← Product Code [PEN](/productcode/PEN) · K190341

# iC-GN iC-Cassette for use on the iC-System (K190341)

_Icubate, Inc. · PEN · Jun 28, 2019 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190341

## Device Facts

- **Applicant:** Icubate, Inc.
- **Product Code:** [PEN](/productcode/PEN.md)
- **Decision Date:** Jun 28, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3365
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K190341 · Jun 28, 2019 | iC-GN iC-Cassette for use on the iC-System | Icubate, Inc. | Leftover de-identified clinical specimens from positive blood cultures | Clinical specimens were used to evaluate the performance of the iC-GN Assay compared to standard-of-care reference methods (culture/MALDI and PCR/sequencing) across five clinical sites. | Method comparison study; Clinical specimens; Positive blood cultures |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| iC-GN Assay Method Comparison Study; Prospective, multi-site, observational method comparison study | Patients with positive blood cultures confirmed by Gram stain to contain Gram-negative bacilli; Sample Size: 1002 evaluable specimens (976 fresh, 26 frozen); Number of Sites: 5 | Reference culture followed by MALDI identification and PCR/bi-directional sequencing | Percent agreement (positive/negative) for bacterial species and resistance markers |

## Indications for Use

The iCubate, Inc. iC-GN Assay™ for use on the iC-System™ is a qualitative, multiplexed, in vitro diagnostic test for the detection and identification of potentially pathogenic gram negative bacteria, which may cause bloodstream infection (BSI). The iC-GN Assay™ is performed directly on positive blood cultures, confirmed by Gram stain gram negative bacilli. Cultures demonstrating mixed Gram stain results should not be tested on the assay. The iC-GN Assay™ is validated for use with select BACTEC™, BacT/ALERT® and VersaTREK® blood culture bottles. The iC-GN Assay™ is indicated for use in conjunction with other clinical and laboratory findings, such as culture, to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor bloodstream infections.

## Device Story

The iC-GN Assay is a qualitative, multiplexed, in vitro diagnostic test for the detection of gram-negative bacteria and resistance markers directly from positive blood cultures. The device uses ARM-PCR (Amplicon Rescued Multiplex PCR) technology for amplification and microarray hybridization for detection. The process is performed in a self-contained, disposable iC-Cassette on the iC-Processor, which automates extraction, amplification, and detection. The iC-Reader then reads the microarray, and iC-Report software interprets the results for display on an iMac. The device is used in clinical laboratories by trained personnel. It aids in the diagnosis of bloodstream infections by identifying specific pathogens and resistance markers (KPC, NDM, CTX-M group 1), allowing for faster clinical decision-making and targeted therapy compared to traditional culture methods. Sub-culturing remains necessary for susceptibility testing and differentiation of mixed growth.

## Clinical Evidence

Clinical study evaluated 1002 prospective specimens (fresh and frozen) and 170 contrived samples across five sites. Compared to reference culture/MALDI and PCR/sequencing. Performance for bacterial targets showed high agreement (e.g., E. coli 98.4% positive agreement, K. pneumoniae 96.8%). Resistance marker detection (CTX-M, KPC, NDM) showed high concordance with sequencing. Study included analysis of mixed cultures and competitive inhibition. Error rate reduced from 2.9% to 0.8% with repeat testing.

## Technological Characteristics

Uses ARM-PCR technology for multiplex amplification and microarray hybridization for detection. Self-contained, disposable iC-Cassette. Automated processing via iC-Processor and iC-Reader. Connectivity: Wired data transmission to iMac computer. Software: iC-Report for result interpretation. Validated for use with various BACTEC, BacT/ALERT, and VersaTREK blood culture bottles.

## Regulatory Identification

A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.

## Special Controls

In combination with the general controls of the FD&C Act, the Verigene® Gram Positive Blood Culture Nucleic Acid Test is subject to the following special controls: The special controls for the BC-GP Assay are contained in the guideline document entitled "Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures."

*Classification.* Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).

## Predicate Devices

- Verigene® Gram Negative Blood Culture Nucleic Acid Test (GC-GN) ([K132843](/device/K132843.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

## I. Background Information:

A. 510(k) Number
K190341

B. Applicant
iCubate, Inc.

C. Proprietary and Established Names
iC-GN iC-Cassette for use on the iC-System
iC-GN Assay

D. Regulatory Information
|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  PEN | Class II | 21 CFR 866.3365 - Multiplex Nucleic Acid Assay For Identification Of Microorganisms And Resistance Markers From Positive Blood Cultures | MI - Microbiology  |

## II. Submission/Device Overview:

A. Purpose for Submission:
The purpose of the submission is to obtain clearance for the iC-GN Assay for use on the iC-System.

B. Measurand:
Bacterial Species: Acinetobacter baumannii complex, Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Pseudomonas aeruginosa, Proteus species, and Serratia marcescens

K190341 - Page 1 of 36

{1}

K190341 - Page 2 of 36

# Resistance Markers

KPC (blaKPC)- associated with resistance to carbapenems

NDM (blaNDM)- associated with resistance to carbapenems

CTX-M group 1 (blaCTX-M group 1)- associated with resistance to extended spectrum beta-lactams

## C. Type of Test:

The iC-GN Assay utilizes polymerase chain reaction (PCR) for the multiplex amplification of specific targets and detects the amplified targets with microarray hybridization. Targets are detected directly from patient positive blood cultures confirmed by Gram stain to contain Gram-negative bacilli. The iC-GN Assay utilizes proprietary ARM-PCR (Amplicon Rescued Multiplex PCR) technology allowing for multiple targets to be amplified in one reaction. Testing is done in a self-contained, automated, disposable cassette using the iCubate processor (iC-Processor). After the reaction is complete, the cassette is read on the iCubate reader (iC-Reader). Results from the iC-Reader are interpreted by iC-Report software and a final report is displayed on the iMac computer. Results are qualitative.

## II. Intended Use/Indications for Use:

### A. Intended Use(s):

See Indications for Use below.

### B. Indication(s) for Use:

The iCubate, Inc. iC-GN Assay for use on the iC-System is a qualitative, multiplexed, in vitro diagnostic test for the detection and identification of potentially pathogenic Gram-negative bacteria which may cause bloodstream infection (BSI). The iC-GN Assay is performed directly on positive blood cultures, confirmed by Gram stain to contain Gram-negative bacilli. Cultures demonstrating mixed Gram stain results should not be tested on the assay. The iC-GN Assay is validated for use with select BACTEC, BacT/ALERT and VersaTREK blood culture bottles. The iC-GN Assay is indicated for use in conjunction with other clinical and laboratory findings, such as culture, to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor bloodstream infections.

The iC-GN Assay detects target DNA and identifies the following:

**Bacterial Genera and Species**

*Acinetobacter baumannii* complex, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Pseudomonas aeruginosa*, *Proteus* species, *Serratia marcescens*

**Resistance Markers**

KPC (blaKPC)- associated with resistance to carbapenems, NDM (blaNDM)- associated with resistance to carbapenems, CTX-M group 1 (blaCTX-M group 1)- associated with resistance to extended spectrum beta-lactams

In mixed growth, the iC-GN Assay does not specifically attribute detection of KPC, NDM, or CTX-M group 1 to a specific genera or species.

{2}

Sub-culturing of positive blood cultures is necessary to recover organisms for susceptibility testing, identification of organisms not detected by the iC-GN Assay, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.

## C. Special Conditions for Use Statement(s):
Rx - For Prescription Use Only

## D. Special Instrument Requirements:
Requires the iC-System

## III. Device/System Characteristics:

### A. Device Description:

The iC-GN Assay utilizes polymerase chain reaction (PCR) for the multiplex amplification of specific targets and detects the amplified targets with microarray hybridization. Targets are detected directly from patient positive blood cultures confirmed by Gram stain to contain Gram-negative bacilli. The iC-GN Assay utilizes proprietary ARM-PCR (Amplicon Rescued Multiplex PCR) technology allowing for multiple targets to be amplified in one reaction. Testing is done in a self-contained, automated, disposable cassette using the iCubate processor (iC-Processor). After the reaction is complete, the cassette is read on the iCubate reader (iC-Reader). Results from the iC-Reader are interpreted by iC-Report software and a final report is displayed on the iMac computer.

To operate, the user opens the iC-Cassette cap and pipettes an aliquot of the diluted positive blood culture sample into the sample/PCR well in the bottom well plate of the cassette. Once inoculated, the cassette cap is closed, and all extraction, amplification and detection processes are completed in the cassette, a closed system. Extraction, amplification and detection sequences are defined by an assay script controlled by the iC-Processor.

The processing script is defined within a barcode label positioned on the top of each iC-Cassette which communicates with the iC-Processor. To access and pierce the foil-sealed reagent wells located in the bottom well plate of the cassette, the processor manipulates the cassette to move the cassette pipette horizontally and vertically. The script directs the transfer of reagents between the wells in the bottom well plate and finally to the array within the cassette. The iC-Processor is capable of processing four (4) iC-Cassettes with random access.

Once processing is complete, the cassette is manually transferred from the iC-Processor to the iC-Reader where the microarray within the cassette is read. The iC-Reader is capable of reading up to four (4) iC-Cassettes at one time. The results are interpreted via the iC-Report software and displayed for the user on the iMac. Raw data and result interpretations are stored within the iMac; raw data is accessible to iCubate service personnel only and not to the end user.

When finished with a loaded iC-GN Cassette, it should be disposed as biohazardous waste.

K190341 - Page 3 of 36

{3}

K190341 - Page 4 of 36

B. Principle of Operation:

Amplicon Rescued Multiplex PCR (ARM-PCR)

C. Instrument Description Information:

|  Modes of Operation | Yes | No  |
| --- | --- | --- |
|  Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐  |
|  Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑  |
|  Software  |   |   |
|  FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐  |

1. Instrument Name:

iC-System which was cleared under K163390.

IV. Substantial Equivalence Information:

A. Predicate Device Name(s):

Verigene Gram-Negative Blood Culture Nucleic Acid Test (bc-gn)

B. Predicate 510(k) Number(s):

K132843

C. Comparison with Predicate(s):

|  Device & Predicate Device(s): | K190341 | K132843  |
| --- | --- | --- |
|  Device Trade Name | iC-GN Assay for use on the iC-System | Verigene Gram-Negative Blood Culture Nucleic Acid Test (CG-GN)  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/ Indications for Use | The iCubate, Inc. iC-GN Assay for use on the iC-System is a qualitative, multiplexed, in vitro diagnostic test for the detection and identification of potentially pathogenic Gram-negative bacteria, which may cause bloodstream infection (BSI). The iC-GN Assay is performed directly on positive blood cultures, confirmed by Gram stain to contain Gram-negative bacilli. Cultures demonstrating mixed Gram stain results should not be tested on the assay. The iC-GN Assay is validated for use with select | The Verigene Gram-Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-results Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous  |

{4}

K190341 - Page 5 of 36

|   | BACTEC, BacT/ALERT and VersaTREK blood culture bottles. The iC-GN Assay is indicated for use in conjunction with other clinical and laboratory findings, such as culture, to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor bloodstream infections.

The iC-GN Assay detects target DNA and identifies the following:

**Bacterial Genera and Species**
Acinetobacter baumannii complex, Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Pseudomonas aeruginosa, Proteus species, Serratia marcescens

**Resistance Markers**
KPC (blaKPC)- associated with resistance to carbapenems, NDM (blaNDM)- associated with resistance to carbapenems, CTX-M group 1(blaCTX-M group 1)- associated with resistance to extended spectrum beta-lactams

In mixed growth, the iC-GN Assay does not specifically attribute KPC, NDM, or CTX-M to a specific genera or species.

iC-GN is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not to be used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for susceptibility testing, identification of organisms not detected by iC-GN, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing. | monitor blood culture system and which contain gram-negative bacteria as determined by Gram stain. BC-GN detects and identifies the following:
Acinetobacter spp.
Citrobacter spp.
Enterobacter spp.
Proteus spp.
Escherichia coli
Klebsiella pneumoniae
Klebsiella oxytoca
Pseudomonas aeruginosa

BC-GN is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, it is not to be used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for susceptibility testing, identification of organisms not detected by BC-GN, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, of for epidemiological typing.  |
| --- | --- | --- |
|  Sample Type | Positive Blood Culture | Positive Blood Culture  |
|  General Device Characteristic Differences |  |   |
|  Instrument Requirements | iC-System | Verigene System  |
|  Test Principle | ARM-PCR | Gold nanoparticle probe-based PCR  |
|  Compatible Blood Culture Bottles | BD BACTEC Standard/10 Aerobic/F
BD BACTEC Standard/10 Anaerobic/F
BD BACTEC Plus Aerobic/F
BD BACTEC Plus Anaerobic/F | BACTEC™ Plus Aerobic/F
BacT/ALERT FA FAN  |

{5}

|   | BD BACTEC Lytic/10 Anaerobic/F
BacT/Alert SA Standard Aerobic
BacT/Alert SN Standard Anaerobic
BacT/Alert FA Aerobic FAN
BacT/Alert FN Anaerobic FAN
BacT/Alert FA Plus Aerobic
BacT/Alert FN Plus Anaerobic
VersaTREK REDOX 1
VersaTREK REDOX 2 |   |
| --- | --- | --- |
|  Throughput | Four (4) samples/iC-Processor | One (1) Sample/Processor  |

## V. Standards/Guidance Documents Referenced:

1. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
2. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
3. Validation of Analytical Procedures: Test and Methodology Q2(R1), ICH Harmonised Tripartite Guideline, Oct. 1994
4. Q2B Validation of Analytical Procedures: Methodology, Guidance for Industry, Nov. 1996

## VI. Performance Characteristics (if/when applicable):

### A. Analytical Performance:

1. **Limit of Detection (LoD):**

A study was performed to determine the limit of detection for each iC-GN Assay target, defined as the lowest concentration (CFU/mL) of analyte that can be detected approximately 95% of the time. For the eleven targets detected by the iC-GN Assay, a panel of twenty-seven representative strains were evaluated, a minimum of three per target. For complex and genus level targets, at least two representative species were evaluated. LoD testing was conducted in two phases: the first phase was to narrow the range for LoD analysis. In phase II, the approximated 95% performance point determined in phase I was confirmed by testing a minimum of twenty replicates on each of three unique cassette lots. Plating and subsequent colony counts were used to determine organism concentrations. The final limit of detection for each target, is provided below.

K190341 - Page 6 of 36

{6}

K190341 - Page 7 of 36

|  Table 1: iC-GN Assay LoD Results  |   |   |   |
| --- | --- | --- | --- |
|  Target | Strain | Concentration (CFU/mL) | Defined Target LoD (CFU/mL)  |
|  *A. baumannii* complex | 307-0294 | 5.3 × 10^{5} | 5.3 × 10^{5} – 5.2 × 10^{6}  |
|   |  CDC-83 | 5.2 × 10^{6}  |   |
|   |  ATCC 23055 | 9.0 × 10^{5}  |   |
|  *E. cloacae* complex | Z101 | 5.0 × 10^{6} | 4.9 × 10^{5} – 5.5 × 10^{6}  |
|   |  CDC-164 | 5.5 × 10^{6}  |   |
|   |  ATCC 700323 | 4.9 × 10^{5}  |   |
|  *E. coli* | ATCC 43895 | 7.7 × 10^{5} | 7.7 × 10^{5} – 8.4 × 10^{5}  |
|   |  ATCC BAA-2326 | 7.9 × 10^{5}  |   |
|   |  CDC-55 | 8.4 × 10^{5}  |   |
|  *K. oxytoca* | Z115 | 6.2 × 10^{5} | 5.4 × 10^{5} – 1.1 × 10^{6}  |
|   |  ATCC 13182 | 5.4 × 10^{5}  |   |
|   |  CDC-147 | 1.1 × 10^{6}  |   |
|  *K. pneumoniae* | ATCC 35657 | 1.9 × 10^{6} | 6.0 × 10^{5} – 4.2 × 10^{6}  |
|   |  CDC-40 | 3.6 × 10^{6}  |   |
|   |  CDC-42 | 1.9 × 10^{6}  |   |
|   |  KPC-2 | 4.2 × 10^{6}  |   |
|   |  LACNY 11 | 6.0 × 10^{5}  |   |
|  *Proteus* species | Z050 | 1.1 × 10^{6} | 6.9 × 10^{5} – 1.1 × 10^{6}  |
|   |  CDC-59 | 9.9 × 10^{5}  |   |
|   |  Z028 | 7.6 × 10^{5}  |   |
|   |  Z129 | 6.9 × 10^{5}  |   |
|  *P. aeruginosa* | Z139 | 1.2 × 10^{6} | 5.0 × 10^{5} – 1.2 × 10^{6}  |
|   |  CDC-231 | 5.0 × 10^{5}  |   |
|   |  CDC-250 | 6.9 × 10^{5}  |   |
|  *S. marcescens* | ATCC 43297 | 7.2 × 10^{5} | 6.4 × 10^{5} – 8.1 × 10^{5}  |
|   |  ATCC 21212 | 8.1 × 10^{5}  |   |
|   |  CDC-91 | 6.4 × 10^{5}  |   |
|  CTX-M group 1 | ATCC BAA-2326 (CTX-M-15) | 7.9 × 10^{5} | 7.9 × 10^{5} – 2.3 × 10^{6}  |
|   |  CDC-40 (CTX-M-15) | 2.3 × 10^{6}  |   |
|   |  CDC-42 (CTX-M-15) | 1.9 × 10^{6}  |   |
|  KPC | CDC-147 (KPC-3) | 2.3 × 10^{6} | 1.5 × 10^{5} – 4.2 × 10^{6}  |
|   |  KPC-2 | 4.2 × 10^{6}  |   |
|   |  CDC-231 (KPC-5) | 1.5 × 10^{5}  |   |
|  NDM | CDC-83 (NDM-1) | 5.2 × 10^{6} | 3.3 × 10^{5} – 5.2 × 10^{6}  |
|   |  CDC-55 (NDM-1) | 4.0 × 10^{6}  |   |
|   |  CDC-250 (NDM-1) | 3.3 × 10^{5}  |   |

{7}

# 2. Bottle Ring:

A study was performed to establish the levels of each iC-GN target organism at two clinically relevant concentrations: initial bottle positivity (bottle "ring") and eight hours beyond initial positivity. Twenty-seven representative organisms were evaluated, a minimum of three per iC-GN target. Organisms were grown in BD BACTEC Plus Aerobic blood culture bottles with human blood added and placed on the BD BACTEC System. Within two hours of initial bottle positivity, the bottles were removed for plating and subsequent colony counts to determine organism concentrations. The bottles were then returned to the incubator and approximately eight hours after initial bottle positivity, the bottles were again removed for plating and subsequent colony counts to determine organism concentrations. Three bottles were grown for each strain. The average concentrations at initial bottle positivity and eight hours beyond initial bottle positivity are provided in below. The concentrations at initial bottle positivity, representative of the lowest levels that may be observed in a clinical setting, are above the limits of detection determined for each strain. The results of the study are as expected and establish that the established LoD is such that the assay has is able to detect clinically relevant levels of bacteria.

|  Table 2: iC-GN Target Organism Concentrations at Bottle “Ring”  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain ID | Initial Bottle Positivity Average Concentration (CFU/mL) | Bottle Positivity + 8 Average Concentration (CFU/mL)  |
|  Acinetobacter baumannii | 307-0294 | 4.24 × 108 | 8.27 × 108  |
|  Acinetobacter baumannii | CDC-83 | 3.39 × 108 | 7.23 × 108  |
|  Acinetobacter calcoaceticus | ATCC 23055 | 6.78 × 107 | 2.93 × 108  |
|  Enterobacter cloacae | Z101 | 2.17 × 108 | 1.97 × 109  |
|  Enterobacter cloacae | CDC-164 | 5.62 × 108 | 2.31 × 109  |
|  Enterobacter hormaechei | ATCC 700323 | 4.36 × 108 | 2.75 × 109  |
|  Escherichia coli | ATCC 43895 | 1.50 × 108 | 9.48 × 108  |
|  Escherichia coli | ATCC BAA-2326 | 6.23 × 108 | 1.52 × 109  |
|  Escherichia coli | CDC-55 | 4.93 × 108 | 1.51 × 109  |
|  Klebsiella oxytoca | Z115 | 5.32 × 108 | 2.07 × 109  |
|  Klebsiella oxytoca | ATCC 13182 | 4.16 × 108 | 4.52 × 109  |
|  Klebsiella oxytoca | CDC-147 | 9.67 × 108 | 1.31 × 109  |
|  Klebsiella pneumoniae | ATCC 35657 | 9.78 × 108 | 1.08 × 109  |
|  Klebsiella pneumoniae | CDC-40 | 2.16 × 108 | 1.36 × 109  |
|  Klebsiella pneumoniae | CDC-42 | 2.55 × 108 | 1.10 × 109  |
|  Klebsiella pneumoniae | KPC-2 | 7.70 × 108 | 1.66 × 109  |
|  Klebsiella pneumoniae | LACNY 11 | 5.43 × 107 | 1.67 × 109  |
|  Proteus mirabilis | Z050 | 1.71 × 108 | 7.40 × 108  |
|  Proteus mirabilis | CDC-59 | 7.37 × 107 | 8.10 × 108  |
|  Proteus penneri | Z028 | 8.88 × 107 | 4.33 × 108  |
|  Proteus vulgaris | Z129 | 4.37 × 107 | 5.00 × 108  |
|  Pseudomonas aeruginosa | Z139 | 9.18 × 107 | 1.37 × 1010  |

K190341 - Page 8 of 36

{8}

K190341 - Page 9 of 36

|  Table 2: iC-GN Target Organism Concentrations at Bottle “Ring”  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain ID | Initial Bottle Positivity Average Concentration (CFU/mL) | Bottle Positivity + 8 Average Concentration (CFU/mL)  |
|  Pseudomonas aeruginosa | CDC-231 | 3.26 × 10^{8} | 7.98 × 10^{8}  |
|  Pseudomonas aeruginosa | CDC-250 | 1.64 × 10^{8} | 8.97 × 10^{8}  |
|  Serratia marcescens | ATCC 43297 | 8.55 × 10^{8} | 2.03 × 10^{9}  |
|  Serratia marcescens | ATCC 21212 | 1.07 × 10^{8} | 8.83 × 10^{8}  |
|  Serratia marcescens | CDC-91 | 7.28 × 10^{8} | 1.67 × 10^{9}  |

3. Blood Culture Bottle Equivalency:

Commonly used blood culture bottle (BCB) media types were evaluated to demonstrate that variability in BCB media composition does not interfere with iC-GN Assay performance. Twenty-seven (27) representative iC-GN target organisms plus one non-target organism were tested in thirteen (13) BCB media types. Target organisms were tested near LoD concentrations (2-3×LoD). Each strain was tested in triplicate in each BCB media type. Target performance is based on all expected targets detected and no false positive targets detected. Non-target performance is based on all expected negative results. In the event of a false negative result, the strain was retested in replicates of ten. In the event of a false positive result or other failure, the strain was retested in triplicate. The results of iC-GN BCB equivalency testing are summarized in the table below. Performance in all bottle types met the acceptance criteria of ≥ 95% performance; all bottle types are validated for use with the iC GN Assay.

|  Table 3: iC-GN Assay BCB Equivalency Results  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  BCB Media Type | Overall Performance (%) | False Negatives (%) | False Positives (%) | PC Check Failures (%) | System Failures (%)  |
|  BACTEC Standard Aerobic | 93/94 (98.9%) | 1/94 (1.1%) | 0/94 (0.0%) | 3/97 (3.1%) | 0/97 (0.0%)  |
|  BACTEC Standard Anaerobic | 85/86 (98.8%) | 0/86 (0.0%) | 1/86 (1.2%) | 0/87 (0.0%) | 1/87 (1.1%)  |
|  BACTEC Plus Aerobic | 93/94 (98.9%) | 1/94 (1.1%) | 0/94 (0.0%) | 2/97 (2.1%) | 1/97 (1.1%)  |
|  BACTEC Plus Anaerobic | 95/96 (98.6%) | 1/96 (1.0%) | 0/96 (0.0%) | 2/100 (2.0%) | 2/100 (2.0%)  |
|  BACTEC Lytic/10 Anaerobic | 81/81 (100.0%) | 0/81 (0.0%) | 0/81 (0.0%) | 0/81 (0.0%) | 0/81 (0.0%)  |
|  BACT/ALERT SA Standard Aerobic | 97/99 (98.0%) | 1/99 (1.0%) | 1/99 (1.0%) | 4/103 (3.9%) | 0/103 (0.0%)  |
|  BACT/ALERT SN Standard Anaerobic | 87/88 (98.9%) | 0/88 (0.0%) | 1/88 (1.1%) | 2/90 (2.2%) | 0/90 (0.0%)  |

{9}

K190341 - Page 10 of 36

|  Table 3: iC-GN Assay BCB Equivalency Results  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  BCB Media Type | Overall Performance (%) | False Negatives (%) | False Positives (%) | PC Check Failures (%) | System Failures (%)  |
|  BACT/ALERT FA Aerobic FAN | 94/96 (97.9%) | 0/96 (0.0%) | 2/96 (2.1%) | 1/97 (1.0%) | 0/97 (0.0%)  |
|  BACT/ALERT FN Anaerobic FAN | 92/94 (97.9%) | 0/94 (0.0%) | 2/94 (2.1%) | 2/97 (2.1%) | 1/97 (1.0%)  |
|  BACT/ALERT FA Plus Aerobic | 94/95 (98.9%) | 1/95 (1.1%) | 0/95 (0.0%) | 1/97 (1.1%) | 1/97 (1.1%)  |
|  BACT/ALERT FN Plus Anaerobic | 87/87 (100.0%) | 0/87 (0.0%) | 0/87 (0.0%) | 2/90 (2.2%) | 1/90 (1.1%)  |
|  VersaTREK REDOX 1 | 81/81 (100.0%) | 0/81 (0.0%) | 0/81 (0.0%) | 0/81 (0.0%) | 0/81 (0.0%)  |
|  VersaTREK REDOX 1 | 92/93 (98.9%) | 1/93 (1.1%) | 0/93 (0.0%) | 1/94 (1.1%) | 0/94 (0.0%)  |

## 4. Inclusivity:

To demonstrate the inclusivity of the iC-GN Assay, eighty-two (82) representative strains were evaluated, a minimum of ten strains for each target analyte. Strains were tested at the lowest level of bottle positivity, considered within two hours of bottle "ring."

Organisms were grown in BD BACTEC Plus Aerobic blood culture bottles with human blood added on the BD BACTEC System. Each strain was tested in triplicate.

Performance calculations are based on all expected targets detected and no false positive targets detected. In the event of a false negative result, the strain was retested in replicates of ten. In the event of a false positive result or other failure, the strain was retested in triplicate. Two strains were not detected by the iC-GN Assay: Acinetobacter

calcoaceticus ATCC 31926 was not detected as A. baumannii complex and Enterobacter kobei ATCC BAA-260 was not detected as E. cloacae complex. The results of iC-GN Inclusivity testing are summarized below. The results of the inclusivity testing support identification of the organisms and resistance markers claimed in the IFU, with the exceptions noted above. It was determined that the risks associated with failure to detect a small number of organisms included in the IFU could be mitigated via labeling. These exceptions are communicated to end users in the Limitations section of the device Package Insert.

|  Table 4: iC-GN Assay Inclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Targets | Performance  |
|  Enterobacter kobei | ATCC BAA-260 | ECX | 0/13³  |
|  Escherichia coli | ATCC 10536 | EC | 3/3  |
|  Escherichia coli | ATCC BAA-2469 | EC, NDM-1 | 3/3  |
|  Escherichia coli | NCTC 9001 | EC | 3/3  |
|  Escherichia coli | NCTC 10538 | EC | 5/5  |
|  Escherichia coli | NCTC 13476 | EC | 3/3  |

{10}

|  Table 4: iC-GN Assay Inclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Targets | Performance  |
|  Escherichia coli | CDC-48 | EC, CTX-M-15, NDM-1 | 3/3  |
|  Escherichia coli | CDC-61 | EC, KPC-3 | 3/3  |
|  Escherichia coli | CDC-104 | EC, KPC-4 | 7/8a  |
|  Escherichia coli | CDC-119 | EC, CTX-M-15, NDM-1 | 3/3  |
|  Escherichia coli | CDC-162 | EC, CTX-M-15, NDM-7 | 3/3  |
|  Klebsiella oxytoca | ATCC 8724 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 43086 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 43165 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 43863 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 49134 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 49334 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 51817 | KO | 3/3  |
|  Klebsiella oxytoca | ATCC 700324 | KO | 3/3  |
|  Klebsiella oxytoca | NCTC 11686 | KO | 3/3  |
|  Klebsiella oxytoca | CDC-71 | KO | 3/3  |
|  Klebsiella pneumoniae | ATCC-13882 | KPN | 3/3  |
|  Klebsiella pneumoniae | ATCC BAA-1705 | KPN, KPC-2 | 3/3  |
|  Klebsiella pneumoniae | NCTC 9633 | KPN | 3/3  |
|  Klebsiella pneumoniae | NCTC 13438 | KPN, KPC-3 | 3/3  |
|  Klebsiella pneumoniae | NCTC 13443 | KPN, CTX-M-15, NDM-1 | 3/3  |
|  Klebsiella pneumoniae | CDC-44 | KPN, CTX-M-15 | 3/3  |
|  Klebsiella pneumoniae | CDC-46 | KPN, CTX-M-15 | 5/5  |
|  Klebsiella pneumoniae | CDC-49 | KPN, CTX-M-15, NDM-1 | 3/3  |
|  Klebsiella pneumoniae | CDC-66 | KPN, CTX-M-15 | 3/3  |
|  Klebsiella pneumoniae subsp. ozaenae | ATCC 11296 | KPN | 3/3  |
|  Proteus mirabilis | ATCC 7002 | Proteus | 3/3  |
|  Proteus mirabilis | ATCC 21100 | Proteus | 3/3  |
|  Proteus mirabilis | ATCC 43071 | Proteus | 3/3  |
|  Proteus mirabilis | NCIMB 13283 | Proteus | 3/3  |
|  Proteus mirabilis | CDC-155 | Proteus, KPC-6 | 3/3  |
|  Proteus mirabilis | CDC-156 | Proteus, KPC-2 | 3/3  |
|  Proteus mirabilis | CDC-159 | Proteus, NDM-1 | 3/3  |
|  Proteus penneri | ATCC 33519 | Proteus | 3/3  |
|  Proteus vulgaris | ATCC 9484 | Proteus | 3/3  |
|  Proteus vulgaris | ATCC 29905 | Proteus | 3/3  |
|  Pseudomonas aeruginosa | ATCC 10145 | PA | 3/3  |
|  Pseudomonas aeruginosa | ATCC 19429 | PA | 3/3  |
|  Pseudomonas aeruginosa | ATCC BAA-1744 | PA | 3/3  |
|  Pseudomonas aeruginosa | CDC-54 | PA | 3/3  |

K190341 - Page 11 of 36

{11}

|  Table 4: iC-GN Assay Inclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Targets | Performance  |
|  Pseudomonas aeruginosa | CDC-64 | PA | 3/3  |
|  Pseudomonas aeruginosa | CDC-90 | PA, KPC-5 | 3/3  |
|  Pseudomonas aeruginosa | CDC-94 | PA | 3/3  |
|  Pseudomonas aeruginosa | CDC-105 | PA | 3/3  |
|  Pseudomonas aeruginosa | CDC-108 | PA | 3/3  |
|  Pseudomonas aeruginosa | CDC-246 | PA, NDM-1 | 5/5  |
|  Serratia marcescens | ATCC 8100 | SM | 3/3  |
|  Serratia marcescens | ATCC 13880 | SM | 3/3  |
|  Serratia marcescens | ATCC 14041 | SM | 3/3  |
|  Serratia marcescens | ATCC 14756 | SM | 3/3  |
|  Serratia marcescens | ATCC 29634 | SM | 3/3  |
|  Serratia marcescens | ATCC 29635 | SM | 3/3  |
|  Serratia marcescens | ATCC 43861 | SM | 3/3  |
|  Serratia marcescens | ATCC 43862 | SM | 3/3  |
|  Serratia marcescens | NCTC 9743 | SM | 3/3  |
|  Serratia marcescens | CDC-99 | SM | 3/3  |

1) 2/2 false negative ABX in initial testing. 7/9 false negative ABX in repeat testing. See limitation.
2) 1/3 false positive ABX in initial testing. 1/3 false positive ABX in repeat testing. Strain repeated in replicates of 10, 10/10 repeats passed.
3) 3/3 false negative ECX in initial testing. 10/10 false negative ECX in repeat testing. See limitation.
4) 1/3 processor error in initial testing. 1/3 false positive KPN in repeat testing. Strain repeated in triplicate, 3/3 repeats passed.

An in silico analysis was performed for each resistance marker. The predicted reactivity of each resistance marker detected by the iC-GN Assay is summarized in the tables below.

K190341 - Page 12 of 36

{12}

|  Table 5: Predicted (in silico) Reactivity for CTX-M group 1  |   |   |   |
| --- | --- | --- | --- |
|  Associated Target Organism | Variant Detected | Associated Target Organism | Variant Detected  |
|  Acinetobacter baumannii complex | CTX-M-3 | Klebsiella oxytoca | CTX-M-3  |
|   |  CTX-M-15 |   | CTX-M-15  |
|  Enterobacter cloacae complex | CTX-M-1 |   | CTX-M-35  |
|   |  CTX-M-3 |   | CTX-M-36  |
|   |  CTX-M-15 |   | CTX-M-162  |
|   |  CTX-M-22 | Klebsiella pneumoniae | CTX-M-1  |
|   |  CTX-M-37 |   | CTX-M-3  |
|   |  CTX-M-55 |   | CTX-M-15  |
|   |  CTX-M-167 |   | CTX-M-22  |
|   |  CTX-M-177 |   | CTX-M-28  |
|   |  CTX-M-187 |   | CTX-M-32  |
|   |  CTX-M-224 |   | CTX-M-54  |
|  Escherichia coli | CTX-M-1 |   | CTX-M-55  |
|   |  CTX-M-2 |   | CTX-M-71  |
|   |  CTX-M-3 |   | CTX-M-72  |
|   |  CTX-M-4 |   | CTX-M-118  |
|   |  CTX-M-5 |   | CTX-M-124  |
|   |  CTX-M-6 |   | CTX-M-129  |
|   |  CTX-M-7 |   | CTX-M-130  |
|   |  CTX-M-8 |  | CTX-M-133  |
|   |  CTX-M-9 |   | CTX-M-135  |
|   |  CTX-M-10 |   | CTX-M-138  |
|   |  CTX-M-11 |   | CTX-M-139  |
|   |  CTX-M-12 |   | CTX-M-173  |
|   |  CTX-M-15 |   | CTX-M-176  |
|   |  CTX-M-28 |   | CTX-M-183  |
|   |  CTX-M-29 |   | CTX-M-188  |
|   |  CTX-M-32 |   | CTX-M-197  |
|   |  CTX-M-33 |   | CTX-M-204  |
|   |  CTX-M-36 |   | CTX-M-208  |
|   |  CTX-M-42 |   | CTX-M-210  |
|   |  CTX-M-55 |   | CTX-M-220  |
|   |  CTX-M-58 | Proteus species | CTX-M-15  |
|   |  CTX-M-69 |   | CTX-M-66  |
|   |  CTX-M-71 |   | CTX-M-116  |
|   |  CTX-M-79 |   | CTX-M-136  |
|   |  CTX-M-82 |   | CTX-M-164  |
|   |  CTX-M-90 |   | CTX-M-167  |
|   |  CTX-M-101 |   | CTX-M-212  |
|   |  CTX-M-102 | Pseudomonas aeruginosa | CTX-M-1  |
|   |  CTX-M-103 |   | CTX-M-15  |
|   |  CTX-M-109 |   | CTX-M-32  |
|  CTX-M-117 |  | CTX-M-3  |   |
|  CTX-M-120 |   | CTX-M-15  |   |

K190341 - Page 13 of 36

{13}

|   | CTX-M-125 | Serratia marcescens | CTX-M-22  |
| --- | --- | --- | --- |
|   |  CTX-M-127 |   | CTX-M-55  |
|   |  CTX-M-128 |   | CTX-M-221  |
|   |  CTX-M-131 |   |   |
|   |  CTX-M-132 |   |   |
|   |  CTX-M-134 |   |   |
|   |  CTX-M-137 |   |   |
|   |  CTX-M-138 |   |   |
|   |  CTX-M-139 |   |   |
|   |  CTX-M-140 |   |   |
|   |  CTX-M-141 |   |   |
|   |  CTX-M-142 |   |   |
|   |  CTX-M-143 |   |   |
|   |  CTX-M-146 |   |   |
|   |  CTX-M-158 |   |   |
|   |  CTX-M-163 |   |   |
|   |  CTX-M-166 |   |   |
|   |  CTX-M-167 |   |   |
|   |  CTX-M-170 |   |   |
|   |  CTX-M-172 |   |   |
|   |  CTX-M-175 |   |   |
|   |  CTX-M-178 |   |   |
|   |  CTX-M-179 |   |   |
|   |  CTX-M-180 |   |   |
|   |  CTX-M-181 |   |   |
|  CTX-M-182 |   |   |   |
|  CTX-M-184 |   |   |   |
|  Table 5a: Predicted (in silico) Reactivity for KPC  |   |   |   |
| --- | --- | --- | --- |
|  Associated Target Organism | Variant Detected | Associated Target Organism | Variant Detected  |
|  Acinetobacter baumannii complex | KPC-2 | Klebsiella pneumoniae | KPC-1  |
|   |  KPC-3 |   | KPC-2  |
|   |  KPC-10 |   | KPC-3  |
|  Enterobacter cloacae complex | KPC-1 |   | KPC-4  |
|   |  KPC-2 |   | KPC-5  |
|   |  KPC-3 |   | KPC-6  |
|   |  KPC-4 |   | KPC-7  |
|   |  KPC-13 |   | KPC-8  |
|   |  KPC-18 |   | KPC-11  |
|   |  KPC-47 |   | KPC-14  |
|  Escherichia coli | KPC-2 |   | KPC-15  |
|   |  KPC-3 |   | KPC-16  |
|   |  KPC-12 |   | KPC-17  |
|   |  KPC-18 |   | KPC-19  |
|   |  KPC-20 |   | KPC-22  |
|   |  KPC-21 |   | KPC-23  |
|   |  KPC-28 |   | KPC-25  |
|  Klebsiella oxytoca | KPC-2 |   | KPC-26  |
|   |  KPC-3 | KPC-27  |   |

K190341 - Page 14 of 36

{14}

|  Table 5b: Predicted (in silico) Reactivity for NDM  |   |
| --- | --- |
|  Associated Target Organism | Variant Detected  |
|  Acinetobacter baumannii complex | NDM-1  |
|   |  NDM-2  |
|   |  NDM-3  |
|   |  NDM-4  |
|   |  NDM-5  |
|   |  NDM-7  |
|   |  NDM-14  |
|   |  NDM-22  |
|  Enterobacter cloacae complex | NDM-1  |
|   |  NDM-4  |
|   |  NDM-5  |
|   |  NDM-6  |
|   |  NDM-7  |
|   |  NDM-9  |
|   |  NDM-11  |
|   |  NDM-12  |
|  Escherichia coli | NDM-13  |
|   |  NDM-15  |
|   |  NDM-16  |
|   |  NDM-17  |
|   |  NDM-18  |
|  Escherichia coli | NDM-19  |
|   |  NDM-20  |
|   |  NDM-21  |
|   |  NDM-27  |

5. Exclusivity:

To demonstrate the exclusivity of the iC-GN Assay, a comprehensive panel of non-target organisms that may be encountered in positive blood cultures was evaluated. A total of 114 strains were tested including organisms phylogenetically related to iC-GN target organisms as well as common blood culture contaminants. Potential cross-reactivity was

K190341 - Page 15 of 36

{15}

evaluated by testing exclusivity panel organisms at the highest possible concentrations, considered eight hours beyond initial bottle positivity or the equivalent. Organisms were grown in BD BACTEC Plus Aerobic blood culture bottles with human blood added on the BD BACTEC System. Each strain was tested in triplicate. Performance is based on the observation of all expected negative results. In the event of a false positive result or other failure, the organism was retested in replicates of three (3) or ten (10). Three (3) strains demonstrated reproducible cross-reactivity with iC-GN Assay targets:

Acinetobacter haemolyticus cross-reacted with Acinetobacter baumannii complex, Klebsiella variicola cross-reacted with Klebsiella pneumoniae, and Serratia odorifera cross-reacted with Serratia marcescens. Exclusivity results are in the table below. Exclusivity testing confirms that the device detects organisms not listed in the IFU at low frequency. It was determined that the risks associated with the detection of organisms not listed in the IFU could be mitigated via labeling. The organisms not in the IFU that were detected are communicated to the end user in the limitations section of the Package Insert.

|  Table 6: iC-GN Assay Exclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Concentration (CFU/mL) | Performance  |
|  Acinetobacter haemolyticus | ATCC 19002 | 7.20 × 108 | 0/31  |
|  Acinetobacter lwoffii | Z141 | 2.45 × 108 | 3/3  |
|  Acinetobacter radioresistens | ATCC 43998 | 5.20 × 107 | 3/3  |
|  Acinetobacter schindleri | ATCC BAA618 | 3.50 × 108 | 3/3  |
|  Acinetobacter ursingii | ATCC BAA617 | 3.80 × 108 | 3/3  |
|  Aerococcus viridans | Z219 | 2.24 × 107 | 3/3  |
|  Aeromonas hydrophila | Z161 | 8.10 × 108 | 3/3  |
|  Alcaligenes faecalis | Z218 | 9.70 × 108 | 3/3  |
|  Aspergillus niger | Z105 | 1.62 × 108 | 3/3  |
|  Bacillus cereus | Z091 | ND | 3/3  |
|  Bacteroides fragilis | Z029 | 8.40 × 109 | 3/3  |
|  Brevundimonas vesicularis | ATCC 11426 | 3.80 × 108 | 5/5  |
|  Burkholderia cepacia | ATCC 25416 | 5.40 × 108 | 3/3  |
|  Campylobacter coli | Z293 | 3.90 × 108 | 3/3  |
|  Campylobacter jejuni | Z086 | 4.60 × 108 | 11/132  |
|  Candida albicans | Z006 | ND | 3/3  |
|  Candida glabrata | Z007 | 3.20 × 107 | 3/3  |
|  Candida krusei | Z009 | 1.90 × 107 | 3/3  |
|  Candida parapsilosis | Z011 | 9.00 × 106 | 3/3  |
|  Candida tropicalis | Z012 | 3.50 × 107 | 4/4  |
|  Cedecea davisae | ATCC 33431 | 6.20 × 108 | 3/3  |
|  Citrobacter amalonaticus | Z051 | 8.4 × 108 | 3/3  |
|  Citrobacter braakii | ATCC 51113 | 4.90 × 108 | 3/3  |
|  Citrobacter freundii | Z064 | 2.25 × 108 | 3/3  |
|  Citrobacter koseri | Z039 | 1.14 × 109 | 3/3  |

K190341 - Page 16 of 36

{16}

K190341 - Page 17 of 36

|  Table 6: iC-GN Assay Exclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Concentration (CFU/mL) | Performance  |
|  Citrobacter sedlakii | ATCC 51115 | 9.80 × 10^{8} | 2/2  |
|  Clostridium difficile (NAP-1 toxigenic) | NAP1 | 4.87 × 10^{7} | 4/4  |
|  Clostridium difficile (non-toxigenic) | Z228 | 5.93 × 10^{7} | 3/3  |
|  Clostridium novyi* | Z179 | 1.14 × 10^{7} | 5/5  |
|  Corynebacterium amycolatum | Z284 | 9.26 × 10^{8} | 3/3  |
|  Corynebacterium genitalium | Z328 | 1.35 × 10^{8} | 3/3  |
|  Corynebacterium jeikeium | Z232 | 8.50 × 10^{8} | 4/4  |
|  Corynebacterium striatum | MCW000 | 2.07 × 10^{9} | 5/6^{3}  |
|  Cronobacter muytjensii | ATCC 51329 | 2.79 × 10^{8} | 3/3  |
|  Cronobacter sakazakii | ATCC 29544 | 6.90 × 10^{8} | 3/3  |
|  Cryptococcus neoformans | Serotype A | 2.15 × 10^{8} | 3/3  |
|  Edwardsiella tarda | Z183 | 8.70 × 10^{7} | 4/5^{4}  |
|  Enterobacter aerogenes | Z052 | 1.77 × 10^{9} | 5/5  |
|  Enterobacter amnigenus | ATCC 51816 | 7.50 × 10^{8} | 3/3  |
|  Enterococcus avium | Z171 | 2.58 × 10^{8} | 5/6^{5}  |
|  Enterococcus casseliflavus | Z002 | 2.44 × 10^{9} | 4/4  |
|  Enterococcus cecorum | Z208 | 1.03 × 10^{9} | 5/6^{6}  |
|  Enterococcus faecalis | ATCC 51299 | 2.13 × 10^{9} | 3/3  |
|  Enterococcus faecium | ATCC 700221 | 7.20 × 10^{8} | 3/3  |
|  Enterococcus gallinarum | Z209 | 1.35 × 10^{9} | 3/3  |
|  Enterococcus hirae | Z193 | 2.37 × 10^{8} | 3/3  |
|  Enterococcus raffinosus | ATCC 49427 | 5.40 × 10^{8} | 3/3  |
|  Escherichia fergusonii | ATCC 35469 | 8.70 × 10^{8} | 3/3  |
|  Escherichia hermannii | Z184 | 1.01 × 10^{9} | 5/5  |
|  Escherichia vulneris | ATCC 33821 | 7.50 × 10^{8} | 3/3  |
|  Fusobacterium varium | Z361 | 2.49 × 10^{9} | 3/3  |
|  Hafnia alvei | ATCC 51815 | 1.37 × 10^{9} | 3/3  |
|  Haemophilus influenzae | ATCC 10211 | 3.09 × 10^{9} | 3/3  |
|  Haemophilus parainfluenzae | ATCC 9796 | 1.33 × 10^{8} | 3/3  |
|  Klebsiella variicola | ATCC 31488 | 4.40 × 10^{8} | 0/3^{7}  |
|  Kluyvera ascorbata (KPC+) | CDC-0144 | 1.40 × 10^{9} | 3/3  |
|  Kocuria kristinae | Z250 | 7.20 × 10^{7} | 3/3  |
|  Kytococcus schroeteri | ATCC BAA2410 | 1.50 × 10^{7} | 3/3  |
|  Lactobacillus acidophilus | Z048 | 6.00 × 10^{8} | 3/3  |
|  Lactobacillus plantarum | 17-5 | 5.30 × 10^{8} | 3/3  |
|  Lactobacillus reuteri | Z333 | 5.80 × 10^{7} | 5/5  |
|  Lactococcus lactis | Z169 | 9.30 × 10^{7} | 3/3  |
|  Leclercia adecarboxylata | ATCC 23216 | 1.01 × 10^{9} | 3/3  |
|  Leminorella grimontii | Z364 | 4.00 × 10^{9} | 3/3  |
|  Leuconostoc mesenteroides | Z197 | 4.00 × 10^{7} | 5/5  |

{17}

|  Table 6: iC-GN Assay Exclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Concentration (CFU/mL) | Performance  |
|  Listeria monocytogenes | ATCC 19115 | 2.03 × 109 | 3/3  |
|  Micrococcus luteus | Z100 | 1.80 × 108 | 3/3  |
|  Moraxella catarrhalis | ATCC 25238 | 1.27 × 109 | 3/3  |
|  Morganella morganii | ATCC 25830 | 1.23 × 109 | 3/3  |
|  Neisseria gonorrhoeae | ATCC 19424 | ND | 3/3  |
|  Neisseria lactamica | ATCC 23970 | 2.90 × 108 | 3/3  |
|  Neisseria meningitidis | Serotype A | 2.55 × 108 | 5/5  |
|  Neisseria mucosa | ATCC 49233 | 5.80 × 108 | 4/4  |
|  Neisseria sicca | ATCC 9913 | 1.43 × 108 | 3/3  |
|  Pantaea agglomerans | ATCC 27155 | 2.00 × 106 | 3/3  |
|  Pasturella multocida | ATCC 12945 | 2.84 × 109 | 2/2  |
|  Pediococcus pentosaceus | Z226 | 1.91 × 108 | 3/3  |
|  Planococcus citreus | ATCC 14404 | 1.95 × 108 | 3/3  |
|  Pluralibacter gergoviae | ATCC 33028 | 1.27 × 109 | 3/3  |
|  Propionibacterium acnes | Z144 | 7.90 × 108 | 5/5  |
|  Providencia alcalifaciens | Z292 | 3.10 × 109 | 3/3  |
|  Providencia rettgeri | Z370 | 2.20 × 109 | 3/3  |
|  Providencia stuartii | Z213 | 1.70 × 109 | 3/3  |
|  Pseudomonas fluorescens | ATCC 13525 | 2.43 × 108 | 3/3  |
|  Pseudomonas luteola | ATCC 43273 | 1.09 × 108 | 3/3  |
|  Pseudomonas mendocina | ATCC 25411 | 1.23 × 109 | 3/3  |
|  Pseudomonas nitroreducens | ATCC 33634 | 5.30 × 108 | 3/3  |
|  Pseudomonas oryzihabitans | ATCC 43272 | 1.70 × 107 | 4/58  |
|  Pseudomonas putida | Z030 | 3.30 × 108 | 3/3  |
|  Pseudomonas stutzeri | ATCC 17588 | 6.20 × 108 | 3/3  |
|  Raoultella planitcola | ATCC 33558 | 1.25 × 109 | 3/3  |
|  Rothia mucilaginosus | Z033 | 5.50 × 107 | 3/3  |
|  Salmonella enterica | ATCC BAA1715 | 2.23 × 109 | 3/3  |
|  Serratia fonticola | ATCC 29844 | 1.18 × 109 | 3/3  |
|  Serratia liquefaciens | ATCC 27592 | 1.24 × 109 | 11/129  |
|  Serratia odorifera | ATCC 33077 | 2.19 × 109 | 11/1310  |
|  Serratia rubidaea | ATCC 19278 | 1.54 × 108 | 5/5  |
|  Staphylococcus aureus | ATCC 700699 | 4.30 × 107 | 3/3  |
|  Staphylococcus capitis | Z192 | 2.13 × 108 | 3/3  |
|  Staphylococcus epidermidis | ATCC 700566 | 5.90 × 107 | 3/3  |
|  Staphylococcus haemolyticus | Z067 | 2.70 × 107 | 5/5  |
|  Staphylococcus hominis | Z031 | 9.90 × 107 | 3/3  |
|  Staphylococcus intermedius | Z112 | 3.30 × 107 | 3/3  |
|  Staphylococcus lugdunensis | Z097 | 2.43 × 108 | 3/3  |
|  Staphylococcus schleiferi | Z294 | 2.52 × 109 | 3/3  |
|  Stenotrophomonas maltophilia | ATCC BAA84 | 1.74 × 109 | 5/5  |

K190341 - Page 18 of 36

{18}

|  Table 6: iC-GN Assay Exclusivity Results  |   |   |   |
| --- | --- | --- | --- |
|  Organism | Strain | Concentration (CFU/mL) | Performance  |
|  Streptococcus agalactiae | Z019 | 5.80 × 108 | 3/3  |
|  Streptococcus anginosus | Z179 | 9.50 × 108 | 3/3  |
|  Streptococcus bovis | Z167 | 8.00 × 108 | 3/3  |
|  Streptococcus dysgalactiae | Z068 | 2.65 × 108 | 3/3  |
|  Streptococcus intermedius | Z126 | 1.40 × 107 | 5/611  |
|  Streptococcus pneumoniae | ATCC 6301 | 3.80 × 108 | 5/612  |
|  Streptococcus pyogenes | Z018 | 4.80 × 107 | 3/3  |
|  Veillonella parvula | Z379 | 6.70 × 109 | 5/613  |

1) 3/3 false positive A. baumannii complex in initial testing. See limitation.
2) 2/3 false positive E. coli in initial testing. 10/10 repeats negative.
3) 1/3 false positive S. marcescens in initial testing. 3/3 repeats negative.
4) 1/3 positive control check failure in initial testing. 1/3 false positive S. marcescens in repeat testing.
5) 1/3 false positive S. marcescens in initial testing. 3/3 repeats negative.
6) 1/3 false positive E. coli in initial testing. 3/3 repeats negative.
7) 3/3 false positive K. pneumoniae in initial testing. See limitation.
8) 1/3 false positive S. marcescens in initial testing. 2/2 repeats negative.
9) 1/3 false positive S. marcescens in initial testing. 9/9 repeats negative.
10) 1/3 false positive S. marcescens in initial testing. 1/10 false positive S. marcescens in repeat testing. See limitation.
11) 1/3 false positive S. marcescens in initial testing. 3/3 repeats negative.
12) 1/3 false positive S. marcescens in initial testing. 3/3 repeats negative.
13) 1/3 false positive E. coli in initial testing. 3/3 repeats negative.

6. Interfering Substances:

iC-GN Assay performance was evaluated in the presence of potentially inhibiting substances that may be encountered in blood and blood culture media. Eight representative target organisms plus one non-target organism were evaluated. Organisms were tested at the lowest levels of bottle positivity, considered within two hours of bottle "ring." Potential interferents were tested at concentrations exceeding the highest concentrations that may be encountered in blood and blood culture media (for concentrations tested see 510(k) Summary for K190341, Table 13 (p.24-25)). Target performance is based on all expected targets detected and no false positive targets detected. Non-target performance is based on all negative results. In the event of a false negative result, the organism/interferent combination was retested in replicates of ten (10). In the event of a false positive result or other failure, the organism/interferent combination was retested in triplicate. If the discordant result was observed in repeat testing, the combination was retested at a decreased inhibitor concentration. below. Interference testing was performed in BD BACTEC Plus Aerobic blood culture bottle media, which has a sodium polyanetholesulfonate (SPS) concentration of  $0.05\%$  w/v. Additional SPS at a concentration greater than  $0.05\%$  w/v was found to interfere with the performance of some iC-GN Assay targets, resulting in increased false negative results and positive control check failures. Interference results are included in the table below. It was determined that the risks associated with the interfering substance, SPS, could be mitigated via labeling. The information regarding interference associated with SPS is provided for the end user in the Package Insert.

K190341 - Page 19 of 36

{19}

|  Table 7: Interfering Substances Test Panel  |   |   |
| --- | --- | --- |
|  Interference Compound | linically Relevant Concentration | Test Concentration  |
|  Hemoglobin | 1-2 g/L | 10 g/L  |
|  Conjugated Bilirubin | 0.1-0.4 mg/dL | 10 mg/dL  |
|  Unconjugated Bilirubin | 0.1-0.8 mg/dL | 10 mg/dL  |
|  Protein (γ-globulin + albumin) | 0.7-1.7 g/dL | 4 g/dL  |
|  Triglyceride | 300-500 mg/dL | 1500 mg/dL  |
|  Human Genomic DNA | NA | 1 × 10^6 cells/mL  |
|  Sodium Polyanetholesulfonate (SPS) | 0.02-0.05% w/v | 0.1% w/v  |
|  Cefepime | 16 μg/mL | 80 μg/mL  |
|  Ceftriaxone | 16 μg/mL | 80 μg/mL  |
|  Fluconazole | 25 μg/mL | 100 μg/mL  |
|  Gentamicin | 20 μg/mL | 80 μg/mL  |
|  Meropenem | 16 μg/mL | 80 μg/mL  |
|  Piperacillin | 32 μg/mL | 160 μg/mL  |
|  Vancomycin | 20 μg/mL | 100 μg/mL  |
|  Table 7a: iC-GN Assay Interfering Substances Performance  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | Target Performance  |   |   |   |   |   |   |   |   |   |   |   |
|  Interference Compound | ABX | ECX | EC | KO | KPN | PM | PA | SM | SE | KPC -2 | NDM-1 | CTX-M -15  |
|  Hemoglobin | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 14/14 | 3/3 | 5/5 | 3/3 | 14/14 | 3/3  |
|  Conjugated Bilirubin | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Unconjugated Bilirubin | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Protein (γ-globulin + albumin) | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Triglyceride | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Human Genomic DNA | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  SPS (0.1%) | 3/12^1 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 2/12^2 | 3/3 | 4/4 | 3/3 | 8/12^3 | 3/3  |

K190341 - Page 20 of 36

{20}

K190341 - Page 21 of 36

|  Table 7a: iC-GN Assay Interfering Substances Performance  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | Target Performance  |   |   |   |   |   |   |   |   |   |   |   |
|  Interference Compound | ABX | ECX | EC | KO | KPN | PM | PA | SM | SE | KPC -2 | NDM-1 | CTX-M -15  |
|  SPS (0.5%) | 3/3 | -- | -- | -- | -- | -- | 3/3 | -- | -- | -- | 3/3 | --  |
|  Cefepime | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Ceftriaxone | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Fluconazole | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Gentamicin | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  Meropenem | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 5/6^{4} | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5  |
|  Piperacillin | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |

7. Microbial Interference:

Potential microbial interference was evaluated by testing high concentrations of Gram-negative exclusivity organisms in combination with low concentrations of iC-GN target organisms. A total of sixty (60) Gram-negative exclusivity strains were tested at the highest possible concentrations, considered eight hours beyond initial bottle positivity or the equivalent. Eight (8) representative iC-GN target organisms were tested at concentrations below the lowest levels of bottle positivity. Each organism combination was tested in triplicate. Performance was based on all expected targets detected and no false positive targets detected. In the event of a false negative result, the combination was retested in replicates of ten (10). In the event of a false positive result or other failure, the combination was retested in replicates of three (3) or ten (10). Microbial interference results are presented in the table below. A small number of organisms initially were observed to cause false negative results, however additional testing showed that the initial results were not reproducible. It was determined that the risks associated with the microbial interference could be mitigated via labeling. The organisms that showed an increased risk of microbial interference are listed in the Limitations section of the Package Insert.

|  Table 8: iC-GN Assay Microbial Interference Results  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Organism | ABX | ECX | EC | KO | KPN | PM | PA | SM | CTX-M-15 | KPC-2 | NDM-1 |   |
|  A. lwoffii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  A. radioresistens | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  A. schindleri | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  A. ursingii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  A. hydrophila | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  A. faecalis | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  B. fragilis | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 4/5^{1} | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  B. vesicularis | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  B. cepacia | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |
|  C. coli | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |   |

{21}

|  Table 8: iC-GN Assay Microbial Interference Results  |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Organism | ABX | ECX | EC | KO | KPN | PM | PA | SM | CTX-M-15 | KPC-2 | NDM-1  |
|  C. jejuni | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. davisae | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. amalonaticus | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. braakii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. freundii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 2/2 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. koseri | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. sedlakii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. muytjensii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  C. sakazakii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  E. tarda | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 5/5  |
|  E. aerogenes | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  E. amnigenus | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3  |
|  E. fergusonii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  E. hermannii | 3/3 | 3/3 | 5/5 | 3/3 | 13/13 | 3/3 | 3/3 | 3/3 | 5/5 | 12/132 | 3/3  |
|  E. vulneris | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  F. varium | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  H. alvei | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  H. influenzae | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  H. parainfluenzae | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  K. ascorbata | 10/133 | 3/3 | 11/124 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 10/124 | 3/3 | 3/3  |
|  L. adecarboxylata | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  L. grimontii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  M. catarrhalis | 12/135 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  M. morganii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 5/5  |
|  N. gonorrhoeae | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  N. lactamica | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 5/5  |
|  N. meningitidis | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3  |
|  N. mucosa | 5/5 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3  |
|  N. sicca | 5/66 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. agglomerans | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. multocida | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. gergoviae | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. alcalifaciens | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. rettgeri | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. stuartii | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. fluorescens | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. luteola | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. mendocina | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3  |
|  P. nitroreducens | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  P. oryzihabitans | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3  |
|  P. putida | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 5/5  |

K190341 - Page 22 of 36

{22}

K190341 - Page 23 of 36

|  Table 8: iC-GN Assay Microbial Interference Results  |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Organism | ABX | ECX | EC | KO | KPN | PM | PA | SM | CTX-M-15 | KPC-2 | NDM-1  |
|  P. stutzeri | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 3/3 | 5/5 | 3/3 | 3/3 | 3/3 | 5/5  |
|  R. planitcola | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  S. enterica | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  S. fonticola | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  S. liquefaciens | 3/3 | 5/5 | 3/3 | 3/3 | 12/13^{7} | 3/3 | 3/3 | 3/3 | 3/3 | 13/13 | 3/3  |
|  S. odorifera | 3/3 | 3/3 | 3/3 | 12/13^{8} | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  S. rubidaea | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  S. maltophilia | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |
|  V. parvula | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3  |

1) 1/3 array registration error in initial testing. 1/3 false positive E. cloacae complex in repeat testing.
2) 1/3 false negative KPC in initial testing. 10/10 repeats passed.
3) 3/3 false negative A. baumannii complex in initial testing; concentration was determined to be below the target limit of detection. 10/10 repeats passed.
4) 1/2 false negative CTX-M in initial testing. 1/10 false positive K. pneumoniae in repeat testing.
5) 1/3 false negative A. baumannii complex in initial testing. 10/10 repeats passed.
6) 1/3 false positive E. coli in initial testing. 3/3 repeats passed.
7) 1/3 false negative K. pneumoniae in initial testing. 10/10 repeats passed.
8) 1/3 false positive S. marcescens in initial testing. 10/10 repeats passed.

## 8. Competitive Inhibition:

iC-GN Assay performance was evaluated with combinations of target analytes that may be found in mixed positive blood cultures. One target organism was prepared at the lowest level of bottle positivity, considered within two hours of bottle "ring", while the second target organism was prepared at the highest possible concentration, considered eight hours after initial bottle positivity. All organisms were grown in BD BACTEC Plus Aerobic blood cultures bottles with human blood added on the BD BACTEC System. The organisms were combined at a ratio of one part "low" to four parts "high". Each low concentration organism was tested in combination with each high concentration organism in triplicate. Performance was based on all expected targets detected. In the event of a false negative result, the organism combination was retested in replicates of ten (10) at the same "low" and "high" organism ratio. In the event of a reproducible false negative result, the organism combination was retested in replicates of ten (10) at a ratio of one part "low" to one part "high." All high concentration iC-GN targets were detected. Due to competitive inhibition, low concentration targets were not detected in 1.7% of tests (3/178). When iC-GN target organisms were present at similar concentrations, all targets were detected. Competitive inhibition results are included in the table below. Based on the results of the competitive inhibition study, the following limitation was included in the Package Insert, "Due to competitive inhibition, target organisms present at low concentrations may not be detected by the iC-GN Assay when a second target organism is present at higher concentrations in a mixed culture".

|  Table 9: iC-GN Assay Competitive Inhibition Performance  |   |
| --- | --- |
|   | Target Performance  |

{23}

|  Low Organism | High Organism | Low Organism | Low Marker | High Organism | High Marker  |
| --- | --- | --- | --- | --- | --- |
|  A. baumannii | ECX | 3/3 | NA | 3/3 | NA  |
|   |  EC (CTX-M-15+) | 3/3 | NA | 3/3 | 3/3  |
|   |  KO | 3/3 | NA | 3/3 | NA  |
|   |  KPN (KPC-2+) | 3/3 | NA | 3/3 | 3/3  |
|   |  PM | 3/3 | NA | 3/3 | NA  |
|   |  PA (NDM-1+) | 3/3 | NA | 3/3 | 3/3  |
|   |  SM | 3/3 | NA | 3/3 | NA  |
|  E. cloacae | ABX | 3/3 | NA | 3/3 | 1 FP KPC  |
|   |  EC (CTX-M-15+) | 3/3 | NA | 3/3 | 3/3  |
|   |  KO | 3/3 | NA | 3/3 | NA  |
|   |  KPN (KPC-2+) | 3/3 | NA | 3/3 | 3/3  |
|   |  PM | 3/3 | NA | 3/3 | NA  |
|   |  PA (NDM-1+) | 3/3 | NA | 3/3 | 3/3  |
|   |  SM | 3/3 | NA | 3/3 | NA  |
|  E. coli (CTX-M-15+) | ABX | 3/3 | 3/3 | 3/3 | NA  |
|   |  ECX | 3/3 | 3/3 | 3/3 | NA  |
|   |  KO | 3/3 | 3/3 | 3/3 | NA  |
|   |  KPN (KPC-2+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  PM | 3/3 | 3/3 | 3/3 | NA  |
|   |  PA (NDM-1+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  SM | 3/3 | 3/3 | 3/3 | NA  |
|  K. oxytoca | ABX | 3/3 | NA | 3/3 | NA  |
|   |  ECX | 3/3 | NA | 3/3 | NA  |
|   |  EC (CTX-M-15+) | 3/3 | NA | 3/3 | 3/3  |
|   |  KPN (KPC-2+) | 3/3 | NA | 3/3 | 3/3  |
|   |  PM | 3/3 | NA | 3/3 | NA  |
|   |  PA (NDM-1+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  SM | 3/3 | 3/3 | 3/3 | NA  |
|  K. pneumoniae (KPC-2+) | ABX | 3/3 | 3/3 | 3/3 | NA  |
|   |  ECX | 3/3 | 3/3 | 3/3 | NA  |
|   |  EC (CTX-M-15+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  KO | 3/3 | 3/3 | 3/3 | NA  |
|   |  PM | 3/3 | 3/3 | 3/3 | NA  |
|   |  PA (NDM-1+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  SM | 3/3 | 3/3 | 3/3 | NA  |
|  P. mirabilis | ABX | 3/3 | NA | 3/3 | NA  |
|   |  ECX | 3/3 | NA | 3/3 | NA  |
|   |  EC (CTX-M-15+) | 3/3 | NA | 3/3 | 3/3  |
|   |  KO | 5/5 | NA | 5/5 | NA  |
|   |  KPN (KPC-2+) | 3/3 | NA | 3/3 | 3/3  |
|   |  PA (NDM-1+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  SM | 3/3 | 3/3 | 3/3 | NA  |
|   | ABX | 3/3 | 3/3 | 3/3 | NA  |

K190341 - Page 24 of 36

{24}

|  P. aeruginosa
(NDM-1+) | ECX | 3/3 | 3/3 | 3/3 | NA  |
| --- | --- | --- | --- | --- | --- |
|   |  EC (CTX-M-15+) | 12/13 | 11/13 | 13/13 | 13/13  |
|   |  EC (CTX-M-15+) 1:1 | 10/10 | 10/10 | 10/10 | 10/10  |
|   |  KO | 3/3 | 3/3 | 3/3 | NA  |
|   |  KPN (KPC-2+) | 3/3 | 3/3 | 3/3 | 3/3  |
|   |  PM | 3/3 | 3/3 | 3/3 | NA  |
|   |  SM | 3/3 | 3/3 | 3/3 | NA  |
|  S. marcescens | KPN (KPC-2+) | 3/3 | NA | 3/3 | 3/3  |
|   |  PM | 3/3 | NA | 3/3 | NA  |
|   |  PA (NDM-1+) | 3/3 | NA | 3/3 | 3/3  |
|   |  ABX | 3/3 | NA | 3/3 | NA  |
|   |  ECX | 3/3 | NA | 3/3 | NA  |
|   |  EC (CTX-M-15+) | 3/3 | NA | 3/3 | 3/3  |
|   |  KO | 3/3 | NA | 3/3 | NA  |

9. Reproducibility:

To confirm the site-to-site, operator-to-operator, system-to-system, and lot-to-lot reproducibility of the iC-GN Assay, a representative panel of target organisms and one non-target organism were evaluated at two clinically relevant concentrations: initial bottle positivity and eight hours beyond initial bottle positivity. Organisms were grown to the appropriate concentrations in BD BACTEC Plus Aerobic blood culture bottles with human blood added on the BD BACTEC System. Testing was performed by two independent operators at each of three sites, two external and one internal. Each operator tested the eighteen-organism panel in triplicate across five, non-consecutive days. Testing was performed on six iC-GN Cassette lots and multiple iC-Systems. Performance is based on all expected targets detected and no false positive targets detected. The table below summarizes the reproducibility results stratified by iC-GN target and concentration. Overall Reproducibility performance was 99.3%, confirming that iC-GN Assay performance is reproducible across sites, operators, systems and lots.

K190341 - Page 25 of 36

{25}

|  Table 10: iC-GN Assay Reproducibility Performance by Target  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Target/Concentration | Overall Performance | Overall Performance % [95% CI] | False Negatives | False Positives | PC Check Failures | System Failures  |
|  A. baumannii complex Bottle Ring | 90/90 | 100.0 [95.91-100.0] | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  A. baumannii complex Bottle Ring + 8 hours | 87/90 | 96.7 [90.65-98.86] | 0/90 (0.00%) | 3/90 (3.33%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  E. cloacae complex Bottle Ring | 86/88 | 97.7 [92.09-99.37] | 1/88 (1.14%) | 1/88 (1.14%) | 2/90 (2.22%) | 0/90 (0.00%)  |
|  E. cloacae complex Bottle Ring + 8 hours | 90/90 | 100.0 [95.91-100.0] | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  E. coli Bottle Ring | 90/90 | 100.0 [95.91-100.0] | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  E. coli Bottle Ring + 8 hours | 89/89 | 100.0 [95.86-100.0] | 0/89 (0.00%) | 0/89 (0.00%) | 0/90 (0.00%) | 1/90 (1.11%)  |
|  K. oxytoca Bottle Ring | 89/90 | 98.9 [93.97-99.80] | 0/90 (0.00%) | 1/90 (1.11%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  K. oxytoca Bottle Ring + 8 hours | 89/89 | 100.0 [95.86-100.0] | 0/89 (0.00%) | 0/89 (0.00%) | 1/90 (1.11%) | 0/90 (0.00%)  |
|  K. pneumoniae Bottle Ring | 90/90 | 100.0 [95.91-100.0] | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  K. pneumoniae Bottle Ring + 8 hours | 89/89 | 100.0 [95.86-100.0] | 0/89 (0.00%) | 0/89 (0.00%) | 1/90 (1.11%) | 0/90 (0.00%)  |
|  Proteus species Bottle Ring | 89/89 | 100.0 [95.86-100.0] | 0/89 (0.00%) | 0/89 (0.00%) | 1/90 (1.11%) | 0/90 (0.00%)  |
|  Proteus species Bottle Ring + 8 hours | 88/88 | 100.0 [95.92-100.0] | 0/88 (0.00%) | 0/88 (0.00%) | 0/90 (0.00%) | 2/90 (2.22%)  |
|  P. aeruginosa Bottle Ring | 88/89 | 98.9 [93.91-99.80] | 1/89 (1.12%) | 0/89 (0.00%) | 1/90 (1.11%) | 0/90 (0.00%)  |
|  P. aeruginosa Bottle Ring + 8 hours | 89/90 | 98.9 [93.97-99.80] | 1/90 (1.11%) | 0/90 (0.00%) | 0/90 (0.00%) | 0/90 (0.00%)  |
|  S. marcescens Bottle Ring | 87/89 | 97.8 [92.17-99.38] | 0/89 (0.00%) | 2/89 (2.25%) | 1/90 (1.11%) | 0/90 (0.00%)  |
|  S. marcescens Bottle Ring + 8 hours | 87/89 | 97.8 [92.17-99.38] | 0/89 (0.00%) | 2/89 (2.25%) | 1/90 (1.11%) | 0/90 (0.00%)  |

K190341 - Page 26 of 36

{26}

|  Table 10 : iC-GN Assay Reproducibility Performance by Target  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Target/Concentration | Overall Performance | Overall Performance % [95% CI] | False Negatives | False Positives | PC Check Failures | System Failures  |
|  CTX-M group 1
Bottle Ring | 90/90 | 100.0
[95.91-100.0] | 0/90
(0.00%) | 0/90
(0.00%) | 0/90
(0.00%) | 0/90
(0.00%)  |
|  CTX-M group 1
Bottle Ring + 8 hours | 89/89 | 100.0
[95.86-100.0] | 0/89
(0.00%) | 0/89
(0.00%) | 0/90
(0.00%) | 1/90
(1.11%)  |
|  KPC
Bottle Ring | 90/90 | 100.0
[95.91-100.0] | 0/90
(0.00%) | 0/90
(0.00%) | 0/90
(0.00%) | 0/90
(0.00%)  |
|  KPC
Bottle Ring + 8 hours | 89/89 | 100.0
[95.86-100.0] | 0/89
(0.00%) | 0/89
(0.00%) | 1/90
(1.11%) | 0/90
(0.00%)  |
|  NDM
Bottle Ring | 89/89 | 100.0
[95.86-100.0] | 0/89
(0.00%) | 0/89
(0.00%) | 1/90
(1.11%) | 0/90
(0.00%)  |
|  NDM
Bottle Ring + 8 hours | 89/90 | 98.9
[93.97-99.80] | 1/90
(1.11%) | 0/90
(0.00%) | 0/90
(0.00%) | 0/90
(0.00%)  |

## B. Comparison Studies:

### 1. Method Comparison with the Predicate Device:

A method comparison study was performed at five (5) geographically dispersed clinical sites. Sites tested 1002 leftover de-identified specimens from anaerobic and aerobic blood culture bottles flagged as positive by their respective continuous monitoring blood culture system. Three of the commonly used blood culture systems were included in the study: Thermo Fisher VersaTREK, BD BACTEC and BioMerieux BacT/ALERT.

Patient positive blood cultures confirmed by Gram stain to be positive for Gram-negative bacilli were enrolled in the study. Any positive blood cultures showing an initial mixed Gram stain were not enrolled or were subsequently withdrawn from the study dataset.

Final performance of the iC-GN Assay organism targets was compared to reference culture followed by MALDI identification per the study protocol. Final performance of the iC-GN Assay resistance marker targets was compared to PCR amplification followed by confirmatory bi-directional sequencing. Phenotypic antimicrobial susceptibility testing (AST) was also performed on all specimens to identify additional samples which required sequencing. Discordant samples were also sequenced.

To supplement performance of observed lower prevalence organisms, 170 contrived samples were prepared using verified strains. Contrived samples were prepared at iCubate using BD BACTEC Plus Aerobic Blood Culture Bottles with 10 mL of human blood added (in accordance with BACTEC instructions). Organisms were spiked into bottles at concentrations of 5-30 CFU/bottle and incubated until bottles were flagged as positive. Aliquots of samples were frozen and provided to the sites (frozen) for testing.

Of the 1107 positive blood culture specimens enrolled in the study, a total of 105 specimens were excluded/withdrawn from the study and excluded from all subsequent performance analyses. Of the 1002 specimens remaining, 976 were fresh prospective specimens and 26 (2.6%) were frozen prospective specimens. Nineteen (19) samples

K190341 - Page 27 of 36

{27}

were excluded from the performance analysis for Proteus mirabilis due to confirmed contamination of BD BACTEC Bottles with non-viable Proteus organisms or nucleic acid, leaving a total of 983 evaluable specimens for Proteus mirabilis.

The total specimens excluded from the iC-GN Assay Method Comparison Study (n=105) are listed by site and the reasons for exclusion are noted in the 510(k) Summary for K190341, Table 12 (p.25). The most common reasons for exclusion included incomplete reference testing and repeat iC-GN errors.

Throughout the course of the study, an initial error rate of 2.9 % (34/1181) was observed. Reasons for error included the following: Positive controls check failure (27), Array registration error (6), and Processor/System error (1). When an error was observed, repeat testing was performed with the iC-GN Assay per the protocol. Upon repeat testing, the error rate was reduced to 0.8% (9/1181). The 510(k) Summary for K190341, Table 13 (p.26) includes a table of the no-call results.

When performance of the iC-GN Assay was compared to reference culture followed by MALDI identification or PCR/bi-directional sequencing, there was no significant difference in performance noted between the five study sites or between the three blood culture systems. Performance for all positive bottle types/systems combined is presented in the tables below for detection of the iC-GN Assay targets as compared to culture and MALDI or PCR/bi-directional sequencing. Results are stratified by prospectively tested fresh specimens, prospectively collected/retrospectively tested frozen specimens and contrived specimens. These data are provided in the tables below.

K190341 - Page 28 of 36

{28}

**1/1 false positive observed was negative for A. baumannii complex by PCR/bi-directional sequencing

|  Table 11a: iC-GN Assay Performance: Acinetobacter baumannii complex (ppa) |   | N= | Percent Agreement |   | Comparator Method  |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type |   |   | Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |
|  Prospective | Fresh | 976 | 100% 7/7 (64.6-100) | 99.9% 968-969** (99.4-100)  |   |
|   |  Frozen | 26 | - 0/0 - | 100% 26/26 (87.1-100)  |   |
|   |  TOTAL | 1002 | 100% 7/7 (64.6-100) | 99.9% 994/995 (99.4-100)  |   |
|  Contrived |   | 170 | 100% 45/45 (92.1-100) | 100% 125/125 (97.0-100) |   |

Table 11b: iC-GN Assay Performance: Enterobacter cloacae complex (ramA)

|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 976 | 94.5% 52/55* (85.1-98.1) |   | 100% 921/921 (99.6-100)  |
|   |  Frozen | 26 | 100% 5/5 (56.6-100) |   | 100% 21/21 (84.5-100)  |
|   |  TOTAL | 1002 | 95.0% 57/60 (86.3-98.3) |   | 100% 942/942 (99.6-100)  |
|  Contrived |   | 170 | 100% 17/17 (81.6-100) | 100% 153/153 (97.6-100) |   |

*1/3 false negatives observed was negative for E. cloacae complex by PCR/bi-directional sequencing; 2/3 were positive for E. cloacae complex by PCR/bi-directional sequencing

K190341 - Page 29 of 36

{29}

K190341 - Page 30 of 36

|  Table 11c: iC-GN Assay Performance: Escherichia coli (uidA)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 976 | 98.4%
480/488*
(96.8-99.2) |   | 100%
488/488
(99.2-100)  |
|   |  Frozen | 26 | 100%
6/6
(61.0-100) |   | 100%
20/20
(83.9-100)  |
|   |  TOTAL | 1002 | 98.4%
486/494
(96.8-99.2) |   | 100%
508/508
(99.2-100)  |
|  Contrived |   | 170 | 100%
15/15
(79.6-100) | 100%
155/155
(97.6-100) |   |

*4/8 false negatives observed were negative for E. coli by PCR/bi-directional sequencing; 3/8 were positive for E. coli by PCR/bi-directional sequencing; 1/8 was not available for sequencing

|  Table 11d: iC-GN Assay Performance: Klebsiella oxytoca (pehX)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 976 | 95.8%
23/24*
(79.8-99.3) |   | 99.7%
949/952**
(99.1-99.9)  |
|   |  Frozen | 26 | -
0/0
- |   | 100%
26/26
(87.1-100)  |
|   |  TOTAL | 1002 | 95.8%
23/24
(79.8-99.3) |   | 99.7%
975/978
(99.1-99.9)  |
|  Contrived |   | 170 | 100%
30/30
(88.6-100) | 100%
140/140
(97.3-100) |   |

*1/1 false negative observed was negative for K. oxytoca by PCR/bi-directional sequencing
**3/3 false positives observed were negative for K. oxytoca by PCR/bi-directional sequencing

{30}

K190341 - Page 31 of 36

|  Table 11e: iC-GN Assay Performance: Klebsiella pneumoniae (parC)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 976 | 96.8%
150/155*
(92.7-98.6) |   | 99.3%
815/821**
(98.4-99.7)  |
|   |  Frozen | 26 | 100%
3/3 (43.9-100) |   | 100%
23/23 (85.7-100)  |
|   |  TOTAL | 1002 | 96.8%
153/158
(92.8-98.6) |   | 99.3%
838/844
(98.4-99.7)  |
|  Contrived |   | 170 | 100%
21/21
(84.5-100) | 99.3%
148/149
(96.3-99.9) |   |

*3/5 false negatives observed were negative for K. pneumoniae by PCR/bi-directional sequencing; 2/3 were positive for K. pneumoniae by PCR/bi-directional sequencing
**6/6 false positives observed were negative for K. pneumoniae by PCR/bi-directional sequencing

|  Table 11f: iC-GN Assay Performance: Proteus mirabilis (rpoB)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 957 | 97.4%
37/38*
(86.5-99.5) |   | 99.5%
914/919**
(98.7-99.8)  |
|   |  Frozen | 26 | 100%
9/9
(70.1-100) |   | 100%
17/17
(81.6-100)  |
|   |  TOTAL | 983*** | 97.9%
46/47
(88.9-99.6) |   | 99.5%
931/936
(98.8-99.8)  |
|  Contrived |   | 170 | 100%
12/12
(75.8-100) | 100%
158/158
(97.6-100) |   |

*1/1 false negative observed was positive for P. mirabilis by PCR/bi-directional sequencing
**3/5 false positives observed were negative for P. mirabilis by PCR/bi-directional sequencing; 2/5 were not available for sequencing
*** 19 samples were excluded from Proteus mirabilis performance analysis due to confirmed Proteus contamination within the BD BACTEC Bottles, leaving a total of 983 evaluable specimens.

{31}

K190341 - Page 32 of 36

|  Table 11g: iC-GN Assay Performance: Pseudomonas aeruginosa (algD)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 976 | 95.1%
78/82*
(88.1-98.1) |   | 99.8%
892/894**
(99.2-99.9)  |
|   |  Frozen | 26 | 100%
1/1
(20.7-100) |   | 100%
25/25
(86.7-100)  |
|   |  TOTAL | 1002 | 95.2%
79/83
(88.3-98.1) |   | 99.8%
917/919
(99.2-99.9)  |
|  Contrived |   | 170 | 100%
10/10
(72.2-100) | 100%
160/160
(97.7-100) |   |

*4/4 false negatives observed were positive for P. aeruginosa by PCR/bi-directional sequencing
**2/2 false positives observed were negative for P. aeruginosa by PCR/bi-directional sequencing

|  Table 11h: iC-GN Assay Performance: Serratia marcescens (gyrB)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | Culture & MALDI  |   |
|  Prospective | Fresh | 976 | 100%
29/29
(88.3-100) |   | 99.6%
943/947**
(98.9-99.8)  |
|   |  Frozen | 26 | 0/0- |   | 100%
26/26
(87.1-100)  |
|   |  TOTAL | 1002 | 100%
29/29
(88.3-100) |   | 99.6%
969/973
(98.9-99.8)  |
|  Contrived |   | 170 | 100%
20/20
(83.9-100) | 99.3%
149/150
(96.3-99.9) |   |

**1/4 false positives observed was positive for S. marcescens by PCR/bi-directional sequencing; 3/4 were negative for S. marcescens by PCR/bi-directional sequencing

{32}

|  Table 11I: iC-GN Assay Performance: CTX-M  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | PCR/Bi-directional sequencing  |   |
|  Prospective | Fresh | 976 | 97.0%
64/66 (89.6-99.2) |   | 99.9%
909/910 (99.4-100)  |
|   |  Frozen | 26 | 100%
1/1 (20.7-100) |   | 100%
25/25 (86.7-100)  |
|   |  TOTAL | 1002 | 97.0%
65/67 (89.8-99.2) |   | 99.9%
934/935 (99.4-100)  |
|  Contrived |   | 170 | 100%
15/15 (79.6-100) | 100%
155/155 (97.6-100) |   |
|  Table 11J: iC-GN Assay Performance: KPC  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | PCR/Bi-directional sequencing  |   |
|  Prospective | Fresh | 976 | 100%
1/1 (20.7-100) |   | 99.9%
974/975 (99.4-100)  |
|   |  Frozen | 26 | -
0/0
- |   | 100%
26/26 (87.1-100)  |
|   |  TOTAL | 1002 | 100%
1/1 (20.7-100) |   | 99.9%
1000/1001 (99.4-100)  |
|  Contrived |   | 170 | 100%
50/50 (92.9-100) | 99.2%
119/120 (95.4-99.9) |   |

K190341 - Page 33 of 36

{33}

K190341 - Page 34 of 36

|  Table 11K: iC-GN Assay Performance: NDM  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Specimen Type | N= | Percent Agreement |   | Comparator Method  |   |
|   |   |  Positive (95% CI) | Negative (95% CI) | PCR/Bi-directional sequencing  |   |
|  Prospective | Fresh | 976 | - 0/0 - |   | 100% 976/976 (99.6-100)  |
|   |  Frozen | 26 | - 0/0 - |   | 100% 26/26 (87.1-100)  |
|   |  TOTAL | 1002 | - 0/0 - |   | 100% 1002/1002 (99.6-100)  |
|  Contrived |   | 170 | 100% 50/50 (92.9-100) | 100% 120/120 (96.9-100) |   |

2. Analysis of Mixed Culture Results:

In the method comparison study, there were thirty (30) mixed culture specimens that were detected by the iC-GN Assay, culture and MALDI, or both. There were twelve (12) discrepant mixed samples for which iC-GN detected a target that was not detected by the comparator assay. There were four (4) discrepant mixed samples for which the comparator assay detected targets that were not detected by iC-GN. Due to competitive inhibition, target organisms present at low concentrations may not be detected by the iC-GN Assay when a second target organism is present at higher concentrations. See the 510(k) Summary, Tables 29 and 30 (p.31-32) which includes tables that list the mixed target combinations detected by iC-GN and the comparator method in the clinical study.

|  Table 12: Multiple Organism Detections by iC-GN as Compared to Culture/MALDI  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Multiple Detections by iC-GN |   |   |   |   | Total Targets Detected by iC-GN | No of Discrepant Targets  |
|  Site | ID | Target 1 | Target 2 | Target 3  |   |   |
|  LAC | 1102 | E. coli | K. pneumoniae |  | 2 | 0  |
|  LAC | 1118 | E. coli | K. pneumoniae |  | 2 | 0  |
|  LAC | 1141 | E. coli | K. pneumoniae |  | 2 | 0  |
|  LAC | 1220 | E. cloacae complex | E. coli |  | 2 | 0  |
|  LAC | 1236 | E. coli | K. oxytoca | K. pneumoniae | 3 | 1  |
|  LAC | 1285 | K. pneumoniae | S. marcescens |  | 2 | 1  |
|  LAC | 1307 | E. coli | K. oxytoca |  | 2 | 0  |
|  LAC | 1378 | E. coli | K. pneumoniae |  | 2 | 1  |
|  LAC | 1382 | K. oxytoca | K. pneumoniae |  | 2 | 1  |
|  MCW | 2023 | E. coli | K. pneumoniae |  | 2 | 0  |

{34}

|  Table 12: Multiple Organism Detections by iC-GN as Compared to Culture/MALDI  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Multiple Detections by iC-GN |   |   |   |   | Total Targets Detected by iC-GN | No of Discrepant Targets | Discrepant Results (Targets Not Detected by culture/MALDI)  |
|  Site | ID | Target 1 | Target 2 | Target 3  |   |   |   |
|  MCW | 2032 | E. cloacae complex | K. oxytoca |  | 2 | 0 |   |
|  MCW | 2038 | K. oxytoca | K. pneumoniae |  | 2 | 0 |   |
|  MCW | 2041 | E. coli | P. mirabilis |  | 2 | 0 |   |
|  MCW | 2104 | E. coli | S. marcescens |  | 2 | 1 | S. marcescens  |
|  MCW | 2193 | K. pneumoniae | S. marcescens |  | 2 | 2 | K. pneumoniae, S. marcescens  |
|  TC | 3015 | K. oxytoca | P. aeruginosa |  | 2 | 2 | K. oxytoca, P. aeruginosa  |
|  TC | 3096 | E. coli | K. pneumoniae |  | 2 | 1 | K. pneumoniae  |
|  TC | 3131 | E. cloacae complex | K. pneumoniae |  | 2 | 0 |   |
|  TC | 3183 | K. pneumoniae | S. marcescens |  | 2 | 1 | S. marcescens  |
|  TGH | 4031 | E. coli | P. mirabilis |  | 2 | 1 | P. mirabilis  |
|  TGH | 4037 | E. coli | P. aeruginosa |  | 2 | 0 |   |
|  TGH | 4124 | E. cloacae complex | P. aeruginosa |  | 2 | 1 | P. aeruginosa  |
|  TGH | 4132 | E. cloacae complex | K. pneumoniae |  | 2 | 0 |   |
|  IU | 5025 | A. baumannii complex | K. pneumoniae |  | 2 | 1 | A. baumannii complex  |
|  IU | 5031 | E. coli | K. pneumoniae |  | 2 | 0 |   |
|  IU | 5042 | E. cloacae complex | K. pneumoniae |  | 2 | 0 |   |
|  LAC | 1102 | K. pneumoniae | E. coli |  | 2 | 0 |   |
|  LAC | 1118 | K. pneumoniae | E. coli |  | 2 | 0 |   |
|  LAC | 1141 | E. coli | K. pneumoniae |  | 2 | 0 |   |
|  LAC | 1220 | E. cloacae complex | E. coli |  | 2 | 0 |   |
|  LAC | 1236 | K. pneumoniae | E. coli |  | 2 | 0 |   |
|  LAC | 1268 | P. aeruginosa | P. mirabilis |  | 2 | 1 | P. aeruginosa  |
|  LAC | 1307 | E. coli | K. oxytoca |  | 2 | 0 |   |
|  LAC | 1338 | E. coli | K. pneumoniae |  | 2 | 1 | K. pneumoniae  |
|  MCW | 2023 | K. pneumoniae | E. coli |  | 2 | 0 |   |
|  MCW | 2032 | K. oxytoca | E. cloacae complex |  | 2 | 0 |   |
|  MCW | 2038 | K. pneumoniae | K. oxytoca |  | 2 | 0 |   |
|  MCW | 2041 | P. mirabilis | E. coli |  | 2 | 0 |   |
|  TC | 3006 | E. coli | K. pneumoniae |  | 2 | 1 | K. pneumoniae  |
|  TC | 3131 | E. cloacae complex | K. pneumoniae |  | 2 | 0 |   |

K190341 - Page 35 of 36

{35}

|  Table 12: Multiple Organism Detections by iC-GN as Compared to Culture/MALDI  |   |   |   |   |   |   |
| --- | --- | --- | --- |…

---

**Source:** [https://fda.innolitics.com/device/K190341](https://fda.innolitics.com/device/K190341)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
