← Product Code [HQF](/productcode/HQF) · K190448

# LightLas Pattern Scanning System - TruScan Pro (K190448)

_Light-Med(Usa), Inc. · HQF · May 24, 2019 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190448

## Device Facts

- **Applicant:** Light-Med(Usa), Inc.
- **Product Code:** [HQF](/productcode/HQF.md)
- **Decision Date:** May 24, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4390
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

There are four selected wavelengths (532nm, 577mm, 810mm) in TruScan Pro Laser Console integrated with single/multi spots of TruScan module integrated with slitlamp or single spot of LIO or endo ocular probes. The 532nm and 810nm laser wavelengths are intended to be used for: - Pan-retinal photocoagulation for proliferative diabetic retinopathy- with Slitlamp or Endoprobe - Laser retinopexy for retinal tear and detachments with Slitlamp or Endoprobe - Focal or grid photocoagulation for clinically significant macular edema (CSME)- with Slitlamp - Focal photocoagulation for choroidal neovascularization (CNV) including but not limited to CNV in the setting of wet age-related macular degeneration (wet AMD)- with Slitlamp - Trabeculoplasty for primary open angle glaucoma (POAG)- with Slitlamp. The 577nm laser wavelength is in the treatment of ocular pathology in the posterior segment; Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structure abnormalities of the retina and choroid including: - Proliferative and Severe and very severe nonproliferative diabetic retinopathy - Clinically Significant Macular edema - Choroidal neovascularization - Branch and central retinal vein occlusion - The treatment of choroidal neovascularization associated with wet age-related macular degeneration - Lattice degeneration - Retinal tears and detachments. The 670 nm laser wavelength is indicated for retinal and pan-retinal photocoagulation.

## Device Story

TruScan Pro is an ophthalmic surgical laser system; delivers four selectable wavelengths (532nm, 577nm, 670nm, 810nm) for photocoagulation and trabeculoplasty. System includes laser console, PC interface, 15-inch LCD touch display, and delivery units (integrated slitlamp, laser indirect ophthalmoscope, or endo-ocular probes). Operated by physicians in clinical settings. Device uses galvanometer-based scanning for single or multi-spot delivery. Physician selects parameters via touch interface; system controls laser energy delivery to target ocular tissue. Benefits include precise treatment of retinal/choroidal vascular abnormalities and glaucoma. System integrates power supplies and electronic control systems to manage laser output and safety.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1, -5, -10, -12 for patient-contact components), electrical safety (ANSI AAMI ES60601-1), EMC (IEC 60601-1-2), and software verification/validation.

## Technological Characteristics

Ophthalmic laser system; 4 wavelengths (532, 577, 670, 810nm); 635nm aiming beam. Galvanometer-based scanning. Fan-cooled with TEC for diodes/crystals. 15-inch LCD touch interface. Standards: ANSI AAMI ES60601-1, IEC 60825-1, IEC 60601-2-22, IEC 60601-1-2, ISO 15004-1. Class IV treatment laser, Class II aiming laser.

## Regulatory Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

## Predicate Devices

- TruScan 577 ([K142172](/device/K142172.md))

## Reference Devices

- LightLas 532/810 ([K152688](/device/K152688.md))
- LightLas 577/670 ([K124043](/device/K124043.md))
- CSO SL980/SL9800 slitlamp ([K992836](/device/K992836.md))
- Keeler All pupil II ([K854244](/device/K854244.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with a stylized human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION".

May 24, 2019

Light-Med (USA), INC. Hedy Chiang Regulatory Affairs Administrator 1130 Calle Cordillera San Clemente, CA 92673

Re: K190448

Trade/Device Name: LightLas Pattern Scanning System - TruScan Pro Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF Dated: February 1, 2019 Received: February 25, 2019

Dear Hedy Chiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190448

Device Name TruScan Pro

Indications for Use (Describe)

There are four selected wavelengths (532nm, 577mm, 810mm) in TruScan Pro Laser Console integrated with single/multi spots of TruScan module integrated with slitlamp or single spot of LIO or endo ocular probes.

The 532nm and 810nm laser wavelengths are intended to be used for:

- · Pan-retinal photocoagulation for proliferative diabetic retinopathy- with Slitlamp or Endoprobe
- · Laser retinopexy for retinal tear and detachments with Slitlamp or Endoprobe
- · Focal or grid photocoagulation for clinically significant macular edema (CSME)- with Slitlamp
- Focal photocoagulation for choroidal neovascularization (CNV) including but not limited to CNV in the setting of wet age-related macular degeneration (wet AMD)- with Slitlamp
- · Trabeculoplasty for primary open angle glaucoma (POAG)- with Slitlamp

The 577nm laser wavelength is in the treatment of ocular pathology in the posterior segment; Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structure abnormalities of the retina and choroid including:

- · Proliferative and Severe and very severe nonproliferative diabetic retinopathy
- · Clinically Significant Macular edema
- · Choroidal neovascularization
- · Branch and central retinal vein occlusion
- · The treatment of choroidal neovascularization associated with wet age-related macular degeneration
- · Lattice degeneration
- · Retinal tears and detachments

The 670 nm laser wavelength is indicated for retinal and pan-retinal photocoagulation.

| Type of Use (Select one or both, as applicable)                                                                                                                     |                                                                                                            |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D) <span style="font-size: 2em; vertical-align: middle;">☑</span> | Over The Counter Use (21 CFR 801 Subpart C) <span style="font-size: 2em; vertical-align: middle;">☐</span> |

|X | Prescription Use (Part 21 CFR 801 Subbart DI ne-Counter Use (21 CFR 801 Subbart Ci

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### K190448

# 510(k) Summary

# I. SUBMITTER

LIGHTMED USA, INC. 1130 Calle Cordillera, San Clemente, CA 92673, U.S.A. T: +1-949-218-9555 F: +1-949-218-9556 Contact Person: Hedy Chiang/Regulatory Affairs Coordinator Date Prepared: Apr. 26. 2019

# II. DEVICE

Trade/Device Name: TruScan Pro Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF

## III. PREDICATE DEVICE

Substantial equivalence to the following predicate device is as follows:

| Main Predicated Device | LIGHTMED Corp.<br>TruScan 577      | K142172 | Decision Date: 02/27/2015 |
|------------------------|------------------------------------|---------|---------------------------|
| Reference Device       | LIGHTMED Corp.<br>LightLas 532/810 | K152688 | Decision Date: 06/15/2016 |
| Reference Device       | LIGHTMED Corp.<br>LightLas 577/670 | K124043 | Decision Date: 9/26/2013  |

{4}------------------------------------------------

#### IV. DEVICE DESCRIPTION

There are four selected laser wavelengths in TruScan Pro Laser console consists of integrated slitlamp for single and multi-spots laser delivery or optional LIO, Endoocular probes laser delivery units. The 532nm or 577nm laser wavelength is the primary treatment laser or configured with 670nm or 810nm secondary treatment laser to be used. Laser console is developed along with the Electronic Control system, Power Supplies, and PC Interface platform inside, and connect with 15'' LCD touch display and ophthalmic Electrical Table for use.

The TruScan Pro compatible Laser delivery units include:

TruScan module Integrated CSO SL980/SL9800 slitlamp (K992836)

Laser Indirect Ophthalmoscope using a Keeler All pupil II (K854244)

Endophotocoagulation handpieces (compatible with 510(k) cleared ophthalmic endo probes)

#### V. INDICATIONS FOR USE

There are four selected wavelengths (532nm, 577nm, 810nm) in TruScan Pro Laser Console integrated with single/multi spots of TruScan module integrated with slitlamp or single spot of LIO or endo-ocular probes.

The 532nm and 810nm laser wavelengths are intended to be used for:

- · Pan-retinal photocoagulation for proliferative diabetic retinopathy- with Slitlamp or Endoprobe
- · Laser retinopexy for retinal tear and detachments with Slitlamp or Endoprobe
- Focal or grid photocoagulation for clinically significant macular edema . (CSME)- with Slitlamp
- . Focal photocoagulation for choroidal neovascularization (CNV) including but not limited to CNV in the setting of wet age-related macular degeneration (wet AMD)- with Slitlamp
- Trabeculoplasty for primary open angle glaucoma (POAG)- with Slitlamp .

The 577nm laser wavelength is in the treatment of ocular pathology in the posterior segment: retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structure abnormalities of the retina and choroid including:

- Proliferative and severe and very severe non-proliferative diabetic retinopathy .

{5}------------------------------------------------

- . Clinically significant macular edema
- . Choroidal neovascularization
- Branch and central retinal vein occlusion .
- The treatment of choroidal neovascularization associated with wet age-related macular . degeneration
- Lattice degeneration .
- . Retinal tears and detachments

The 670nm laser wavelength is indicated for retinal and pan-retinal photocoagulation.

{6}------------------------------------------------

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A summary of TruScan Pro design characteristics that are identical as and differ from

| Device                         | Subject Device                                                                                                                          | Main predicated<br>device                                                                                                               | Reference Device                                                                       | Reference<br>Device                                                                       | Comparison<br>of Same or<br>Difference |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|----------------------------------------|
| Model name                     | TruScan Pro                                                                                                                             | TruScan 577                                                                                                                             | LightLas532/810                                                                        | LightLas577/670                                                                           |                                        |
| Company Name                   | LIGHT-MED(USA),<br>INC.                                                                                                                 | LIGHTMED Corp.                                                                                                                          | LIGHTMED Corp.                                                                         | LIGHTMED<br>Corp.                                                                         |                                        |
| 510(k) Number                  | K190448                                                                                                                                 | K142172                                                                                                                                 | K152688                                                                                | K124043                                                                                   |                                        |
| Laser wavelength               | Treatment: 532nm,<br>577nm, 670nm,<br>810nm<br>Aiming: 635nm                                                                            | Treatment: 577nm<br>Aiming: 635nm                                                                                                       | Treatment: 532nm,<br>810nm<br>Aiming: 635nm                                            | Treatment:<br>577nm, 670nm<br>Aiming: 635nm                                               | Same                                   |
| Power Output                   | 532nm: 2000mW<br>577nm: 2000mW<br>670nm: 700mW<br>810nm: 3000mW<br>635nm: <1mW                                                          | 577: 2000mW<br>635nm: <1mW                                                                                                              | 532nm: 2000mW<br>810nm: 3000mW<br>635nm: <1mW                                          | 577nm:<br>2000mW<br>670nm: 700mW<br>635nm: <1mW                                           | Same                                   |
| Laser Safety<br>Classification | Treatment Laser: 4/IV<br>Aiming Laser: 2/II                                                                                             | Treatment Laser: 4/IV<br>Aiming Laser: 2/II                                                                                             | Treatment Laser: 4/IV<br>Aiming Laser: 2/II                                            | Treatment Laser:<br>4/IV<br>Aiming Laser:<br>2/II                                         | Same                                   |
| Intended use                   | Ophthalmic Surgical<br>Laser System                                                                                                     | Ophthalmic Surgical<br>Laser System                                                                                                     | Ophthalmic Surgical<br>Laser System                                                    | Ophthalmic<br>Surgical Laser<br>System                                                    | Same                                   |
| Laser Delivery Unit            | Galvanometer<br>Integrated Slitlamp<br>LIO<br>Endo probe                                                                                | Galvanometer<br>Integrated Slitlamp<br>LIO<br>Endo probe                                                                                | Integrated Slitlamp<br>LIO<br>Endo probe                                               | Integrated<br>Slitlamp<br>LIO<br>Endo probe                                               |                                        |
| Spot size                      | Single spot: 50μm,<br>100μm, 200μm,<br>300μm, 400μm,<br>500μm, 600μm,<br>800μm, 1000μm<br>Multi spots: 100μm,<br>200μm, 300μm,<br>400μm | Single spot: 50μm,<br>100μm, 200μm,<br>300μm, 400μm,<br>500μm, 600μm,<br>800μm, 1000μm<br>Multi spots: 100μm,<br>200μm, 300μm,<br>400μm | Single spot: 50μm,<br>100μm, 200μm,<br>300μm, 400μm,<br>500μm, 600μm,<br>800μm, 1000μm | Single spot:<br>50μm, 100μm,<br>200μm, 300μm,<br>400μm, 500μm,<br>600μm, 800μm,<br>1000μm | Same                                   |
| Exposure time                  | Single Spot 0.01s to<br>3S continuous<br>Multi Spots 0.01s &<br>0.02s                                                                   | Single Spot 0.01s to<br>3S continuous<br>Multi Spots 0.01s &<br>0.02s                                                                   | 0.01s to 3S continuous                                                                 | 0.01s to 3S<br>continuous                                                                 | Same                                   |
| Pattern Type                   | 532nm: single spot,<br>Straight Line, Square,<br>Triangle, Circle, Arc                                                                  | 577nm: single spot,<br>Straight Line, Square,<br>Triangle, Circle, Arc                                                                  | 532nm: single spot<br>810nm: single spot                                               | 577nm: single<br>spot 670nm:<br>single spot                                               | Same                                   |
| Type of scanning system        | Galvanometer                                                                                                                            | Galvanometer                                                                                                                            | NA                                                                                     | NA                                                                                        | Same as Main predicated device         |
| Power requirements Voltage     | 100 to 230Vac                                                                                                                           | 100 to 230Vac                                                                                                                           | 100 to 230Vac                                                                          | 100 to 230Vac                                                                             | Same                                   |
| Dimensions                     | 40.5cm L x 40.5cm W x 13 cm H                                                                                                           | 33cm L x 37cm W x 13cm H                                                                                                                | 33cm L x 37cm W x 13cm H                                                               | 33cm L x 37cm W x 13cm H                                                                  | Different                              |
| Weight                         | 22Kg                                                                                                                                    | 33Kg                                                                                                                                    | 33Kg                                                                                   | 33Kg                                                                                      | Different                              |
| Computer Control               | YES the PC interface board is attached inside the laser console and connect with 15" touch display                                      | YES the PC is attached with 11" LCD touch panel                                                                                         | YES the PC is attached with 11" LCD touch panel                                        | YES the PC is attached with 11" LCD touch panel                                           | Different                              |
| Cooling method                 | Fan cooled and TEC's for Laser Diodes and Crystal                                                                                       | Fan cooled and TEC's for Laser Diodes and Crystal                                                                                       | Fan cooled and TEC's for Laser Diodes and Crystal                                      | Fan cooled and TEC's for Laser Diodes and Crystal                                         | Same                                   |
| Applicable Standards           | ANSI AAMI60601-1<br>EMC<br>IEC60825-1<br>IEC60601-2-22<br>ISO15004-1                                                                    | IEC60601-1<br>EMC<br>IEC60825-1<br>IEC60601-2-22<br>ISO15004-1                                                                          | IEC60601-1<br>EMC<br>IEC60825-1<br>IEC60601-2-22<br>ISO15004-1                         | IEC60601-1<br>EMC<br>IEC60825-1<br>IEC60601-2-22<br>ISO15004-1                            | Same                                   |

the predicate is provided below:

577nm: single spot, Straight Line, Square, Triangle, Circle, Arc 670nm: single spot 810nm: single spot

{7}------------------------------------------------

TruScan Pro is an extended model to compare the main predicated device and reference devices designed and manufactured by LIGHTMED.

The indications for Use statement and Laser delivery unit for the TruScan Pro is similar with predicated devices.

The major difference between TruScan Pro and predicated devices is the hardware development in order to drive maximum four wavelengths in one laser console; however, the difference do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. The performance data to prove the safety and performance of TruScan Pro are provides in Section VII of this summary below.

## VII. PERFORMANCE DATA

The following performance data were provided to support substantial equivalence determination.

#### Biocompatibility testing

There are no direct patient contact components in TruScan Pro Laser System but the integrated slitlamp supplied by original manufacturer with patient contact component 

{8}------------------------------------------------

such as chinrest and headrest were subjected to In vitro Cytotoxity, Irritation, and skin sensitization testing in accordance with FDA guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."

The patient contact materials for Chinrest and headrest are surface contact/skin device only. Contact duration is limited to less than 30 minutes. (which is shorten than 24 hours mentioned in ISO 10993 A-limited.) The Biocompatibility testing results showed patient contact materials of integrated slitlamp conform to ISO 10993-1, 10993-5, ISO 10993-10, and ISO 10993-12 standards.

{9}------------------------------------------------

## Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on TruScan Pro. according to

applicable federal and international safety and performance standards:

- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 ●
- IEC 60825-1: 2014 •
- IEC 60601-2-22:2007+A1:2012 ●
- IEC 60601-1-2:2014
- IEC 60601-1-6:2013
- EN ISO 14971:2012
- ISO15004-1:2006 ●

### The testing results demonstrate that TruScan Pro comply with all the required standards.

### Software verification and validation testing

According to the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005)", the level of concern of TruScan Pro and its software is determined to be Major. The Software of TruScan Pro controls the delivery of laser energy. A failure of operation or latent flaw of TruScan Pro could directly result in serious injury to the patient or operator.

The software verification and validation results confirm the fulfillment of software requirement specifications.

## VII. CONCLUSIONS

The TruScan Pro is substantially equivalent to the predicate devices in technical characteristics, design features, operating principles, functional and performance characteristics, and for the intended uses in the stated medical specialties. The TruScan Pro is designed to comply with applicable federal and international safety and performance standards. There are no new safety and effectiveness issues being raised.

---

**Source:** [https://fda.innolitics.com/device/K190448](https://fda.innolitics.com/device/K190448)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
