K190654 · Institut Straumann AG · NHA · Jun 28, 2019 · Dental
Device Facts
Record ID
K190654
Device Name
Straumann CI RD Ceramic Healing Caps
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Jun 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Healing Caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Healing Caps should be used only with suitable implant connections. The healing components are intended to be used up to 6 months.
Device Story
Straumann CI RD Ceramic Healing Caps are two-piece, screw-retained components used in dental implant procedures. They are placed into the dental implant to protect the internal configuration and manage soft tissue during the healing phase. The device is used in a clinical setting by a dentist or oral surgeon. The caps are made of yttria-stabilized tetragonal zirconia (Y-TZP) and titanium grade 4. They are placed out of occlusion and do not support prosthetic restorations. The device is intended for short-term use (up to 6 months) to facilitate transmucosal healing. By maintaining soft tissue architecture, the device supports the preparation of the site for subsequent prosthetic restoration.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation per ISO 11135 (Half Cycle Overkill Approach) to achieve SAL 10^-6, LAL Endotoxin Analysis (limit 20 EU/device), ISTA 2A transportation testing, and packaging stability studies. Biocompatibility was assessed per ISO 10993-1, including cytotoxicity (ISO 10993-5) and chemical characterization (ISO 10993-18).
Technological Characteristics
Materials: Yttria-stabilized tetragonal zirconia (Y-TZP) and titanium grade 4. Form factor: Two-piece screw-retained healing cap. Dimensions: 5.2 mm diameter; 6.2, 7.2, and 8.7 mm heights. Sterilization: Ethylene oxide (ISO 11135) to SAL 10^-6. Connection: Regular Diameter (RD) interface.
Indications for Use
Indicated for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize, and form soft tissue during the healing process. Intended for use up to 6 months. Contraindicated for use with unsuitable implant connections.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 28, 2019
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K190654
Trade/Device Name: Straumann® CI RD Ceramic Healing Caps Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 31, 2019 Received: June 3, 2019
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190654
Device Name
Straumann® CI RD Ceramic Healing Caps
### Indications for Use (Describe)
Healing Caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing Caps should be used only with suitable implant connections. The healing components are intended to be used up to 6 months.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span>☒</span><span>Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span>☐</span><span>Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Straumann® CI RD Ceramic Healing Caps
510(k) Summary
# 510(k) Summary
#### Submitter 1.1
Straumann USA, LLC (on behalf of Institut Straumann AG) 60 Minuteman Road Andover, MA 01810 Contact Person: Jennifer M. Jackson, MS - Director, Regulatory Affairs Phone Number: +1 (978) 747-2509 Fax Number: +1 (978) 747-0023 E-mail: jennifer.jackson@straumann.com Prepared By: Pierre-Yves Calinon - Regulatory Affairs & Compliance Manager Date of Submission: June 28, 2019
### 1.2 Device
| Trade Name: | Straumann® CI RD Ceramic Healing Caps |
|----------------------|------------------------------------------------------|
| Common Name: | Endosseous Dental Implant Abutment |
| Classification Name: | Endosseous Dental Implant Abutment (21 CFR 872.3630) |
| Regulatory Class: | II |
| Product Code: | NHA |
| Review Panel: | Dental |
#### Predicate Device 1.3
| Primary Predicate: | K180477 – Straumann PURE Ceramic Implant System |
|--------------------|--------------------------------------------------------------|
| Reference Device: | K130808 – Straumann Healing Abutments, Healing Caps, Closure |
| Screws | |
#### Device Description 1.4
The Healing Caps are screws machined into 2 pieces and come in three gingival heights to accommodate individual gingival thickness. The material of the devices is yttria-
{5}------------------------------------------------
Straumann® CI RD Ceramic Healing Caps
510(k) Summary
stabilized tetragonal zirconia (Y-TZP) and titanium grade 4. Healing caps are screwed into the implant to protect the inner configuration in cases of transmucosal healing protocols and are placed out of occlusion and do not support a prosthetic restoration.
### 1.5 Indications for Use
Healing Caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Healing Caps should be used only with suitable implant connections. The healing components are intended to be used up to 6 months.
#### Technological Characteristics 1.6
The proposed Straumann® CI RD Ceramic Healing Caps intended use and fundamental operating principals are identical to the primary predicated device, K180477. The additional components mentioned in the primary predicate device indications for use statement are not subject to this submission as the change only impacts the Healing Caps. The technological characteristics of the subject devices are compared to the primary predicate and reference devices in Table 1 below.
{6}------------------------------------------------
# Straumann® CI RD Ceramic Healing Caps
510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number and<br>name of the<br>device | K190654<br>Straumann® CI RD<br>Ceramic Healing caps | K180477<br>Straumann PURE Ceramic<br>Implants | K130808<br>Straumann Sterile Healing<br>Solution |
| Indications for<br>Use | Healing Caps are intended<br>for use with the Straumann<br>Dental Implant System<br>(SDIS) to protect the inner<br>configuration of the implant<br>and maintain, stabilize and<br>form the soft tissue during<br>the healing process.<br>Healing Caps should be<br>used only with suitable<br>implant connections.<br>The healing components<br>are intended to be used up<br>to 6 months. | Straumann PURE Ceramic<br>lmplant:<br>The Straumann PURE Ceramic<br>Implant is indicated for the<br>restoration of single-tooth gaps and<br>in edentulous or partially edentulous<br>jaws. The prosthetic restorations<br>used are single crowns, fixed partial<br>or full dentures, which are<br>connected to the implants through<br>the corresponding components.<br>Closure and healing caps:<br>Closure and Healing caps are<br>intended for use with the<br>Straumann Dental Implant System<br>(SDIS) to protect the inner<br>configuration of the implant and<br>maintain, stabilize and form the soft<br>tissue during the healing process.<br>Closure and Healing caps should<br>be used only with suitable implant<br>connections. Do not use healing<br>components for longer than 6<br>months.<br>Temporary Abutments:<br>The provisional components are<br>intended to serve as a base for<br>temporary crown or bridge<br>restoration out of occlusion for the<br>Straumann® PURE Ceramic Implant<br>System. The Straumann®<br>Temporary Abutment VITA CAD-<br>Temp® for the Straumann® PURE<br>Ceramic Implant is indicated for<br>temporary usage of up to 180 days.<br>CI RD Straumann PUREbase<br>Abutments:<br>CI RD Straumann PUREbase<br>abutment is a titanium base placed<br>onto Straumann ceramic dental<br>implants to provide support for<br>customized prosthetic restorations<br>and is indicated for screw-retained<br>single tooth or cement-retained<br>single tooth and bridge restorations.<br>All digitally designed copings and/or<br>crowns for use with the Straumann®<br>Variobase Abutment system are<br>intended to be sent to Straumann<br>for manufacture at a validated<br>milling center. | Closure Screws, healing<br>caps, and healing<br>abutments, are intended for<br>use with the Straumann<br>Dental Implant System<br>(SDIS) to protect the inner<br>configuration of the implant<br>and maintain, stabilize and<br>form the soft tissue during<br>the healing process.<br>Customizable healing<br>abutments made of PEEK<br>are for use up to six months. |
| Healing cap<br>material | Y-TZP | Y-TZP<br>Closure and Healing Caps<br>Titanium Grade 4 | Titanium Grade 4 |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE |
| K Number and<br>name of the<br>device | K190654<br>Straumann® CI RD<br>Ceramic Healing caps | K180477<br>Straumann PURE Ceramic<br>Implants | K130808<br>Straumann Sterile Healing<br>Solution |
| Implant to cap<br>connection | Regular Diameter (RD)<br>with a Y-TZP implant<br>fixture and a Y-TZP cap/Ti<br>screw | RD with a Y-TZP implant fixture and<br>a Ti cap/screw | Regular Neck (RN)/Wide<br>Neck (WN)/ Narrow Neck<br>CrossFit (NNC)/ Narrow<br>CrossFit (NC)/Regular<br>CrossFit (RC) connections<br>with Titanium-Zirconium<br>(TiZr) or Ti implant fixture<br>and Ti cap/screw |
| Implant to cap<br>mode of<br>attachment | Screw retained | Screw retained | Screw retained |
| Diameter | Healing Cap:<br>Ø5.2 mm | Closure Cap:<br>Ø4.8 mm<br>Healing Cap:<br>Ø5.2 mm | Closure & Healing Caps:<br>Ø3.5 mm – 6.5 mm |
| Overall height | Healing Cap:<br>6.2, 7.2, and 8.7 mm | Closure Cap:<br>4.6 mm<br>Healing Cap:<br>6.2 and 7.2 mm | Closure Cap:<br>6.2 and 6.8 mm<br>Healing Cap:<br>8.1, 9.6, and 11.1 mm |
| Gingival height | Healing Cap:<br>2.0, 3.0, and 4.5 mm | Closure Cap:<br>0.5 mm<br>Healing Cap:<br>2.0 and 3.0 mm | Closure Cap:<br>0 mm and 0.5 mm<br>Healing Cap:<br>2.0, 3.0, and 4.5 mm |
| Sterilization<br>method | End user receives product<br>sterilized per Ethylene<br>oxide to an SAL of 10-6 | <b>Straumann PURE Ceramic<br/>Implants</b><br>Ethylene oxide to an SAL of 10-6<br>Closure and Healing Caps<br>End user receives product sterilized<br>per Irradiation to an SAL of 10-6 | End user receives product<br>sterilized per Irradiation to<br>an SAL of 10-6 |
{7}------------------------------------------------
### Straumann® CI RD Ceramic Healing Caps
510(k) Summary
Table 1 – Comparison of Subject Device with Primary Predicate and Reference Devices
#### Performance Data 1.7
The following performance data were provided in support of the substantial equivalence determination.
### Sterilization
A sterilization validation was performed per ISO 11135, Sterilization of healthcare products – Ethylene Oxide – Requirements for development, validation and routine control of a sterilization process for medical devices, using the Half Cycle Overkill Approach. The validation demonstrates that the sterilization process and equipment is capable of reliably and consistently sterilizing the subject device to a minimum Sterility Assurance Level (SAL) of 10-6.
{8}------------------------------------------------
Straumann® CI RD Ceramic Healing Caps
510(k) Summary
The method used to make the determination that the device meets endotoxin limit specifications is LAL Endotoxin Analysis. The testing limit is 20 EU/device. The testing limit was chosen based on a device that is blood contacting and/or implanted, as outlined in the FDA Guidance Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile, issued on 21 January 2016. In all cases, the testing was performed on the completed package from a production lot after sterilization.
A transportation study has been performed per ISTA 2A to demonstrate the integrity of the sterile barrier system considering the new sterilization process.
A packaging stability study was performed to demonstrate the integrity of the packaging system and the sterile barrier system after sterilization process via Ethylene Oxide, handling, distribution, transport and storage up to the defined product shelf life.
### Biocompatibility Testing
The biological safety of the CI RD Ceramic Healinq caps was assessed according to the ISO 10993-1 and tests were conducted to assess the following:
- Cytotoxicity (ISO 10993-5) ●
- Chemical Characterization (ISO 10993-18) .
### 1.8 Conclusion
Based upon assessment of the design and applicable performance data, the subject devices have been determined to be substantially equivalent to the identified predicate and/or reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.