PREVENA PLUS Incision Management System (No Ag); PREVENA PLUS DUO Incision Management System (No Ag)
K190697 · Kci USA, Inc. · OMP · Nov 22, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190697
Device Name
PREVENA PLUS Incision Management System (No Ag); PREVENA PLUS DUO Incision Management System (No Ag)
Applicant
Kci USA, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Nov 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PREVENA PLUS Incision Management System (No Ag) is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closed environment and removing exudate via the application of negative pressure wound therapy. The PREVENA PLUS DUO Incision Management System (No Ag) is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.
Device Story
System manages surgical incision environment; applies continuous -125 mmHg negative pressure via therapy unit (PREVENA PLUS 125, ACTIVAC, V.A.C. ULTA, or V.A.C. RX4) to dressing placed over sutured/stapled incision. System components include therapy unit, disposable canister, tubing, and dressing with occlusive drape. Dressing protects incision from external contamination; negative pressure draws incision edges together and removes fluid into canister. Applied in operating room immediately post-surgery; used for up to 7 days. Healthcare providers monitor fluid collection and system function. Benefits include exudate removal and maintenance of closed environment to support healing.
Clinical Evidence
Bench testing only. Evaluations included system performance testing, biocompatibility testing per ISO 10993-1, and material specification equivalency testing for horizontal wicking performance to ensure safety, efficacy, and conformance to design specifications.
Technological Characteristics
Powered suction pump delivering continuous -125 mmHg negative pressure. Components: therapy unit, disposable canister, tubing, and dressing with occlusive drape. Interface layer modified to remove silver. SENSAT.R.A.C. technology integrated. Biocompatibility per ISO 10993-1. System is non-sterile/sterile components as applicable. Connectivity via tubing sets.
Indications for Use
Indicated for patients with surgical incisions closed via sutures or staples that continue to drain, requiring negative pressure wound therapy to manage the incision environment and remove exudate.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
PREVENA PLUS DUO Incision Management Systems with PEEL & PLACE Dressings (K161897)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2019
KCI USA, Inc. Patricia Lopez Sr. Regulatory Affairs Specialist 6203 Farinon Drive San Antonio, Texas 78249
Re: K190697
Trade/Device Name: PREVENA PLUS Incision Management System (No Ag) PREVENA PLUS DUO Incision Management System (No Ag) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 23, 2019 Received: October 24, 2019
Dear Patricia Lopez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kimberly M. Ferlin, Ph.D. Assistant Director (acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name PREVENA PLUS™ Incision Management System (No Ag)
Indications for Use (Describe)
The PREVENA PLUS Incision Management System (No Ag) is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closed environment and removing exudate via the application of negative pressure wound therapy.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
Device Name PREVENA PLUS DUO™ Incision Management System (No Ag)
Indications for Use (Describe)
mulations for ood (Socially)
The PREVENA PLUS DUO Incision Management System (No Ag) is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for KCI, an Acelity company. The logo features a stylized sphere on the left, with a portion of it colored in a gradient of purple and pink. To the right of the sphere are the letters "KCI" in a bold, sans-serif font, colored in gray. Below the letters, in a smaller font, is the text "AN ACELITY COMPANY".
# 510(K) SUMMARY
| Submitter Information [21 CFR 807.929(a)(1)] | |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | KCI USA, Inc. (Kinetic Concepts, Inc.) |
| Address | 6203 Farinon Drive<br>San Antonio, TX 78249 |
| Establishment<br>Registration<br>Number | 1625774 |
| Contact person,<br>by name, title,<br>phone number,<br>fax number and e-<br>mail address | Shannon Scott<br>Senior Director, Regulatory Affairs<br>Telephone: (210) 515-7433<br>E-mail: Shannon.Scott@Acelity.com |
| Alternate contact<br>person, by name,<br>title, phone<br>number, fax<br>number and e-<br>mail address | Terrie McDaniel<br>Interim VP, Global Regulatory Affairs<br>Telephone: (210) 515-4248<br>E-mail: Terrie.McDaniel@Acelity.com |
| Date prepared | 22 Nov 2019 |
| Name of the device [21 CFR 807.92(a)(2)] | |
| Trade or<br>proprietary name | PREVENA PLUS Incision Management System (No Ag) PREVENA PLUS DUO Incision Management System (No Ag) |
| Common or usual<br>name | Negative Pressure Wound Therapy System |
| Classification<br>name | Negative Pressure Wound Therapy Powered Suction Pump (and<br>components) |
| Classification<br>panel | General and Plastic Surgery |
| Regulation | 878.4780 |
| Product Code(s) | OMP |
| Legally marketed<br>device(s) to which<br>equivalence is<br>claimed<br>[21 CFR<br>807.92(a)(3)] | PREVENA PLUS Incision Management System (K173426) PREVENA PLUS DUO Incision Management Systems with PEEL &<br>PLACE Dressings (K161897) |
| Device<br>description<br>[21 CFR<br>807.92(a)(4)] | PREVENA PLUS Incision Management System with PREVENA PEEL<br>& PLACE Dressing (No Ag) with SENSAT.R.A.C. Technology PREVENA PEEL & PLACE Dressing (No Ag) with SENSAT.R.A.C.<br>Technology |
| | |
| | • PREVENA PLUS Incision Management System with PREVENA PLUS<br>CUSTOMIZABLE Dressing (No Ag) |
| | • PREVENA PLUS CUSTOMIZABLE Dressing (No Ag) with<br>SENSAT.R.A.C. Technology |
| | • PREVENA PLUS DUO Incision Management System with PEEL &<br>PLACE Dressings (No Ag) with SENSATRAC Technology |
| | Negative pressure wound therapy system for application to surgically<br>closed incisions. The system consists of a therapy unit (PREVENA PLUS™<br>125 Therapy Unit, ACTIVAC™ Therapy Unit, V.A.C. ULTA™ Therapy Unit, or<br>V.A.C. RX4™ Therapy Unit), dressing and canister. The pump in the therapy<br>unit delivers continuous negative pressure at -125 mmHg through tubing to<br>a dressing placed over the incision site. Application of negative pressure<br>wound therapy to an incision site that is closed via staples or sutures helps<br>draw the incision edges together and remove fluid from the incision site into<br>a canister fitted to the therapy unit. The occlusive drape of the dressing<br>provides a negative pressure environment and protects the incision from<br>external contamination. The systems are applied to the incision site<br>immediately after surgery for up to seven (7) days depending on the<br>surgeon's preference. |
| Indications for<br>use<br>[21 CFR<br>807.92(a)(5)] | The PREVENA PLUS Incision Management System (No Ag) (or PREVENA<br>PLUS DUO Incision Management System (No Ag)) is intended to manage<br>the environment of surgical incisions that continue to drain following<br>sutured or stapled closure by maintaining a closed environment and<br>removing exudate via the application of negative pressure wound therapy. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KCI, an Acelity company. The logo features a sphere with a purple swirl at the bottom left. To the right of the sphere are the letters "KCI" in a sans-serif font. Below the letters is the text "AN ACELITY COMPANY" in a smaller font.
Comparison of the Technological Characteristics (i.e., design, material, chemical composition, energy source) with the Predicate Device [21 CFR 807.92(a)(6)]
Negative Pressure Wound Therapy is the technological principal for both the subject and predicate devices
# At a high level, the subject device and predicate device are based on the following same technological elements:
- . The dressings are applied over the incision site in the operating room.
- . The dressing is connected to the selected therapy unit (PREVENA PLUS™ 125 Therapy Unit, ACTIVAC™ Therapy Unit, V.A.C. ULTA™ Therapy Unit, or V.A.C. RX4™ Therapy Unit) via a disposable canister and, where appropriate, tubing connectors.
- . The selected negative pressure therapy unit provides -125 mmHg of negative pressure continuously to the dressing for a maximum of 7 days.
- A tubing set connects the dressing to the canister to deliver negative pressure from the ● pump and for removal of incision fluids.
- Incision fluid is collected into the disposable canister in the therapy unit.
# The following technological differences exist between the subject and predicate device:
- The dressings have been modified by not adding silver to the patient/wound contacting . interface layer (wicking interface layer)
- Incorporation of the SENSAT.R.A.C. Pad Tubing set associated modifications are, .
- . SENSAT.R.A.C. Pad Tubing set has replaced the PREVENA Tubing set and PREVENA V.A.C. Connector
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for KCI, an Acelity company. The logo features a stylized sphere on the left, with a purple and white gradient. To the right of the sphere are the letters "KCI" in a sans-serif font, rendered in a metallic gray color. Below the letters, in a smaller font size, is the text "AN ACELITY COMPANY".
- V.A.C. Y-Connector has replaced the PREVENA Y-Connector, a system ● component of the PREVENA PLUS DUO Incision Management Systems (No Ag)
### Performance Data [21 CFR 807.92(b)]
### Summary of tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]
The PREVENA PLUS Incision Management System (No Aq) and PREVENA PLUS DUO Incision Management Systems (No Ag) were evaluated to assure safety, efficacy, conformance to design specifications and equivalence to the predicate device. The following tests were conducted:
- . System Performance Testing
- . Biocompatibility testing in accordance with ISO 10993-1
- . Material specification equivalency testing with respect to Horizontal Wicking performance
### Conclusions drawn [21 CFR 807.92(b)(3)]
The PREVENA PLUS Incision Management System (No Ag) and PREVENA PLUS DUO Incision Managements Systems (No Ag) are substantially equivalent to the predicate device systems with respect to indications for use and technology.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.