K190278 · Auragen Aesthetics, LLC · MUU · May 3, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190278
Device Name
AuraGen 123 Suction Lipoplasty System (A123)
Applicant
Auragen Aesthetics, LLC
Product Code
MUU · General, Plastic Surgery
Decision Date
May 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The A123 is used for aspiration, harvesting, filtering and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is to be used without any additional manipulation. The A123 is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
Device Story
AuraGen 123 (A123) is a sterile, single-use, disposable suction lipoplasty system used in operating rooms. It connects to a user-provided liposuction cannula, vacuum source, and waste canister. The device enables surgeons to harvest, wash, filter, concentrate, and transfer autologous adipose tissue for immediate reinjection into the same patient. Components include a collection chamber, mesh basket, concentration chamber with fluid-absorbing pads, AuraClens powder packets, tubing, and a mixing spatula. The system processes fat to improve concentration for body contouring. By facilitating efficient fat processing during the same procedure, it benefits patients by providing autologous tissue for reconstruction or cosmetic enhancement without additional manipulation.
Clinical Evidence
Bench testing only. Evaluated cell viability (Trypan Blue), fat volume, fat concentration, time-to-graft, device usability, canister implosion resistance, tubing collapse, system leak, and tubing tensile strength. Biocompatibility testing performed per ISO 10993; sterilization validated per ISO 11137. All tests met acceptance criteria.
Technological Characteristics
Sterile, single-use disposable system. Components: collection chamber, mesh basket, concentration chamber with fluid-absorbing pads, AuraClens powder, tubing, and spatula. Mechanical suction principle; requires external vacuum source. Materials biocompatible per ISO 10993. Sterilized via Gamma irradiation per ISO 11137. Mechanical performance tested per ISO 10079-1:2015.
Indications for Use
Indicated for patients undergoing aesthetic body contouring or surgical procedures requiring soft tissue aspiration across plastic, reconstructive, neuro, GI, urological, general, orthopedic, gynecological, thoracic, and laparoscopic specialties.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 3, 2019
AuraGen Aesthetics LLC % Mr. Randy Prebula Partner Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K190278
Trade/Device Name: AuraGen 123 Suction Lipoplasty System (A123) Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: February 8, 2019 Received: February 8, 2019
Dear Mr. Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
David Krause, Ph.D. Acting Division Director Division of Infection Control and Plastic Surgery Devices Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMANSERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page
510(k) Number (if known) K190278
#### Device Name
The AuraGen 123TM Suction Lipoplasty System (A123)
Indications for Use (Describe)
The A123 is used for aspiration, harvesting, filtering and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is to be used without any additional manipulation.
The A123 is intended for use in the following surgical speciation of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
# AuraGen Aesthetics' AuraGen 123TM Suction Lipoplasty System (A123)
### Submitter
AuraGen Aesthetics LLC 11 Dellbrook Road, Weston, MA 02493
Phone: 617-818-4008 Facsimile: 857-999-3929 Contact Person: Yiannis Monovoukas, Ph.D. Date Prepared: April 26, 2019
Name of Device: AuraGen 123TM Suction Lipoplasty System Common or Usual Name: A123 Classification Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU (General and Plastic Surgery Panel)
### Predicate Device
510(K) # K120902 Trade name: GID 700 Tissue Canister Manufacturer: The GID Group, USA (Distributed in the US as REVOLVE™ System by LifeCell Inc.)
# Reference Device: N/A
# Device Description
The A123 is a suction lipoplasty system designed to be used in the operating room in conjunction with, and attached to, a user-provided liposuction cannula, a vacuum source, and a waste canister. The A123 allows the surgeon to conveniently and accurately harvest, wash, filter, concentrate, and transfer autologous adipose tissue for reinjection into the same patient for body contouring in cosmetic and reconstructive surgery applications during the same procedure in which autologous adipose tissue is collected. The system is a sterile, disposable unit for single patient use.
The A123 consists of the following components:
- Collection chamber ●
- . Collection mesh basket
- . Concentration chamber with fluid-absorbing pads
- AuraClens™ powder packets ●
- Outlet tube and drain valve with tube clamp
- Mixing spatula
{5}------------------------------------------------
The A123 is to be used together with FDA-cleared devices (such as a lipoplasty device [Product Code MUU, Requlation Number 21 CFR 878.50401, liposuction cannula, high vacuum tubing, waste container, syringes [e.g. 60 cc Toomev-tip svringes, Luer-Lock, Product Code KYZ1, and roomtemperature, sterile 0.9% normal saline solution, all provided by the user.
# Intended Use / Indications for Use
The A123 is used for aspiration, harvesting, filtering and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is to be used without any additional manipulation.
The A123 is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
### Performance Data
The Company conducted biocompatibility testing for the A123 and the AuraClens powder in accordance with the requirements of ISO 10993 for external communicating devices having contact < 24hours. In all instances, the test articles were found to be biocompatible.
The A123 and the AuraClens powder were sterilized by Gamma irradiation and both met the sterility requirements per ISO 11137 - Sterilization of Health Care Products.
In addition, the following table summarizes the bench testing performed with the device:
| Bench Test | Methods |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cell Viability | Trypan Blue Dye Exclusion test |
| Fat Volume | Processed adipose tissue was centrifuged to separate the oil, fat, and aqueous phases. The volume of the fat layer was measured. |
| Fat Concentration | Processed adipose tissue was centrifuged to separate the oil, fat, and aqueous phases. The fat concentration was calculated as: volume of the fat layer/volume of adipose tissue processed. |
| Time-to-Graft (TTG) | Time from the start of the washing step to the end of the transfer/extraction step was measured.) |
| Device Usability | Usability of the device was evaluated. |
| Canister implosion | A123 units were tested for medical vacuum suction canister implosion test requirements per ISO 10079-1:2015(E) Medical Suction Equipment Part 1, Section 6.1.3 and Annex A.3 |
| Tubing collapse | A123 units were tested for medical vacuum suction canister tubing collapse test requirements per ISO 10079-1:2015(E) Medical Suction Equipment Part 1, Section 6.3.1 and Annex A.4. |
| System leak | A123 units were tested for vacuum seal to determine leakage per the |
{6}------------------------------------------------
| Bench Test | Methods |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | specification in ISO 10079-1:2015(E) Medical Suction Equipment Part 1, Section 7.7.1 and Annex A.8.1. |
| Tubing tensile strength<br>(pull-off force) (to assess<br>tubing connections) | A123 units were tested to measure the tensile strength (pull-off force) of the A123 tubing assembly and the user-supplied liposuction and aspiration tubing from the A123 tubing connectors. |
The A123 meets the acceptance criteria for all tests.
### Substantial Equivalence
The A123 is as safe and effective as the GID 700 Tissue Canister (the "Predicate Device") that FDA has already cleared (K120902). The A123 has the same intended uses and indications for use, as well as similar technological characteristics and principles of operation as its predicate device. In addition, the minor technological differences between the A123 and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the A123 is as safe and effective as the GID 700 Tissue Canister. Thus, the A123 is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.