← Product Code [HTY](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HTY) · K190287

# Arthrex DynaNite® PIP (Hammertoe) Implant (K190287)

_Arthrex, Inc. · HTY · May 3, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HTY/K190287

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [HTY](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HTY.md)
- **Decision Date:** May 3, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Arthrex DynaNite® PIP (Hammertoe) Implant is intended to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges.

## Device Story

The Arthrex DynaNite® PIP (Hammertoe) Implant is a permanent, single-use, sterile bone fixation device. It consists of a Nickel Titanium (Nitinol) implant featuring a threaded end and a barbed end, available in 12mm, 14mm, and 16mm lengths in straight or bent configurations. The device is provided on a handled inserter for surgical use. It is implanted by a physician to stabilize phalangeal fractures, fusions, or osteotomies. The device functions as a mechanical fastener to maintain bone alignment during the healing process, providing stability to the affected phalanges.

## Clinical Evidence

Bench testing only. Performance data included barb pull-out, thread pull-out, static/fatigue cantilever bend, insertion/failure torque, corrosion resistance, barb compression force, and transformation temperature testing. Bacterial Endotoxins Test (BET) was performed using the Kinetic Chromogenic Method per ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, and EP 2.6.14, confirming compliance with pyrogen limit specifications.

## Technological Characteristics

Material: Nickel Titanium (Nitinol). Design: Threaded and barbed bone fixation fastener. Configurations: 12mm, 14mm, 16mm lengths; straight or bent. Delivery: Handled inserter. Sterilization: Sterile, single-use. No software or electronic components.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- dynaMX Intramedullary Implant ([K170326](/device/K170326.md))
- DePuy Sterile Kirschner Wires and Steinmann Pins ([K960385](/device/K960385.md))
- Arthrex DynaNite Nitinol Staple ([K172052](/device/K172052.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Build Correspondence

Image /page/0/Picture/2 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and features an abstract design of human figures. The FDA logo is on the right and includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 3, 2019

Arthrex Inc. Ms. Rebecca R. Homan Senior Regulatory Affairs Associate 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K190287

Trade/Device Name: Arthrex DynaNite® PIP (Hammertoe) Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: April 15, 2019 Received: April 17, 2019

Dear Ms. Homan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for CAPT Raquel Peat, PhD, MPH, USPHS Director Office of Health Technology 6 Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K190287

Device Name Arthrex DynaNite® PIP (Hammertoe) Implant

Indications for Use (Describe)

The Arthrex DynaNite® PIP (Hammertoe) Implant is intended to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary or 510(k) Statement

| Date Prepared         | April 12, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter             | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Contact Person        | Rebecca R. Homan<br>Senior Regulatory Affairs Associate<br>1-239-643-5553, ext. 73429<br>rebecca.homan@arthrex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Name of Device        | Arthrex DynaNite® PIP (Hammertoe) Implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name           | Bone Fixation Fasteners                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code          | HTY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name   | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulatory Class      | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device      | K170326: dynaMX Intramedullary Implant (Primary Predicate)<br>K960385: DePuy Sterile Kirschner Wires and Steinmann Pins (Reference Predicate)<br>K172052: Arthrex DynaNite Nitinol Staple (Reference Predicate)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Purpose of Submission | This Traditional 510(k) premarket notification is submitted to obtain clearance for<br>the Arthrex DynaNite PIP (Hammertoe) Implant.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description    | The Arthrex DynaNite PIP (Hammertoe) Implant is a Nickel Titanium (Nitinol)<br>bone fixation device intended to be permanently implanted. The implant has a<br>threaded end and a barbed end. The implant will be offered in 12mm, 14mm and<br>16 mm lengths, each available in straight and bent configurations. The implant is<br>provided on a handled inserter and is sold as sterile, single-use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications for Use   | The Arthrex DynaNite PIP (Hammertoe) Implant is intended to stabilize and aid in<br>the fixation of fractures, fusions, and osteotomies of the phalanges.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Performance Data      | Barb Pull-out, Thread Pull-out, Static/Fatigue Cantilever Bend, Insertion<br>Torque/Failure Torque, and Corrosion resistance testing were conduct to<br>demonstrate that the Arthrex DynaNite PIP (Hammertoe) Implant performed<br>statistically equivalent to the predicate device cleared under K960385. Barb<br>Compression Force and Transformation Temperature tester were also conducted.<br><br>Bacterial Endotoxins Test (BET) was performed on the Arthrex DynaNite PIP<br>(Hammertoe) Implant utilizing the Kinetic Chromogenic Method in accordance<br>with ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, EP 2.6.14. Testing<br>was performed in compliance with US FDA good manufacturing practice (GMP)<br>regulations 21 CFR Parts 210, 211 and 820. The testing conducted demonstrates<br>that the Arthrex DynaNite PIP (Hammertoe) Implant meets pyrogen limit<br>specifications. |
| Conclusion            | The Arthrex DynaNite PIP (Hammertoe) Implant is substantially equivalent to the<br>predicate devices in which the basic design features and intended uses are the<br>same. Any differences between the proposed device and the predicate devices<br>are considered minor and do not raise different questions of safety or<br>effectiveness.<br><br>Based on the indications for use, technological characteristics, and the summary<br>of data submitted, Arthrex Inc. has determined that the Arthrex DynaNite PIP<br>(Hammertoe) Implant is substantially equivalent to the currently marketed<br>predicate device.                                                                                                                                                                                                                                                                                |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HTY/K190287](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HTY/K190287)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
