← Product Code [FDF](/productcode/FDF) · K190669

# Endotics (K190669)

_Era Endoscopy S.R.L. · FDF · Jan 3, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190669

## Device Facts

- **Applicant:** Era Endoscopy S.R.L.
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** Jan 3, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Endotics System is intended to provide 360° visualization and diagnostic/therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, transverse colon, cecum and ileocecal valve) for endoscopy and endoscopic surgery. The Endotics Probe (colonoscope component of the Endotics Colonoscope System) is a single use disposable device. The Endotics Probe cannot be reprocessed.

## Device Story

Endotics is a self-propelled, disposable robotic colonoscopy system. It consists of a single-use flexible probe and an electro-pneumatic workstation with a command console. The probe features a CMOS digital camera, LED light source, and a steering system (180° flexion). It utilizes a 'caterpillar' motion principle, anchoring to the intestinal wall via pneumatic suction to advance without external pushing. The operator controls movement via a joystick at the workstation. The system includes a working channel for surgical tools. The workstation displays real-time video and probe status. Used in clinical settings by physicians, the device provides diagnostic and therapeutic access to the colon. Benefits include reduced patient discomfort compared to traditional colonoscopy, as the self-propelling mechanism minimizes trauma and the need for sedation.

## Clinical Evidence

Clinical study (n=20) compared Endotics to traditional colonoscopy. Primary endpoints: efficacy (diagnostic accuracy), safety, and tolerability. Results: Endotics demonstrated non-inferiority to traditional colonoscopy. Cecal intubation rate was 100%. No adverse events (perforations, bleeding) were recorded. Patient-reported discomfort (1.45 vs 6.5) and pain (1.3 vs 7.4) were significantly lower for Endotics compared to traditional colonoscopy. Animal studies (n=3 pigs) confirmed the clamping mechanism causes no significant bowel wall lesions.

## Technological Characteristics

Materials: Silicone, ABS, PVC. Sensing: CMOS 1M pixel camera, LED illumination. Energy: Pneumatic propulsion, 100-240V AC power. Dimensions: 17.5mm insertion diameter, 2030mm working length. Connectivity: Electro-pneumatic interface to workstation. Sterilization: Ethylene Oxide. Software: Moderate level of concern, ISO 62304 compliant.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Aer-o-scope Colonoscope System ([K141286](/device/K141286.md))
- Invendo C25 Colonoscopy System ([K121582](/device/K121582.md))
- OLYMPUS CF-Q160 ([K001241](/device/K001241.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 3, 2020

Era Endoscopy s.r.1. Alberto Arena, Eng Via Boccioni, 1 56037 Peccioli, ITALY

K190669 Re: Trade/Device Name: Endotics Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: November 25, 2019 Received: December 2, 2019

Dear Alberto Arena:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling

{1}------------------------------------------------

(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, PhD Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

### K190669

Device Name Endotics

Indications for Use (Describe)

The Endotics System is intended to provide 360° visualization and diagnostic/therapeutic access to the adult lower volvo for and eases and and limited to, the anus, rectum, sigmoid colon, transvers colon, cecum additioner valve) for endoscopy and endoscopic surgery.

The Endotics Probe (colonoscope component of the Endotics Colonoscope System) is a single use disposable device. The Endotics Probe cannot be reprocessed.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for regarding this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5 510(k) Summary

#### I. SUBMITTER

| Applicant Information | Era Endoscopy s.r.l.<br>Via Boccioni, 1 56037<br>Peccioli (Pi) - Italy                                                                       |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact               | ing. Alberto Arena (Quality and Regulatory Manager)<br>Phone: +39 0587 635235<br>Fax Number: +39 0587 829970<br>e-mail: qualita@endotics.com |
| Date Prepared         | 1st February 2019                                                                                                                            |
|                       | <b>II. DEVICE</b>                                                                                                                            |
| Device Trade Name     | ENDOTICS                                                                                                                                     |
| Common Name           | Colonoscope And Accessories                                                                                                                  |
| Device Class          | II                                                                                                                                           |
| Classification Name   | Endoscope and accessories                                                                                                                    |
| Regulation            | 21 CFR 876.1500                                                                                                                              |
| Product Code          | FDF                                                                                                                                          |

#### III. PREDICATE DEVICE

## Legally Marketed Predicate Device (s)

| 510(k) Number | Product Code | Regulation Number | Trade Name                        | Applicant                              |
|---------------|--------------|-------------------|-----------------------------------|----------------------------------------|
| K141286       | FDF          | 21 CFR § 876.1500 | Aer-o-scope<br>Colonoscope System | GI View, Ltd.                          |
| K121582       | FDF          | 21 CFR § 876.1500 | Invendo C25<br>Colonoscopy System | Invendo Medical<br>GmbH                |
| K001241       | FDF          | 21 CFR § 876.1500 | OLYMPUS CF-Q160                   | Olympus Medical<br>Systems Corporation |

#### IV. Device Description

Endotics is an endoscopic system to be used for the examination of the final section of the intestine, the ends of which are marked by the anal sphincter and by the ileocecal valve, known as the colonrectum.

{4}------------------------------------------------

The system comprises two parts, which can be sold separately, namely: the"disposable probe", and the "workstation", equipped with a command console.

Thanks to its flexibility, the disposable probe can easily adapt its shape to that of the colon. The motion, based on the "caterpillar" principle, is ensured by an anchoring system, placed by the ends of a flexible body.

The colonoscope includes on its distal segment a viewing system, featuring a CMOS digital camera and a LED light source.

The probe is further equipped with a steering system, which enhances the efficiency in terms of both motion and vision (maximum flexion angle: 180° in every direction).

The endoscope is activated pneumatically.

This device allows the use of specific surgical tools through a working channel.

The exit of the working channel is centred compared to the robot and it is located near the camera, which is therefore slightly decentred. When the tool exits the probe, it is immediately detected and the operator can then visually pilot it over the zone to be treated.

The anchoring system performs a double action. By extracting the air from inside the intestine it enables the intestine linings to approach the robot precisely at the point in which they will then be mechanically anchored. This system produces a steady hold on the intestine lining but prevents any damage or trauma from occurring.

Endotics Workstation is composed of an electro-pneumatic system, controlled by a Panel PC which displays the different progress phases of the device and the images delivered by the internal camera, and of a command console for controlling the probe's movements.

Only the Endotics disposable probe can be connected to this workstation. The probe is linked to the workstation by means of an electro-pneumatic connector is placed on a twistable arm which is connected to the machine's main block. This arm is fitted on the inside with pneumatic tubes and electric cables which connect the probe to the workstation's electro-pneumatic system and which connect the camera (placed by the probe's head) to a video system. The arm is also equipped with an electronic board which acts as an opto-insulator between the camera and the electric features of the workstation. The electro-pneumatic distributor is controlled by a computerised system, on which specifically designed command software has been installed.

The monitor also displays the probe's status, indicating its current motion phase e the captured pictures. The associated accessories include:

{5}------------------------------------------------

- Stiffening accessory
- -Waterjet accessory

#### V. Indication for Use

The Endotics System is intended to provide 360° visualization and diagnostic/therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, transverse colon, cecum and ileocecal valve) for endoscopy and endoscopic surgery.

The Endotics Probe (colonoscope component of the Endotics Colonoscope System) is a single use disposable device. The Endotics Probe cannot be reprocessed.

### VL. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

Colonoscopy is the technological principle for both the Endotics and predicate devices. It is based on the use of endoscopic instrumentation for the examination of the final section of the intestine, the ends of which are marked by the anal sphincter and by the ileocecal valve, known as the colonrectum.

The Endotics and predicate devices (K141286 and K121582) are based on the following same technological elements:

- । A controlling workstation
- -A single use disposable colonoscope

Like the predicate Aer-o-scope Colonoscope System and Invendo, the ENDOTICS Disposable Probe is not directly manually manipulated by the operator and moves through the colon under the direction and control of the operator using a joystick.

The only difference between Endotics and the predicates consists in propulsion mechanism inside the intestine: Endotics probe is not a pushed device, while the predicates result in externally pushed devices.

In the follow table are reported a side-by-side comparison of technological characteristics of Endotics device and the predicate devices

|                                                              | Endotics<br>K190669                  | Aer-o-scope<br>K141286                            | Invendo C25<br>K121582                          | Olympus CF-Q160<br>K001241      |
|--------------------------------------------------------------|--------------------------------------|---------------------------------------------------|-------------------------------------------------|---------------------------------|
| Viewing direction                                            | Forward                              | OMNY                                              | Forward                                         | Forward                         |
| Field of view                                                | 140°                                 | Front 90°<br>Angular 360°                         | 114°                                            | Forward 140°                    |
| Depth of Field                                               | 3-100 mm                             | Omni- 11-50 mm<br>Front- 10 to 60 mm              | -                                               | 3-100 mm                        |
|                                                              | Endotics<br>K190669                  | Aer-o-scope<br>K141286                            | Invendo C25<br>K121582                          | Olympus CF-Q160<br>K001241      |
| Chip                                                         | CMOS 1M<br>pixels                    | CMOS 1.3 M pixels                                 | CMOS<br>0.25M<br>Pixels                         | CCD color                       |
| Illumination                                                 | LEDs                                 | LEDs                                              | LEDs                                            | Light guide                     |
| Max. diameter<br>insertion point                             | 17.5 mm                              | 13 mm                                             | 18 mm                                           | 12.8 mm                         |
| Multilumen max<br>diameter                                   | 7.5 mm                               | 7.5 mm                                            | 18 mm                                           | 12.8 mm                         |
| Working length                                               | 2030 mm                              | 2000 mm                                           | 2100 mm                                         | 1680 mm                         |
| Total length                                                 | 2390 mm                              | 2300 mm                                           | 3600 mm                                         | -                               |
| Bending capacity                                             | 180° in all<br>directions            | 160° in all<br>directions                         | 180° in all<br>directions                       | 180° Up-down<br>160° right-left |
| Diameter working<br>channel                                  | 3 mm                                 | -                                                 | 3.1 mm                                          | 3.7 mm                          |
| N. of working<br>channels                                    | 1                                    | -                                                 | 1                                               | 1                               |
| N. of other<br>channels<br>(suction/insufflati<br>on Fluids) | 1 air/water<br>2 air                 | 1                                                 | 1                                               | 1 air/water 1 water             |
| Power supply base<br>unit                                    | 100-240 V<br>50/60Hz                 | 100-240 V 50/60Hz                                 | 100-240 V<br>50/60Hz                            | 100-240 V<br>50/60Hz            |
| Max internal<br>pressure<br>base unit                        | 450 KPa                              | 450 KPa                                           | 600 KPa                                         | -                               |
| Max internal<br>vacuum base unit                             | -75Kpa                               | -                                                 | -75KPa                                          | -                               |
| Workstation<br>weight                                        | 37 Kg<br>w/Panel PC                  | 167 Kg                                            | -                                               | -                               |
| Workstation<br>dimensions                                    | 280 (H) x<br>650 (W) x<br>580 (D) mm | 1200 (H) x 400 (W)<br>x 600 (D) mm                | -                                               | -                               |
| Main materials in<br>contact with<br>patient                 | Silicone,<br>ABS and<br>PVC          | Polyurethane,<br>PEBAX and<br>hydrophilic coating | Silicone,<br>polycarbonat<br>e and<br>lubricant | -                               |
| Type of<br>sterilization                                     | Ethylene<br>Oxide                    | -                                                 | Ethylene<br>Oxide                               | -                               |
| Disposable                                                   | YES                                  | YES                                               | YES                                             | NO                              |

{6}------------------------------------------------

Moreover, the results of the bench tests and clinical tests show that differences in propulsion systems provide equivalent levels of safety and efficacy compared to the predicate devices and that clinical 

{7}------------------------------------------------

performance of Endotics compared to conventional colonoscopy systems (see predicate K001241), have the similar characteristics.

The product specification, functionality and indication for use of the ENDOTICS are the same or very similar to the legally marketed, claimed predicate devices for the purpose of this 510(k) submission.

#### VII. Performance Data

The ENDOTICS has been tested and is in compliance with:

Biocompatibility (according to ISO 10993)

The battery of testing included the following tests:

- Cytotoxicity Test Report Nº2249-1/13 (Attach 13) B Cytotoxicity Test) -
- -Sensitization - Test Report Nº2249-4/13 and Nº2249-3/13 (Attach 13 C Sensitization test Apolar and Attach 13 D Sensitization test Polar)
- Irritation/Intracutaneous Test Report N°2249-2/13 (Attach 13) E Irritation Test) -

The disposable probe is considered tissue contacting for a duration of less than 24 hours.

Electrical Safety, Electromagnetic Compatibility EMC (according to EN 60601-1 and EN 60601-1-2)

The battery of testing included the following tests:

- Electrical Safety Test: Test Report R-EL-27-0414-01A (see attachment 15 A) -
- Electromagnetic Compatibility: Test Report R-EM-227-0119-01A (see attachment 15 B) -

Sterilization Process Validation (According to ISO 11135)

The battery of testing included the following tests:

- Validation Protocol of Sterilization Process (see Attachment 12) A.) -
- -Validation Report of Sterilization Process (see Attachment 12 B)
- Sterilization Process Validation ERA Endoscopy and Steril Verona Documentation Annex 1 -(see Attachment 12 C)
- Sterilization Process Validation Result of Physical Performance 2 -(see Annex Attachment 12 D)
- -Sterilization Process Validation - Result of Microbiological Performance - Annex 3 (see Attachment 12 E)

## Shelf life (According to ISO 11607-1)

- Medical Devices Shelf-Life (3 years) Final Report (see Attach 12 F Shelf life 3 years validation) -

{8}------------------------------------------------

ETSI EN 300-019-2.2 V2.2.1 Enviromental condition and enviromental test for telecomunications equipment; Part 2-2 Specification of enviromental tests; Transportation

Test Report R-EL-227- 1014 - 02A (Attach 12 G Enviromental Conditions and trasportation test for Workstation)

Software Validation (According to ISO 62304)

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or operator.

The documentation of Software Validation can be found in the annexes of Section 14

Usability (According to IEC 62366-1:2015)

The usability of the product was conducted in accordance with IEC 62366-1:2015 and concerns various aspects of the device use.

The documentation can be found in the annexes of Section 16:

- Usability File (see Attachment 16 A) -
## Mechanical Testing

Bench Test of the Endotics were performed to support the efficacy/safety of the device and substantial equivalence.

Tests carried out have covered the following areas (see Attachment 16 B Bench Testing):

- Mechanical resistance test -
- Fatigue strength test -
- Breaking Pressure Test -
- -Tensile Strength Test
- Mechanical Characterization Test -
- -Simulated use testing

## ANIMAL STUDIES

Era Endoscopy during the development of the Endotics System performed a clinical study on animals (Attach 17 A Animal test). The study was designed to provide objective parametric data that allow excluding that the device is imposing relevant trauma with acute or delayed response of the colon tissue.

{9}------------------------------------------------

The study was carried out in the domestic pig model (n=3), female pigs, the post-operative followup survival period was 1 week.

The clamping mechanism does not create significant lesions in the bowel wall, such as mucosal lacerations. No bleeding was found as a result of the interaction of the clamping mechanism with the colonic wall. Also in case of prolonged clamping of 5 minutes duration no relevant lesions were seen. The marks on the colonic wall inherent to the procedure proved to be temporary and left no visible remains after a follow-up period of 7 days.

In these regards the clamping system of the device can be considered safe in the animal model. Bowel wall thickness in the adult pig correlates well with the human, therefore we do not anticipate the device clamping mechanism should be harmful for the human colon.

## CLINICAL STUDIES

Era Endoscopy, during the Endotics project development, carried out thr following clinical investigations to demonstrate the safety and efficacy of the device.

- -Functional evaluation of the Endotics System, a new disposable self-propelled robotic colonoscope: in vitro (see Attachment 18 A).
- -Endotics system vs colonoscopy for the detection of polyps (see Attachment 18 B).
- Functional assessment in terms of safety and efficacy of the Endotics system in endoscopic colorectal studies (Attachment 18 C).

| Title                  | Functional assessment in terms of safety and efficacy of the Endotics® system in        |
|------------------------|-----------------------------------------------------------------------------------------|
|                        | endoscopic colorectal studies                                                           |
| Methodology            | Single center, controlled and randomized NO PROFIT study                                |
| Study Description      | A controlled and randomized NO PROFIT study for the functional evaluation in            |
|                        | terms of safety, effectiveness and efficacy of the Endotics® system, a single-use       |
|                        | semi-automatic robotic colonoscope, in the identification of colorectal diseases.       |
| Number of patients     | Planned: 20; Analysed: 20                                                               |
| Objectives             | To evaluate the diagnostic accuracy (efficacy), safety and tolerability for the patient |
|                        | of the Endotics system in comparison with the traditional Colonoscopy                   |
| Study Population       | 20 (twenty) patients, of both sexes, aged between 18 and 65, meeting the inclusion      |
|                        | / exclusion criteria, have been recruited, consecutively selected among those in        |
|                        | whom routine colonoscopy was indicated.                                                 |
| Criteria for inclusion | Patients aged between 18 and 65 in which a colonoscopy examination was                  |
|                        | indicated.                                                                              |
| Study period           | 5 months from May to September 2018                                                     |
| Participant Duration   | 3 days for bowel preparation and 2 hours for both procedures (ENDOTICS and CT).         |
| Statistical methods    | The statistical analysis has been conducted according to the principles of the PP/AT    |
|                        | analysis (per protocol / as treated). All participants with a recorded exam have been   |
|                        | randomized.                                                                             |
| Primary Endpoints      | Efficacy                                                                                |

The results of this latest study are summarized in the table below:

{10}------------------------------------------------

|                        | Statistical comparison between the PDR/ADR of the TC and that of the ENDOTICS.<br>ENDOTICS have to result not inferior if compared to TC with pvalue< 0.05.<br>Moreover, ADR (ENDOTICS) should be greater or equal than 25% (according to<br>ASGE guidelines).<br>Safety<br>Statistical comparison between EA/SEA occurring with the traditional colonoscopy<br>(TC) compared to those occurred with the ENDOTICS. ENDOTICS have to be not<br>inferior to TC with P<0.05.                                                                                                                                                                                                       |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Secondary<br>Endpoints | Cecal intubation rate:<br>Statistical comparison between CIR (ENDOTICS) and CIR (TC). CIR (ENDOTICS)<br>has to be not inferior to CIR (TC) with pvalue< 0.05. Moreover, CIR (ENDOTICS)<br>should be greater or equal than 90%.<br>Tolerability: Statistical comparison of parameters related to pain and discomfort                                                                                                                                                                                                                                                                                                                                                             |
|                        | between ENDOTICS and TC. ENDOTICS has to be superior compared to TC with<br>pvalue< 0.05.<br>Cecal intubation time (CIT) and procedure time (PT): Statistical comparison<br>between CIT/PT of the CT and that of ENDOTICS. CIT/PT (ENDOTICS) must not<br>be inferior compared to CT with p-value<0.05.                                                                                                                                                                                                                                                                                                                                                                          |
| Summary Results        | Efficacy<br>Based on the results of this study, statistically Endotics colonoscopy is at least not<br>less effective and efficient than traditional colonoscopy (see predicate device<br>K001241). Previous studies had already demonstrated the not inferior diagnostic<br>capacity of Endotics if compared to TC, and these data are confirmed by this study.<br>Safety                                                                                                                                                                                                                                                                                                       |
|                        | No adverse events, minor or serious, or deaths were recorded in this study, so it<br>makes no sense to calculate the statistical parameters that were set (mean,<br>standard deviation and confidence interval). No patient requested sedation either<br>during robotic or traditional procedures. Oxygen saturation was monitored for each<br>patient and there were no cases of vagal crisis. No adverse events occurred during<br>polyp's removal, such as perforation with both types of instruments used in the<br>study. Therefore, it can be concluded that Endotics robotic colonoscopy is no less<br>safe than traditional colonoscopy (see predicate device K001241). |

Endotics robotic endoscopic system proved to be effective (ADR compatible with ASGE and CIR guidelines as of 100%), safe (no adverse events, minor or serious, were recorded during the procedure or within the following 30 days) and very well tolerated by patients (on a scale of 0 to 10, it recorded an average discomfort for the entire procedure of 1.45 and an average pain of 1.3). Also compared with the predicate colonoscopy system, Endotics proved to be not inferior and more tolerated by the patient (annoyance ES 1.45 versus 6.5 for TC and pain ES 1.3 versus 7.4 for TC).

#### VIII. Conclusions

Era Endoscopy s.r.l. has determined, by using comparisons and tests, that Endotics is substantially equivalent to the listed predicate devices in terms of intended use, typical clinical use, operational characteristics, and fundamental technological characteristics.

---

**Source:** [https://fda.innolitics.com/device/K190669](https://fda.innolitics.com/device/K190669)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
