Cochlear Osia OSI100 Implant, Cochlear Osia Sound Processor, Osia Intraoperative Test Software, Osia Fitting Software, Cochlear Osia Surgical Instruments
Cochlear Osia OSI100 Implant, Cochlear Osia Sound Processor, Osia Intraoperative Test Software, Osia Fitting Software, Cochlear Osia Surgical Instruments
Applicant
Cochlear Americas
Product Code
PFO · Ear, Nose, Throat
Decision Date
Jul 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3340
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cochlear Osia System uses bone conduction to transmit sounds to the cochlea (inner ear). Cochlear Osia implants are single use devices intended for long term implantation under the skin in the mastoid region of either side of the head. They are for professional use only.
Device Story
Active implantable bone conduction hearing system; consists of OSI100 implant (receiver/stimulator and piezoelectric actuator) and external sound processor. Sound processor captures/digitally processes environmental sound; transmits power/data via transcutaneous RF link to implant. Implant converts digital signal to analog; piezoelectric actuator generates mechanical vibrations on mastoid bone; bypasses middle ear to stimulate inner ear. Used in clinical/home settings; operated by patient after surgical implantation by physician. Output provides mechanical amplification to compensate for hearing loss. Benefits include improved speech perception in noise/quiet and audiometric thresholds.
Clinical Evidence
Open, two-armed, prospective multi-center clinical trial. Primary analysis at 3 months, safety at 6 months. Compared to pre-implant condition, Osia System showed significant improvements in audiometric thresholds, speech in noise, and speech in quiet. Patient-reported outcomes demonstrated significant benefits. Performance improvements expected to be similar to predicate device.
Technological Characteristics
Active implantable bone conduction system. Components: OSI100 implant (titanium, silicone, platinum), external sound processor, fitting software. Actuator: piezoelectric transducer (piezo bender with tungsten masses). Fixation: BI300 osseointegrated implant. Energy: RF inductive link for power/data; external battery-powered processor. Biocompatibility per ISO 10993-1:2009. Software: minor level of concern.
Indications for Use
Indicated for patients 12+ years old with conductive or mixed hearing loss (PTA BC threshold <= 55 dB HL) or single-sided deafness (SSD) with profound sensorineural loss in one ear and normal hearing in the opposite ear (PTA AC threshold <= 20 dB HL). Bilateral fitting indicated for symmetrical conductive/mixed loss. Recommended for patients with prior experience with air or bone conduction hearing aids.
Regulatory Classification
Identification
An active implantable bone conduction hearing system is a prescription device consisting of an implanted transducer, implanted electronics components, and an audio processor. The active implantable bone conduction hearing system is intended to compensate for conductive or mixed hearing losses by conveying amplified acoustic signals to the cochlea via mechanical vibrations on the skull bone.
Special Controls
In combination with the general controls of the FD&C Act, the active implantable bone conduction hearing system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must characterize any adverse events observed during implantation and clinical use, and must also demonstrate that the device performs as intended under anticipated conditions of use.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:
(i) Performance data must validate force output in a clinically relevant model.
(ii) Impact testing in a clinically relevant anatomic model must be performed.
(iii) Mechanical integrity testing must be performed.
(iv) Reliability testing consistent with expected device life must be performed.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(6) Performance data must demonstrate the wireless compatibility, electromagnetic compatibility, and electrical safety of the device.
(7) Software verification, validation, and hazard analysis must be performed.
(8) Labeling must include:
(i) A summary of clinical testing conducted with the device that includes a summary of device-related complications and adverse events;
(ii) Instructions for use;
(iii) A surgical guide for implantation, which includes instructions for imaging to assess bone dimensions;
(iv) A shelf life, for device components provided sterile;
(v) A patient identification card; and
(vi) A patient user manual.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2019
Cochlear Americas Laura Blair Director Regulatory Affairs 13059 E. Peakview Avenue Centennial, Colorado 80111
Re: K190589
Trade/Device Name: Cochlear Osia OSI100 Implant, Cochlear Osia Sound Processor, Osia Intraoperative Test Software, Osia Fitting Software, Cochlear Osia Surgical Instruments Regulation Number: 21 CFR 874.3340 Regulation Name: Active Implantable Bone Conduction Hearing System Regulatory Class: Class II Product Code: PFO Dated: June 6, 2019 Received: June 6, 2019
Dear Laura Blair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190589
Device Name Cochlear™ Osia™ System
#### Indications for Use (Describe)
The Osia System is intended for the following patients and indications:
· Patients 12 years of age or older.
· Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5. 1, 2, and 3 kHz) should be better than or equal to 55 dB HL.
· Bilateral fitting of the Osia System is intents having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 3 kHz, or less than 15 dB at individual frequencies.
· Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e., singlesided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2, and 3 kHz).
• The Osia System for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
· Prior to receiving the device, it is recommended that an individual have experience with appropriately fitted air conduction or bone conduction hearing aids.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 1. 510(k) Summary
# A. Submitter Information
| A. Submitter Information | |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Cochlear Americas<br>13059 East Peakview Ave.<br>Centennial, CO 80111 |
| On behalf of the manufacturer: | Cochlear Ltd – Macquarie<br>1 University Avenue<br>Macquarie University, NSW 2109<br>Australia<br>(Establishment Number 9076254) |
| Contact: | Laura Blair<br>Director, Regulatory Affairs<br>Cochlear Americas<br>lblair@cochlear.com<br>(303) 264-2752 (o)<br>(303) 524-6825 (f) |
| B. Date Prepared | 7/2/2019 |
| C. Device Name and Classification | |
| Trade/Proprietary Name: | Cochlear™ Osia™ System |
| Common/Usual Name: | Osia System |
| Classification Name: | Active implantable bone conduction hearing system<br>21 CFR 874.3340, Class II |
| Classification Panel | Ear, Nose, and Throat |
| Product Code: | PFO |
| D. Predicate Device | |
| Trade/Proprietary Name: | Bonebridge™ System |
| Common/Usual Name: | Bonebridge System |
| Classification Name: | Active implantable bone conduction hearing system<br>21 CFR 874.3340, Class II |
| Classification Panel | Ear, Nose, and Throat |
| Product Code: | PFO |
{4}------------------------------------------------
510(k):
#### DEN170009
### E. Purpose of Submission
This Traditional 510(k) seeks clearance for an active implantable bone conduction hearing system, known as the Cochlear™ Osia™ System. The Osia System can be used by individuals aged 12 and older with mixed or conductive hearing loss, or single sided deafness to provide improved hearing.
## F. Device Description
Cochlear's Osia System mechanically vibrates the skull bone and subsequently the cochlea to compensate for conductive hearing loss, mixed hearing loss, or single-sided deafness (SSD).
The Osia System is made up of several components. The Osia Implant (OSI100) consists of a receiver/stimulator and an actuator (vibrator) which is surgically implanted on the skull bone. The external component of the Osia System is a sound processor, worn off-the-ear, which picks up the sound from the environment, and sends, after processing, the information to the implant via a transcutaneous inductive link. This link is also referred to as radiofrequency (RF) link. Each Osia System is configured to meet an individual's hearing needs, using dedicated fitting software. The Osia System is illustrated in Figure 1 below.
Image /page/4/Figure/7 description: The image shows a medical device implanted in a bone. The device consists of four labeled parts. Part 1 is a gray, oval-shaped component positioned above the bone. Parts 2, 3, and 4 are connected and appear to be implanted within the bone structure.
Figure 1: Overview of the Osia System, including the Osia Sound Processor
In normal operation, the Osia System functions as follows (referring to Figure 1):
- The external sound processor captures and digitally processes sound. 1.
- 2. The sound processor transmits power and digital information to the implant.
- The implant receiver-stimulator converts the digital information into an electric analogue 3. signal.
- This electric signal is converted to vibrations by the implant piezoelectric actuator. 4.
{5}------------------------------------------------
The actuator converts the electrical signal into an amplified mechanical stimulation, bypassing the impaired middle ear (origin of the conductive part of the hearing loss) and providing some level of mechanical amplification in order to compensate for the damaged inner ear (sensorineural part of the hearing loss, in case of mixed hearing loss).
#### G. Intended Use
The Cochlear Osia System uses bone conduction to transmit sounds to the cochlea (inner ear). Cochlear Osia implants are single use devices intended for long term implantation under the skin in the mastoid region of either side of the head. They are for professional use only.
#### H. Indications for Use
The Osia System is intended for the following patients and indications:
- Patients 12 years of age or older. .
- Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 55 dB HL.
- . Bilateral fitting of the Osia System is intended for patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 3 kHz, or less than 15 dB at individual frequencies.
- Patients who have profound sensorineural hearing loss in one ear and normal hearing in the ● opposite ear (i.e., single-sided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2, and 3 kHz).
- . The Osia System for SSD is also indicated for any patient who is indicated for an airconduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
- Prior to receiving the device, it is recommend that an individual have experience with ● appropriately fitted air conduction or bone conduction hearing aids.
#### I. Technological Characteristics
Like other active implantable bone conduction hearing systems, the Osia System is comprised of multiple components, including: an implant, sound processor, fitting software, and other cables and accessories. The Osia System is intended to compensate for conductive or mixed hearing loss or single sided deafness by conveying amplified acoustic signals to the cochlea via mechanical vibrations on the skull bone. These vibrations bypass the damaged parts of the outer and/or middle ear to stimulate the inner ear hair cells, allowing patients to clearly hear sounds and speech around them. Both the Osia System and the predicate are surgically implanted in the mastoid bone and an external audio processor is held in place on the patients scalp by magnetic attraction between the implant and audio sound processor.
{6}------------------------------------------------
The implant coil, magnet, case and lead assembly reuse the Freedom Cochlear Implant (CI24RE) platform (P970051/S028). This platform has been used in Cochlear's implantable systems for over ten years. The implant receiver-stimulator contains dedicated electronics whose power is supplied by the batteries in the external sound processor. The actuator is connected to the implant receiver-stimulator via a standard Cochlear lead, which contains a bundle of conductors grouped into two conductive paths. Just like the predicate device, an electromagnetic inductive link (also known as radio-frequency link) between the external sound processor coil and the implant coil allows sound data and power to be transmitted to the implant. A magnet located at the center of the implant coil allows the coil of the external sound processor to be placed in the correct position, over the receiver-stimulator coil, so that the inductive link is operational.
The actuator for the OSI100 Implant is a piezoelectric transducer which converts the signal into corresponding vibrations. The actuator design concept is conceived of a piezo bender connected to two identical tungsten moving masses. When the piezo is electrically driven by the stimulator electronics, a vibration is generated at the central fixation clamp, the location where the actuator is fixed to the bone by the Baha® B1300 Implant (K100360). The fixation system consists of the BI300 Bone Implant and the fixation screw to fix the piezoelectric actuator to the BI300 Bone Implant. The Osia Implant is attached via the osseointegrated BI300 Implant and the transducer of the predicate device is attached via the osseointegrated cortical screws. Although the method of attachment is not the same, both attachments (screw and implant) are made of titanium and are attached and dependent upon osseointegration. Both devices have an implanted transducer. The energy used and delivered by the Osia System is the same as the predicate device.
Both the Osia Sound Processor and the audio processor from the predicate device are worn on the head behind the ear and held in place by magnets. Both are powered by standard hearing aid batteries and have multiple color variants. Neither the Osia Sound Processor nor the Audio Processor of the predicate device contain software or are sterile.
Both the Osia System and the predicate have dedicated fitting software. Although the software platforms are different, both software platforms have a minor level of concern and are used for the same purpose.
Like the predicate device, the Osia Implant is provided sterile and is implanted in a surgical setting. Following programming of the sound processor, both the Osia System and the predicate system are used by the recipient in their daily life.
The Osia System has the same intended use, the same functional characteristics, the same fundamental operating principles, and is made of biocompatible materials like the predicate device. The Osia System can accommodate greater degrees of hearing loss (45 versus 55 dB). Force output measurements made at the same frequencies are higher for the Osia System, enabling the expanded fitting range of the Osia Sound Processor versus the predicate.
### J. Materials
{7}------------------------------------------------
Like the predicate device, the Osia Implant is implanted in the mastoid bone and is a device in permanent tissue/bone contact. The Osia Sound Processor is classified as an intact-skin contacting device for permanent use.
Both the Osia Implant and Osia Sound Processor is made of medical grade plastics and metals that have been shown to be biocompatible and safe for human use. The Osia Implant is attached via the osseointegrated BI300 Implant and the transducer of the predicate device is attached via the osseointegrated cortical screws. Although the method of attachment is not the same, both attachments (screw and implant) are made of titanium and are attached and dependent upon osseointegration. Direct contact materials in the OSI100 Implant include, titanium, silicone, and platinum. Materials from the BCI 601 predicate device that are in contact with the body are silicone, titanium and polypropylene. Both the Osia System and the predicate device had biocompatibility testing completed in accordance with ISO 10993-1:2009.
### K. Performance Data
Testing of the Osia System included testing to confirm patient safety, product usability and reliability, interoperability, and electromagnetic compatibility. Patient safety testing included biocompatibility testing as well as basic medical electrical safety testing. Usability was demonstrated through simulated use testing in a variety of use cases. Reliability testing was conducted through exposure of the device to mechanical stresses, and environmental extremes, and simulated long-term use followed by functional testing. Simulated testing has demonstrated that the lifetime of the OSI100 actuator is a minimum of six years. Interoperability testing was performed, including functional testing with existing Baha fixation systems, as well as the ability to perform a prescribed fitting through the Osia Fitting Software. Electromagnetic compatibility testing established that the device did not emit excessive amounts of electromagnetic energy, and that the device operated as intended in the presence of interference sources. Coexistence testing was performed, with the device paired and in a streaming configuration while exposed to various sources of in-band interference. The results of the testing indicate that the Osia System is substantially equivalent to the predicate device.
Clinical performance data was gathered in a multi-center trial. The investigation was an open, two-armed prospective clinical investigation that had one site in the United States. The study was not blinded. The investigation had a primary analysis endpoint at 3 months with an additional 9 months of follow up. The primary safety analysis was at 6 months. When compared to the preimplant condition, data gathered with the Osia Device showed significant improvements in audiometric thresholds, speech in noise, and speech in quiet. Patient reported outcomes also demonstrated significant benefits. Although some test measures varied, based on the clinical data gathered with the Osia System, performance improvements are expected to be similar to the predicate device.
As required by the Special Controls identified for this device type, a summary of the clinical data is provided in the product labelling.
### L. Conclusion
Despite the physical and material differences of the Osia System and the predicate device, the Osia System and the predicate have substantially equivalent function and technology, and
{8}------------------------------------------------
intended use. Based on the indications for use, technological characteristics, and substantial equivalence comparison to the predicate device, and supported by clinical and non-clinical data, the Cochlear Osia System has been shown to be as safe and effective for its intended use as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.