Applied Medical Laparoscopic Linear Cutter

K190331 · Applied Medical Resources Corporation · GDW · Nov 22, 2019 · General, Plastic Surgery

Device Facts

Record IDK190331
Device NameApplied Medical Laparoscopic Linear Cutter
ApplicantApplied Medical Resources Corporation
Product CodeGDW · General, Plastic Surgery
Decision DateNov 22, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Applied Medical Laparoscopic Linear Cutter has applications in open and/ or minimally invasive general, urologic, gynecologic, pediatric and thoracic surgery for resection, transection and/or creation of anastomosis. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. It can be used with staple line or tissue buttressing materials.

Device Story

Sterile, single-use laparoscopic stapling system; used in open/minimally invasive surgery for tissue transection, resection, and sealing. Device features rotating shaft and articulating jaws; places six staggered rows of titanium staples (three per side). Available in three shaft lengths (20cm, 37cm, 50cm) and three reload lengths (30mm, 45mm, 60mm) with color-coded staple heights (1.0mm, 1.5mm, 2.0mm) for varying tissue thicknesses. Operated by surgeon; requires five trigger strokes to complete firing. Provides mechanical tissue sealing and division; benefits patient by facilitating surgical access and ensuring hemostasis in complex procedures.

Clinical Evidence

No human clinical studies performed. Evidence based on bench testing (firing force, leak pressure, articulation, staple formation, tissue compression) and preclinical porcine studies (21-day chronic survival study for hemostasis/histopathology; acute study on high-pressure arteries and thin-walled veins). Biocompatibility testing performed per ISO 10993 standards.

Technological Characteristics

Single-use, sterile, mechanical stapler. Materials: Titanium staples. Dimensions: 20cm, 37cm, 50cm shaft lengths; 30mm, 45mm, 60mm reload lengths. Staple heights: 1.0mm, 1.5mm, 2.0mm. Biocompatibility per ISO 10993-5, -6, -7, -10, -11, -18 and USP 40-NF 35:<151>. Sterilization: Ethylene Oxide.

Indications for Use

Indicated for patients undergoing open or minimally invasive general, urologic, gynecologic, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis, including liver, pancreas, kidney, and spleen procedures.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it. November 22, 2019 Applied Medical Resources Corporation Aditi Iyengar Regulatory Affairs Specialist II 22872 Avenida Empresa Rancho Santa Margarita, California 92688 Re: K190331 Trade/Device Name: Applied Medical Laparoscopic Linear Cutter Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: October 23, 2019 Received: October 24, 2019 # Dear Ms. Iyengar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {2}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K190331 Device Name Applied Medical Laparoscopic Linear Cutter #### Indications for Use (Describe) Applied Medical Laparoscopic Linear Cutter has applications in open and/or minimally invasive general, urologic, gynecologic, pediatic and thoracic surgery for resection and or creation of anastomosis. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. It can be used with staple line or tissue buttressing materials. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMP ETED FORM TO THE PRA STAFF EMAI ADDRESS BE OW * The urden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to > epartment of Health and Human Services Food and rug Administration ffice of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@f a.hhs.gov ap rso is otr quir to, a coll ctio of A ag ot co uct or sportsor a tor spo C a i for atio u l ss it ispla s a curr tl vvali OMB u b r." {4}------------------------------------------------ # 510(k) Summary | 510(K) Submitter | Applied Medical Resources Corp.<br>22872 Avenida Empresa<br>Rancho Santa Margarita, CA, 92688<br>(949) 713-8000 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Aditi Iyengar<br>Regulatory Affairs Specialist II<br>Applied Medical Resources Corp.<br>aiyengar@appliedmedical.com<br>Tel: (949) 713-8007<br>Fax: (949) 713-8205 | | Date of Preparation | February 12, 2019 | | Trade Name | Applied Medical Laparoscopic Linear Cutter | | Common Name | Surgical Stapler/Implantable Staple | | Classification | Regulation: 21 CFR 878.4750<br>Regulation Name: Implantable Staple<br>Device Class: Class II<br>Product Code: GDW<br>Regulation: 21 CFR 878.4800<br>Regulation Name: Surgical Stapler<br>Device Class: Class I<br>Product Code: GAG | | Predicate Device | Ethicon Echelon Flex Endopath<br>510(k)#: K081146<br>Product Code: GDW, GAG | | Intended Use | The Applied Medical Laparoscopic Linear Cutter has applications<br>in open and/ or minimally invasive general, urologic, gynecologic,<br>pediatric and thoracic surgery for resection, transection and/or<br>creation of anastomosis. The instruments may also be used for<br>transection and resection of liver parenchyma (hepatic vasculature<br>and biliary structures), pancreas, kidney and spleen. It can be used<br>with staple line or tissue buttressing materials. | {5}------------------------------------------------ Device Description The Applied Medical Laparoscopic Linear Cutter (LLC) is a sterile, single-use stapling system that is intended to be used for mechanically transecting, resecting, and sealing tissue during laparoscopic and open surgical procedures. The stapler places six staggered rows of titanium staples, three on either side of the transection line. The shaft can rotate freely in both directions and the jaws articulate to facilitate access to the operative site. > The Applied LLC is available in 3 different shaft lengths (20cm, 37cm, 50cm) that can each accommodate 3 different industrystandard reload lengths (60mm, 45mm, 30mm). Each of the reload lengths is available in three industry-standard reload colors and formed staple heights to accommodate various tissue thicknesses: 1.0mm (white reload), 1.5mm (blue reload), and 2.0mm (green reload). # Summary of Technological Characteristics with the Predicate Device The subject Applied Medical Laparoscopic Linear Cutter (LLC) and the predicate Ethicon Echelon Flex Endopath are similar in technological characteristics, materials, performance, and method of operation. Both devices are intended to mechanically seal and transect tissue by placing two, triple-staggered rows of titanium staples and simultaneously dividing the stapled tissue. The following technological differences exist between the subject and predicate staplers: - The subject device has 3 modular components: 1) handle 2) shaft 3) reload. The predicate device has 2 components: 1) handle and shaft 2) reload. - The subject device requires 5 trigger strokes to complete a firing while the predicate device requires 3 trigger strokes to complete a firing. - . The compact and long versions of the subject shaft differ in length from the predicate. However, this difference in shaft length does not impact staple formation. # Discussion of Performance Testing Submitted Evidence of safety and effectiveness of the subject device in support of substantial equivalence was obtained from bench testing, preclinical testing, and biocompatibility testing. No human clinical studies were required to demonstrate the safety and effectiveness of the subject device in support of this premarket notification. A summary of the performance tests used to evaluate the technological and performance characteristics in support of substantial equivalence is presented in Table 5-2. All performance testing acceptance criteria were met. {6}------------------------------------------------ ## Table 5-1 Bench Tests | Test | Test Objective | |-----------------------------------------|-------------------------------------------------------------------------------| | Firing Force | To evaluate the input force required to actuate the trigger | | Leak Test | To evaluate staple line leak pressures | | Articulation Testing | To evaluate device performance when fully articulated | | Reinforcement Material<br>Compatibility | To evaluate device performance with reinforcement material | | Shelf Life Testing | To evaluate device performance after simulated transit and storage conditions | | Staple Formation | To evaluate staple formation in media and out of media | | Formed Staple Height | To measure the formed staple height | | Radiograph Imaging | To evaluate location and quality of staple formation in media | | Tissue Compression Evaluation | To evaluate tissue effects associated with application of the device | | Crossed Staple Line | To evaluate device performance when fired across existing staple line | # Table 5-2 Preclinical Tests | Test | Test Objective | |------------------------------------------|----------------------------------------------------------------------------------------------------------| | 21-Day Porcine Chronic Survival<br>Study | To evaluate in vivo device performance including in<br>vivo hemostasis and histopathological evaluations | | Acute Porcine In-Vivo Study | To evaluate device performance on high pressure<br>arteries and thin-walled veins | ## Biocompatibility Tests The following biological evaluations were performed in support of substantial equivalence: - Ethylene Oxide Residual (per ISO 10993-7) ● - Material Mediated Pyrogenicity (per USP 40-NF 35:<151>) ● - Cytotoxicity (per ISO 10993-5) ● - Sensitization (per ISO 10993-10) - Acute Systemic Toxicity (per ISO 10993-11) - Intracutaneous Reactivity (per ISO 10993-10) - Intramuscular Implant Testing (2 week and 13 week, per ISO 10993-6) A chemical characterization and subsequent toxicological risk assessment (per ISO 10993-18) was performed to evaluate the safety of the implantable staples. Results from the testing were used to assess the following endpoints: - Reproductive and Developmental Toxicity {7}------------------------------------------------ - Subacute Toxicity . - Subchronic Toxicity ● - . Chronic Toxicity - Genotoxicity - . Carcinogenicity # Conclusion Based on the intended use, technological characteristics, and performance testing results, the subject Applied Laparoscopic Linear Cutter is substantially equivalent to the predicate Ethicon Echelon Flex Endopath and does not raise any new issues of safety and efficacy.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...