K190382 · Company Elamed · HKX · Dec 6, 2019 · Ophthalmic
Device Facts
Record ID
K190382
Device Name
Intraocular Pressure Tonometer EASYTON
Applicant
Company Elamed
Product Code
HKX · Ophthalmic
Decision Date
Dec 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The Tonometer Easyton is indicated for the measurement of intraocular pressure in human eyes.
Device Story
Easyton is a portable, battery-powered, transpalpebral tonometer used to measure intraocular pressure (IOP) through the eyelid. The device features an electromagnetic vibrator rod that contacts the eyelid; the rod applies static pressure (approx. 10g) and dynamic vibration (approx. 150 Hz). The system forms an elastic coupling with the eye; the device measures the frequency of free damped vibrations of this system, which correlates with IOP. The device is operated by healthcare professionals; it provides a digital IOP reading on an LCD screen after a 2-second measurement cycle. It does not require anesthesia or direct corneal contact. The device benefits patients by providing a non-invasive, rapid method for IOP assessment, suitable for various anatomical eyelid types.
Clinical Evidence
Three prospective comparative clinical studies were conducted in Russia, Kazakhstan, and India (total N=381 eyes) comparing Easyton to Goldmann Applanation Tonometry per ANSI Z80.10-2014. Populations included varying eyelid stiffness and epicanthal folds. Results showed measurement errors within ±5% tolerance. Bias ranged from -0.88 to +0.32 mmHg with SDs between ±2.10 and ±4.01 mmHg. Studies confirmed performance across anatomical variations.
Technological Characteristics
Portable, battery-powered (2x AAA) device. Patient-contacting materials: Alloy D16 (tonometer stock) and BRAZH9-4 (ring). Sensing principle: Inductive transducer measuring vibration frequency of an electromagnetic rod. Dimensions: 173x27x21 mm; Weight: 88g. Not sterilized; reusable parts require disinfection. No connectivity/network features. Software level of concern: Moderate.
Indications for Use
Indicated for measurement of intraocular pressure in human eyes. No specific age or gender contraindications noted.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2019
Company Elamed Mikhail Kuzin Certification Engineer Yelatma instrument-making enterprise JSC 391351, 25 Janina St. Yelatma, RU Ryazan region
Re: K190382
Trade/Device Name: Intraocular Pressure Tonometer EASYTON Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: Class II Product Code: HKX Dated: October 29, 2019 Received: November 4, 2019
Dear Mikhail Kuzin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general control's provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidanceregulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-devicesafety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally. you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-andconsumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tieuvi Nguyen, Ph.D. Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190382
Device Name Easyton
Indications for Use (Describe)
The Tonometer Easyton is indicated for the measurement of intraocular pressure in human eyes.
Type of Use (Select one or both, as applicable)
| <label> <input type="checkbox"/> Residential Use (Per CAI Statutes, Please see CAI Law Center at CAIonline.org) </label> |
|-----------------------------------------------------------------------------------------------------------------------------|
| <label> <input type="checkbox"/> Commercial Use (Per CAI Statutes, Please see CAI Law Center at CAIonline.org) </label> |
|× | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for ELAMED. The logo is green and features a stylized tree symbol to the left of the company name. The company name is in all caps and is also green. The logo is set against a white background with horizontal green lines.
All for your health. Health for you.
Yelatma Instrument Making Enterprise, JSC
25 Yanin st., Yelatma, Ryazan region, 391351, Russia. Tel./fax: +007 49131 / 2-04-57, 2-21-09, 4-38-29
# 510(k) Summary Complying with 21 CFR 807.9
### I. SUBMITTER
Company Elamed Yelatma instrument-making enterprise, JSC 391351, 25 Janina st., Yelatma, Ryazan region, Russia
Phone: +7 (4912) 513-565 Fax: +7 (49131) 204-57
Contact Persons: Mr. Mikhail Kuzin Certification Engineer Company "ELAMED" Phone:: +74912776059 x 2256 or +79537428930 Address: Russia,Ryazan,Vysokovoltnaya street 48 E-mail: kuzin_mn@elamed.com
Alternate Only: Mr. Borovkov Oleg Registration Team leader Company "ELAMED" E-mail: borovkov@elamed.com Phone: +74912776059 x 2254 Address: Russia,Ryazan,Vysokovoltnaya street 48
## II. DEVICE
Name of Device: Intraocular Pressure Tonometer EASYTON Common or Usual Name: tonometer, ac-powered Regulation: 886.1930 Regulatory Class: II Product Code: HKX
### III. PREDICATE DEVICE
Diaton, K060780 This predicate has not been subject to a design-related recall.
{5}------------------------------------------------
## IV. DEVICE DESCRIPTION
### Device Description
### A. Principles of Operation
Measurement Method
The principle of EASYTON measurement is based on simultaneous use of the two types of measuring actions – static and dynamic. Both actions are carried out by the vibrator's rod on the eyelid.
Static action is carried out during the time of measurement and determined by the weight of the vibrator.
Dynamic exposure represents itself during the vibrating action with frequency: about 150 HZ and amplitude: the millimeter hundredth parts and tactile is felt as soft vibration. Vibrator's rod is elastically movable in its axial direction and is set in oscillatory motion by electromagnetic way. While carrying out the measurement, the rod is put up on the eyelid, sagging (pressing the eyelid down with its weight about 10g and is fixed on sclera or an eye cornea forming elastic system of eye-vibrator with total mechanical rigidity. This system is disturbed from the equilibrium state by the rod's short term electromagnetic removal. In a system within tan equilibrium state restoration there are arising free damped vibration.
Functional connection between elastic system rigidity and the period of is own vibration is known. This period is measured by the tonometer and is used for IOP calculation being shown on the display.
| Image: [Schematic representation of the tonometer rod movements.]<br>Fig. 1<br>Schematic representation of the tonometer<br>rod movements. | When the EASYTON rod is placed on the<br>eyelid and slightly pressed on the device,<br>the rod slightly plunges into the device, and<br>the generation of a measuring vibration<br>effect begins.<br><br>This is schematically shown in Figure 1.<br><br>The vibration frequency of the "rod-eye"<br>depends on the IOP.<br><br>The higher the IOP, the greater the vibration<br>frequency.<br><br>The device registers the vibration frequency,<br>recalculates it into the IOP value and<br>displays it on its indicator. |
|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### EASYTON measures only the intraocular pressure. Operating principle of the tonometer.
{6}------------------------------------------------
### Technical Specifications
| Specification | Value |
|------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Measuring range of IOP, mmHg | 7 to 50 |
| Limits of permissible absolute error of measurement: | |
| From 7 to 23 mm Hg | +/- 2 |
| More than 23 mm Hg | +/- 5 |
| Measurement time (sec) | 2 |
| Repeatability (coefficient of variation), % | ≤8,1 |
| applanation area, contact area, or area of corneal deformation | None |
| type of pressure transducer (e.g., optical detector, strain<br>gauge) and specifications | Inductive |
| Power supply: | |
| Number of elements x voltage, V | 2 x 1,5 |
| Type of battery | AAA |
| Power supply voltage, V | 2 - 3,3 |
| Current consumption in measurement mode, mA, not more | 100 |
| Display | LCD |
| Data output | Display |
| data storage capability (e.g., measurement results, patient<br>information) | Store of last measurement<br>result until power off |
| interface with other equipment (e.g., printer, computer<br>network) | None |
| dimensions (L x H x W) mm, not more | 173 x 27 x 21 |
| weight (g) | 88 |
| Operating temperature range, C (need F) | +10 to +35 |
| Relative air humidity, %, not more | 80 |
| Atmospheric pressure, kPa | 84 to 106.7 |
## V. INDICATIONS FOR USE
The Tonometer Easyton is indicated for the measurement of intraocular pressure in human eyes.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Item | Intraocular<br>Pressure<br>Tonometer<br>EASYTON | DIATON | Substantially<br>Equivalent or Different |
|---------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------|
| Indications for Use | The Tonometer<br>Easyton is indicated<br>for the measurement<br>of intraocular | The Diaton Tonometer<br>is intended to measure<br>intraocular pressure<br>(IOP). The device is | Substantially Equivalent |
{7}------------------------------------------------
| | pressure in human<br>eyes. | intended for use as an<br>aid in the diagnosis of<br>glaucoma and for<br>monitoring IOP. | |
|--------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------|
| Classification | product code HKY<br>(regulation number<br>886.1930) | product code HKY<br>(regulation number<br>886.1930) | Substantially Equivalent |
| Materials | Biocompatible<br>patient contacting<br>materials | Biocompatible patient<br>contacting materials | Substantially Equivalent |
| Sterilization method | not sterilized | not sterilized | Substantially Equivalent |
| Technological<br>Characteristics: | | | |
| Measurement on base | Goldman | Maklakov | Different |
| Measurement range (Hg) | from 7 to 50 | from 5 to 60 | Substantially Equivalent |
| Time of single measurement<br>(sec) | 2 | 3 | Substantially Equivalent |
| Power supply type and<br>voltage | 1,5 V AAA x2 | 1,5 V AAA x2 | Substantially Equivalent |
| Size/dimensions (mm) | 173 x 27 x 21 | 174 x 26 x 20 | Substantially Equivalent |
| Weight | 88 g | 89 g | Substantially Equivalent |
| Features: | | | |
| No direct affect on the<br>cornea | Yes | Yes | Substantially Equivalent |
| Portability | Yes | Yes | Substantially Equivalent |
| Displays independent from<br>corneas crookedness | Yes | Yes | Substantially Equivalent |
| Digital IOP Indication | Yes | Yes | Substantially Equivalent |
| Short time measurement | Yes | Yes | Substantially Equivalent |
| Sterilization is not required | Yes | Yes | Substantially Equivalent |
| Anesthesia is not required | Yes | Yes | Substantially Equivalent |
#### Discussion of substantial equivalence
The indications for use, features, materials and technical parameters (with the exception of the measurement method) are either the same or substantially equivalent to each other. The difference between the two devices are the measurement methods. The difference in the measurement method does not impact safety or effectiveness
All tonometers are known to take IOP measurements indirectly, by measuring the response of the eye coats to a mechanical measuring stimulus.
The main feature of transpalpebral tonometry is the damping effect of the eyelid. The effect is balanced out by compressing the eyelid before the procedure, in the same manner as an eyelid is compressed (pushed against) by fingers before palpation.
The operating principle of DIATON is based on elastic interaction of the freely falling rod with the eyeball through eyelid.
Interaction of the rod with eyeball is the only part that is essential. The duration of this contact is less than 0.1 sec. The mechanical interference (movement of the eye, tremor of the operator's hand, etc.) in this fraction of time has a significant
{8}------------------------------------------------
impact on the measurement result, which is a defect common for all dynamic pulse methods of measurement.
Meanwhile, EASYTON enables a measuring process of a different kind. In this case, the evelid is compressed before the measurement is started, and remains in this compressed state up until the measurement completion. The measuring process implies continuous active interaction of the rod with the eyeball (average time of contact is 1 sec.).
The amplitude of the free vibrations is of minor importance: it is the frequency that counts, which depends on the mass of the rigidity of the eye (IOP). Stability of the result is obtained by filtering 10-20 measuring periods produced during one measurement cycle.
Thus, the measuring method provided by EASYTON successfully combines the advantages of both static and dynamic methods.
The measuring action of EASYTON, just as that of DIATON, is exerted through eyelid in the scleral area corresponding to corona ciliaris in the meridian of 12 hours, and is perceived as mild vibration. No anesthesia is required prior to the measurement, which is taken in one second.
In addition, an improved method of disinfection with Rapicide RA raster registered in the US market was analyzed and laboratory confirmed.
Reference (comparative) studies were carried out on the measurement of IOP with the Tonometer Easyton and Tonometer Goldman in accordance with the requirements of ANSI Z80.10-2014 (ISO 8612) on eyes of representative groups that take into account options for the anatomical features of the eyelids. Studies were conducted on representatives of:
-with low stiffness of the eyelids, in the Glaucoma Community of Russia, IPO (Moscow, Russia);
-with epicanthus, in the Regional Medical Center (Karaganda, Kazakhstan) and TENSV Clinic, LLP (Karaganda, Kazakhstan);
- with the stiffness rigidity of the eyelids, in Hospital Park (Gurgaon, New Delhi, India).
The study showed that the results of measurements with the tonometer Easyton and the Goldman tonometer are in the acceptable ranges of measurement error (no more than ± 5%) for all types of eyes examined.
Based on the above comparisons we have determined that the difference in measuring does not impact safety or effectiveness. Therefore we conclude that the subject device and predicate device are substantially equivalent.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
{9}------------------------------------------------
#### Biocompatibility testing
We have assessed all of our patient contacting materials for biocompatibility requirements in accordance with the Mav 1, 1995 FDA Biocompatibility Guidance, the FDA-modified matrix of the "International Standard ISO-10993", "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", including the flow chart entitled "Biocompatibility Flow Chart for the Selection of Toxicity Tests of 510(k)'s.
We have determined that the patient contacting portion of the device is a body surface contacting device for less than 24 hours.
The patient contacting materials are:
Tonometer stock made of Alloy D16 complies with requirements of GOST 21488-97 with a chemical composition according to GOST 4784 and there is a chance of contact Tonometer ring made of BRAZH9-4 complies with requirements of GOST 1628-78 with chemical composition according to GOST 18175.- working part tonometer (1).
Additionally, an investigation was carried out on non-contacting parts of the Tonometer to completely eliminate hazards. ABS RESIN HI-121- cover, case (2).
The device passed all biocompatibility tests.
#### Validation of Cleaning and Sterization Methods
The device is not sterilized. Infection measures are analyzed of investigation relating to the validation process.
Disinfection instructions for the buffer ring and of the rod is included in the owner's manual. These parts are reusable.
Electrical safety and EMC testing were conducted on the complies with: the IEC 60601-1, IEC 60601-2-2 standards for safety and the IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-8, CISPR 11 standards for EMC.
#### Software Verification
The software level of concern for this device type is moderate based on the answers to the questions in the guidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
#### Clinical Performance
To account for anatomical differences of eyelid structures in the indicated patient population, three comparative studies were carried out for measurements of IOP using the Easyton and Goldmann Applanation Tonometers. All three studies were conducted in accordance with the requirements of ANSI Z80.10-2014 and two of the studies included additional measurements of high astigmatic eves. Brief descriptions and results of these comparative studies are provided below:
{10}------------------------------------------------
- 1) "Low stiffness eyelids study". In this study we enrolled a Caucasian population with low stiffness of the eyelids. The study was conducted in the Glaucoma Community of Russia, IPO (Moscow, Russia). The study was conducted in 156 patients (eyes) distributed in 3 IOP ranges, as follows:
| IOP Range | #of eyes (≤3D Cyl) | #of eyes (>3D Cyl) |
|------------|---------------------|---------------------|
| 7 to 16 | 46 | 10 |
| >16 to <23 | 40 | 10 |
| >23 | 40 | 10 |
The comparative study results are shown in the following graphs:
{11}------------------------------------------------
Image /page/10/Figure/1 description: The image contains two plots comparing TVGD-02 with Goldman. The top plot is a scatter plot with a regression line, showing the relationship between the two variables. The bottom plot is a Bland-Altman plot, which shows the difference between the two variables against their mean. The Bland-Altman plot includes lines indicating the bias, as well as the +95% confidence level and +1.96 standard deviations.
Appendix 1. The regression and Bland-Altman graphs studies for representatives with low stiffess of the eyelids.
{12}------------------------------------------------
- 2) "Epicanthus study". In this study we enrolled an Asian population with epicanthal fold of the eyelid. The study was conducted in Regional Medical Center (Karaganda, Kazakhstan) and TENSV Clinic, LLP (Karaganda, Kazakhstan). The study was conducted in 150 patients (eyes) distributed in 3 IOP ranges, as follows
| IOP Range | #of eyes (≤3D Cyl) | #of eyes (>3D Cyl) |
|------------|--------------------|--------------------|
| 7 to 16 | 40 | 10 |
| >16 to <23 | 40 | 10 |
| ≥23 | 40 | 10 |
The results of the "Epicanthus" comparative study are shown in the following graphs:
{13}------------------------------------------------
Image /page/12/Figure/1 description: The image contains two scatter plots comparing Eazyton and Goldman measurements. The top plot shows Eazyton measurements plotted against Goldman measurements, with a regression line and equation displayed as Eazyton = -0.1722 + 1.0245*x. The correlation coefficient (r) is 0.9765, and the p-value is 0.0000. The bottom plot is a Bland-Altman plot comparing the difference between Goldman and Eazyton measurements against the mean of the two measurements, including lines for bias, +1.96SD, and -1.96SD.
Appendix 2. The regression and Bland-Altman graphs studies for representatives with epicanthus.
{14}------------------------------------------------
- 3. "Rigid stiffness eyelids study". In this study we enrolled an Indian population with rigid stiffness of eyelids. The study was conducted in Hospital Park (Gurgaon, New Delhi, India). The study was conducted in 78 patients (155) distributed in 3 IOP ranges, as follows:
| IOP Range | #of eyes (≤3D Cyl) |
|------------|--------------------|
| 7 to 16 | 53 |
| >16 to <23 | 60 |
| ≥23 | 42 |
The results of the "Rigid stiffness eyelids" comparative study are shown in the following graphs:
{15}------------------------------------------------
Image /page/14/Figure/1 description: The image shows two scatter plots. The top scatter plot shows a positive correlation between two variables. The bottom scatter plot shows the difference between the two variables plotted against the average of the two variables, with a bias of -0.88 and limits of agreement of +1.96 SD: 3.12 and -1.96 SD: -4.89.
Appendix 3. The regression and Bland-Altman graphs studies for representatives with the stiffness of the eyelids
{16}------------------------------------------------
| Comparison<br>Study | Main Group | | | Astigmatic Group | | |
|-----------------------------------------|---------------|-------|--------------------|------------------|-------|--------------------|
| | Bias | Slope | $\pm$ 1.96 SD | Bias | Slope | $\pm$ 1.96 SD |
| Russia<br>(Caucasian, thin<br>eyelids) | -0.25<br>mmHg | 0.978 | $\pm$ 3.78<br>mmHg | +0.83<br>mmHg | 1.035 | $\pm$ 3.13<br>mmHg |
| Kazakhstan<br>(Asian, fatty<br>eyelids) | +0.32<br>mmHg | 1.02 | $\pm$ 2.59<br>mmHg | +0.77<br>mmHg | 0.948 | $\pm$ 2.10<br>mmHg |
| India (stiff<br>eyelids) | -0.88<br>mmHg | 0.923 | $\pm$ 4.01<br>mmHg | | | |
The table below shows the summary of all three comparative studies:
Subjects with the rigid eyelids have a larger scatter of values and a lower correlation coefficient. This suggests that the biomechanical characteristics and anatomical features of the eyelid influence the IOP measurement result. However, it should be noted that this does not significantly affect the measurement results. The results from all three the studies showed that for all type of tested eyelids requirements of Z80.10-2014 are met, because not more than 5% of the paired differences between the reference tonometer reading and the test tonometer reading for each pressure range are greater than the tolerance for that range.
It can be concluded that IOP measurements by the tonometer «EASYTON» correspond to the declared accuracy for all representative groups that take into account options for the anatomical features of the eyelids.
#### VIII. CONCLUSIONS
Substantial equivalence comparison and bench performance tests support the conclusion of substantial equivalence of the Easyton Tonometer to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.