Ethyl Chloride Medium Jet Stream, Ethyl Chloride Fine Pinpoint spray, Ethyl Chloride Mist, Ethyl Chloride Accustream 360* Medium Spray, Ethyl Chloride Accustream 360* Fine Spray
K190161 · Gebauer Company · MLY · May 7, 2019 · PM
Device Facts
| Record ID | K190161 |
| Device Name | Ethyl Chloride Medium Jet Stream, Ethyl Chloride Fine Pinpoint spray, Ethyl Chloride Mist, Ethyl Chloride Accustream 360* Medium Spray, Ethyl Chloride Accustream 360* Fine Spray |
| Applicant | Gebauer Company |
| Product Code | MLY · PM |
| Decision Date | May 7, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
Gebauer's Ethyl Chloride Topical Anesthetic Spray (Mist Spray, Fine Spray and Medium Spray): A vapocoolant (skin refrigerant) intended for topical application to control pain associated with injections (starting IV's and venipuncture), minor surgical procedures (such as lancing boils, or incision and drainage of small abscesses), and the temporary relief of minor sports injuries. The Fine and Medium Sprays are also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.
Device Story
Device is a vapocoolant spray delivering ethyl chloride to intact skin. Input: manual pressure on actuator. Transformation: rapid evaporation of volatile propellant on skin surface extracts thermal energy, providing cooling effect. Output: localized skin cooling. Used in clinical settings by healthcare providers. Effect: provides temporary topical anesthesia to manage pain during minor procedures or musculoskeletal conditions. Benefits: non-invasive pain management for injections and minor surgeries.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1 for surface devices with limited contact (<24 hours). USP <51> Antimicrobial Effectiveness Testing confirmed the product does not support microbial growth.
Technological Characteristics
Pressurized aerosol delivery system using steel cans or pharmaceutical glass bottles. Vapocoolant: 100% Ethyl Chloride. Energy source: thermal energy via refrigerant evaporation. Application: manual actuator. Biocompatibility: ISO 10993-1 compliant. Preservative testing: USP <51>.
Indications for Use
Indicated for patients requiring topical anesthesia for pain control during injections (IV starts, venipuncture), minor surgical procedures (lancing boils, incision/drainage of abscesses), and temporary relief of minor sports injuries. Fine and Medium sprays are also indicated for myofascial pain, restricted motion, and muscle tension.
Predicate Devices
- Gebauer's Ethyl Chloride (Fine Spray and Medium Spray) (K991514)
- Gebauer's Pain Ease Skin Refrigerant (Mist Spray and Medium Spray) Topical Anesthetic (K032671)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
May 7, 2019
Gebauer Company Brittney Cooper Sr. Regulatory Affairs and Quality Specialist 4444 East 153rd Street Cleveland, Ohio 44128
Re: K190161
Trade/Device Name: Gebauer's Ethyl Chloride Topical Anesthetic Spray Regulatory Class: Unclassified Product Code: MLY Dated: January 29, 2019 Received: January 30, 2019
Dear Brittney Cooper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# VivekJ. Pinto -S
for Carlos Peña, PhD, MS Director OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190161
Device Name
Gebauer's Ethyl Chloride Topical Anesthetic Spray
Indications for Use (Describe)
Gebauer's Ethyl Chloride Topical Anesthetic Spray . Fine Spray and Medium Spray): A vapocoolant (skin refrigerant) intended for topical application to control pain associated with injections (starting IV's and venipuncture), minor surgical procedures (such as lancing boils, or incision and drainage of small abscesses), and the temporary relief of minor sports injuries. The Fine and Medium Sprays are also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Gebauer Company. The logo consists of a blue stylized letter "G" above the words "GEBAUER COMPANY", which are also in blue. The letter "G" is designed with a curved shape and a horizontal line extending from the top.
# 510(k) Summary—K190161
This 510(k) Summary is being submitted in accordance of 21 CFR Part 807.92.
#### I. SUBMITTER
II.
III.
Secondary:
| Owner: | Gebauer Company<br>4444 East 153rd Street<br>Cleveland, OH 44128<br>(216) 581-3030 |
|----------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person: | Brittney Cooper<br>Gebauer Company<br>4444 East 153rd Street<br>Cleveland, OH 44128<br>(216) 581-3030 |
| Date Prepared: | May 6, 2019 |
| DEVICE | |
| Trade Name: | Gebauer's Ethyl Chloride Topical<br>Anesthetic Spray |
| Common Name: | Cold Spray |
| Classification Name: | Refrigerant Topical, Vapocoolant |
| Product Code: | MLY |
| PREDICATE DEVICE: | |
| Primary: | Gebauer's Ethyl Chloride (Fine Spray and<br>Medium Spray)<br>K991514 |
e Ecgaily marketed medical device
Gebauer's Pain Ease Skin Refrigerant (Mist Spray and Medium Spray) Topical Anesthetic
{4}------------------------------------------------
K032671
#### IV. DESCRIPTION
Gebauer's Ethyl Chloride Topical Anesthetic Spray is a prescription device designed to deliver ethyl chloride in a mist, fine or medium spray. This chemical self-propels itself from the delivery system, which is designed to account for its low vapor pressure. The device is packaged in either a pharmaceutical glass bottle or steel aerosol can with several variations of nozzles. The patient contact is less than ten seconds and the skin is cooled through rapid evaporation of the non-medicated volatile propellant. The new device, Gebauer's Ethyl Chloride is identical in all aspects to the predicate device, Gebauer's Ethyl Chloride 510(k) K991514, with the exception of revised precautions and warnings and information regarding biocompatibility and USP <51> Antimicrobial Effectiveness Testing.
#### V. INDICATIONS FOR USE OF DEVICE
The Indications for Use is similar to the predicate devices.
Gebauer's Ethyl Chloride Topical Anesthetic Spray (Mist Spray, Fine Spray and Medium Spray): A vapocoolant (skin refrigerant) intended for topical application to control pain associated with injections (starting IV's and venipuncture), minor surgical procedures (such as lancing boils, or incision and drainage of small abscesses), and the temporary relief of minor sports injuries. The Fine and Medium Sprays are also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.
#### VI. TECHNICAL SUMMARY
The new device has the same technical characteristics as the predicate device including materials, design, and energy source. There are no technological differences between the predicate and the new device. As with the predicate device, the cooling action experienced by the patient is caused by the evaporation of the chemical from the patient's skin. The user applies pressure to the nozzle to dispense the aerosol product onto the skin. The material is contained in a steel aerosol can or pharmaceutical glass bottle, filled under pressure, and dispensed using standard aerosol nozzle technology. The only difference is product fill volume which is 3.9 f1. oz in the new device versus the predicate device which is 3.5 fl. oz. This change does not impact the performance of the device, only the number of uses per can or bottle. See table below for the comparison between the new device and the predicate:
{5}------------------------------------------------
| | Comparison Chart – Technological Characteristics | | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Gebauer's Ethyl<br>Chloride Predicate<br>Device | Gebauer's Pain<br>Ease Skin<br>Refrigerant (Mist<br>Spray and Medium<br>Spray) Topical<br>Anesthetic<br>Predicate Device | Gebauer's Ethyl<br>Chloride Topical<br>Anesthetic Spray<br><i>New Device</i> |
| Product Design | Pressurized<br>dispensing<br>container which<br>includes the<br>vapocoolant, steel<br>aerosol can, or<br>pharmaceutical<br>glass bottle, valve<br>(if a can) and<br>actuator. | Pressurized<br>dispensing<br>container which<br>includes the<br>vapocoolant,<br>aerosol can, valve<br>and actuator. | Pressurized<br>dispensing container<br>which includes the<br>vapocoolant, steel<br>aerosol can, or<br>pharmaceutical glass<br>bottle, valve (if a<br>can) and actuator. |
| Indication for Use<br>and Intended Use | Gebauer's Ethyl<br>Chloride (Fine and<br>Medium Nozzles):<br>Gebauer's Ethyl<br>Chloride is a<br>vapocoolant (skin<br>refrigerant)<br>intended for<br>topical application<br>to control pain<br>associated with<br>minor surgical<br>procedures (such<br>as lancing boils,<br>incisions and<br>drainage of small<br>abscesses),<br>injections and the<br>temporary relief of<br>minor sports<br>injuries. It is also<br>intended for the<br>treatment of<br>restricted motion<br>associated with<br>myofascial pain<br>caused by trigger<br>points. | A vapocoolant<br>(skin<br>refrigerant)<br>intended for<br>topical<br>application to<br>skin, intact<br>mucous<br>membranes<br>(oral cavity,<br>nasal<br>passageways<br>and the lips)<br>and minor open<br>wounds.<br>Gebauer's Pain<br>Ease Skin<br>Refrigerant<br>controls pain<br>associated with<br>minor surgical<br>procedures<br>(such as<br>lancing boils,<br>incisions,<br>drainage of<br>small<br>abscesses, and | Gebauer's Ethyl<br>Chloride Topical<br>Anesthetic Spray<br>(Mist Spray, Fine<br>Spray and Medium<br>Spray): A<br>vapocoolant (skin<br>refrigerant) intended<br>for topical<br>application to control<br>pain associated with<br>injections (starting<br>IV's and<br>venipuncture), minor<br>surgical procedures<br>(such as lancing<br>boils, or incision and<br>drainage of small<br>abscesses), and the<br>temporary relief of<br>minor sports<br>injuries. <i>The Fine<br/>and Medium Sprays<br/>are also intended for<br/>the treatment of</i> |
| | | sutures),<br>injections<br>(venipuncture,<br>IV starts,<br>cosmetic<br>procedures)<br>and the<br>temporary<br>relief of minor<br>sports injuries<br>(sprains,<br>bruising, cuts<br>and abrasions).<br>The Medium<br>Spray is also<br>intended for the<br>treatment of<br>myofascial pain<br>caused by<br>trigger points,<br>restricted<br>motion and<br>muscle tension. | myofascial pain<br>caused by trigger<br>points, restricted<br>motion and muscle<br>tension. |
| Product Fill Volume | 3.5 fl. oz. (103.5<br>mL) | 3.5 flo. Oz. (103.5<br>mL) | 3.9 fl. oz. (116 mL) |
| Vapocoolant<br>Composition | Ethyl Chloride<br>(100%) | 1,1,1,3,3<br>Pentafluoropropane<br>(HFC 245fa high<br>purity) and<br>1,1,1,2<br>Tetrafluoroethane<br>(HFC 134a<br>pharmaceutical<br>grade). | Ethyl Chloride<br>(100%) |
| Energy Delivered | Thermal energy<br>via Refrigerant<br>Spray. | Thermal energy via<br>Refrigerant Spray. | Thermal energy via<br>Refrigerant Spray. |
| Vapocoolant<br>Discharge Method | Depress the<br>actuator to release<br>the vapocoolant. | Depress the<br>actuator to release<br>the vapocoolant. | Depress the actuator<br>to release the<br>vapocoolant |
| Method of<br>Application | Direct spray<br>application to<br>intact skin | Direct spray<br>application to<br>intact skin, intact | Direct spray<br>application to intact<br>skin. |
| | | membranes and<br>minor open<br>wounds. | |
| Environmental<br>Compatibility | Flammable. | Non-Flammable. | Flammable. |
| Mechanical Safety | Mechanism has<br>positive shut-off<br>release. | Mechanism has<br>positive shut-off<br>release. | Mechanism has<br>positive shut-off<br>release. |
{6}------------------------------------------------
{7}------------------------------------------------
#### VII. DETERMINATION OF SUBSTANTIAL EQUIVELANCE
This premarket notification 510(k) "change being effected" is being submitted to revise, remove or include specific information on the labeling identified and addressed within this pre-market submission. There is demonstrated equivalency in basic product design and technology, in indications for use, target population, and risk factors.
As stated above, the new device is identical in formulation, delivery system and packaging to the predicate device Gebauer's Ethyl Chloride, 510(k) K991514.
# Labeling
The new device has revised, removed or added information on the device labeling. The predicate device contained precautions and warnings not representative of clinical use of the device whereas the new device is labeled in accordance with the biocompatibility requirements of ISO 10993-1. Additionally, the labeling of the new device was revised to include information regarding USP <51> Antimicrobial Effectiveness Testing. Both the new and predicate devices are indicated for use to control pain associated with minor surgical procedures (such as lancing boils, incisions and drainage of small abscesses), injections and the temporary relief of minor sports injuries. The stream applications are also intended for use as a counterirritant in the management of myofascial pain, restricted motion and muscle tension. The new device labeling is substantially equivalent to the predicate device.
#### VIII. PERFORMANCE DATA
The predicate and new devices use the same materials, design and energy source and no technological differences exist. No changes were made to the device, use of the device, production of the device or output of the device. To support the labeling changes and to demonstrate that changes do not impact substantial equivalence, the following tests were performed:
# Biocompatibility
Gebauer's Ethyl Chloride Topical Anesthetic Spray is categorized according to ISO 10993-1 as a surface device that has limited (< 24 hours) contact with intact skin.
{8}------------------------------------------------
The relevant endpoints for the device are cytotoxicity, sensitization and irritation. Testing included in the submission support the biocompatibility of the device.
USP <51> Antimicrobial Effectiveness Testing (Preservative Effectiveness) Testing was performed in accordance with USP <51> to demonstrate that the product acts as its own preservative and does not support growth of microorganisms during use. All test method acceptance criteria were met.
#### IX. CONCLUSION
Based on the information above and within this submission, it is concluded that the new Ethyl Chloride device is safe and effective for its intended use and is substantially equivalent to the predicate device.