← Product Code [HAW](/productcode/HAW) · K190250

# Microscope Navigation (K190250)

_Brainlab AG · HAW · Sep 6, 2019 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K190250

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [HAW](/productcode/HAW.md)
- **Decision Date:** Sep 6, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Software as a Medical Device

## Indications for Use

The Brainlab Navigation System - Microscope Navigation Software module, that when used with a Brainlab navigation system and compatible instrument accessories, is intended as image guided planning and navigation system to enable open and minimally invasive surgery. It links an instrument and the view of the surgical field (e.g., video, view through surgical microscope) to a virtual computer image space on patient image data being processed by the navigation. The system is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy.

## Device Story

Microscope Navigation software module for Brainlab navigation systems; integrates with surgical microscopes to provide image-guided surgery. Inputs: patient medical imaging (CT, MR, etc.), microscope position data, and surgical field video. Operation: links microscope view and instrument position to virtual 3D patient image space; provides registration verification/correction; enables motorized microscope control (focus depth, trajectory alignment, instrument tip tracking). Used in OR by surgeons; output displayed on navigation workstation monitor or through microscope eyepiece. Benefits: enhances surgical precision by overlaying navigation data onto the surgical field, facilitating minimally invasive procedures.

## Clinical Evidence

Bench testing only. Verification included testing of motorized movement safety limits, UI functionality, and communication protocol compliance with microscope manufacturers. Validation included usability testing to ensure safe and effective operation. All tests passed.

## Technological Characteristics

Software-only module running on Brainlab navigation workstation (computer, display, IR tracking camera). Interfaces with external motorized surgical microscopes via manufacturer-specific communication protocols. Features include focus depth control, bookmarking, trajectory alignment, and instrument tip tracking. No patient-contacting materials; software-based.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Microscope Navigation ([K172820](/device/K172820.md))
- VECTORVISION CRANIAL ([K082060](/device/K082060.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 6, 2019

Brainlab AG Hugo Morales Regulatory Affairs Manager Olof-Palme-Str. 9 81829 Munich Germany

Re: K190250

Trade/Device Name: Microscope Navigation Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: August 7, 2019 Received: August 9, 2019

Dear Hugo Morales:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K190250

Device Name Microscope Navigation

### Indications for Use (Describe)

The Brainlab Navigation System - Microscope Navigation Software module, that when used with a Brainlab navigation system and compatible instrument accessories, is intended as image guided planning and navigation system to enable open and minimally invasive surgery.

It links an instrument and the view of the surgical field (e.g., video, view through surgical microscope) to a virtual computer image space on patient image data being processed by the navigation. The system is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY MICROSCOPE NAVIGATION 1.5

IN ACCORDANCE WITH REQUIREMENTS OF 21 CFR PART 807.92

| Manufacturer:        | Brainlab AG<br>Olof-Palme-Straße 9<br>81829 Munich<br>Germany                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:           | Oliver Fleig                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Contact person:      | Alexander Schwiersch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Summary date         | December 14, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device               | Microscope Navigation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade name           | Microscope Navigation Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Common name:         | Frameless Stereotaxic Navigation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification name: | Neurological Stereotaxic Instrument (21 CFR 882.4560, Product Code HAW)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate Device:    | K172820 Microscope Navigation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Regulatory Class:    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation Number:   | 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Intended use:        | The Brainlab Navigation System - Microscope Navigation Software is a software<br>module, that when used with a Brainlab navigation system and compatible instrument<br>accessories, is intended as image guided planning and navigation system to enable<br>open and minimally invasive surgery.<br>It links an instrument and the view of the surgical field (e.g. video, view through surgical<br>microscope) to a virtual computer image space on patient image data being processed<br>by the navigation workstation. The system is indicated for any medical condition in<br>which a reference to a rigid anatomical structure can be identified relative to images<br>(CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy. |

{4}------------------------------------------------

| Device description:              | The Microscope Navigation Software (also referred to as subject device or Microscope<br>App) is a software. It runs on a Brainlab navigation system consisting of a computer, a<br>display and an IR tracking camera (referred to as platform) and a Brainlab Image<br>Guided Surgery software (referred to as IGS software).<br>The Microscope Navigation encapsulates microscope specific functionality and<br>separates it from the IGS software.<br>The device interfaces with the IGS software and e.g. utilizes the registration provided<br>by the IGS software.<br><br>The device assists surgeries where a surgical microscope is used. It provides<br>information based on the field of view through the microscope, the microscope position<br>relative to the patient and the medical imaging data of the patient.<br>The subject device provides functionality to verify and correct a patient registration.<br>The Microscope Navigation does not provide its own registration. |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reason for 510(k)<br>submission: | Microscope Navigation Version 1.0 has been cleared under K172820, with K082060<br>VECTORVISION CRANIAL as predicate device. Compared to its predicate device,<br>Microscope Navigation Version 1.0 provides limited interface implementation and<br>compatibilities of the microscope protocols. These interface implementation and<br>compatibilities have been present in K082060 VECTORVISION CRANIAL and have<br>been re-implemented in Microscope Navigation Version 1.5.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

#### Substantial Microscope Navigation has been verified and validated using non-clinical data equivalence: according to Brainlab procedures for product design and development. The information provided by Brainlab in this Special 510(k) application supports the claim of substantial equivalence to the predicate device.

Microscope Navigation separates microscope specific functionality in an separate application. Microscope specific views are bundled in this application.

### Comparison to the Predicate Device

Motorized movement functionality has been re-implemented in the Subject Device to provide full interface implementation to motorized microscopes. All features have already been present in K082060 VECTORVISION CRANIAL:

- Set focus depth to instrument tip ●
- Bookmarking ●
- . Align to trajectory
- Follow instrument tip
- Align to instrument axis

{5}------------------------------------------------

### Verification summary:

| Test                                 | Test Method Summary                                                                                                                                                                                                                | Results                                                                                                                   |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Safety Measures                      | Verification of implemented limits and automatic<br>safety measures for the motorized movements                                                                                                                                    | All tests passed                                                                                                          |
| Motorized movement<br>user interface | Interactive testing of user interface                                                                                                                                                                                              | All tests passed                                                                                                          |
| Communication<br>protocols           | Verification of the communication between the<br>subject device and the integrated microscopes. The<br>tests verify compliance of the subject device to the<br>communication protocols provided by the microscope<br>manufacturer. | Protocols and acceptance<br>criteria are unchanged<br>between subject device and<br>predicate device. All tests<br>passed |

### Validation summary:

The validation comprises usability tests which ensure that the user interface can be used safely and effectively. All tests were rated as successfully passed according to their acceptance criteria. The non-clinical validation has been performed with software and equipment that are identical or equivalent to the final version of the product.

---

**Source:** [https://fda.innolitics.com/device/K190250](https://fda.innolitics.com/device/K190250)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).