K190476 · FUJIFILM Sonosite, Inc. · IYN · Mar 15, 2019 · Radiology
Device Facts
Record ID
K190476
Device Name
FUJIFILM Sonosite Vevo MD Imaging System
Applicant
FUJIFILM Sonosite, Inc.
Product Code
IYN · Radiology
Decision Date
Mar 15, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.
Device Story
High-frequency diagnostic ultrasound system; acquires/displays real-time 2D, Color Doppler, Power Doppler, PW Doppler, and M-Mode data. Comprised of UHF transducers (UHF22, UHF48, UHF70) and main processing unit. Used by physicians in clinical settings for anatomical evaluation and procedural guidance (e.g., needle placement, nerve blocks). System processes acoustic data to generate high-resolution images for clinical decision-making. Benefits include high-resolution visualization of superficial structures and small organs. Operates via software-controlled interface; supports DICOM 3.0 for connectivity.
Clinical Evidence
No clinical data required; substantial equivalence supported by bench testing, including electrical, thermal, mechanical, and EMC safety, biocompatibility, and acoustic output characterization per NEMA UD 2 and IEC 62359.
Technological Characteristics
Track 3 diagnostic ultrasound system. Transducers: Linear Array (UHF22, UHF48, UHF70; 15-49MHz). 64 digital transmit/receive channels. Modes: 2D, Color Doppler, Power Doppler, PW Doppler, M-Mode. Connectivity: DICOM 3.0 (Store, Print, Worklist, PPS, Storage Commitment). Biocompatible patient-contact materials per ISO 10993-1. Software lifecycle per IEC 62304.
Indications for Use
Indicated for use by qualified physicians for ultrasound imaging or fluid flow analysis in abdominal, dermatological, musculoskeletal (conventional/superficial), neonatal cephalic, ophthalmic, pediatric, peripheral vessel, and small organ (breast, thyroid, testicles, prostate) applications. Includes needle/catheter placement and peripheral nerve block guidance. Use limited by transducer-specific maximum imaging depth.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
FUJIFILM SonoSite Vevo MD Imaging System (K160674)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
FUJIFILM SonoSite, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO MN 55313
March 15, 2019
Re: K190476
Trade/Device Name: FUJIFILM SonoSite Vevo MD Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 25, 2019 Received: February 27, 2019
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190476
Device Name FUJIFILM SonoSite Vevo MD Imaging System
Indications for Use (Describe)
The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.
Specific clinical applications and exam types include:
Abdominal Dermatological Musculoskeletal (conventional) Musculoskeletal (superficial) Neonatal Cephalic Ophthalmic Pediatric Peripheral vessel Small Organ (breast, thyroid, testicles, prostate)
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
## Table 1.3.1: Diagnostic Ultrasound Indications for Use Form - FUJIFILM SonoSite Vevo MD Imaging System
| System: | Vevo MD Imaging System | | | | | | | |
|-------------------------------------------------------|------------------------------------------------------------------------------------|---|-----|-----|--------------------------|---------------------------|--------------------------------|------------------|
| Transducer: | UHF22, UHF48, UHF70 | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| | Mode of Operation | | | | | | | |
| Clinical Application | 2D | M | PWD | CWD | Color<br>Doppler<br>(CD) | Power<br>Doppler<br>(CPD) | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | N | | | | | | | 1,2,3 |
| Fetal | | | | | | | | |
| Abdominal | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2,3 |
| Intra-operative (Abdominal<br>organs and vascular) | | | | | | | | |
| Intra-operative (Neuro.) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
| Small Organ (breast,<br>thyroid, testicles, prostate) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
| Neonatal Cephalic | N | N | N | | N | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | |
| Musculo-skel. (Convent.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
| Musculo-skel. (Superfic.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
| Intra-luminal | | | | | | | | |
| Other (spec.) | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Trans-esophageal (card.) | | | | | | | | |
| Other (spec.) | | | | | | | | |
| Peripheral vessel | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
| Other (Dermatology) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 |
N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination
#### Additional Comments/Other (Spec.):
I: The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging of abdominal organs or non-superficial musculoskeletal structure may be limited to neonatal and small pediative is limited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.
2: Includes imaging to assist in the placement of needler or other anatomical structures and imaging guidance for peripheral nerve block procedures.
3: Abdominal and Ophthalmic clinical applications are indications for use for the UHF22 transducer only
Page 1 of 4
VOL 002 Section 1 General Information Section 003 Indications for Use
{4}------------------------------------------------
| System: | Vevo MD Imaging System | | | | | | | | |
|-------------------------------------------------------|------------------------------------------------------------------------------------|---|-----|-----|--------------------------|---------------------------|--------------------------------|------------------|--|
| Transducer: | UHF22 | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | Combined<br>(Spec.) | Other<br>(Spec.) | |
| | B<br>(2D) | M | PWD | CWD | Color<br>Doppler<br>(CD) | Power<br>Doppler<br>(CPD) | | | |
| Ophthalmic | N | | | | | | | 1,2,3 | |
| Fetal | | | | | | | | | |
| Abdominal | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2,3 | |
| Intra-operative (Abdominal<br>organs and vascular) | | | | | | | | 1,2,3 | |
| Intra-operative (Neuro.) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Small Organ (breast,<br>thyroid, testicles, prostate) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Neonatal Cephalic | N | N | N | | N | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Convent.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Musculo-skel. (Superfic.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Intra-luminal | | | | | | | | | |
| Other (spec.) | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esophageal (card.) | | | | | | | | | |
| Other (spec.) | | | | | | | | | |
| Peripheral vessel | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Other (Dermatology) | N | N | N | | N | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination
#### Additional Comments/Other (Spec.):
I: The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging depth. For this reason imaging of abdominal organs or non-superficial musculoskeletal structure may be limited to neonatal and small imaging is limited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.
2. Includes imaging to assist in the placement of needles and cather anatomical structures and imaging guidance for peripheral nerve block procedures.
3: Abdominal and Ophthalmic clinical applications are indications for use for the UHF22 transducer only
{5}------------------------------------------------
| System: | Vevo MD Imaging System | | | | | | | | |
|-------------------------------------------------------|------------------------------------------------------------------------------------|---|-----|-----|--------------------------|---------------------------|--------------------------------|------------------|--|
| Transducer: | UHF48 | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | Other<br>(Spec.) | |
| | B<br>(2D) | M | PWD | CWD | Color<br>Doppler<br>(CD) | Power<br>Doppler<br>(CPD) | Combined<br>(Spec.) | | |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Abdominal<br>organs and vascular) | | | | | | | | | |
| Intra-operative (Neuro.) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Small Organ (breast,<br>thyroid, testicles, prostate) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Neonatal Cephalic | N | N | N | | N | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Convent.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Musculo-skel. (Superfic.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Intra-luminal | | | | | | | | | |
| Other (spec.) | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esophageal (card.) | | | | | | | | | |
| Other (spec.) | | | | | | | | | |
| Peripheral vessel | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
| Other (Dermatology) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD,<br>2D+PWD | 1,2 | |
N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination
#### Additional Comments/Other (Spec.):
19 101 820 800000000 1 100 16
1: The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging of abdominal organs or non-superficial musculoskelal structure may be limited to neonatal and small pediatric patients and all imaging is fimited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.
2: Includes imaging to assist in the placement of needler or other anatomical structures and imaging guidance for peripheral nerve block procedures.
FINAL TRANCE
{6}------------------------------------------------
| System: | Vevo MD Imaging System | | | | | | | |
|----------------------------------------------------|------------------------------------------------------------------------------------|---|-----|-----|--------------------|---------------------|-----------------------------|---------------|
| Transducer: | UHF70 | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | | | | | | | |
| | B (2D) | M | PWD | CWD | Color Doppler (CD) | Power Doppler (CPD) | Combined (Spec.) | Other (Spec.) |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | | | | | | | | |
| Intra-operative (Abdominal organs and vascular) | | | | | | | | |
| Intra-operative (Neuro.) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
| Small Organ (breast, thyroid, testicles, prostate) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
| Neonatal Cephalic | N | N | N | | N | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | |
| Musculo-skel. (Convent.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
| Musculo-skel. (Superfic.) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
| Intra-luminal | | | | | | | | |
| Other (spec.) | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Trans-esophageal (card.) | | | | | | | | |
| Other (spec.) | | | | | | | | |
| Peripheral vessel | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
| Other (Dermatology) | P | P | N | | P | N | 2D+CD, 2D+M, 2D+CPD, 2D+PWD | 1,2 |
N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination
#### Additional Comments/Other (Spec.):
I The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging depth. For this reason imaging of abdominal organs or non-superficial musculoskeletal structure may be limited to neonatal and small pediatric patients and all imaging is limited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.
2: Includes imaging to assist in the placement of needles and cather anatomical structures and imaging guidance for peripheral nerve block procedures.
{7}------------------------------------------------
SonoSite
# 510(K) Summary-K190476
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
#### 1) Submitter's name, address, telephone number, contact person;
FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904
| Corresponding Official: | Anoush Frankian |
|-------------------------|---------------------------------|
| | Sr. Manager, Regulatory Affairs |
| E-mail: | anoush.frankian@fujifilm.com |
| Telephone: | (425) 951-6824 |
| Facsimile: | (425) 491-8356 |
| Date prepared: | February 11, 2019 |
#### 2) Name of the device. including the trade or proprietary name if applicable. the common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
#### Proprietary Name
FUJIFILM SonoSite Vevo MD Imaging System
Classification Names
| Name | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO |
#### 3) Identification of the predicate or legally marketed device:
## FUJIFILM SonoSite Vevo MD Imaging System (K160674) (Primary Predicate) SonoSite X-Porte ultrasound system (K171437) (Reference Predicate)
#### 4) Device Description:
The Vevo MD Imaging System is a high frequency general purpose, software controlled, diagnostic imaging system used to acquire ad display high-resolution, real-time ultrasound data in 2D, Color Doppler, Power Doppler, PW Doppler and M-Mode. The Vevo MD Imaging System is comprised of
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red square with a white triangle cut out of it, which is placed above the "I" in Fujifilm. Below the logo, in a smaller font, is the text "Value from Innovation."
Image /page/8/Picture/1 description: The image shows the word "SonoSite" in a bold, sans-serif font. The letters are all black and evenly spaced. The dot above the "i" in "Site" is a perfect circle.
transducers responsible for ultrasound signal generation and recording, and a main unit that controls the transducers, processes the acoustic data, and processes and displays images.
#### 5) Intended Use:
The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.
Specific clinical applications and exam types include:
Abdominal Dermatological Musculoskeletal (conventional) Musculoskeletal (superficial) Neonatal Cephalic Ophthalmic
Pediatric
Peripheral vessel Small Organ (breast, thyroid, testicles, prostate)
#### 6) Technological Characteristics:
The Vevo MD Imaging System subject of this submission and its primary predicate Vevo MD Imaging System (K160674) and reference predicate X-Porte (K171437) are all Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.
| Feature | SonoSite VevoMD<br>Ultrasound System<br>(this submission) | SonoSite<br>VevoMD<br>Ultrasound<br>System<br>(K160674) | SonoSite X-Porte<br>Ultrasound System<br>(K171437) |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Diagnostic<br>ultrasound imaging<br>or fluid flow analysis<br>of the human body | Diagnostic<br>ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body | Diagnostic ultrasound<br>imaging or fluid flow<br>analysis of the human<br>body |
| Transducer<br>Types<br>Available | Linear Array | Linear Array | Linear Array<br>Curved Linear Array<br>Intracavitary<br>Phased Array |
| Transducer<br>Center<br>Frequency | 15-49MHz | 15-49MHz | 1.0 - 15.0 MHz |
| Feature | SonoSite VevoMD<br>Ultrasound System<br>(this submission) | SonoSite<br>VevoMD<br>Ultrasound<br>System<br>(K160674) | SonoSite X-Porte<br>Ultrasound System<br>(K171437) |
| Acoustic<br>Output<br>Display &<br>FDA Limits | MI Output Display<br>TI Output Display | MI Output<br>Display<br>TI Output Display | Display Feature for<br>Higher Outputs<br>MI Output Display<br>TI Output Display |
| Modes of<br>Operation | 2D Mode, Color<br>Doppler, M-Mode,<br>Power Doppler, PW<br>Doppler | B-mode (2-D<br>Grayscale<br>Imaging<br>Color Doppler<br>Combination<br>Modes<br>M-mode | B-mode Grayscale<br>Imaging<br>Tissue Harmonic<br>Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom<br>Combination Modes<br>Pulsed Wave (PW)<br>Doppler<br>Continuous Wave<br>(CW) Doppler<br>SonoHD2 Noise<br>Reduction<br>SonoMB/MBe Image<br>Compounding<br>Steered CW Doppler<br>Velocity Color<br>Doppler<br>Tissue Doppler<br>Imaging (TDI) |
| | | | |
| Feature | SonoSite VevoMD<br>Ultrasound System<br>(this submission) | SonoSite<br>VevoMD<br>Ultrasound<br>System<br>(K160674) | SonoSite X-Porte<br>Ultrasound System<br>(K171437) |
| DICOM | DICOM 3.0 Store,<br>Print, Modality<br>Worklist, Perform<br>Procedure Step<br>(PPS), Storage<br>Commitment | DICOM 3.0 Store,<br>Print, Modality<br>Worklist, Perform<br>Procedure Step<br>(PPS), Storage<br>Commitment | DICOM 3.0 Store,<br>Print, Modality<br>Worklist, Perform<br>Procedure Step (PPS),<br>Storage Commitment |
| IMT<br>Measurement | Manual<br>Measurement<br>available on the<br>ultrasound system<br>itself. | Manual<br>Measurement<br>available on the<br>ultrasound system<br>itself. | Not available |
| #Transmit<br>Channels | 64 digital channels | 64 digital<br>channels | 128 digital channels |
| #Receive<br>Channels | 64 digital channels | 64 digital<br>channels | 64 digital channels<br>(128 digital channels<br>using Synthetic<br>Aperture) |
| 510(k) Track | Track 3 | Track 3 | Track 3 |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square above the letter "I". Below the logo, the words "Value from Innovation" are written in a smaller, black font. The logo is simple and modern.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in bold, black letters. The "I" in Fujifilm has a red square on the top right corner. Below the word "FUJIFILM" is the tagline "Value from Innovation".
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red accent on the "I" in Fujifilm. Below the logo is the text "Value from Innovation" in black.
#### 7) Determination of Substantial Equivalence:
### Summary of Non-Clinical Tests:
The Vevo MD Imaging System has been evaluated for electrical, thermal, mechanical and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated, and the device has been found to conform to applicable mandatory medical device safety standards. Assurance of quality was established by employing the following elements of product development: Design Phase Reviews, Risk Assessment, Requirements Development, System and Software Verification, Hardware Verification, Safety Compliance Verification, Clinical Validation. All patient contact materials are biocompatible and are materials that are already used in the predicate device or meet 10993.
The Vevo MD Imaging System is designed to comply with the following voluntary standards.
| Reference No. | Recognition<br>No. | Title |
|--------------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI/ANSI/ISO<br>10993-1 | 2-220 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process |
| IEC 60601-1 | 19-4 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance |
| IEC 60601-1-2 | 19-8 | Medical Electrical Equipment - Part 1-2: General<br>Requirements For Basic Safety And Essential<br>Performance - Collateral Standard: Electromagnetic<br>Disturbances - Requirements And Tests (Edition 4) |
| IEC 60601-2-37 | 12-293 | Particular Requirements for the basic safety and<br>essential performance of ultrasonic medical<br>diagnostic and monitoring equipment (Edition 2.1) |
| IEC 62359 | 12-316 | Ultrasonic - Field Characterization - Test Methods For<br>The Determination Of Thermal And Mechanical<br>Indices Related To Medical Diagnostic Ultrasonic<br>Fields |
| IEC 62304 | 13-79 | Medical Device Software - Software Life Cycle<br>Processes |
| ISO 14971 | 5-40 | Medical devices - Application of risk management to<br>medical devices |
| NEMA UD 2 | 12-105 | Acoustic Output Measurement Standard for<br>Diagnostic Ultrasound Equipment |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red square on top of the letter 'I'. Below the logo is the text 'Value from Innovation' in black font. The logo is simple and modern.
Image /page/12/Picture/1 description: The image shows the word "SonoSite" in a bold, sans-serif font. The letters are black against a white background. The word is horizontally oriented and centered in the image.
#### Summary of Clinical Tests:
The Vevo MD Imaging System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.
#### 8) Conclusion:
Intended uses and other key features are consistent with traditional clinical practice and FDA guidance. The Vevo MD device and predicate both conform to applicable electromedical device safety standards with compliance verified through independent evaluation. The Vevo MD device and predicate both meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence and are manufactured using the same ISO 13485 quality system. FUJIFILM SonoSite, Inc. believes that the Vevo MD Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.