NanoBone® SBX Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® SBX Putty resorbs and is replaced with bone during the healing process.
Device Story
NanoBone® SBX Putty/QD is a resorbable bone void filler; consists of non-sintered nanocrystalline hydroxyapatite (HA) embedded in a porous silica gel matrix with a poloxamer carrier. Supplied in sterile applicators or cartridges; used by surgeons to fill osseous defects in extremities, pelvis, or posterolateral spine. Device does not set in-situ; acts as an osteoconductive scaffold; resorbs over time and is replaced by natural bone during healing. Provides a stable environment for bone regeneration; intended for non-load-bearing applications. No complex electronics or software involved.
Clinical Evidence
Bench testing included chemical/elemental analysis, XRD phase purity, dissolution, biocompatibility, pyrogenicity, sterilization validation, and shelf-life testing. Additional animal testing was conducted to support the posterolateral fusion indication. Results demonstrate the device functions as intended with a safety and effectiveness profile similar to predicate and reference devices.
Technological Characteristics
Material: Non-sintered nanocrystalline hydroxyapatite (HA) in silica gel matrix; Poloxamer carrier. Structure: Highly porous (nano, micro, macro pores) mimicking human cancellous bone. Complies with ASTM F1185. Sterilization: Radiation. Form factor: Putty supplied in sterile applicators/cartridges (0.25mL to 10mL).
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities, posterolateral spine, pelvis) resulting from surgical creation or traumatic injury, where the defect is not intrinsic to bony structure stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
APATECH Actifuse ABX E-Z-Fil Putty Bone Graft Substitute (K071206)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
ARTOSS GmbH Walter Gerike, Ph.D. Managing Partner Friedrich-Barnewitz-Strasse 3 18119 Rostock. Germany
Re: K190110
Trade/Device Name: NanoBone® SBX PUTTY, NanoBone® QD Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: January 21, 2019 Received: January 23, 2019
Dear Dr. Gerike:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190110
Device Name NanoBone® SBX PUTTY
#### Indications for Use (Describe)
NanoBone® SBX Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® SBX Putty resorbs and is replaced with bone during the healing process.
Type of Use (*Select one or both, as applicable*)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190110
Device Name NanoBone® QD
Indications for Use (Describe)
NanoBone® QD Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® QD Putty resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "NanoBone" in a stylized font. The word "Nano" is in pink, while the word "Bone" is in green. There is a circled R symbol to the right of the word "Bone", indicating that it is a registered trademark.
510(k) Summary [as required by 21 CFR 807.92(c)]
NanoBone® Bone Graft Substitute NanoBone® SBX Putty NanoBone® QD
510(k) 190110
| DATE PREPARED: | 18 March 2019 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | ARTOSS GmbH<br>Friedrich-Barnewitz-Strasse 3<br>18119 Rostock, Germany<br>Telephone: +49 (0) 381 5 43 45 - 701<br>Fax: +49 (0) 381 5 43 45 - 702 |
| CONTACT: | Walter Gerike, PhD<br>Managing Partner<br>ARTOSS GmbH<br>gerike@artoss.com |
| TRADE NAME: | NanoBone® bone graft substitutes<br>• NanoBone® SBX Putty<br>• NanoBone® QD |
| COMMON NAME: | Bone Void Filler |
| CLASSIFICATION | Resorbable calcium salt bone void filler |
| REGULATION: | 21 CFR 888.3045, Product Code MQV |
| DEVICE CLASS: | 2 |
| PANEL CODE: | Orthopedic / 888 |
| PREDICATE DEVICE: | ARTOSS NanoBone® bone graft substitutes<br>NanoBone® SBX Putty (K161351) |
| REFERENCE DEVICE: | APATECH Actifuse ABX E-Z-Fil Putty Bone Graft<br>Substitute (K071206) |
### INTENDED USE/INDICATIONS FOR USE:
NanoBone® SBX Putty and NanoBone® QD have the same intended use and indications for use:
NanoBone® SBX Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® SBX Putty resorbs and is replaced with bone during the healing process.
{6}------------------------------------------------
NanoBone® OD Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® OD Putty resorbs and is replaced with bone during the healing process.
## DEVICE DESCRIPTION:
NanoBone® SBX Putty consists of NanoBone® granulate embedded in an aqueous gel. NanoBone granulate consists of phase-pure non-sintered nanocrystalline osteoconductive hydroxyapatite (HA) embedded in a highly porous silica gel matrix. The high porosity of the product includes nano pores, micro pores, and macro pores. The interconnected and open porous structure of the macro pores of the NanoBone® is similar to human cancellous bone. NanoBone® SBX Puttv does not set in-situ following implantation.
NanoBone® SBX material is supplied in two different style applicators. The NanoBone® SBX Putty is supplied in a sterile applicator with an attached plunger. NanoBone® QD is an alternate packaging of the NanoBone® SBX Putty. NanoBone® OD is supplied in a sterile cartridge with a separate sterile plunger.
| Name of<br>Devices | NanoBone® SBX Putty<br>/ NanoBone® QD | NanoBone® SBX Putty | Actifuse ABX Putty |
|--------------------|-----------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------|
| | Subject | Predicate | Reference |
| 510(k) | | K161351 | K071206 |
| Intended Use | Bone void filler | Bone void filler | Bone void filler |
| Target Location | Extremities, pelvis, spine | Extremities, pelvis | Extremities, pelvis, spine |
| Formulation | Putty | Putty | Putty |
| Material | Hydroxyapatite (HA)<br>embedded in a silica gel<br>matrix | Hydroxyapatite (HA)<br>embedded in a silica gel<br>matrix | Silicate substituted<br>calcium phosphate<br>(hydroxyapatite) |
| Carrier | Poloxamer | Poloxamer | Poloxamer |
| Granule Size | 0.6mm x 2mm | 0.6mm x 2mm | 1mm-2mm |
| How Supplied | In applicators ranging in<br>size from 0.25mL to 10<br>mL | In applicators ranging in<br>size from 0.25mL to 10<br>mL | Pack size:<br>1.5mL to 20mL<br>E-Z Prep applicator:<br>10 mL<br>MIS cartridge: 7.5mL |
| Sterilization | Radiation | Radiation | Radiation |
#### COMPARISON WITH PREDICATE AND REFERENCE DEVICES:
The NanoBone® SBX Putty/NanoBone® OD and the predicate and reference devices have the same intended use, to fill bony voids and gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or those created from traumatic injury to the bone.
The subject device was not been previously submitted for a posterolateral indication. The subject device and the predicate device have identical design and technological characteristics. NanoBone® SBX Putty/NanoBone® OD have the same indications, contraindications, risks, basic technology, equivalent materials, and potential adverse events as the NanoBone® predicate (K161351). The only difference between the predicate ARTOSS
{7}------------------------------------------------
NanoBone® bone graft substitute and the subject device is the addition of posterolateral fusion indication, and the addition of the alternate applicator.
## NON-CLINICAL TESTING/PERFORMANCE DATA:
Previously submitted bench testing has shown the NanoBone® products meet all relevant requirements for calcium salt bone void filler devices, including ASTM F1185. Testing was performed on the predicate device to characterize and evaluate the performance of the NanoBone® products. The testing included:
- Chemical / elemental analysis ●
- Phase purity / XRD ●
- . Dissolution testing
- Animal testing ●
- Biocompatibility assessment / testing
- Pyrogenicity ●
- Sterilization validation ●
- Shelf life
The in vitro bench tests demonstrated that the NanoBone® bone graft substitutes met all acceptance criteria and performed similarly to the predicate devices.
Additional animal testing was performed to show the performance for use in posterolateral fusion. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate and reference devices.
# BIOCOMPATIBILITY:
The NanoBone® bone graft substitutes were compared to the predicate and reference devices. Based on the use of identical materials in the subject device and predicates devices, the biocompatibility of the NanoBone® bone graft substitutes was verified to be the same as those of the predicates.
# CONCLUSION:
The NanoBone® bone graft substitutes (NanoBone® SBX Putty and NanoBone® OD) have the same intended use and the same technological characteristics such as components, design, materials, sterilization method, and operating principles as the predicate. Performance data demonstrates that the device functions as intended.
Therefore, the NanoBone® SBX Putty and NanoBone® QD are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.