K190336 · Soniquence, LLC · GEI · Jun 20, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190336
Device Name
Soniquence Bipolar Electrodes
Applicant
Soniquence, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Soniquence Bipolar Electrode is intended for use by a physician familiar with bipolar coagulation with electrosurgery where coagulation/contraction of soft tissue is needed.
Device Story
Soniquence Bipolar Electrodes are disposable, single-use electrosurgical devices consisting of a plastic handle, insulated shaft, bipolar tip, and integrated cable. Operated by physicians in surgical settings, the device connects to a Soniquence high-frequency RF generator. The generator delivers RF energy to the active tip, which is applied to soft tissue to achieve coagulation or contraction. The device functions via standard bipolar electrosurgical principles, where RF energy is transferred to the target tissue site. The device is provided sterile via ethylene oxide. Clinical benefit is derived from the ability to perform controlled soft tissue coagulation during surgical procedures. The healthcare provider controls the energy delivery via the generator settings, allowing for precise thermal effects on target tissue.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601 series), biocompatibility (ISO 10993 series), and performance testing on ex vivo tissue (liver, kidney, muscle) to evaluate thermal damage zones and temperature profiles during maximum energy delivery.
Technological Characteristics
Disposable bipolar electrosurgical device; plastic handle, insulated shaft, bipolar tip. Energy source: external RF generator. Sterilization: Ethylene oxide (ISO 11135). Biocompatibility: ISO 10993. Electrical safety: IEC 60601-1, IEC 60601-2-2. Packaging: ASTM F1886, F1929, F1980. Shelf life: 3 years.
Indications for Use
Indicated for use by physicians familiar with bipolar electrosurgery for coagulation or contraction of soft tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 20, 2019
Soniquence, LLC Suzanne Lucas Sr. Regulatory Affairs Specialist 2473 Grand Avenue Baldwin, New York 11510
#### Re: K190336
Trade/Device Name: Soniquence Bipolar Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: March 22, 2019 Received: March 25, 2019
Dear Suzanne Lucas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190336
Device Name Soniquence Bipolar Electrodes
Indications for Use (Describe)
The Soniquence Bipolar Electrode is intended for use by a physician familiar with bipolar coagulation with electrosurgery where coagulation/contraction of soft tissue is needed.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a blue square with rounded corners, outlined in white and gray. Inside the square is a white, stylized double helix shape, reminiscent of a DNA strand. The background is a gradient of white and gray, with a blue line extending from the top left corner.
Image /page/3/Picture/1 description: The image shows the logo for Soniquence. The logo is blue and features a stylized DNA double helix on the left side. The text "oniquence" is written in a sans-serif font to the right of the helix, with a small registered trademark symbol in the upper right corner.
K190336
# Exhibit 4. 510(k) SUMMARY
(As required by 21 CFR 807.92(a))
Date Prepared
May 6, 2019
# Submitter's Information (807.92(a)(1))
Company Name and Address:
Soniquence, LLC 2473 Grand Avenue Baldwin, NY 11510 Phone: (516) 654-4000 www.soniquence.com
Establishment Registration for Soniquence, LLC will be completed once this 510(k) has been FDA cleared.
### Contact Information:
Suzanne Lucas - Regulatory Affairs Phone: (516) 654-8017 Fax: (516) 654-8000 Email: slucas@soniquence.com
Paul D. Buhrke IV- QA/RA Manager Phone: (516) 277-9010 Fax: (516) 277-9011 Email: pbuhrke@elliquence.com
{4}------------------------------------------------
#### K190336
#### Device Information (807.92(a)(2))
Trade Name: Soniquence Bipolar Electrodes
#### Common/Usual Name
Electrosurqical, Cutting & Coaqulation Device & Accessories
#### Classification Name and Requlation
Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400
#### Classification Panel
Division of General, Restorative, and Neurological Devices
#### Device Class/Product Code
FDA Classification: Class 2 FDA Product Code: GEI
#### Predicate Devices (807.92(a)(3))
- Meddusa Bipolar System (K130110)
- Triager Flex Electrodes (K003126) .
### Device Description (807.92(a)(4))
The Soniquence Bipolar Electrodes family is a line of disposable bipolar electrosurgical devices which is intended for use by a physician familiar with electrosurgery in bipolar coagulation for general surgery where coagulation of soft tissue is needed. The basic design of the Soniquence Bipolar Electrodes is a plastic handle, an elongated, insulated electrode shaft, a bipolar electrode tip, and an integrated cable and plug which is designed to connect exclusively to the Soniquence line of high-frequency RF energy sources.
The operational principle of the device is common to all active bipolar electrosurgical devices; the active tip of the Soniquence Bipolar Electrode emits RF energy generated according to settings established on the Soniquence RF generator which emits high frequency, low temperature radiowaves directed to the electrode tip. The electrical power operating at radio frequency (RF) is transferred to tissue at the surgical site. This energy is capable of cutting or coagulating target tissue depending on the waveform associated with the mode and power setting of the generator. The devices are disposable and intended for single-use only. The devices are intended for transient contact with the patient, and are provided sterilized by ethylene oxide and has a shelf life of three years from the date of sterilization.
{5}------------------------------------------------
### K190336
### Intended Use:
The Soniquence Bipolar Electrode is intended for use by a physician familiar with bipolar coagulation with electrosurgery where coagulation/contraction of soft tissue is needed.
#### Substantial Equivalence Comparison (807.92(a)(6))
The Soniquence Bipolar Electrodes are substantially equivalent in intended use and identical in technological characteristics to the Meddusa Bipolar System (K130110), and the Bipolar Trigger-Flex Electrode (K003126). An extensive comparison chart concerning key characteristics is provided in the Substantial Equivalence section.
### Non-Clinical Testing (807.92(b)(1))
The Soniquence Bipolar Electrodes will be manufactured in accordance with the design control requirements of 21 CFR 820.30. Appropriate non-clinical verification and validation activities were planned and conducted in order to address identified risks and ensure the safety and effectiveness of the device. The following tests were successfully performed:
| • IEC 60601-1:2005/(R)2012 and A1:2012 - Medical Electrical Equipment, Part I: General requirements for basic safety and essential performance |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • IEC 60601-1-6:2010 (Third Edition) + A1 2013: General requirements for basic safety and essential performance – Collateral standard: Usability |
| • IEC 62366:2007 (First Edition) + A1:2014: Medical devices – Application of usability engineering to medical devices |
| • IEC 60601-1-2:2014 - Medical Electrical Equipment, Part I-2: General requirements for basic safety and essential performance. Collateral Standard: Electromagnetic Compatibility |
| • IEC 60601-2-2: 2009 +A11:2011 - Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories |
| • ISO 11135:2014 - Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices |
| • ISO 11737-1:2006 Sterilization of medical products--Microbiological methods--Part 1: Determination of population of microorganisms on product |
| • ISO 11737-2:2009 Sterilization of medical devices--Microbiological methods--Part 2: Tests of sterility performed in the validation of a sterilization process |
| • ISO 11138-1:2006 Sterilization of health care products--Biological indicators--Part1: General Requirements |
| • ISO 11138-2:2006 Sterilization of health care products--Biological indicators--Part2: Biological indicators for ethylene oxide sterilization processes |
| • ISO 10993-7:2008 Biological evaluation of medical device--Part: 7: Ethylene Oxide sterilization residual. |
| • ISO 10993-4:2002+Amd 1 :2006: Biological Evaluation of Medical Devices, Part 4: Selection of test for interactions with blood. |
| • ISO 10993-1 - Biological evaluation of medical devices -- Part 1: Evaluation and testing |
- 0993-1 -Bloiogical evalua within a risk management process
{6}------------------------------------------------
- ISO 10993-5 Biological evaluation of medical devices Part 5: Tests for in vitro . cytotoxicity
- ISO 10993-10 Biological evaluation of medical devices Part 10: Tests for irritation and . skin sensitization
- ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for systemic ● toxicitv
- ASTM F1886/F1886M -Standard Test Method for Determining Integrity of Seals for ● Flexible Packaging by Visual Inspection
- ASTM F1929 Standard Test Method for Detecting Seal Leaks in Porous Medical ● Packaging by Dye Penetration
- ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for . Medical Devices
### Performance Data
The subject device is composed of biocompatible materials, has passed dielectric testing, and performs identical to the predicate devices. Bench tests used ex vivo tissue that included liver, kidney, and muscle tissue. The subject devices were tested to demonstrate the thermal effect on tissue by measuring the width and depth of the thermally damaged zones in relation to tissue type, intensity setting, and duration of activation. The temperature profile of the subject device applicator and cable was recorded during simulation use for the maximum energy delivery duration at the maximum power to demonstrate the maximum applicator surface area and maximum cable surface temperature will be safe to the user and/or to the patient. The peak temperatures of the electrode tips and target tissue/vessels when the device is used for the maximum recommended duration and generator output settings was compared. Based on the results of the various bench tests, it was determined that the subject device-is safe and effective.
### Technological Characteristics
The devices are substantially equivalent to the predicate devices based on a comparison of physical and performance characteristics.
### Clinical Testing (807.92(b)(2))
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
# Conclusion (807.92(b)(3))
ln accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate devices in terms of technology, performance, and indications for use, Soniquence, LLC concludes that the subject device, Soniquence Bipolar Electrodes does not raise any issues of safety or effectiveness and are substantially equivalent to the predicate devices as described above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.