Zio XT ECG Monitoring System, Zio AT ECG Monitoring System
K190593 · iRhythm Technologies, Inc. · DQK · Aug 23, 2019 · Cardiovascular
Device Facts
Record ID
K190593
Device Name
Zio XT ECG Monitoring System, Zio AT ECG Monitoring System
Applicant
iRhythm Technologies, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Aug 23, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
AI/ML
Indications for Use
The Zio ECG Utilization Service (ZEUS) System is intended to analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring by Zio XT or Zio AT Patch, a final report is generated based on the beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients.
Device Story
ZEUS System processes/analyzes continuous single-lead ECG data recorded by Zio XT or Zio AT patches. System downloads, stores, and aggregates ECG data; automated analysis identifies cardiac events. Certified Cardiographic Technician (CCT) reviews automated findings to generate final clinical report. Used in clinical settings by trained technicians to assist physicians in cardiac diagnosis. Output provides comprehensive ECG analysis to support clinical decision-making; enables long-term rhythm monitoring for transient symptom evaluation.
Clinical Evidence
No clinical testing was performed. Evidence consists of bench testing, system performance testing, and software verification testing to ensure compliance with design specifications and consensus standards.
Technological Characteristics
Programmable diagnostic computer for ECG processing. Complies with AAMI/ANSI ISO 14971:2007(R)2010, AAMI/ANSI IEC 60601-2-47:2012, and AAMI/ANSI EC57:2012. System performs automated ECG analysis and aggregation of single-lead data. Cybersecurity managed per FDA guidance.
Indications for Use
Indicated for patients 18 years or older, symptomatic (palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, anxiety) or asymptomatic, requiring long-term ECG monitoring. Not for critical care patients.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
iRhythm Technologies, Inc. Zio® AT ECG Monitoring System (K181502)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
iRhythm Technologies, Inc. Gabrielle Logan Senior Regulatory Affairs Specialist 650 Townsend Street, Ste 500 San Francisco, California 94103
Re: K190593
Trade/Device Name: Zio® ECG Utilization Service (ZEUS) System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DQK Dated: July 19, 2019 Received: July 23, 2019
## Dear Gabrielle Logan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Nicole Goodsell Assistant Director (Acting) External Heart Rhythm and Rate Devices Team Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190593
Device Name Zio® ECG Utilization Service (ZEUS) System
#### Indications for Use (Describe)
The Zio ECG Utilization Service (ZEUS) System is intended to analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring by Zio XT or Zio AT Patch, a final report is generated based on the beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Notification K190593
#### l. General Information
#### Applicant:
iRhythm Technologies, Inc. 650 Townsend Street, Suite 500 San Francisco, CA 94103, USA Phone: 415-632-5700 Fax: 415-632-5701
### Contact Person:
Gabrielle Logan Sr. Regulatory Affairs Specialist Phone: 415-214-7092 Email: glogan@irhythmtech.com
Date Prepared: August 22, 2019
#### ll. Device Information
Trade Name: Zio® ECG Utilization Service (ZEUS) System
### Generic/Common Name:
Programmable diagnostic computer
### Classification Names:
Programmable diagnostic computer [21CFR§870.1425]
### Regulatory Class: Class II
Product Codes: DQK
#### III. Predicate Devices
The following predicate device has been selected:
- iRhythm Technologies, Inc. Zio® AT ECG Monitoring System [K181502] |
No reference devices were used in this submission.
{4}------------------------------------------------
### Section 1b 510(k) Summary
#### IV. Indications for Use
The Zio ECG Utilization Service (ZEUS) System is intended to analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring. After patient monitoring by Zio XT or Zio AT Patch, a final report is generated based on the beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients.
#### V. Device Description
The ZEUS System is an electrocardiogram (ECG) processing and analysis system, designed to handle continuously recorded, single-lead ECG data. It downloads, stores, analyzes and aggregates the ECG data for a Certified Cardiographic Technician (CCT) to review and generate a report of the findings contained within the data; thereby enabling the provision of a complete ECG processing and analysis service.
Automated ECG analysis performance was quantified for any claimed analysis metrics. The resulting statistics demonstrate sensitivity and positive predictivity levels which satisfy requirements and minimize safety or efficacy concerns.
{5}------------------------------------------------
### Section 1b 510(k) Summary
#### VI. Substantial Equivalence
The proposed indications for use statement for the Zio Utilization Service (ZEUS) System is substantially equivalent to the intended use in the cleared Indications for Use statement for the predicate device. The performance testing results demonstrate that the differences in the technological characteristics between the devices do not raise any new issues of safety or efficacy. Therefore, the ZEUS System is determined to be substantially equivalent to the predicate device.
#### VII. Nonclinical Testing in Support of Substantial Equivalence Determination
There are no required FDA performance standards for the ZEUS System. Overall system performance testing was conducted as part of the verification activities on incremental changes to the ZEUS System to ensure performance as intended per specifications and to support a determination of substantial equivalence to the predicate device.
Nonclinical testing included:
- . System performance testing
- . Software verification testing
The results confirm by examination and provision of objective evidence that the design output met the design input requirements in conformance with the following list of recognized consensus standards:
| FDA# | Body | Number / Version | Title |
|-------|------------------|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5-70 | AAMI<br>ANSI ISO | 14971:2007(R)2010<br>(Corrected 4 October 2017) | Medical Devices – Applications Of Risk<br>Management To Medical Devices |
| 3-127 | AAMI<br>ANSI IEC | 60601-2-47:2012 | Medical Electrical Equipment -- Part 2-<br>47: Particular Requirements For The<br>Basic Safety And Essential Performance<br>Of Ambulatory Electrocardiographic<br>Systems |
| 3-118 | AAMI<br>ANSI | EC57:2012 | Testing And Reporting Performance<br>Results Of Cardiac Rhythm And ST-<br>Segment Measurement Algorithms |
| N/A | U.S. FDA | October 2, 2014 | Guidance for Industry and FDA Staff –<br>Content of Premarket Submissions for<br>Management of Cybersecurity in<br>Medical Devices |
No clinical testing was performed in support of this premarket notification.
{6}------------------------------------------------
Section 1b 510(k) Summary
#### VIII. Conclusion
The ZEUS System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.