K190234 · Cnd Global Technologies, Inc. · MUH · Mar 1, 2019 · Dental
Device Facts
Record ID
K190234
Device Name
ODI HD Dental Sensor
Applicant
Cnd Global Technologies, Inc.
Product Code
MUH · Dental
Decision Date
Mar 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
ODI HD Dental Sensor is intended to be used by dentists and other qualified professionals for providing diagnostic x-ray radiographs of dentition, jaws and other oral structures.
Device Story
ODI HD Dental Sensor is an intraoral digital x-ray detector used by dentists, radiologists, and dental assistants. It captures incident x-rays using a Cesium Iodide (CsI) scintillator, which converts radiation into visible light. This light is transmitted via a fiber optic plate to a CMOS-based light detection imager, which converts the signal into digital data. The sensor connects via USB 2.0 to a PC running Apteryx XrayVision software. The device acts as a passive receiver and does not control the x-ray generator. The captured images are displayed on a monitor for the clinician to assist in diagnosing dental and oral diseases. The device benefits patients by enabling digital radiographic imaging for dental assessment.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), detective quantum efficiency (IEC 62220-1), and enclosure protection (IEC 60529).
Technological Characteristics
Intraoral digital x-ray sensor using Cesium Iodide (CsI) scintillator, fiber optic plate, and CMOS technology. Dimensions: 600mm² (Size 1) and 884mm² (Size 2). Pixel size: 20x20 µm. Resolution: 20 Lp/mm. Connectivity: USB 2.0. Power: 5V via USB. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62220-1, IEC 60529. Software: Apteryx XrayVision.
Indications for Use
Indicated for dentists and qualified professionals to provide diagnostic x-ray radiographs of dentition, jaws, and oral structures in patients requiring dental imaging.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
CND Global Technologies, Inc. % Ms. Lisa Jager Regulatory Affairs Consultant Regulatory Science Solutions LLC 11576 W. M179 Hwy MIDDLEVILLE MI 49333
March 1, 2019
## Re: K190234
Trade/Device Name: ODI HD Dental Sensor Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: February 1, 2019 Received: February 6, 2019
### Dear Ms. Jager:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 4. Indications for Use Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K190234
Device Name ODI HD Dental Sensor
Indications for Use (Describe)
ODI HD Dental Sensor is intended to be used by dentists and other qualified professionals for providing diagnostic x-ray radiographs of dentition, jaws and other oral structures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
## K190234
# 510(k) Summary
| Submitted By: | CND Global Technologies, Inc.<br>16 Branch Avenue<br>Cumberland, RI 02864<br>319-333-6767 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Lisa Jager<br>Regulatory Affairs Consultant<br>11576 W. M179 Hwy<br>Middleville, MI 49333<br>regulatorysciencesolutions@gmail.com<br>616-802-2364 |
| Date of Summary<br>Preparation | January 30, 2019 |
| Trade Name | ODI HD Dental Sensors |
| Submission Number | K190234 |
| Common Name | Intraoral Digital x-ray Sensors |
| Classification Name | Extraoral source x-ray system |
| Classification | Class II per 21 CFR 872.1800 |
| Product Code | MUH |
| Classification Panel | Dental |
#### Predicate Device Information
| Submitted By: | Masterlink LLC |
|----------------------|---------------------------------|
| Trade Name | Apex Dental Sensors |
| Submission Number | K163282 |
| Common Name | Intraoral Digital x-ray Sensors |
| Classification Name | Extraoral source x-ray system |
| Classification | Class II per 21 CFR 872.1800 |
| Product Code | MUH |
| Classification Panel | Dental |
#### Device Description
The ODI HD Dental Sensors are electronic medical device intra-oral radiographic images. The sensor can be operated by Radiologists, Dental Assistants and other healthcare professionals, who are both trained and competent to take Dental X-ray radiographs. Intra-oral positioning of the sensor is accomplished by the use of dedicated intra-oral positioning devices that facilitate the accurate alignment of the sensor may also be aligned with the assistance of the patient. The ODI HD Dental Sensors is an indirect light converting digital x-ray detector. A scintillating device composed of Cesium lodide (Csl) converts incident x-rays into visible light that is optically coupled to a light detection imager based on CMOS technology. The ODI HD Dental Sensors allow for automatic detection of such incident x-rays in order to generate data. Software interprets this data into images used for dental applications. The ODI HD Dental Sensors support USB 2.0 direct connectivity to personal computers and or laptops with dedicated electronics and a sensor software driver. The subject device does not control the generator, it is only a receiver. The X-ray system and the software used are not a part of this submission.
The following hardware and software requirements need to be taken into account to ensure successful integration of ODI HD Dental Sensors with the X-ray systems.
{4}------------------------------------------------
#### Hardware requirements:
The type of x-ray systems that integrate with the Apex Dental Sensors are wall-mounted x-ray generators (both AC and DC) with a tube current between 1 and 15 mA inclusive, and with a tube voltage between 50 and 100 kV inclusive, with in-built controls to set exposure parameters. Generators allow variable mA/K/ to be selected, all will control the exposure time.
The subject device is only a receiver, and not a controller. The subject device and the software cannot act as the controller of an x-ray generators are controlled by built-in controllers and are not part of the subject device.
Premium Plus Disposable Barrier Sleeve (K163447) - Premium Plus Disposable Barrier Sleeve is intended to be used as a disposable barrier sleeve for ODI HD Dental Sensors. The Premium Plus Disposable Barrier Sleeve is non-sterile and intended for single patient use only.
#### Software requirements:
The Apteryx XrayVision software needs assistance from a TWAIN driver to recognize X-ray images. The recommended requirements for PC hardware for the sensor and software combination would be:
- . Processor: Intel core i5-250m 2.5GHz or Higher
- . Memory: 4G or higher
- Hard disk: Above 20G, 60G recommended
- Interface: USB 2.0 ●
- Display: Resolution 1024 × 758 (15") or above .
- Operating System: .
- O Microsoft Windows 10(64/32 bit)
- o Microsoft Windows 8.1(64-bit)
- Microsoft Windows 8(64-bit) O
- Windows 7 SP1 32 and 64-bit o
- . The computer connected to system shall be in accordance with IEC 60950-1:2005
#### Indications for Use
ODI HD Dental Sensor is intended to be used by dentists and other qualified professionals for providing diagnostic xray radiographs of dentition, jaws and other oral structures.
#### Technological Characteristics
The ODI HD Dental Sensors (Subject Device) makes use of a Predicate Device, Apex Dental Sensors (K163282). The detectors in the ODI HD Dental Sensors are the detectors in the predicate device. The subject device and the predicate device are identical and sourced from the same supplier Hamamatsu. The model numbers of the ODI HD Dental Sensors are:
> S11684-12 is Size #1 S11685-12 is Size #2
| Attribute | Subject Device-<br>ODI HD Dental Sensors | Predicate Device-<br>Apex Dental Sensors (K163282) | Equivalence | Difference and<br>Justification |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | | | | |
| Manufacturer | CND Global Technologies Inc. | Masterlink LLC | | |
| Classification and<br>Product Code | Class II<br>21 CFR 872.1800<br>MUH | Class II<br>21 CFR 872.1800<br>MUH | Equivalent | No difference |
| Common Name | Intraoral Digital X-ray Sensors | Intraoral Digital X-ray Sensors | Equivalent | No difference |
| Classification Name | Extraoral source X-ray System | Extraoral source X-ray System | Equivalent | No difference |
| Classification Panel | Dental | Dental | Equivalent | No difference |
| Intended Use | Providing diagnostic x-ray radiographs of<br>dentition, jaws and other oral structures. | Radiographic examination<br>to assist with diagnosis of<br>diseases of the teeth, jaw,<br>and oral structure. | Equivalent | No difference |
| Indications for Use | ODI HD Dental Sensors is intended to be used by<br>dentists and other qualified professionals for<br>providing diagnostic x-ray radiographs of<br>dentition, jaws and other oral structures. | The Apex Dental Sensors is<br>intended to be used for a<br>radiographic examination by<br>a dental professional to assist<br>in the diagnosing of diseases<br>of the teeth, jaw and oral<br>structures. | Equivalent | No difference |
| Number of Sensors | 2 | 2 | Equivalent | No difference |
| Cable Length | 2m | 2m | Equivalent | No difference |
| Pixel Size | 20 x 20 µm | 20 x 20 µm | Equivalent | No difference |
| Resolution | 20 Lp/mm typ<br>(S11684-12) | 20 Lp/mm typ<br>(S11684-12) | Equivalent | No difference |
| Technology | CMOS chip +optical fiber<br>plate + Csl scintillator | CMOS chip +optical fiber plate +<br>Csl Scintillator | Equivalent | No difference |
| Matrix dimensions<br>(mm²) | Sensor Active Area:<br>600mm² (Size 1)<br>884mm² (Size 2) | Sensor Active Area:<br>600mm² (Size 1)<br>884mm² (Size 2) | Equivalent | No difference |
| Principles of<br>Operation | X-ray (radiation) =><br>scintillator (convert to<br>light) => fiber optic<br>(filtering) => CMOS<br>(convert to digital)=><br>electronics=> PC (capture<br>& display image) | X-ray (radiation) =><br>scintillator (convert to<br>light) => fiber optic<br>(filtering) => CMOS<br>(convert to digital)=><br>electronics=> PC (capture<br>& display image) | Equivalent | No difference |
| | Other Requirements | | | |
| Recommended PC<br>Requirements | Processor: Intel core i5-250m 2.5GHz or<br>Higher or above; Memory: Above 4G;<br>Hard disk: Above 20G, Interface: USB<br>2.0;<br>Display: Resolution 1024 × 768 (15") or<br>above<br>Operating System: Microsoft Windows<br>10(64/32 bit)<br>Microsoft Windows 8.1(64-bit)<br>Microsoft Windows 8(64-bit)<br>Windows 7 SP1 32 and 64-bit<br>The computer connected to<br>system shall be in accordance with IEC<br>60950-1:2005. | Processor: Intel 1.2GHz chip or<br>above; Memory: Above 1G;<br>Hard disk: Above 40G; Interface:<br>USB 2.0;<br>Display: Resolution 1024 ×<br>758 (15") or above<br>Operating System: Windows<br>ХР<br>The computer connected to<br>system shall be in accordance<br>with IEC<br>60950-1:2005. | Equivalent | The Processor,<br>Memory, Hard Disk<br>and Operating<br>System<br>requirements for<br>supporting use of<br>the subject and<br>predicate device<br>differ slightly but are<br>found to be<br>equivalent. The<br>outside supporting<br>PC requirements do<br>not impact the<br>devices intended<br>use, technical<br>requirements or<br>safety requirements.<br>No differences are<br>noted for Display<br>and Computer IEC<br>requirements. |
| Connection to Imaging<br>Practice PC | USB 2.0 Interface | USB 2.0 Interface | Equivalent | No difference |
| Software-Image<br>Management | Apteryx XrayVision | Apteryx XrayVision | Equivalent | No difference |
| Sensor Board | All control electronics<br>directly integrated on<br>CMOS sensor chip | All control electronics<br>directly integrated on<br>CMOS sensor chip | Equivalent | No difference |
| Sensor input voltage<br>and current | USB2.0 (5V, 4.25min) | USB2.0 (5V, 4.25min) | Equivalent | No difference |
| Operating<br>Temperature | 0°C~+35°C | 0°C~+35°C | Equivalent | No difference |
| Electrical Safety | | | | |
| Standards of<br>Conformity | IEC 60601-1 (Electrical)<br>IEC 60601-1-2 (EMC)<br>IEC 62220-1 (Performance)<br>IEC 60529 (Performance) | IEC 60601-1-1 (Electrical)<br>IEC 60601-1-2 (EMC)<br>62220-1 (Performance)<br>IEC 61326-1 (EMC)<br>IEC 60601-2-65 (Electrical) | Equivalent | Cited standards<br>differ slightly,<br>however the devices<br>are still found to be<br>equivalent as the<br>main Electrical, EMC<br>and Performance<br>standards cited are<br>the same. |
{5}------------------------------------------------
{6}------------------------------------------------
#### Similarities
- . The ODI HD Dental Sensors and Predicate Device both are available in two sizes
- The interface to PC of both ODI HD Dental Sensors and the Predicate is USB ●
- Both ODI HD Dental Sensors and the Predicate Sensor use CMOS technology, the pixel size for both sensors is 20 x ● 20 um
- . The Sensor Active Area for ODI HD Dental Sensors and Predicate Device is the same i.e., 600mm2 (Size 1) and 884mm2 (Size 2)
- The principles of operation of ODI HD Dental Sensors and Predicate Device is the same, i.e., X-ray (radiation)=> . scintillator (convert to light) => fiber optic (filtering) => CMOS (convert to idigital)=> electronics=> PC (capture & display image)
- . The Software-Image Management of ODI HD Dental Sensors and Predicate Device is the same, i.e., Apteryx XrayVision which has been cleared by the FDA under (K983111)
- Both ODI HD Dental Sensors and Predicate Device have all control electronics directly integrated on CMOS sensor chip .
- The sensor input voltage is 5V in both ODI HD Dental Sensors and Predicate Device ●
- The operating temperature of ODI HD Dental Sensors and Predicate Device is also comparable, i.e., 0°C~+35°C ●
- The ODI HD Dental Sensors and the Predicate Device are tested per IEC 60601-1-2 for Electrical Safety ● and Electromagnetic compatibility
- . The recommended PC requirements in ODI HD Dental Sensors and the Predicate Device is also comparable. The processor that is recommended for ODI HD Dental Sensors is Intel core i5-250m 2.5GHz or Higher, whereas the processor for the Predicate Device is Intel 1.2GHz chip or above
- The ODI HD Dental Sensors and the Predicate Device have the same sensor resolution ●
- The ODI HD Dental Sensors and the Predicate have the same cable lengths. The sensor cable length in ODI HD Dental . Sensors is 2m and in Predicate Device is also 2m.
- . The interface for ODI HD Dental Sensors requires USB 2.0, and the Predicate Device also utilizes USB 2.0.
#### Performance Data
The following Performance testing has been performed on the Subject Device in accordance with appropriate FDA guidance documents and relevant standards, to support the determination of substantial equivalence. The test data showed that the Subject device is safe and effective for its intended use.
- Electrical and safety requirements established in AAMI ES 60601-1:2005/(R) 2012 .
- Electromagnetic compatibility requirements in IEC 60601-1-2 Edition 3-2007 .
- . IEC 62220-1:2015 – Determination of the Detective Quantum Efficiency Detectors used in Radiographic Imaging
- . IEC 60529:2001 - Degrees of Protection Provided by Enclosures (IP Code)
- The Subject Device is also in compliance with FDA guidance for the submission of 510(k)s for Solid ● State X-ray Imaging Devices
#### Clinical Performance
Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterizes all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on the ODI HD Dental Sensors.
#### Conclusion
The ODI HD Dental Sensors are substantially equivalent to the Predicate, and do not raise any new or different questions of safety or effectiveness. The subject and the predicate devices are identical; the only difference exists in the trade names (for marketing purposes).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.