | Test to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force.                                                                       |
| Reference Guidance: FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term<br>and Long-Term Intravascular Catheters, 1995 |                                                                                                                                                                             |
| Catheter Collapse Test                                                                                                                       | Test to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum.                                                             |
| Shaft Tensile Test                                                                                                                           | Test to evaluate the maximum catheter strain and modulus at break.                                                                                                          |
| Suture Wing Integrity<br>Test                                                                                                                | Test to measure the maximum force a catheter junction suture wing can withstand prior to break.                                                                             |
| Priming Volume                                                                                                                               | Test to measure the volume required to prime a full-length catheter.                                                                                                        |
| OD Swell                                                                                                                                     | Test to confirm that the catheter does not swell beyond twice the size of the labeled OD during power injection.                                                            |
| Tip Stability Test                                                                                                                           | Test to confirm that the catheter tip remains in the same orientation during power injection (tip pointing in direction of venous flow) at the maximum indicated flow rate. |
| Guidewire Drag Test                                                                                                                          | Test to ensure that the guidewire used to place the catheter can be removed without difficulty.                                                                             |

{8}------------------------------------------------

| Reference Standards: ISO 594: Conical Fittings for a 6% (Luer) Taper for Syringes, Needles, and Certain<br>Other Medical Equipment - Part 1: General Requirements and Part 2: Lock Fittings |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Luer Testing                                                                                                                                                                                | Testing to ensure that luer connectors meet requirements for:<br>• Leak<br>• Leak Decay<br>• Stress Cracking<br>• Resistance to Separation from Axial Load<br>• Resistance to Separation from Unscrewing<br>• Resistance to Overriding<br>• Gauging                                                                                                                                                                                                                                    |
| Reference Standard: USP<788>: Sizing and Counting Particulate Matter                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Particulate Testing                                                                                                                                                                         | Testing to ensure that particulate matter on the catheter post-manufacture is not<br>exceeded for prescribed particle sizes.                                                                                                                                                                                                                                                                                                                                                           |
| Reference Standard: FDA Guidance on Establishing Safety and Compatibility of Passive Implants in the<br>Magnetic Resonance (MR) Environment, 2014                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| MR Safety                                                                                                                                                                                   | Testing to demonstrate that the subject device is safe for use in an MR environment.                                                                                                                                                                                                                                                                                                                                                                                                   |
| Diagram                                                                                                                                                                                     | Image: Diagram of Catheter with labels: Female Luer Hubs, Thumb Clamp, Trifurcation/Suture Wing, Soft Tip, Catheter Shaft Tubing, Extension Legs                                                                                                                                                                                                                                                                                                                                       |
| Technological<br>Comparison<br>to Predicate<br>Device                                                                                                                                       | Technological characteristics of the subject BD Acute Central Line are substantially equivalent with regard<br>to the basic design and function of the predicate device, Arrow Central Venous Catheter (K071538). Thepriming volumes, differ from the predicate device. However, these differences do not alter the intended use<br>of the subject device, and do not raise any new or different questions regarding safety or effectiveness<br>when compared to the predicate device. |
| Summary of<br>Substantial<br>Equivalence                                                                                                                                                    | Based on the risk management activities and testing, the subject BD Acute Central Line has been<br>demonstrated to be substantially equivalent to the cited predicate device.                                                                                                                                                                                                                                                                                                          |

Based on the indications for use, technological characteristics, and results of performance testing, the Conclusion: subject BD Acute Central Line has been demonstrated to be substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K190855](https://fda.innolitics.com/device/K190855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
