K190542 · Soliton, Inc. · GEX · May 24, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190542
Device Name
Soliton Acoustic Wave Device
Applicant
Soliton, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 24, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Soliton Acoustic Wave Device (AWD) is indicated for use as an accessory to laser tattoo removal procedures using a 1064 nm Q-Switched laser in Fitzpatrick Skin Type I-III patients.
Device Story
AWD is an accessory to 1064 nm Q-Switched laser tattoo removal. Device console supplies saline to a handpiece, which generates rapidly rising acoustic waves. Waves are applied to the skin via an acoustically transparent window and hydrogel pad. Principle: acoustic waves dissipate gas vacuoles (whitening) induced by laser treatment, allowing multiple laser passes in a single session. Used in clinical settings by trained operators. Benefits: enables immediate subsequent laser passes, reducing total office visits required for tattoo removal. Clinical decision-making is supported by the ability to perform multiple passes per session, improving efficiency compared to standard single-pass protocols.
Clinical Evidence
Prospective, single-center study (n=24 subjects, 35 tattoos). Subjects were Fitzpatrick Skin Type I-III. Primary endpoint: number of laser passes per session. Results: mean of 4.16 laser passes achieved with AWD (95% CI lower limit 3.97). Safety profile: all adverse events were mild or moderate and resolved; no serious or severe events reported. Animal study (Gottingen mini-pigs) confirmed histological safety and efficacy in clearing dermal whitening without residual epidermal damage.
Technological Characteristics
System consists of a console, handpiece, and connecting cable. Operates by generating acoustic waves in saline, transmitted through an acoustically transparent window and hydrogel pad. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Biocompatibility per ISO 10993. Software-controlled operation.
Indications for Use
Indicated for Fitzpatrick Skin Type I-III patients undergoing laser tattoo removal using a 1064 nm Q-Switched laser.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
ON Light Science's DeScribe® Transparent PFD Patch (K150212)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Soliton Inc. % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
May 24, 2019
Re: K190542
Trade/Device Name: Soliton Acoustic Wave Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: QHF Dated: March 4, 2019 Received: March 4, 2019
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Jennifer Stevenson Acting Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
### K190542
Device Name
Soliton Acoustic Wave Device
Indications for Use (Describe)
The Soliton Acoustic Wave Device (AWD) is indicated for use as an accessory to laser tattoo removal procedures using a 1064 nm Q-Switched laser in Fitzpatrick Skin Type I-III patients.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K190542 510(k) SUMMARY
### Soliton Acoustic Wave Device
| Submitted by: | Soliton, Inc.<br>5304 Ashbrook Drive<br>Houston, TX 77081 |
|-------------------|----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Leslie Honda<br>VP, Requlatory Affairs and Quality Systems<br>Tel: 206.375.8586 |
| Date Prepared: | May 14, 2019 |
| Trade Name: | Soliton Acoustic Wave Device |
| Common Name: | Dermatology Laser System |
| Classification: | Class II<br>Laser surgical instrument for use in general and plastic surgery and in<br>dermatology (21 CFR 878.4810) |
| Predicate Device: | ON Light Science's DeScribe® Transparent PFD Patch (K150212) |
### Device Description
The Soliton Acoustic Wave Device (AWD) is designed as an accessory to laser treatments for tattoo removal. The AWD uses repeated, rapidly rising acoustic waves, releasing pigment particles from the pigment laden macrophage (PLM) and dissipating the laserinduced whitening. This allows multiple laser passes in a single session, resulting in fewer office visits.
# Intended Use / Indications for Use:
The Soliton Acoustic Wave Device (AWD) is indicated for use as an accessory to laser tattoo removal procedures using a 1064 nm Q-Switched laser in Fitzpatrick Skin Type I-III patients.
### Summary of Technological Characteristics
The Soliton AWD is composed of three parts: the Console, the Hand Piece and the connecting Cable. The Console supplies saline to the Hand Piece to enable formation of the shock wave within the acoustic pulse chamber. The Hand Piece generates acoustic waves in the saline. The acoustic waves pass through the acoustically transparent Window and acoustic ultrasound gel or similar hydrogel pad, which when placed against the surface of the tattoo to be removed, penetrates the skin to the typical depth of tattoo ink.
The shape, frequency and repetition rate of these acoustic waves are optimized to diffuse gas bubbles (vacuoles) caused by the use of a Q-switched laser without damaging adjacent non-pigmented tissue. Removing them by using the Soliton AWD allows another Q-switched laser treatment to be performed immediately instead of waiting for the body to remove the gas pockets and heal over a period of 4-6 weeks between standard laser treatments.
{4}------------------------------------------------
### Performance Data
Electrical safety and electromagnetic compatibility (EMC) testing for the AWD was conducted by an independent test laboratory in accordance with IEC 60601-1, Medical electrical equipment, Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests.
The biocompatibility of the AWD is established by biocompatibility evaluations consistent with the guidelines in FDA's "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing'" Guidance Document.
Software verification and validation testing was conducted and the testing results were found acceptable for software release.
All performance testing demonstrated that the AWD performs according to specifications and functions as intended.
# Animal Study Data
The AWD was evaluated using Gottingen Mini-pigs tattooed with multiple circular black tattoo spots. Tattoo sites were treated with either a single-pass Q-switched Nd: YAG laser treatment (Laser Only), or multiple-pass laser treatment with each laser treatment followed by an AWD application (Laser+AWD). Dermal clearing was assessed histologically and colorimetrically using a handheld colorimeter. The dermal effects of using AWD as an accessory to the laser were assessed histologically both immediately and at 48-72 hours post treatment. Additionally, the treatment sites were assessed photographically at 10-12 weeks.
The results of this study demonstrate that when used as an accessory to a Q-switched Nd:YAG laser in the treatment of tattoos, AWD clears laser-induced dermal whitening. The use of Laser+AWD was also shown to be safe. After 10 to 12 weeks of healing the photographic appearance of the epidermis at the treatment sites was normal, showing no residual epidermal changes.
# Clinical Study Data
The Soliton AWD was evaluated in a single-center, prospective study to assess the safety and efficacy of the device for its indicated use in tattoo removal. The clinical study enrolled 24 subjects with 35 tattoos at an investigational site in the U.S. The majority of subjects were female, Fitzpatrick Skin Type I-III, with a mean age of 34 years.
Each tattoo was divided into 3 approximately equal areas. Each end area was designated for treatment with either a single laser pass (Laser Only) or multiple consecutive laser passes alternating with AWD (Laser+AWD). The middle area of the tattoo between the treated areas served as the control and was not treated. Each subject was assessed at baseline, post-treatment, and during follow up after treatment. For the primary endpoint analysis, the number of laser treatment passes delivered in a single session was recorded.
{5}------------------------------------------------
In addition, the subjects' tattoos were evaluated by blinded independent evaluators, as well as the study investigator. Any potential safety effects were recorded.
The primary effectiveness endpoint of the study was met as the mean number of laser passes delivered in a single treatment session with AWD treatment was 4.16, with a lower limit of 95% confidence interval of 3.97.
The safety of the AWD was evaluated based on the adverse events reported during the study. All adverse events observed during the study were categorized as mild or moderate and were expected. All symptoms were completely resolved by the end of the study. No adverse events were identified as serious or severe. No patients asked to stop the treatment or dropped out of the study due to any adverse event.
Based on the clinical performance as documented in the pivotal clinical study, the Soliton AWD was found to have a safety and effectiveness profile that is similar to the predicate device.
### Summary of Substantial Equivalence
The AWD and the predicate device have the same intended use and similar indications for use. The use of an accessory to a Q-switched laser to dissipate whitening and allow multiple laser passes in a single session has previously been cleared by FDA for the predicate device. Both the AWD and the PFD patch cause the dissipation and absorption of the gas bubbles that cause whitening. As demonstrated in the performance testing, the AWD device performs similarly compared to the predicate device. Clinical testing confirms that the differences in technology compared to the predicate do not adversely impact performance. Therefore, the AWD is substantially equivalent to its predicate device.
# Conclusions
Nonclinical and clinical testing of the AWD demonstrated that the device performs as intended with a favorable safety profile. Results in the clinical study were comparable to those reported for the predicate device, in support of substantial equivalence. Therefore, the AWD is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.