← Product Code [GCJ](/productcode/GCJ) · K190831

# E240AG LCD Monitor, E240A LCD Monitor, JUSHA-E240AG LCD Monitor, JUJSHA-E240A LCD Monitor, C230A LCD Monitor, JUSHA-C230A LCD Monitor, C230M LCD Monitor, JUSHA-C230M LCD Monitor (K190831)

_Nanjing Jusha Display Technology Co., Ltd. · GCJ · May 31, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190831

## Device Facts

- **Applicant:** Nanjing Jusha Display Technology Co., Ltd.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** May 31, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The E240AG, C230A, C230M, E240A, JUSHA-E240AG, JUSHA-C230A, JUSHA-C230M, JUJSHA-E240A LCD Monitor is intended to provide 2D color video displays of images from surgical endoscopic/laparoscopic camera systems and other compatible medical imaging systems. The E240AG is a widescreen, high-definition, medical grade monitor for real-time use during minimally invasive surgical procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments.

## Device Story

Medical-grade LCD monitor for real-time visualization during minimally invasive surgery. Inputs: analog and digital video signals from endoscopic/laparoscopic camera systems and other medical imaging sources. Operation: accepts various input formats via built-in option slots; converts analog to digital signals; displays 2D color video on 23.8-inch TFT LCD panel. Used in ORs, surgical centers, and clinics by surgeons and clinical staff. Output: high-definition visual display of surgical field. Benefits: provides real-time, high-quality imaging to assist clinicians in surgical decision-making and procedural guidance.

## Clinical Evidence

Bench testing only. No clinical or animal studies performed. Bench testing included luminance non-uniformity, chromaticity non-uniformity, contrast ratio, luminance stability, electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).

## Technological Characteristics

23.8-inch TFT LCD panel; LED backlight; 1920x1080 resolution; 1000:1 contrast ratio; 300cd/m² brightness; 14ms response time; 1024 grayscale levels. Connectivity: VGA, DVI-A/D, SDI, S-Video, YPbPr, RGBS, CVBS. Power: 100-240 Vac, 47-63 Hz. Includes backlight sensor. Standards: IEC 60601-1:2012, IEC 60601-1-2:2014.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Sony LMD-2451MT LCD Monitor ([K113203](/device/K113203.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 31, 2019

Nanjing Jusha Display Technology Co., Ltd Zilong Liang Certification Manager Suite A, 8/F, Bldg 1, No. 301, Hanzhongmen St. Nanjing International Service Outsourcing Mansion Nanjing, China 210036

Re: K190831

Trade/Device Name: E240AG LCD Monitor, E240A LCD Monitor, JUSHA-E240AG LCD Monitor, JUJSHA-E240A LCD Monitor, C230A LCD Monitor, JUSHA-C230A LCD Monitor, C230M LCD Monitor, JUSHA-C230M LCD Monitor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: April 1, 2019 Received: April 1, 2019

Dear Zilong Liang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of 

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jennifer R. Stevenson Acting Division Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190831

Device Name

E240AG, C230A, C230M, E240A, JUSHA-E240AG, JUSHA-C230M, JUJSHA-E240A LCD Monitor

Indications for Use (Describe)

The E240AG LCD Monitor is intended to provide 2D color video displays of images from surgical endoscopic/ laparoscopic camera systems and other compatible medical imaging systems. The E240AG is a widescreen, high definition, medical grade monitor for real-time use during minimally invasive surgical procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments.

Type of Use (Select one or both, as applicable)

| <span style="font-size:100%;">Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-----------------------------------------------------------------------------------|
| <span style="font-size:100%;">Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

F

| Date:                     | December 2,2018                                                                                                                                                             |  |  |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Submitter:                | Nanjing Jusha Display Technology Co., Ltd<br>Add: 301, Hanzhongmen Street, 8F Block A, No.1,<br>Nanjing International Service Outsourcing Mansion,<br>Nanjing, 210036 China |  |  |
| Primary Contact Person:   | Zilong Liang<br>Certification Manager<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271                                        |  |  |
| Secondary Contact Person: | Dongdong Li<br>Quality Manager<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271                                               |  |  |
| Device Trade Name:        | E240AG, C230A, C230M, E240A, JUSHA-E240AG,<br>JUSHA-C230A, JUSHA-C230M, JUJSHA-E240A LCD Monitor                                                                            |  |  |
| Common/Usual Name:        | LCD Monitor                                                                                                                                                                 |  |  |
| Classification Name:      | Endoscope and Accessories                                                                                                                                                   |  |  |
| Product Code:             | GCJ                                                                                                                                                                         |  |  |
| Predicate Device(s):      | Sony LMD-2451MT LCD Monitor, K 113203                                                                                                                                       |  |  |
| Device Description:       | The E240AG, C230A, C230M, E240A, JUSHA-E240AG,<br>JUSHA-C230A, JUSHA-C230M, JUJSHA-E240A LCD<br>Monitor is intended primarily for use in a medical                          |  |  |

{4}------------------------------------------------

| environment for displaying images captured during<br>minimally invasive surgical procedures. The monitor features<br>a 23.8-inch widescreen LCD display panel and employs full<br>WUXGA high-definition (HID) performance. In addition<br>to digital signals, the E240AG accepts analog signals and<br>converts them to digital signals.<br>The E240AG LCD Monitor is designed with the flexibility to<br>support a variety of formats with two built-in option slots to<br>select, expand, and change input/output signals. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The E240AG, C230A, C230M, E240A, JUSHA-E240AG,<br>JUSHA-C230A, JUSHA-C230M, JUJSHA-E240A LCD Monitor<br>is intended to provide 2D color video displays of images from<br>surgical endoscopic/laparoscopic camera systems and other<br>compatible medical imaging systems. The E240AG is a<br>widescreen, high-definition, medical grade monitor for real-time<br>use during minimally invasive surgical procedures and is suitable<br>for use in hospital operating rooms, surgical centers, clinics,<br>doctors' offices and similar medical environments. |
| Technological<br>characteristic<br>and Substantial<br>equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The E240AG, C230A, C230M, E240A, JUSHA-E240AG,<br>JUSHA-C230A, JUSHA-C230M, JUJSHA-E240A LCD Monitor<br>has the same overall purpose and function as the predicate device<br>cited above. Both of the devices are intended to provide color<br>video displays of images from surgical camera systems, primarily<br>in endoscopic/laparoscopic applications.                                                                                                                                                                                                 |
| Performance<br>Testing:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Testing of the E240AG, C230A, C230M, E240A,<br>JUSHA-E240AG, JUSHA-C230A, JUSHA-C230M,<br>JUJSHA-E240A LCD Monitor demonstrated that the device<br>is in compliance with applicable requirements of recognized<br>standards for electromagnetic compatibility and electrical<br>safety.                                                                                                                                                                                                                                                                     |
| Conclusion:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Based on the similarities in overall purpose and function, the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | E240AG LCD Monitor has demonstrated substantial equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | to the cited predicate device and any differences do not affect the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | safety or effectiveness of the device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{5}------------------------------------------------

## 12.1 Product Comparison

This comparison identifies the similarities and differences of the proposed E240AG LCD Monitor device to the legally marketed predicate Sony LMD-2451MT LCD Monitor device to which substantial equivalency is claimed.

| Attributes                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Discussion of<br>Differences                                                                                                                                                                                                                                   |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product                                | Sony LMD-2451MT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | E240AG, C230A, C230M,<br>E240A, JUSHA-E240AG,<br>JUSHA-C230A,<br>JUSHA-C230M,<br>JUJSHA-E240A LCD<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                |
| 510(k) Number                          | K113203                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                |
| Display Performance/Specifications     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                |
| Screen technology                      | 24" TFT LCD Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 23.8" TFT LCD Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                           |
| Viewing angle<br>(H,V)                 | Horizontal<br>178°, Vertical 178° (the<br>contrast ratio ≥ 10 : 1 )                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Horizontal 178°, Vertical<br>178° (the contrast ratio ≥ 10 :1)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                           |
| Resolution                             | $1920 \times 1200$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | $1920 \times 1080$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different Resolution<br>design due to the<br>different panel size.                                                                                                                                                                                             |
| Display area                           | $518.4 \times 324.0$ mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | $527.04 \times 296.46$ mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Different Display<br>area design due to the<br>different panel size.                                                                                                                                                                                           |
| Contrast<br>Ratio(typ)                 | 1000:1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1000:1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                           |
| Attributes                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Discussion of<br>Differences                                                                                                                                                                                                                                   |
| Product                                | Sony LMD-2451MT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | E240AG, C230A, C230M,<br>E240A, JUSHA-E240AG,<br>JUSHA-C230A,<br>JUSHA-C230M,<br>JUJSHA-E240A LCD<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                |
| 510(k) Number                          | K113203                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                |
| Recommended<br>brightness              | 300cd/m²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 300cd/m²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                           |
| Pixel Pitch                            | 0.27×0.27 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 0.27×0.27 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                           |
| Backlighting                           | LED backlight                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | LED backlight                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                           |
| grayscale                              | 1024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 1024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                           |
| Response time                          | 14ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 14ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                           |
| Sensor                                 | Backlight sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Backlight sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                           |
| Input signals                          | BNC、DVI、SDI、<br>Y/C(S-Video)、YPbPr、<br>RGBS、HD15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | VGA、DVI-A/D、SDI、<br>Y/C(S-Video)、YPbPr、RGBS、<br>CVBS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | The differences<br>caused by different<br>signals interface type<br>used in the medical<br>display, the E240AG<br>has<br>VGA&CVBS ,This<br>only shows the<br>signals interface type<br>is<br>different, nothing<br>to do with the display<br>function          |
| Input<br>terminational                 | 1 DVI-D Dual Link<br>1 HDMI Port                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 1 DVI-D Dual Link<br>1 HDMI Port                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                           |
| Display controller                     | Off the shelf                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Off the shelf                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                           |
| Video Signal Output                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                |
| Attributes                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Discussion of<br>Differences                                                                                                                                                                                                                                   |
| Product                                | Sony LMD-2451MT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | E240AG, C230A, C230M,<br>E240A, JUSHA-E240AG,<br>JUSHA-C230A,<br>JUSHA-C230M,<br>JUJSHA-E240A LCD<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                |
| 510(k) Number                          | K113203                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                |
| Output signals                         | BNC\u3001DVI\u3001SDI\u3001<br>Y/C(S-Video)\u3001YPbPr\u3001<br>RGBS\u3001HD15                                                                                                                                                                                                                                                                                                                                                                                                                                                             | DVI-D\u3001SDI\u3001Y/C<br>(S-Video)\u3001YPbPr\u3001RGBS\u3001<br>CVBS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The differences<br>caused by different<br>signals interface type<br>used in the medical<br>display, the E240AG<br>has<br>VGA&CVBS ,This<br>only shows the<br>signals interface type<br>is different, nothing<br>to do with the display<br>function             |
| Power<br>Requirement                   | 100 V - 240 Vac \u00b110%,<br>50/60 Hz 1.53 A - 0.58 A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 100 - 240 Vac \u00b110%, 47-63<br>Hz 1.4 - 0.6A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                           |
| Power<br>Consumptions/Sav<br>e Mode    | 125W/less than 0.7W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 35W/less than 0.5W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The differences<br>caused by different<br>adapter and<br>components used in<br>the medical<br>display,the E240AG<br>has lower power<br>consumptions,This<br>only shows the power<br>consumption is<br>different, nothing to<br>do with the display<br>function |
| Attributes                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Discussion of<br>Differences                                                                                                                                                                                                                                   |
| Product                                | Sony LMD-2451MT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | E240AG, C230A, C230M,<br>E240A, JUSHA-E240AG,<br>JUSHA-C230A,<br>JUSHA-C230M,<br>JUJSHA-E240A LCD<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                |
| 510(k) Number                          | K113203                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                |
| Dimensions w/o<br>stand<br>(W x H x D) | With stand:<br>$602.4\u00d7454\u00d7302$ mm<br>Without stand:<br>$602.4\u00d7386.2\u00d7110$ mm                                                                                                                                                                                                                                                                                                                                                                                                                                            | With stand:<br>565.3x238x460.3mm<br>Without stand:<br>565.3x65x375.3mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Different housing<br>design due to the<br>different panel size.                                                                                                                                                                                                |
| Indication for use                     | The Sony LMD-2451MT<br>LCD Monitor is intended<br>to provide 3D and 2Dcolor<br>video<br>displays of images from<br>surgical<br>endoscopic/laparoscopic<br>camera systems and other<br>compatible medical<br>imaging systems. The<br>LMD-2451MT is a<br>widescreen,<br>high-definition, medical<br>grade monitor for<br>real-time use during<br>minimally invasive<br>surgical procedures and is<br>suitable for use in hospital<br>operating rooms, surgical<br>centers, clinics, doctors'<br>offices and similar medical<br>environments. | The E240AG, C230A,<br>C230M, E240A,<br>JUSHA-E240AG,<br>JUSHA-C230A,<br>JUSHA-C230M,<br>JUJSHA-E240A LCD<br>Monitor is intended to<br>provide 2D color video<br>displays of images from<br>surgical<br>endoscopic/laparoscopic<br>camera systems and other<br>compatible medical<br>imaging systems. The<br>E240AG is a widescreen,<br>high-definition, medical<br>grade monitor for real-time<br>use during minimally<br>invasive surgical<br>procedures and is suitable<br>for use in hospital<br>operating rooms, surgical<br>centers, clinics, doctors'<br>offices and similar medical<br>environments. | Same                                                                                                                                                                                                                                                           |
| Attributes                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Discussion of<br>Differences                                                                                                                                                                                                                                   |
| Product                                | Sony LMD-2451MT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | E240AG, C230A, C230M,<br>E240A, JUSHA-E240AG,<br>JUSHA-C230A,<br>JUSHA-C230M,<br>JUJSHA-E240A LCD<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                |
| 510(k) Number                          | K113203                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                |
| Applicable<br>standard                 | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>1988+A1 : 1991 + A2:1995<br>2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2: General<br>requirements for basic safety<br>and essential performance -<br>Collateral standard:<br>Electromagnetic compatibility<br>Requirements and tests                                                                                                                 | 1 IEC 60601-1:2012 Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance.<br>2 IEC 60601-1-2:2014,<br>Medical electrical<br>equipment-Part 1-2:General<br>requirements for basic safety<br>and essential<br>performance-Collateral<br>Standard:Electromagnetic<br>disturbances-Requirements and<br>tests                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                           |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

### PERFORMANCE DATA:

The following performance data were provided in support of the substantial equivalence determination.

Bench testing:

Bench testing was conducted to demonstrate the E240AG meets all performance standards as follows:

Measurement of the luminance non-uniformity characteristics of the display screen

Measurement of the chromaticity non-uniformity characteristics of the display screen

Measurement of contrast ratio.

Performance data on luminance stability

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the E240AG The device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.

Animal and clinical study

The subject of this premarket submission, E240AG, does not require animal or clinical studies to support substantial equivalence.

## CONCLUSIONS

E240AG, C230A, C230M, E240A, JUSHA-E240AG, JUSHA-C230A, JUSHA-C230M,

JUJSHA-E240A LCD Monitor is substantially equivalent to the predicate device with respect to technical characteristics, performance, application and intended use. The non-clinical data support the safety of the device. The device should perform as intended in the specified use conditions. Nanjing Jusha Display Technology Co., Ltd considers the E240AG Medical Display does not raise any new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K190831](https://fda.innolitics.com/device/K190831)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).