Immucise

K190427 · Terumo Corporation · FMF · Jul 18, 2019 · General Hospital

Device Facts

Record IDK190427
Device NameImmucise
ApplicantTerumo Corporation
Product CodeFMF · General Hospital
Decision DateJul 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Immucise Intradermal Injection System is injections of FDA approved drugs. The system is to be used in the deltoid region for infants aged two months (excluding low birth weight and/or preterm birth) to adults.

Device Story

Single-use, manual intradermal injection system; consists of needle and syringe. Input: FDA-approved drugs. Operation: Manual injection into deltoid region. Output: Intradermal drug delivery. Used in clinical settings by healthcare providers. Benefits: Facilitates precise intradermal drug administration across expanded age range (infants 2+ months to adults).

Clinical Evidence

No human clinical data. Evidence consists of bench testing (from K181369) and animal studies. Animal studies evaluated wheal formation success rate compared to reference device and histopathological injection depth validation to support pediatric indication.

Technological Characteristics

Manual injection system. Materials: Stainless steel cannula, polypropylene hub/barrel/luer lock, styrene-based TPE, silicone oil lubricant, polyacrylate adhesive, polystyrene plunger. Sterilization: Electron beam radiation. Dimensions: 33G (0.2 mm) needle, 1.15 mm length, 0.4 mL syringe capacity. No software/electronics.

Indications for Use

Indicated for intradermal injections of FDA approved drugs in the deltoid region for infants aged two months (excluding low birth weight and/or preterm birth) to adults.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 18, 2019 Terumo Corporation % Rudima Jackson Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Ave Suite 320 Somerset, New Jersey 08873 Re: K190427 Trade/Device Name: Immucise Intradermal Injection System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI, FMF Dated: June 19, 2019 Received: June 20, 2019 Dear Rudima Jackson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190427 #### Device Name Immucise Intradermal Injection System Indications for Use (Describe) The Immucise Intradermal Injection System is injections of FDA approved drugs. The system is to be used in the deltoid region for infants aged two months (excluding low birth weight and/or preterm birth) to adults. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary: K190427 | Manufacturer: | Kofu Factory of Terumo Corporation (510(k) applicant)<br>Registration Number: 9681835 | |-----------------------------|-------------------------------------------------------------------------------------------------------------------| | Address: | 1727-1 Tsuijiarai, Showa-Cho<br>Nakakoma-Gun, Yamanashi, Japan 409-3853 | | Corresponding official: | Rudima L Jackson<br>Regulatory Affairs Specialist<br>Terumo Medical Corporation | | Telephone Number:<br>Email: | (732)302-4900<br>rudima.jackson@terumomedical.com | | Summary Date: | July 8, 2019 | | Trade Name: | Immucise Intradermal Injection System | | Common or Usual Name: | Intradermal Needle and Syringe | | Regulation Number: | 21 CFR 880.5570 | | Regulation Name: | Hypodermic Single Lumen Needle | | Product Code: | FMI (Needle) and FMF (Syringe) | | Class: | II | | Panel: | General Hospital | | Primary Predicate Device: | Immucise Intradermal Injection System, Manufactured by<br>Kofu Factory of Terumo Corporation (Japan)<br>(K181369) | | Reference Device: | Intradermal Adapter, Manufactured by West<br>Pharmaceutical Services, Inc.<br>(K123588) | # Device Description: The Immucise Intradermal Injection System is a single use, electron beam radiation sterilized device that is designed to be used for intradermal injections of FDA approved drugs. This system consists of an Intradermal Injection Needle and an Immucise Syringe. The system is made of common materials used in needles and syringes. The Immucise Intradermal Injection Needle is comprised of a needle base. The Immucise Syringe is comprised of a barrel, a gasket and a plunger. The sterile Immucise Intradermal Injection Needle and Immucise Syringe are packed separately and assembled prior to use. The Immucise Intradermal Injection System is operated by a manual process. {4}------------------------------------------------ # Indications for Use: The Immucise Intradermal Injection System is indicated for intradermal injections of FDA approved drugs. The system is to be used in the deltoid region for infants aged two months (excluding low birth weight and/or preterm birth) to adults. ## Technological Characteristics: The Immucise Intradermal Injection System with the pediatric indication is the same as its primary predicate device, K181369. The only change is the expanded indication to include infants aged two months (excluding low birth weight and/or preterm birth) to adults. Therefore, all technological characteristics are the exact same. ## Summary of Comparative Information This traditional 510(k) is being submitted for the Immucise Intradermal Injection System with the intent of expanding the indication to include infants aged two months (excluding low birth weight and/or preterm birth) to adults. The subject device is the same device as the primary predicate cleared under K181369 except for the Indications for Use. It has the same intended use, operating principle, design, construction, materials, and sterilization method. The proposed device also has the same design specification to the needle gauge, needle length, and syringe nominal capacity. | Feature of the<br>device | Subject Device<br><br>Immucise<br>Intradermal<br>Injection System | Primary Predicate<br>(K181369)<br>Immucise<br>Intradermal<br>Injection System | Reference Device<br>(K123588)<br>Intradermal<br>Adapter | Discussion /<br>Comment to<br>Primary<br>Predicate | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | FMI, FMF | FMI, FMF | FMF | Same as<br>K181369 | | Classification | II | II | II | Same as<br>K181369 | | Number of<br>uses | Single use Rx only | Single use Rx only | Single use Rx only | Same as<br>K181369 | | Materials | The subject device is<br>constructed of<br>materials commonly<br>used in medical<br>devices.<br><br>Needle<br>• Needle Tube<br>• Stainless steel<br>cannula*<br>• Silicone oil lubricant*<br>• Needle base<br>• Polypropylene hub* | The subject device is<br>constructed of<br>materials commonly<br>used in medical<br>devices.<br><br>Needle<br>• Needle Tube<br>• Stainless steel<br>cannula*<br>• Silicone oil lubricant*<br>• Needle base<br>• Polypropylene hub* | The West<br>Pharmaceutical<br>device is<br>constructed of<br>materials<br>commonly used in<br>medical devices.<br><br>Adapter<br>• Polycarbonate | Same as<br>K181369 | | | • Polypropylene luer<br>lock connector*<br>• Styrene-based TPE w/<br>pigment elastic spacer*<br>• Polyacrylate needle<br>hub adhesive<br><br>Syringe<br>• Barrel<br>• Polypropylene barrel*<br>• Silicone oil lubricant*<br>• Black ink barrel<br>printing<br>• Gasket<br>• Styrene-based TPE w/<br>pigment*<br>• Silicone oil lubricant*<br><br>• Plunger<br>• Polystyrene<br><br>*Patient body<br>contacting material | • Polypropylene luer<br>lock connector*<br>• Styrene-based TPE w/<br>pigment elastic spacer*<br>• Polyacrylate needle<br>hub adhesive<br><br>Syringe<br>• Barrel<br>• Polypropylene barrel*<br>• Silicone oil lubricant*<br>• Black ink barrel<br>printing<br>• Gasket<br>• Styrene-based TPE w/<br>pigment*<br>• Silicone oil lubricant*<br><br>• Plunger<br>• Polystyrene<br><br>*Patient body<br>contacting material | | | | Sterilization<br>method | Electron Beam<br>radiation | Electron Beam<br>radiation | Ethylene Oxide | Same as<br>K181369 | | Biocompatibility | Biocompatible<br>The finished device's<br>patient contacting<br>parts were assessed in<br>accordance with tests<br>recommended in the<br>FDA Guidance - Use<br>of International<br>Standard ISO-10993-<br>1, "Biological<br>evaluation of medical<br>devices - Part 1:<br>Evaluation and testing<br>within a risk<br>management<br>process." | Biocompatible<br>The finished device's<br>patient contacting<br>parts were assessed in<br>accordance with tests<br>recommended in the<br>FDA Guidance - Use<br>of International<br>Standard ISO-10993-<br>1, "Biological<br>evaluation of medical<br>devices - Part 1:<br>Evaluation and testing<br>within a risk<br>management<br>process." | Information not<br>available | Same as<br>K181369 | | Indications for<br>Use | The Immucise<br>Intradermal Injection<br>System is indicated<br>for use in intradermal<br>injections of FDA<br>approved drugs.<br><br>The system is to be<br>used in the deltoid<br>region for infants<br>aged two months<br>(excluding low birth<br>weight and/or<br>preterm birth) to<br>adults | The Immucise<br>Intradermal Injection<br>System is indicated<br>for use in intradermal<br>injections of FDA<br>approved drugs.<br><br>The system is to be<br>used in the deltoid<br>region for adults. | The Intradermal<br>Adapter is an<br>accessory to a 1<br>ml, ½ inch fixed-<br>needle allergy<br>syringe indicated<br>for use as a guide<br>for performing<br>intradermal<br>injections. | The proposed<br>indications for<br>use for the<br>subject device<br>have been<br>modified to<br>include a<br>specific user<br>population<br>(pediatric) to<br>the primary<br>predicate. The<br>differences in<br>the indications | | | | | | | | Operating<br>principles | Manual | Manual | Manual | a new intended<br>use. As the<br>intended use in<br>the subject and<br>predicate<br>devices has not<br>changed, the<br>modification to<br>the indications<br>for use does not<br>raise different<br>questions of<br>safety and<br>effectiveness. | | | | | | Same as<br>K181369 | | Design<br>specifications | Needle Gauge<br>33G (0.2 mm)<br>Needle Length<br>1.15 mm<br>Syringe Nominal<br>Capacity<br>0.4 mL | Needle Gauge<br>33G (0.2 mm)<br>Needle Length<br>1.15 mm<br>Syringe Nominal<br>Capacity<br>0.4 mL | Adapter<br>Designed to adapt<br>to a disposable 1cc<br>piston syringe with<br>needle (1 ml of<br>capacity, 1/2"<br>needle length, 27 –<br>29G needle gauge). | Same as<br>K181369 | | Package<br>(Primary) | Needle<br>Individual package<br>Syringe<br>Blister package | Needle<br>Individual package<br>Syringe<br>Blister package | Adapter<br>Blister package | Same as<br>K181369 | | Shelf life | Needle: 36 months<br>Syringe: 12 months | Needle: 36 months<br>Syringe: 12 months | Information not<br>available | Same as<br>K181369 | ### Summary of comparative information {5}------------------------------------------------ {6}------------------------------------------------ # Performance Data: # Bench Test Bench testing for Immucise Intradermal Injection System was performed in K181369 using non-aged and aged samples and met the predetermined acceptance criteria. No additional bench tests were deemed necessary to support the expanded indication, since the subject device is identical to the primary predicate K181369. # Animal Test Terumo conducted an animal study to demonstrate the intended use of intradermal injection for adults in K181369. In order to expand the indications for use from adults only to infants (aged two months) to adults, Terumo conducted an additional animal study. The test method and acceptance criteria are same as the study performed previously except for the skin thickness. {7}------------------------------------------------ | Test Name | Test Description / Standard | |---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Animal study<br>(Functionality test) | In-house standard: This study was conducted to evaluate<br>the efficacy of Immucise Intradermal Injection System for<br>the indication for use. The test compared the post-injection<br>wheal formation success rate between the Immucise<br>Intradermal Injection System and the reference device of<br>West Intradermal Adapter. | | Animal study<br>(Histopathological<br>evaluation) | In-house standard: This study was conducted to validate<br>the injection depth required for the indications for use. | Based on these animal studies, Terumo concludes that the Immucise Intradermal Injection System has been found to be substantially equivalent for the intended use and for the target patient population without raising different questions of safety and effectiveness. The test results support a determination of substantial equivalence. # Conclusion: In summary, the intended use, technology/principle of operation, materials, and performance showed that the subject device did not raise different questions of safety and effectiveness when compared to the predicate devices. Therefore, Immucise Intradermal Injection System is substantially equivalent to the legally marketed Predicate Device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...