← Product Code [KTT](/productcode/KTT) · K190486

# Pitkar External Fixation System (K190486)

_S.H.Pitkar Orthotools Pvt. , Ltd. · KTT · Oct 16, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190486

## Device Facts

- **Applicant:** S.H.Pitkar Orthotools Pvt. , Ltd.
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Oct 16, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Pitkar External Fixation System is indicated for both adults and pediatric patients for: 1. Open and closed fracture fixation 2. Pseudoarthrosis or non-union of long bones 3. Limb lengthening by epiphyseal or metaphyseal distraction 4. Correction of bony or soft tissue deformities 5. Correction of segmental or nonsegmental bony or soft tissue defects 6. Post-Traumatic joint contracture which has resulted in loss of range of motion

## Device Story

The Pitkar External Fixation System is a metallic orthopedic device used for bone stabilization and deformity correction. It consists of implantable components manufactured from stainless steel. The system is used by surgeons in clinical settings to stabilize fractures, lengthen limbs, or correct bony/soft tissue defects. It functions as a mechanical fixation frame; the surgeon attaches the device to the patient's bone to provide structural support during the healing or correction process. The device is provided non-sterile and requires steam sterilization by the user prior to application. By providing rigid or adjustable external support, the system facilitates bone alignment and healing, potentially improving patient mobility and functional outcomes.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing conducted in accordance with ASTM F1541-02, "Standard Specification and Test Methods for External Skeletal Fixation Devices."

## Technological Characteristics

Metallic external fixation system; materials: stainless steel per ASTM F899; mechanical fixation principle; non-sterile, requires user steam sterilization; design features similar to predicate external fixation systems.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- ACE-Fischer External Fixation System ([K083789](/device/K083789.md))
- Taylor Spatial Frame ([K970748](/device/K970748.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

January 14, 2020

S H Pitkar Orthotools Pvt Ltd. Vivek Mangalwedhekar Head of Firm Plot No. EL 32, J Block, MIDC Bhosari Pune, Maharashtra, 411026 INDIA

Re: K190486

Trade/Device Name: Pitkar External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: July 13, 2019 Received: July 18, 2019

Dear Vivek Mangalwedhekar:

This letter corrects our substantially equivalent letter of October 16, 2019.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Ting Song

Ting Song Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K190486

Device Name Pitkar External Fixation System

#### Indications for Use (Describe)

- The Pitkar External Fixation System is indicated for both adults and pediatric patients for:
- 1. Open and closed fracture fixation
- 2. Pseudoarthrosis or non-union of long bones
- 3. Limb lengthening by epiphyseal or metaphyseal distraction
- 4. Correction of bony or soft tissue deformities
- 5. Correction of segmental or nonsegmental bony or soft tissue defects
- 6. Post-Traumatic joint contracture which has resulted in loss of range of motion

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Pitkar External Fixation System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.

| Sponsor:                        | S H Pitkar Orthotools Pvt Ltd.<br>Plot No. EL 32, J Block, MIDC Bhosari<br>Pune<br>Maharashtra 411026<br>India                                                                                                                                                            |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                 | Vivek Mangalwedhekar<br>Head of Firm<br>Telephone: +912040706464<br>Fax: +912046768107                                                                                                                                                                                    |
| Date:                           | 15 July 2019                                                                                                                                                                                                                                                              |
| Subject Device:                 | Trade Name: Pitkar External Fixation System                                                                                                                                                                                                                               |
|                                 | Classification Name:<br>• KTT - Single/multiple component metallic bone<br>fixation appliances and accessories (21 CFR 888.3030)<br>• JDW - Smooth or threaded metallic bone fixation<br>fastener (21 CFR 888.3040)                                                       |
| Predicate Device(s):            | K083789 ACE-Fischer External Fixation System DePuy Orthopaedics<br><br>K970748 Taylor Spatial Frame Smith & Nephew, Inc.                                                                                                                                                  |
| Purpose and Device Description: | The purpose of this submission is to request clearance for<br>the new Pitkar External Fixation System. The<br>implantable components are manufactured from<br>Stainless Steel per ASTM F899. The system will be<br>provided in non-sterile configuration and will require |

{4}------------------------------------------------

## Intended Use and Indications for Use:

## Summary of Technological Characteristics:

The Pitkar External Fixation System is indicated for both adults and pediatric patients for:

- 1. Open and closed fracture fixation
- 2. Pseudoarthrosis or non-union of long bones
- 3. Limb lengthening by epiphyseal or metaphyseal distraction
- Correction of bony or soft tissue deformities 4.
- Correction of segmental or nonsegmental bony or soft 5. tissue defects
- 6. Post-Traumatic joint contracture which has resulted in loss of range of motion

The rationale for substantial equivalence is based on consideration of the following characteristics:

- 0 Intended Use: The intended use is similar to the intended use of the predicate systems.
- Indications for Use: The indications for use are ● similar to the indications for use cleared in predicate systems.
- o Materials: The new Pitkar Extremal Fixation System implants are manufactured from Stainless Steel (per ASTM F899). Implant grade stainless steels are commonly used materials in orthopedic implants.
- Design Features: The design features for new Pitkar External Fixation System is similar to those in currently marketed predicate devices. The design differences have not identified any issues that would impact the safety and effectiveness of the device.
- . Sterilization: The system is offered to the user in nonsterile configuration and will required to be steam sterilized by the user prior to use. The non-sterile packaging configuration is similar to the predicate devices currently marketed.

Summary of Performance Data (Nonclinical and/or Clinical)

#### Non-Clinical Tests: 0

- o ASTM F1541-02 "Standard Specification and Test Methods for External Skeletal Fixation Devices"

{5}------------------------------------------------

#### Clinical Tests: ●

o N/A

Substantial Equivalence Conclusion

The Pitkar External Fixation System has shown to be substantially equivalent to the predicate devices. Results of the non-clinical tests indicate that the device will perform within the intended uses and no new issues of safety and effectiveness have been raised.

---

**Source:** [https://fda.innolitics.com/device/K190486](https://fda.innolitics.com/device/K190486)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
