← Product Code [HKY](/productcode/HKY) · K190316

# Icare ic200 (K190316)

_Icare Finland OY · HKY · Jan 14, 2020 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190316

## Device Facts

- **Applicant:** Icare Finland OY
- **Product Code:** [HKY](/productcode/HKY.md)
- **Decision Date:** Jan 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1930
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Icare ic200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

## Device Story

Hand-held, battery-operated rebound tonometer; measures intraocular pressure (IOP) without topical anesthesia. Uses sterile, single-use 1.8 mm diameter magnetized probes. Device records probe speed upon initiation; probe decelerates upon corneal contact; deceleration rate and rebound time correlate to IOP. Calculates IOP via internal algorithm. Operable at any angle from 0° (sitting) to 90° (supine). Used by clinicians in professional healthcare settings. Output displayed on OLED screen; data transferable via Bluetooth to PC or printer. Facilitates IOP monitoring for glaucoma management or general ophthalmic assessment. Benefits include non-invasive measurement, no anesthesia requirement, and versatility in patient positioning.

## Clinical Evidence

Clinical study (n=152 eyes) compared Icare ic200 to Goldmann Applanation Tonometer (GAT) reference. In sitting position, mean IOP was 20.44 mmHg (ic200) vs 19.88 mmHg (GAT). In supine position, ic200 compared against Tonopen Avia and Perkins. ANSI Z80.10:2014 compliance demonstrated; 1.3% of measurements fell outside ±5 mmHg difference from GAT. No adverse events reported.

## Technological Characteristics

Rebound tonometer; battery-powered (4x AA). Uses sterile, gamma-sterilized single-use plastic probes (26.5 mg). Measurement range 7-50 mmHg (display 1-99 mmHg). Connectivity: Bluetooth (OTS module). Software: Embedded microcontroller, IEC 62304 compliant. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ANSI Z80.10 (tonometry).

## Regulatory Identification

A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.

## Predicate Devices

- Icare ic100 tonometer (type TA011), [K153694](/device/K153694.md)

## Reference Devices

- Goldmann Applanation Tonometer (GAT)
- Tonopen Avia
- Perkins

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 14, 2020

Icare Finland Oy Hannes Hyvonen Regulatory Affairs Manager Ayritie 22, Vantaa, FI 01510

Re: K190316

Trade/Device Name: Icare ic200 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: Class II Product Code: HKY Dated: December 4, 2019 Received: December 6, 2019

Dear Hannes Hyvonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Tieuvi Nguyen, Ph. D. Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K190316

Device Name Icare ic200

Indications for Use (Describe)

The Icare ic200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 (K)SUMMARY

#### SUBMITTER NAME AND ADDRESS I.

Icare Finland Oy Äyritie 22 Vantaa, Finland FI-01510

Contact: Hannes Hyvönen 358 9 8775 1150 Phone: Fax: 358 9 728 6670 Email: regulatory@icarefinland.com Date Prepared: January 9, 2020

#### II. DEVICE NAME

| Trade Name:              | Icare ic200               |
|--------------------------|---------------------------|
| Common/Usual Name:       | Tonometer                 |
| Classification Name:     | Tonometer and Accessories |
| Regulation No:           | 21 CFR 886.1930           |
| Device Regulatory Class: | II                        |
| Review Panel:            | Ophthalmic                |
| Product Code:            | HKY                       |

Premarket Notification (510(k)) Number: K190316

#### III. PREDICATE DEVICE

Icare ic100 tonometer (type TA011), K153694

This predicate has not been subject to a design-related recall.

#### IV. INDICATIONS FOR USE / INTENDED USE

"The Icare ic200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye."

The intended use of the Icare ic200 (type TA031) as well as the method used by the clinician to obtain a measurement remain unchanged from the predicate tonometer Icare ic100 (TA011) by this design modification.

{4}------------------------------------------------

## V. DEVICE DESCRIPTION

The Icare ic200 (TA031) tonometer is a hand-held, battery operated device, which measures intraocular pressure (IOP) without the need for topical anesthesia. The Icare ic200 tonometer measures the IOP by utilizing the rebound method, where a small (1.8 mm diameter), light (26.5 mg) probe makes brief contact with the eve. The Icare ic200 probes are sterile (gamma sterilized), single-use accessories of the device. All other accessories of the Icare ic200 are non-sterile and do not have direct patient contact.

Icare ic200 allows IOP measurement of patients in all positions between sitting position (direction of probe movement 0°) and supine position (direction of probe movement 90°).

Identical to the predicate device, the Icare ic200 (TA031) tonometer records the speed of the probe upon initiation of a measurement sequence. When the moving probe contacts the cornea, the probe decelerates at a rate which depends on the intraocular pressure. The tonometer measures the deceleration of the magnetized probe and the rebound time during contact with the eye and calculates the IOP from these parameters.

A single measurement sequence includes six measurements. After the six measurements are completed, the tonometer calculates the final IOP and the result is provided on the display. JOP measurement result can be transferred to printer or to a PC via Bluetooth® connection.

The measurement method, the IOP measurement algorithm and rebound technology (including sterile, single-use probe) of Icare ic200 are identical to the predicate device Icare ic100 (TA011, K153694).

#### Software

The Icare ic200 (TA031) Tonometer has embedded software utilizing a microcontroller. The software performs the following functions:

- Provides a user interface utilizing the user operable button and an OLED . display to control the measurement cycle
- Provides control to LED Probe base light
- Controls the probe's movement ●
- Calculates the intraocular pressure from the probe speed during impact ●
- Handles possible error conditions, such as recognizing if probe didn't contact . cornea or if the speed of the probe is abnormal
- . Calculates the standard deviation of the measurements during the measurement cycle and instructs user to repeat the measurement if the variability is unacceptable

{5}------------------------------------------------

- . Provides time setting to the internal clock
- Retrieves the serial number for identification of the device
- Stores measurement history in an external flash memory IC.
- . Controls the device hardware based on device position (inclination)

In addition to the embedded software, the Icare ic200 includes an OTS (Off the Shelf) Bluetooth® module. which enables wireless communication.

Icare proprietary software Icare EXPORT and Icare CLINIC are designed to work with the Icare ic200 as software accessory. Icare EXPORT is intended for transferring measurement data from Icare tonometers into the Icare CLINIC software's database. Icare CLINIC performs the following functions:

- Stores measurement data received from Icare devices ●
- Shows and prints charts based on the measurement data
- Exports measurement data from Icare CLINIC to external systems ●
- Imports measurement data from external systems
- Maintains patient database and measurement history ●

#### Calibration

As is the case for the predicate device, the design and principle of operation of the Icare ic200 (TA031) tonometer is such that maintenance calibration is not required. The tonometer does not include any mechanical or electrical parts subject to wear, which would cause an out of tolerance condition. Only the probe base, which as an accessory, can be replaced or cleaned by user as instructed in the Instructions for Use.

## Environment of use

The Icare ic200 tonometer is intended to be used in professional healthcare environment.

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Technological principle of both the subject and predicate devices is rebound tonometry. The table below summarizes the similarities in technological characteristics of the new device compared to the predicate device.

|    | Icare ic100 (TA011, K153694)                                                      | Icare ic200 (TA031, K190316)                                               |
|----|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| 1. | Indication for Use<br>The Icare TA011 Tonometer is<br>intended to be used for the | The Icare TA031<br>Tonometer is intended to be<br>used for the measurement |

{6}------------------------------------------------

|     |                                                          | measurement of intraocular pressure<br>of the human eye.                                                                | of intraocular pressure of<br>the human eye. |
|-----|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| 2.  | Design                                                   | Hand-held microprocessor based                                                                                          | Same                                         |
| 3.  | Measurement<br>technique                                 | Rebound tonometry, result calculated<br>based on IOP measurement algorithm                                              | Same                                         |
| 4.  | Calibration                                              | No maintenance calibration required                                                                                     | Same                                         |
| 5.  | Contact tip<br>(probe)                                   | Lightweight, disposable, single use,<br>plastic probe (26.5mg)                                                          | Same                                         |
| 6.  | Contact tip<br>sterilization                             | Gamma-sterilized                                                                                                        | Same                                         |
| 7.  | Force applied to<br>eye during<br>rebound<br>measurement | 8-16 mN                                                                                                                 | Same                                         |
| 8.  | Display<br>                                              | 1.50" OLED Display 128x128 Full<br>color                                                                                | Same                                         |
| 9.  | Range of<br>measurement                                  | 7-50 mmHg (display range 1-99<br>mmHg)                                                                                  | Same                                         |
| 10. | Measurement<br>accuracy (bench)                          | Accuracy: ± 1.2 mmHg (IOP ≤ 20<br>mmHg)<br>Accuracy: ± 2.2 mmHg (IOP > 20<br>mmHg)<br>Coefficient of variation CV < 8 % | Same                                         |
| 11. | Anesthesia<br>required                                   | No                                                                                                                      | Same                                         |
| 12. | Power source                                             | 4 x 1.5 Volt AA batteries                                                                                               | Same                                         |

The following technological differences exist between the subject and predicate devices:

|                     | Icare ic100 (TA011, K153694)                                 | Icare ic200 (TA031)                                                                   |
|---------------------|--------------------------------------------------------------|---------------------------------------------------------------------------------------|
| 1. Device structure | Display and navigation buttons<br>on the side                | Display and navigation buttons<br>towards the user, on the back side of<br>the device |
|                     | Weight                                                       | Weight                                                                                |
|                     | 4.94 oz. without batteries (140 g)                           | 5.82 oz. without batteries (165 g)                                                    |
|                     | Dimensions 1.14" x 3.74" x<br>8.46"<br>(29mm x 95mm x 215mm) | Dimensions 1.69" x 4.10" x 8.43"<br>(43mm x 104mm x 214mm)                            |

{7}------------------------------------------------

| 2. | Versatility of<br>measurement<br>direction | Tonometer must be oriented<br>horizontally (0°, patient in<br>sitting position) | Tonometer can be used in any angle<br>between 0° (sitting) and 90° (patient in<br>supine position)                                                      |
|----|--------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3. | Wireless data<br>transfer                  | Not available                                                                   | Wireless measurement result transfer<br>to a printer or PC via Bluetooth®<br>connection.                                                                |
| 4. | Graphical<br>user interface                | Includes all basic functions.                                                   | Includes all basic functions similar to<br>ic100, and possibility to add eye side<br>information and patient ID and to<br>select Bluetooth® connection. |

The primary modification within this 510(k) submission is the improvement in versatility of the measurement direction (sitting and supine). Modifications of external device structure and user interface are secondary modifications, which support this primary modification and improve the usability of the device.

Wireless measurement result transfer via Bluetooth® is a state-of-the-art feature, which has been added as a separate option that can also be disabled. Wireless connection doesn't affect the clinical use, measurement software, primary result display or any features related to the clinical safety and effectiveness of the device.

## VII. PERFORMANCE TESTING

The following performance testing has been conducted to confirm that the performance and safety aspects of the modified tonometer Icare ic200 are substantially equivalent with the Icare ic100 cleared for marketing under 510(k) K153694.

#### Software Verification and Validation Testing

The software used in the Icare ic200 (TA031) is based on predicate device Icare ic100 software and the modifications have been developed and maintained per the requirements IEC 62304 Medical device software - Software life cycle processes. Software verification and validation testing were conducted, and documentation was generated as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered to cause "moderate" level of concern, since a failure or latent flaw in the software could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care.

Bluetooth® software (OTS) included in Icare ic200 was considered to cause "minor" level of concern, since the software does not influence the intended use or clinical functions of the device and the measurement result display enabled by the Bluetooth® is additional to the primary measurement result display method.

{8}------------------------------------------------

The Icare CLINIC and EXPORT software are accessories to ic200 and therefore they are also considered to have "moderate" level of concern. They have been developed and maintained per the requirements of IEC 62304 standard and software verification and validation testing has been conducted and documented per FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

#### Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Icare ic200 tonometer. The device complies with the IEC 60601-1standard for electrical safety and the IEC 60601-1-2 standard for EMC.

#### Bench performance testing

Icare ic200 tonometer performance in IOP measurement was compared to predicate device Icare ic 100 tonometer performance. The testing was performed with a simulated IOP model, which was also used during the original predicate device submission (K153694).

- Icare ic200 met the predetermined acceptance criteria for accuracy and bias over the . measurement range ensuring substantial equivalence to the previously cleared Icare ic100 tonometer.
- Repeatability of the Icare ic200 tonometer was assessed in bench test. Test was done . by measuring a manometrically controlled test cornea. Test pressures (7, 10, 20, 30, 40 and 50 mmHg) covered the specified measurement range of the Icare ic200 tonometer. To assess repeatability, 10 measurements were performed with the test tonometer in three different angles (0, 45 and 90 degrees).

Icare ic200 tonometer demonstrated an agreement with true manometric pressure, R-squared values being over 99.7 %, regardless the angle of measurement (0, 45 or 90 degrees). On average, the Icare ic200 tonometer underestimated the pressure by 0.33 mmHg with respect to true manometric pressure, standard deviation being 0.82 mmHg.

- Reproducibility was assessed by test in which two operators performed three . measurements with three different ic200 units in each measurement setup, resulting 162 measurements per operator. The mean difference between operators was 0.10 mmHg with standard deviation of 0.98 mmHg. The R-squared value in regression analysis was over 99.5%, which indicates high reproducibility across the operators and across the devices.
Please be advised that bench testing conditions do not cover all the error sources within a clinical setting and thus higher variability is expected in clinical use.

{9}------------------------------------------------

#### Clinical performance testing

Clinical study was to compare the agreement and precision of the intraocular pressure (IOP) results measured by the Icare ic200 tonometer device with the Goldmann Applanation Tonometer (GAT) reference device per ANSI Z80.10:2014 "Ophthalmic Instruments - Tonometers" standard (in accordance with FDA's extent of recognition).

In addition, GAT and Tonopen Avia were compared with Icare ic200 in the sitting position and Tonopen Avia and Perkins in the supine position. Possible individual patient discomfort and potential device complications or adverse events were recorded. No adverse events (including corneal abrasions) were reported in this study.

Per the ANSI Z80.10 standard the differences between ic200 and GAT measurements overall and within three IOP subgroups are summarized in Table 1 below.

| Group                                     | N (eyes) | Outside<br>±5 mmHg |
|-------------------------------------------|----------|--------------------|
| GAT group = low<br>≤ 16 mmHg              | 45       | 0/45 (0.0%)        |
| GAT group =<br>medium<br>> 16 to <23 mmHg | 67       | 2/67 (3.0%)        |
| GAT group = high<br>≥ 23 mmHg             | 40       | 0/40 (0.0%)        |
| Overall                                   | 152      | 2/152<br>(1.3%)    |

Table 1. Differences between ic200 and GAT, overall and within IOP subgroups

The descriptive statistics (n, Mean, SD, Median, Min, Max) for the IOP measurements from ic200, GAT and Tonopen Avia in sitting position are summarized in Table 2 below.

Table 2. Summary of IOP measurements ic200, GAT and Tonopen Avia in sitting position

|                    | ic200 | GA<br>T | Tonopen Avia |
|--------------------|-------|---------|--------------|
| n (eyes)           | 152   | 152     | 152          |
| Mean IOP<br>(mmHg) | 20.44 | 19.88   | 20.26        |
| SD (mmHg)          | 5.82  | 5.49    | 5.68         |
| Median<br>(mmHg)   | 20.30 | 20.00   | 20.50        |
| Min (mmHg)         | 10.8  | 10.0    | 10.5         |
| Max (mmHg)         | 44.6  | 43.0    | 42.5         |

{10}------------------------------------------------

The descriptive statistics for the IOP measurements from ic200, Tonopen Avia and Perkins in supine position are summarized in Table 3 below.

|                    | ic200 | Tonopen Avia | Perkins |
|--------------------|-------|--------------|---------|
| n (eyes)           | 152   | 152          | 152     |
| Mean IOP<br>(mmHg) | 21.01 | 20.50        | 20.15   |
| SD (mmHg)          | 5.59  | 5.66         | 5.61    |
| Median<br>(mmHg)   | 21.38 | 20.50        | 20.50   |
| Min<br>(mmHg)      | 10.2  | 11.5         | 10.5    |
| Max<br>(mmHg)      | 41.9  | 42.0         | 40.5    |

Table 3. Summary of IOP measurements ic200, Tonopen Avia and Perkins in supine position

The results of the study demonstrated that Icare ic200 is safe and performed as intended and conformed with the ANSI Z80.10 standard compared to GAT.

#### Conclusions

The overall performance testing (bench and clinical) demonstrated that the Icare ic200 provides equivalent performance as the predicate device Icare ic100.

#### VIII. SUBSTANTIAL EQUIVALENCE

The Icare ic200 tonometer uses the same operating principle found in the predicate device and in all rebound type tonometers. The intended use of the Icare ic200 as well as the method used by the clinician to obtain a measurement remain unchanged by the modifications. The technological comparison demonstrates that the use of Icare ic200 does not raise additional questions of safety and effectiveness.

In conclusion, it has been demonstrated that the Icare ic200 (TA031) tonometer is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization (of probes), operating principle, performance and intended use/ indication for use as the previously cleared Icare ic100 tonometer described in the original 510(k) submission (K153694).

---

**Source:** [https://fda.innolitics.com/device/K190316](https://fda.innolitics.com/device/K190316)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
