The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft. The Camber Spine Technologies SPIRA Open Matrix LLIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA Open Matrix LLIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA Open Matrix LLIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.
Device Story
SPIRA Open Matrix ALIF and LLIF are intervertebral body fusion devices designed for spinal stabilization. ALIF features arched support members; LLIF features spiral support members; both utilize an open matrix design to facilitate bone growth throughout the implant. Surfaces are roughened to minimize migration. Devices are implanted by surgeons during spinal fusion procedures to treat DDD. They require use with supplementary fixation systems and autogenous or allogenic bone graft. The primary purpose of this submission is to expand the indications to include the use of allogenic bone graft.
Clinical Evidence
Bench testing only. Finite Element Analysis (FEA) was conducted to evaluate design changes, specifically assessing static axial compression, shear, and surface area for subsidence and expulsion equivalence. Results confirmed that design changes did not create new worst-case scenarios.
Technological Characteristics
Intervertebral body fusion devices with open matrix architecture. ALIF utilizes arched support members; LLIF utilizes spiral support members. Roughened surface texture on non-perimeter surfaces. Designed for use with autogenous or allogenic bone graft.
Indications for Use
Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade I spondylolisthesis or retrolisthesis, who have failed 6 months of nonoperative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
September 23, 2019
Camber Spine Technologies % Matthew Carey Junior Regulatory Affairs/Quality Consultant RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913
Re: K190483
Trade/Device Name: SPIRA Open Matrix ALIF and LLIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 16, 2019 Received: August 27, 2019
Dear Matthew Carey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190483
Device Name SPIRA Open Matrix ALIF and LLIF
### Indications for Use (Describe) SPIRA Open Matrix ALIF
The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.
### SPIRA Open Matrix LLIF
The Camber Spine Technologies SPIRA Open Matrix LLIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA Open Matrix LLF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA Open Matrix LLIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------|---------------------------------------------|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K190483 Page 1 of 3
## 510(k) SUMMARY
Camber Spine Technologies SPIRA Open Matrix ALIF and LLIF
#### SUBMITTER l.
Camber Spine Technologies 501 Allendale Road King of Prussia, PA 19406
Phone: 484.427.7060
Contact Person: Matthew Carey RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, MA 01913 Phone: 1-(978) 834-5703 Fax: 1 (978) 477-0206
Date Prepared: 19 September 2019
#### II. DEVICE
| Trade/Device Name: | SPIRA Open Matrix ALIF and SPIRA Open Matrix LLIF |
|-----------------------|------------------------------------------------------|
| Common or Usual Name: | Intervertebral Fusion Device with Bone Graft, Lumbar |
| Classification Name: | Intervertebral body fusion device |
| Regulation Number: | 21 CFR 888.3080 |
| Regulatory Class: | Class II |
| Product codes | MAX |
#### lll. PREDICATE DEVICES
| Primary Predicates: | SPIRA Open Matrix LLIF (K180724)<br>SPIRA Open Matrix ALIF (K162986) |
|-----------------------|-------------------------------------------------------------------------------------------|
| Additional Predicate: | Centinel Spine STALIF TT70, STALIF MIDLINE®, MIDLINE II™, and<br>MIDLINE II-Ti™ (K150643) |
| Reference Predicate: | SPIRA-C Open Matrix Cervical Interbody, Camber Spine<br>Technologies (K172446) |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
SPIRA Open Matrix ALIF is an interbody fusion device with an open matrix design consisting primarily of arched support members to permit bone growth (i.e., interbody fusion) throughout the implant, SPIRA Open Matrix ALIF. With the exception of the perimeter, all surfaces have a roughened texture to help prevent movement of the device.
SPIRA Open Matrix LLIF is an interbody fusion device with an open matrix design consisting primarily of spiral support members to permit bone growth (i.e., interbody fusion) throughout the implant, SPIRA-L Open Matrix LLIF
The purpose of this submission is the addition of the allogenic bone graft indication to both SPIRA ALIF and SPIRA LLIF devices that have been cleared under K162986 and K180724 respectively.
## Indications for Use:
## SPIRA™ Open Matrix ALIF
The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.
### SPIRA™ Open Matrix LLIF
The Camber Spine Technologies SPIRA Open Matrix LLIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA Open Matrix LLIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA Open Matrix LLIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.
{5}------------------------------------------------
#### V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject SPIRA Open Matrix LLIF and SPIRA Open Matrix ALIF device was demonstrated to be substantially equivalent to primary predicates SPIRA Open Matrix LLIF (K180724) and SPIRA Open Matrix ALIF (K162986) with respect to indications, design, materials, function, manufacturing, and/or performance.
#### VI. PERFORMANCE DATA
A Finite Element Analysis (FEA) was performed to evaluate the effects of design changes to the SPIRA Open Matrix LLIF on the worst-case implant and determine if the design changes created new worst-case scenarios. The testing performed was static axial compression shear, and surface area comparison (subsidence and expulsion equivalence). Testing proved that the new design did not create a new worst-case under any testing condition.
#### VII. CONCLUSION
Camber Spine Technologies provided sufficient information to demonstrate the SPIRA Open Matrix LLF and SPIRA Open Matrix ALIF device is substantially equivalent to predicates SPIRA Open Matrix LLIF (K180724), SPIRA Open Matrix ALIF (K162986), and Centinel Spine STALIF TT70, STALIF MIDLINE®, MIDLINE II™, and MIDLINE II-Ti™ (K150643) with respect to indications, design, materials, function, manufacturing, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.