The clinical study utilized data from patients undergoing routine clinical investigation for IBD/IBS symptoms to establish clinical sensitivity and specificity of the BÜHLMANN fCAL turbo assay.
Prospective enrollment of patients presenting with symptoms suggestive of IBS or IBD referred for endoscopy as part of standard clinical practice.
337 subjects (274 adults, 63 pediatric); 135 IBD, 130 IBS, 72 other GI disorders.; Sample Size: 337; Number of Sites: 13
Not applicable for this study
Clinical sensitivity and specificity for IBD diagnosis and IBD vs. IBS discrimination.
Indications for Use
The BÜHLMANN fCAL turbo is an in vitro diagnostic assay intended for the quantitative measurement of fecal calprotectin, a neutrophilic protein that is a marker of intestinal mucosal inflammation, in human stool. The BÜHLMANN fCAL turbo aids in the diagnosis of inflammatory bowel disease (IBD), specifically Crohn's disease (CD) and ulcerative colitis (UC) and aids in the differentiation of IBD from irritable bowel syndrome (IBS) in conjunction with other laboratory and clinical findings.
Device Story
Particle-enhanced turbidimetric immunoassay (PETIA) for fecal calprotectin quantification; utilizes patient stool extracts; calprotectin mediates immunoparticle agglutination on polystyrene beads; turbidity proportional to calprotectin concentration; light absorbance measured at 546 nm on clinical chemistry analyzers (e.g., Roche cobas c501/c502); results calculated via interpolation of calibration curve; used in clinical laboratories by technicians; output aids clinicians in diagnosing IBD and differentiating from IBS; benefits include automated, rapid processing compared to manual ELISA.
Clinical Evidence
Clinical study included 337 subjects (274 adults, 63 pediatric, age ≥ 4). Final diagnosis determined by endoscopy/histology/Rome III criteria: 135 IBD, 130 IBS, 72 other GI disorders. At 80 μg/g cutoff: sensitivity 91.1% (95% CI: 85.0-95.3%), specificity 74.3% (95% CI: 67.7-80.1%) for IBD vs. non-IBD. At 160 μg/g cutoff: sensitivity 80.0% (95% CI: 72.3-86.4%), specificity 85.1% (95% CI: 79.5-89.8%). Method comparison with predicate (n=220) showed Passing-Bablok slope 1.025 and R=0.972.
Technological Characteristics
Particle-enhanced turbidimetric immunoassay (PETIA). Materials: polystyrene nanoparticles coated with polyclonal antibodies against human calprotectin. Energy: optical (546 nm absorbance). Form factor: kit for automated clinical chemistry analyzers (Roche cobas c501/c502). Connectivity: integrated into clinical chemistry analyzer workflow. Standards: CLSI EP05-A3, EP6-A, EP07-A2, EP17-A2, EP25-A.
Indications for Use
Indicated for quantitative measurement of fecal calprotectin in human stool to aid in diagnosis of IBD (Crohn's disease and ulcerative colitis) and differentiation of IBD from IBS in adults.
Regulatory Classification
Identification
A fecal calprotectin immunological test system is an in vitro diagnostic device that consists of reagents used to quantitatively measure, by immunochemical techniques, fecal calprotectin in human stool specimens. The device is intended for in vitro diagnostic use as an aid in the diagnosis of inflammatory bowel diseases (IBD), specifically Crohn's disease and ulcerative colitis, and as an aid in differentiation of IBD from irritable bowel syndrome.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.5180 with special controls. The special control guidance document, "Class II Special Controls Guidance Document: Fecal Calprotectin Immunological Test Systems" will be available shortly.
*Classification.* Class II (special controls). The special control for these devices is FDA's guidance document entitled “Class II Special Controls Guidance Document: Fecal Calprotectin Immunological Test Systems.” For the availability of this guidance document, see § 866.1(e).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K190784
B. Purpose for Submission:
New Device
C. Measurand:
Fecal calprotectin
D. Type of Test:
Quantitative, Particle-enhanced turbidimetric immunoassay (PETIA)
E. Applicant:
BÜHLMANN Laboratories AG
F. Proprietary and Established Names:
BÜHLMANN fCAL turbo
G. Regulatory Information:
1. Regulatory Section:
21 CFR 866.5180, Fecal calprotectin immunological test system
2. Classification:
Class II
3. Product Code:
NXO, Calprotectin, Fecal
{1}
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use:
The BÜHLMANN fCAL turbo is an in vitro diagnostic assay intended for the quantitative measurement of fecal calprotectin, a neutrophilic protein that is a marker of intestinal mucosal inflammation, in human stool. The BÜHLMANN fCAL turbo aids in the diagnosis of inflammatory bowel disease (IBD), specifically Crohn's disease (CD) and ulcerative colitis (UC) and aids in the differentiation of IBD from irritable bowel syndrome (IBS) in conjunction with other laboratory and clinical findings.
2. Indication for use:
Same as Intended Use
3. Special conditions for use statement:
Prescription use only
4. Special instrument requirements:
Roche cobas c501/c502 platforms
I. Device Description:
The BÜHLMANN fCAL turbo kit contains the following materials (96 tests):
- Reaction Buffer (R1) (1 vial, 35 mL)
- Immunoparticles (R2) (1 vial, 7 mL)
The following materials are required for the assay but not included in the BÜHLMANN fCAL turbo kit:
- BÜHLMANN fCAL turbo Calibrator Kit: Calibrator 1-6 (1 vial/calibrator, 1 mL each, total 6 vials)
- BÜHLMANN fCAL turbo Control Kit: Low and High Control (3 vials/control, 1 mL each, total 6 vials)
- Stool Extraction buffer (125 mL/bottle)
2
{2}
J. Substantial Equivalence Information:
1. Predicate device name:
BÜHLMANN fCAL ELISA
2. Predicate 510(k) number:
K181012
3
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | BÜHLMANN fCAL turbo | BÜHLMANN fCAL ELISA (K181012) |
| Intended Use | The BÜHLMANN fCAL turbo is an in vitro diagnostic assay intended for the quantitative measurement of fecal calprotectin, a neutrophilic protein that is a marker of intestinal mucosal inflammation, in human stool. The BÜHLMANN fCAL turbo aids in the diagnosis of inflammatory bowel disease (IBD), specifically Crohn's disease (CD) and ulcerative colitis (UC) and aids in the differentiation of IBD from irritable bowel syndrome (IBS) in conjunction with other laboratory and clinical findings. | The BÜHLMANN fCAL ELISA is an in vitro diagnostic assay intended for the quantitative measurement of fecal calprotectin in human stool. The BÜHLMANN fCAL ELISA aids in the diagnosis of inflammatory bowel disease (IBD), specifically Crohn's disease (CD) and ulcerative colitis (UC) and aids in the differentiation of IBD from irritable bowel syndrome (IBS) in conjunction with other laboratory and clinical findings. |
| Analyte | Human fecal calprotectin (MRP8/14) | Same |
| Assay format | Quantitative | Same |
| Specimen type | Human stool | Same |
| Extraction Method | Manual Weighing (1:50 dilution in Extraction Buffer) | Same |
| Clinical Decision Thresholds | Normal: < 80 μg/g Gray-zone: 80–160 μg/g Elevated: > 160 μg/g | Same |
{4}
| Differences | | |
| --- | --- | --- |
| Item | BÜHLMANN fCAL turbo | BÜHLMANN fCAL ELISA (K181012) |
| Method | PETIA | ELISA |
| Automation | Automated | Manual |
| Solid Phase | Polystyrene nanoparticles (beads) | 96-well polystyrene microtiter plate |
| Detection Method | Automated clinical chemistry analyzer read at 546 nm | Microtiter plate reader read at 450 nm |
| Analyte-specific antibody components | Polyclonal antibodies against human calprotectin coated on polystyrene beads | Capture antibodies: monoclonal antibodies against human calprotectin coated on microtiter plates
Detection antibodies: monoclonal antibodies against human calprotectin conjugated to HRP |
| Measuring range | 30–2,000 μg/g | 30–1,800 μg/g |
| Expected Measuring Range (1:10 dilution) | 30–10,000 μg/g | N/A |
| Calibrators | 6 levels: Target values: 0, 50, 200, 500, 1000, 2000 μg/g | 5 levels: 4, 12, 40, 120, and 240 ng/mL |
| Calibrator Source | Native human calprotectin from human granulocyte extract | Native human calprotectin from human serum |
K. Standard/Guidance Documents Referenced:
- CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition
- CLSI EP6-A Evaluation of The Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- CLSI EP07-A2 Interference Testing in Clinical Chemistry; Approved Guideline – Second
{5}
Edition
- CLSI EP10-A2 Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition
- CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
L. Test Principle:
The BÜHLMANN fCAL turbo test is a particle-enhanced turbidimetric immunoassay (PETIA) which allows for automated quantification of calprotectin in fecal extracts on clinical chemistry analyzers. Fecal samples are extracted with extraction buffer using the manual weighing method and applied at a final dilution of 1:500. The extracts are incubated with reaction buffer and mixed with polystyrene nanoparticles coated with calprotectin-specific antibodies (immunoparticles). Calprotectin available in the sample mediates immunoparticle agglutination. Sample turbidity, measured by light absorbance, increases with calprotectin-immunoparticle complex formation and is proportional to calprotectin concentration. The detected light absorbance allows quantification of calprotectin concentration via interpolation on an established calibration curve.
M. Performance Characteristics:
1. Analytical performance: All studies presented below met the manufacturer’s predetermined acceptance criteria.
a. Precision/Reproducibility:
Precision: Eight extracted stool samples of various concentrations of calprotectin were tested for 20 days with two runs per day in two replicates per run, for a total of 80 BÜHLMANN fCAL turbo test results per sample. The mean BÜHLMANN fCAL turbo test result was computed and the within-run, between-run, between-day, and total standard deviations (SD) and associated coefficients of variation (%CV) were calculated.
6
{6}
| Sample | Mean (μg/g) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 42.9 | 3.6 | 8.3 | 1.2 | 2.7 | 1.1 | 2.5 | 3.9 | 9.1 |
| 2 | 98.4 | 2.5 | 2.6 | 1.8 | 1.8 | 2.2 | 2.2 | 3.7 | 3.8 |
| 3 | 166.5 | 4.3 | 2.6 | 0.8 | 0.5 | 1.9 | 1.2 | 4.8 | 2.9 |
| 4 | 267.6 | 3.9 | 1.4 | 2.5 | 0.9 | 1.8 | 0.7 | 5.0 | 1.9 |
| 5 | 642.0 | 20.1 | 3.1 | 14.9 | 2.3 | 0.0 | 0.0 | 25.1 | 3.9 |
| 6 | 1414.2 | 19.6 | 1.4 | 11.1 | 0.8 | 3.5 | 0.2 | 22.8 | 1.6 |
| 7 | 3251.4 | 40.8 | 1.3 | 21.4 | 0.7 | 19.7 | 0.6 | 50.1 | 1.5 |
| 8 | 5405.6 | 40.2 | 0.7 | 56.6 | 1.0 | 34.5 | 0.6 | 77.5 | 1.4 |
Reproducibility: To evaluate between-operator imprecision and between-site reproducibility, eight extracted stool samples at various concentrations of calprotectin were tested for five days with one run per day in five replicates per run at three sites, allowing for a total of 75 BÜHLMANN fCAL turbo test results per sample. The mean BÜHLMANN fCAL turbo test result was computed and the within-run, between-day, between-site, and total standard deviations (SD) and associated coefficients of variation (%CV) were calculated.
| Sample | Mean (μg/g) | Within-run | | Between-day | | Between-site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 47.2 | 3.6 | 7.6 | 2.4 | 5.0 | 0.0 | 0.0 | 4.3 | 9.1 |
| 2 | 91.1 | 3.5 | 3.8 | 3.5 | 3.8 | 2.8 | 3.1 | 5.7 | 6.2 |
| 3 | 185.4 | 5.1 | 2.7 | 2.7 | 1.4 | 5.5 | 3.0 | 7.9 | 4.3 |
| 4 | 276.9 | 6.4 | 2.3 | 4.5 | 1.6 | 9.7 | 3.5 | 12.5 | 4.5 |
| 5 | 674.5 | 12.9 | 1.9 | 1.2 | 0.2 | 22.8 | 3.4 | 26.3 | 3.9 |
| 6 | 1519.6 | 25.3 | 1.7 | 17.8 | 1.2 | 62.3 | 4.1 | 69.6 | 4.6 |
| 7 | 3343.8 | 54.6 | 1.6 | 35.6 | 1.1 | 100.0 | 3.0 | 119.4 | 3.6 |
| 8 | 5475.6 | 72.1 | 1.3 | 35.8 | 0.7 | 154.2 | 2.8 | 173.9 | 3.2 |
{7}
To evaluate between-lot imprecision, eight extracted stool samples of various concentrations of calprotectin were tested for five days in five replicates and three lots, for a total of 75 BÜHLMANN fCAL turbo test results per sample. The mean BÜHLMANN fCAL turbo test result was computed and the within-run, between-run, between-day, and total standard deviations (SD) and associated coefficients of variation (%CV) were calculated.
| Sample | Mean (μg/g) | Within-run | | Between-day | | Between-lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 45.2 | 3.2 | 7.1 | 1.4 | 3.0 | 3.7 | 8.2 | 5.1 | 11.3 |
| 2 | 86.4 | 3.7 | 4.3 | 1.2 | 1.4 | 5.7 | 6.6 | 6.9 | 7.9 |
| 3 | 175.8 | 5.0 | 2.9 | 0.3 | 0.2 | 9.9 | 5.6 | 11.1 | 6.3 |
| 4 | 263.9 | 7.6 | 2.9 | 0.0 | 0.0 | 10.0 | 3.8 | 12.5 | 4.7 |
| 5 | 647.4 | 15.5 | 2.4 | 0.0 | 0.0 | 15.3 | 2.4 | 21.7 | 3.4 |
| 6 | 1460.7 | 33.7 | 2.3 | 11.6 | 0.8 | 41.0 | 2.8 | 54.3 | 3.7 |
| 7 | 3234.5 | 71.2 | 2.2 | 8.9 | 0.3 | 130.3 | 4.0 | 148.8 | 4.6 |
| 8 | 5303.1 | 97.4 | 1.8 | 11.2 | 0.2 | 163.9 | 3.1 | 191.0 | 3.6 |
To evaluate extraction reproducibility, stool samples in three different categories of stool consistency with one liquid (Sample 6), four semi-solid (Sample 1, 4, 7, and 10), and five solid samples (Sample 2,3,5,8, and 9) were extracted manually extracted by two operators for two days, tested in four extractions, and measured in five replicates per extraction, for a total of 80 BÜHLMANN fCAL turbo test results obtained per sample. For each sample, the means BÜHLMANN fCAL turbo test results were calculated and the within-extraction, between-extraction, and total standard deviations (SD) and associated coefficients of variation (%CV) were calculated.
{8}
| Sample | Mean (μg/g) | Within-run | | Between-extraction | | Between-day | | Between-operator | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 47.7 | 2.8 | 5.9 | 1.1 | 2.4 | 0.7 | 1.5 | 1.4 | 2.9 | 3.4 | 7.2 |
| 2 | 72.3 | 3.8 | 5.2 | 3.9 | 5.4 | 4.2 | 5.8 | 0.0 | 0.0 | 6.8 | 9.5 |
| 3 | 96.1 | 3.8 | 3.9 | 2.2 | 2.3 | 1.4 | 1.4 | 0.0 | 0.0 | 4.6 | 4.8 |
| 4 | 170.6 | 4.0 | 2.4 | 2.5 | 1.5 | 8.7 | 5.1 | 0.0 | 0.0 | 9.9 | 5.8 |
| 5 | 277.0 | 3.7 | 1.4 | 27.9 | 10.1 | 10.0 | 3.6 | 11.0 | 4.0 | 31.8 | 11.5 |
| 6 | 421.1 | 9.8 | 2.3 | 5.9 | 1.4 | 15.3 | 3.6 | 0.0 | 0.0 | 19.1 | 4.5 |
| 7 | 573.9 | 5.4 | 0.9 | 39.5 | 6.9 | 0.0 | 0.0 | 0.0 | 0.0 | 39.9 | 6.9 |
| 8 | 1387.4 | 39.1 | 2.8 | 75.1 | 5.4 | 159.9 | 11.5 | 0.0 | 0.0 | 180.9 | 13.0 |
| 9 | 3264.9 | 87.2 | 2.7 | 236.2 | 7.2 | 256.9 | 7.9 | 0.0 | 0.0 | 359.7 | 11.0 |
| 10 | 3330.4 | 89.8 | 2.7 | 92.4 | 2.8 | 75.7 | 2.3 | 0.0 | 0.0 | 149.4 | 4.5 |
# b. Linearity/assay reportable range:
Linearity: A linearity study was conducted in accordance with CLSI guideline EP6-A. Two sample pools were prepared to obtain 15 concentration levels per sample pool to cover the full range of the assay from $30~\mu \mathrm{g / g}$ to $2,000~\mu \mathrm{g / g}$ ( $30~\mu \mathrm{g / g}$ to $10,000~\mu \mathrm{g / g}$ for samples with an additional dilution) calprotectin. Four replicates were obtained at each concentration level to cover the assay measuring range.
| Sample | Range (μg/g) | Slope (95% CI) | Intercept (95% CI) | R² |
| --- | --- | --- | --- | --- |
| 1 | 37.6–12,216.0 | 1.06 (1.04, 1.08) | 5.7 (1.6, 16.9) | 0.998 |
| 2 | 33.5–13,339.5 | 1.03 (1.01, 1.04) | 3.8 (-0.4, 13.3) | 0.998 |
Accuracy/Recovery: To evaluate the accuracy of the BÜHLMANN fCAL turbo, seven stool specimen extracts were spiked with calprotectin calibrators at a defined concentration of $500~\mu \mathrm{g / g}$ for Sample 1-4 and $2,000~\mu \mathrm{g / g}$ for Samples 5-7 in $10\%$ specimen extract volume, corresponding $50~\mu \mathrm{g / g}$ for Sample 1-4 and $200~\mu \mathrm{g / g}$ for Samples 5-7. Four replicates were obtained per each sample.
{9}
| Specimen Extract | Mean Baseline Result (μg/g) | Spiked Value (μg/g) | Expected Post-Spike Result (μg/g) | Observed Post-Spike Result (μg/g) | Recovery (%) |
| --- | --- | --- | --- | --- | --- |
| 1 | 44.1 | 56.9 | 101.0 | 94.6 | 93.6 |
| 2 | 65.5 | 56.9 | 122.4 | 114.5 | 93.6 |
| 3 | 116.4 | 56.9 | 173.3 | 170.2 | 98.2 |
| 4 | 138.5 | 56.9 | 195.4 | 186.9 | 95.7 |
| 5 | 230.9 | 227.8 | 458.7 | 453.1 | 98.8 |
| 6 | 510.8 | 227.8 | 738.6 | 753.2 | 102.0 |
| 7 | 1076.3 | 227.8 | 1304.1 | 1309.3 | 100.4 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
No international reference material or reference measurement procedures are available for calprotectin. Therefore, metrological traceability to international (SI) units cannot be established. The BÜHLMANN fCAL turbo is traceable to an internal standard generated by mixing a recombinant calprotectin antigen (MRP8/14 dimer) with the BÜHLMANN B-CAL-EX extraction buffer. Final value assignment was performed with the BÜHLMANN fCAL turbo using multiple reagent lots.
Stability:
Real-time reagent stability: Three reagent lots and four stool extracts (one below 80 μg/g, one between 80 μg/g and 160 μg/g, and two above 160 μg/g) were used to establish reagent stability. Reagents were stored at 2–8°C throughout the study. Stool extracts were stored in accordance with the instructions for use. The results of the study support the claim that BÜHLMANN fCAL turbo reagents are stable for six months at 2–8°C.
Transport stability: Four stool samples and two levels of control material were used. Reagents were stored at 28°C and 37°C to mimic temperature conditions that may be experienced during shipping. Reagent/sample/control combinations were tested in duplicates at baseline, and at seven days, 14 days, 21 days, and 28 days post-baseline. The results of the study support the claim that BÜHLMANN fCAL turbo reagents are stable for 28 days at 37°C.
On board stability: One reagent lot, four stool samples, and two levels of control material were used to evaluate on board stability. The reagents were opened and reconstituted prior to testing at baseline then stored at 5–12°C and used at 2, 4, 6, 8, 10, 12, and 13 weeks after the baseline time point. The results of the study support the claim that the BÜHLMANN fCAL turbo reagents are stable for 12 weeks in open kit/in-use.
{10}
Analyte stability: Analyte stability was evaluated under the following storage conditions of:
- raw stool specimens at 2–8°C
- stool extracts at -20°C and 2–8°C
The demonstrated stability duration is four days at 2–8°C for raw stool specimen, two months at -20°C for stool extract, and 11 days at 2–8°C for stool extract.
d. Detection limit:
Limit of Blank (LoB)
Four blank samples (extraction buffer) were tested over a three-day period using five replicates in two reagent lots. The study design generated a total of 60 test determinations with each reagent lot and the LoB was estimated as the 95th percentile of the distribution of test results using a non-parametric method. The higher estimate obtained with one lot was used as the claimed LoB. The LoB value was determined to be 16.7 µg/g.
Limit of Detection (LoD)
Four specimen extracts with low levels of calprotectin were tested over three days with each sample tested in five replicates. The study design generated a total of 60 test determinations with each reagent lot. The LoD was calculated using parametric analysis. The study was conducted independently with two reagent lots and the higher estimate obtained with one lot was used as the claimed LoD. The LoD value was determined to be 23.7 µg/g.
Limit of Quantitation (LoQ)
The LoQ was established with six specimen extracts at low calprotectin levels. The samples were measured over three days in five replicates each day to generate a total of 90 replicates. The study was conducted independently with two reagent lots. The mean test result and associated coefficient of variation (%CV) were computed for the lowest calprotectin concentration with imprecision below 20%CV. The LoQ value was determined to be 30.0 µg/g.
e. Analytical specificity:
A total of 19 potential interferents were investigated in this study, including six microorganisms and 13 chemical interferents (e.g., oral pharmaceuticals and nutritional supplements) with four samples with calprotectin concentrations at: near 30.0 µg/g, 100.0 µg/g, 300.0 µg/g, and 550.0 µg/g are listed in the tables below:
11
{11}
Microorganism interferences:
| Microorganism | Concentration (cfu/mL) |
| --- | --- |
| Escherichia coli | 3.3 x 10^{7} |
| Salmonella enterica subsp. enterica | 9.0 x 10^{7} |
| Klebsiella pneumoniae subsp. pneumonia | 5.3 x 10^{7} |
| Citrobacter freundii | 12.9 x 10^{7} |
| Shigella flexneri | 5.0 x 10^{7} |
| Yersinia enterocolitica subsp. enterocolitica | 9.8 x 10^{7} |
Drugs and endogenous factor interferences:
| Trade name | Active component | Concentration (mg/50 mg stool) |
| --- | --- | --- |
| gyno-Tardyferon | Iron (II) sulfate | 0.11 |
| Prednisone | Prednisone | 0.31 |
| Imurek | Azathioprine | 0.19 |
| Salofalk | Mesalamine; 5-ASA | 5.21 |
| Agopton | Lansoprazole | 0.18 |
| Asacol | Mesalamine; 5-ASA | 2.50 |
| Vancocin | Vancomycin | 2.00 |
| Sulfamethoxazol | Sulfamethoxazole | 1.60 |
| Trimethoprim | Trimethoprim lactate | 0.35 |
| Ciproxine | Ciprofloxacin | 1.25 |
| Vitamin E | DL-α Tocopherol Acetate | 0.30 |
| Bion 3 | Multivitamin | 1.06 |
| Hemoglobin | Hemoglobin | 1.25 |
The tested potential interferents exhibited a change of <10% in the mean BÜHLMANN fCAL turbo test results when added to stool specimen extracts at various levels of fecal calprotectin.
f. Assay cut-off:
The assay cut-offs for the BÜHLMANN fCAL turbo are:
| Calprotectin Concentration | Interpretation | Follow-Up |
| --- | --- | --- |
| < 80 μg/g | Normal | None |
| 80 μg/g ≤ x ≤ 160 μg/g | Gray-zone/Borderline | Retest within 4–6 weeks |
| > 160 μg/g | Elevated | Repeat as needed |
{12}
g. Carry over:
Three sample pools (low, medium, and high) were tested over 5 days, in one run per day, with three results generated per sample pool per run. One reagent lot was used in the study. No sample carry-over with the BÜHLMANN fCAL turbo test on Roche cobas 6000 c501 instrument was detected.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 248 samples were obtained for the method comparison study; samples with the test results within the measuring range for both BÜHLMANN fCAL ELISA and BÜHLMANN fCAL turbo (n = 220) were included in the regression analysis. The study subject population consisted of: 76 samples from patients with IBD (61 adults and 15 pediatric); 56 samples from patients with IBS (43 adults and 13 pediatric); 32 samples from patients with other GI conditions (30 adults and 2 pediatric); and 56 samples from normal healthy adults. The linear regression analysis obtained from Passing-Bablok regression analysis comparing BÜHLMANN fCAL turbo test results to corresponding BÜHLMANN fCAL ELISA test results is shown below:
| Lot | N | Range (μg/g) | Slope (95% CI) | Intercept (95% CI) | R | % Bias at 80 μg/g (95% CI) | % Bias at 160 μg/g (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| All | 220 | 30.6–1,763.9 | 1.025 (0.990–1.058) | -4.5 (-8.7–0.3) | 0.972 | -3.1 (-7.2–0.5) | -0.3 (-2.4–2.7) |
| 1 | 74 | 30.6–1,120.4 | 1.026 (0.951–1.086) | -7.3 (-16.4–0.5) | 0.975 | -6.5 (-13.4 to -0.2) | -1.9 (-7.1–3.2) |
| 2 | 73 | 32.1–1,421.6 | 0.996 (0.951–1.052) | -3.2 (-12.4–1.6) | 0.974 | -4.5 (-14.7 to -0.6) | -2.5 (-8.0–0.5) |
| 3 | 73 | 35.1–1,763.9 | 1.045 (0.987–1.120) | 0.1 (-8.0–6.6) | 0.975 | 4.6 (-1.4–11.4) | 4.5 (0.4–10.2) |
All 248 samples were used in a qualitative assessment of the assay. The study subject population consisted of 77 samples from patients with IBD, 65 samples from patients with IBS, 37 samples from patients with other GI conditions, and 69 samples from normal healthy adults and were additionally tested for agreement analysis using
{13}
BÜHLMANN fCAL turbo and the predicate assay (BÜHLMANN fCAL ELISA) in accordance with each assay's instructions for use. The comparison analysis was performed using cut-offs of both assays and the number of study subjects at each level.
| | Subject number in BÜHLMANN fCAL ELISA range (μg/g) | | | | |
| --- | --- | --- | --- | --- | --- |
| | | < 80 | 80–160 | > 160 | Total |
| Subject number in BÜHLMANN fCAL turbo range (μg/g) | < 80 | 84 | 10 | 0 | 94 |
| | 80–160 | 8 | 41 | 6 | 55 |
| | > 160 | 0 | 7 | 92 | 99 |
| Total | | 92 | 58 | 98 | 248 |
Estimates of percent agreement, positive percent agreement (PPA) and negative percent agreement (NPA), between BÜHLMANN fCAL ELISA and BÜHLMANN fCAL turbo assay results area shown below:
| Metric | Estimate | 95% C.I. |
| --- | --- | --- |
| PPA at 80μg/g cutoff | 146/156 = 93.6% | [88.5%, 96.9%] |
| NPA at 80μg/g cutoff | 84/92 = 91.3% | [83.6%, 96.2%] |
| PPA at 160μg/g cutoff | 92/98 = 93.9% | [87.1%, 97.7%] |
| NPA at 160μg/g cutoff | 143/150 = 95.3% | [90.6%, 98.1%] |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity/clinical specificity:
The study subject population consisted of 337 subjects including 274 adults (above age 21) and 63 pediatric (age between 4 and 21) patients. Per standard clinical practice, all patients presenting with signs and symptoms suggestive of IBS or IBD, including patients previously diagnosed with IBD and referred for endoscopic investigation due to a suspected flare or routine exam, were enrolled and referred for endoscopy. The final clinical diagnosis of IBD, IBS, or other GI disorder was determined by the clinical investigator based on the results of the endoscopy as well
{14}
as other clinical findings, e.g. histology and Rome III criteria. The final diagnoses of the subjects were as follows: 135 IBD (102 adult and 33 pediatric), 130 IBS (102 adult and 28 pediatric), and 72 other gastrointestinal disorders other than IBD or IBS (70 adults and 2 pediatric). Thirteen clinical sites in the U.S. enrolled and determined the final clinical diagnoses.
| Final Diagnosis | Subject number and percent in BÜHLMANN fCAL turbo range (μg/g) | | | |
| --- | --- | --- | --- | --- |
| | <80 | 80-160 | >160 | Total |
| IBD | 12 (8.9%) | 15 (11.1%) | 108 (80.0%) | 135 (100%) |
| IBS | 99 (76.2%) | 15 (11.5%) | 16 (12.3%) | 130 (100%) |
| GI Other | 51 (70.8%) | 7 (9.7%) | 14 (19.4%) | 72 (100%) |
## IBD vs. non-IBD comparison:
Estimates of sensitivity and specificity along with 95% confidence intervals were calculated for the BÜHLMANN fCAL turbo as an aid in diagnosis of IBD by comparing the test results between IBD and non-IBD (IBS and other gastrointestinal disorders) patients at ≥ 4 years of age, using test cutoffs at 80 and 160 μg/g.
| Positive cases considered at ≥80 μg/g | Clinical Diagnosis | | Total | |
| --- | --- | --- | --- | --- |
| | | IBD | | Non-IBD |
| BÜHLMANN fCAL turbo | Positive | 123 | 52 | 175 |
| | Negative | 12 | 150 | 162 |
| | Total | 135 | 202 | 337 |
| Sensitivity = 91.1%; 95% C.I. (85.0%, 95.3%) | | | | |
| Specificity = 74.3%; 95% C.I. (67.7%, 80.1%) | | | | |
| Positive cases considered at ≥160 μg/g | Clinical Diagnosis | | Total | |
| --- | --- | --- | --- | --- |
| | | IBD | | Non-IBD |
| BÜHLMANN fCAL turbo | Positive | 108 | 30 | 138 |
| | Negative | 27 | 172 | 199 |
| | Total | 135 | 202 | 337 |
| Sensitivity = 80.0%; 95% C.I. (72.3%, 86.4%) | | | | |
| Specificity = 85.1%; 95% C.I. (79.5%, 89.8%) | | | | |
## IBD vs. IBS diagnosis:
Estimates of sensitivity and specificity along with 95% confidence intervals were calculated for the BÜHLMANN fCAL turbo as an aid in the discrimination between
{15}
IBD and IBS in all subjects (≥4 years of age), using test cutoffs at 80 and 160 μg/g.
| Positive cases considered at ≥80 μg/g | Clinical Diagnosis | | Total | |
| --- | --- | --- | --- | --- |
| | | IBD | | IBS |
| BÜHLMANN fCAL turbo | Positive | 123 | 31 | 154 |
| | Negative | 12 | 99 | 111 |
| | Total | 135 | 130 | 265 |
| Sensitivity = 91.1%; 95% C.I. (85.0%, 95.3%) | | | | |
| Specificity = 76.2%; 95% C.I. (67.9%, 83.2%) | | | | |
| Positive cases considered at ≥160 μg/g | Clinical Diagnosis | | Total | |
| --- | --- | --- | --- | --- |
| | | IBD | | IBS |
| BÜHLMANN fCAL turbo | Positive | 108 | 16 | 124 |
| | Negative | 27 | 114 | 141 |
| | Total | 135 | 130 | 265 |
| Sensitivity = 80.0%; 95% C.I. (72.3%, 86.4%) | | | | |
| Specificity = 87.7%; 95% C.I. (80.8%, 92.8%) | | | | |
b. Other clinical supportive data (when a. is not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Stool samples were obtained from 141 healthy normal adults (≥ 21 years of age) with no symptoms or signs of gastrointestinal disease. The test results were categorized by the assay cut-offs below:
| | Calprotectin level by BÜHLMANN fCAL turbo | | | |
| --- | --- | --- | --- | --- |
| | < 80 μg/g | 80 - 160 μg/g | > 160 μg/g | Total |
| Subject number (%) | 106 (75.2%) | 18 (12.8%) | 17 (12.0 %) | 141 (100 %) |
{16}
N. Proposed Labelling:
The labelling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.