K190075 · Rebound Therapeutics · GEI · Feb 12, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190075
Device Name
Aurora Evacuator
Applicant
Rebound Therapeutics
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 12, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AURORA Evacuator is a powered instrument with a handpiece intended for removal of soft tissue and fluids under direct visualization. Types of direct visualization may include laparoscopic, pelviscopic, endoscopic, percutaneous, and open. Applications include those when access to the surgical site is limited, such as Neurosurgical/Spinal and ENT/Otolaryngological.
Device Story
AURORA Evacuator is a sterile, single-use, handheld aspiration device for removing soft tissue and fluids in minimally invasive surgery. Device consists of a stainless-steel suction wand with an aspiration window and internal rotating wire loops, connected to a plastic handle containing a battery-powered micro gear motor and suction control vent. Surgeon operates device by connecting the vacuum tube to an OR suction source; positioning the aspiration window against target tissue; covering the suction vent to draw tissue into the wand; and depressing a motor control button to rotate wire loops, mechanically breaking up tissue for evacuation into a collection receptacle. Used in OR settings by surgeons. Benefits include efficient tissue removal in limited-access surgical sites.
Clinical Evidence
Bench testing only. Performance validated using surrogate soft tissue and fluid materials. Verification testing confirmed product specifications including physical characteristics, rotational speeds, and operational temperatures. Biocompatibility, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and sterilization (ISO 11135-1) data were adopted from the predicate device.
Technological Characteristics
Handheld aspiration device; stainless-steel wand, plastic handle. Battery-powered micro gear motor drives internal rotating wire loops. Suction via external OR vacuum source. Dimensions: 7cm, 10cm, 13cm lengths; 3.0mm-3.6mm OD. Sterile, single-use. Biocompatibility per ISO 10993-1; Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2; Sterilization per ISO 11135-1.
Indications for Use
Indicated for removal of soft tissue and fluids during minimally invasive surgical procedures, including neurosurgical/spinal and ENT/otolaryngological applications, where access to the surgical site is limited and direct visualization (laparoscopic, endoscopic, percutaneous, or open) is available.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2019
Rebound Therapeutics Corporation Mr. Donald Atienza Director of Quality Assurance and Regulatory Affairs 13900 Alton Parkway, Suite 120 Irvine, California 92618
Re: K190075
Trade/Device Name: Aurora Evacuator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 9, 2019 Received: January 16, 2019
Dear Mr. Atienza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Radiological Health
Digitally signed by Long H. Chen Long H. Chen -S Date: 2019.02.12 14:42:41 -05'00' fOI Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190075
Device Name AURORA Evacuator
#### Indications for Use (Describe)
The AURORA Evacuator is a powered instrument with a handpiece intended for removal of soft tissue and fluids under direct visualization. Types of direct visualization may include laparoscopic, endoscopic, percutaneous, and open. Applications include those when access to the surgical site is limited, such as Neurosurgical/Spinal and ENT/ Otolaryngological.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1) SUBMITTER
Rebound Therapeutics Corporation 13900 Alton Parkway, STE 120 Irvine, CA 92618
Contact Person: Jane Metcalf, Vice President of Regulatory Affairs Telephone: (949) 525-1493 Email: jmetcalf@reboundtx.com
Contact Person: Donald Atienza, Director Quality Assurance and Regulatory Affairs Telephone: (949) 305-8111 ext. 108 Email: datienza@reboundtx.com
Date Prepared: February 11, 2019
# 2) DEVICE
Name of Device: AURORA Evacuator Common or Usual Name: Aspirator Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Regulatory Class: II Product Code: GEI
## 3) PREDICATE DEVICE
Primary Predicates: AURORA Evacuator-7, AURORA Evacuator-10, AURORA Evacuator-13, cleared under 510(k) #K180372 Secondary Predicate(s): None
## 4) PURPOSE OF SUBMISSION
The purpose of this submission is to achieve market clearance for two (2) modifications to the AURORA Evacuator. The primary modification is a product line extension to include two (2) larger diameter suction wands with an elongated whisk. This change adds six (6) new configurations to the original 3 configurations. See Table 5-1 below. The other modification is to change the identification label on the device handle to include French size.
Note that model numbers have also changed. The original three (3) model numbers submitted in K180372 were "Aurora Evacuator-13", "Aurora Evacuator-10" and "Aurora Evacuator-7". These model numbers have been changed to "AURORA Evacuator-9Fx13cm", "AURORA Evacuator-9Fx10cm", and "AURORA Evacuator-9Fx7cm", respectively
## 5) DEVICE DESCRIPTION
There is a total of nine (9) devices in the AURORA Evacuator product line; three (3) from the original 510(k), K180372, and six (6) in this 510(k). All members of the product line have the same design and mode of operation. See Table 5-1 below.
{4}------------------------------------------------
| | Model # | Wand<br>Working<br>Length<br>(cm) | Wand OD<br>(mm) | Wand ID<br>(mm) |
|----------------------------------|---------------------------|-----------------------------------|-----------------|-----------------|
| Predicates | AURORA Evacuator-9Fx7cm | 7 | 3.0 | 2.6 |
| | AURORA Evacuator-9Fx10cm | 10 | 3.0 | 2.6 |
| | AURORA Evacuator-9Fx13cm | 13 | 3.0 | 2.6 |
| Subject of<br>this<br>Submission | AURORA Evacuator-10Fx7cm | 7 | 3.4 | 3.0 |
| | AURORA Evacuator-10Fx10cm | 10 | 3.4 | 3.0 |
| | AURORA Evacuator-10Fx13cm | 13 | 3.4 | 3.0 |
| | AURORA Evacuator-11Fx7cm | 7 | 3.6 | 3.2 |
| | AURORA Evacuator-11Fx10cm | 10 | 3.6 | 3.2 |
| | AURORA Evacuator-11Fx13cm | 13 | 3.6 | 3.2 |
The AURORA Evacuator is provided sterile, for single-use only. It is a disposable, handheld, aspiration device designed to remove fluids and soft tissue during minimally invasive surgical procedures. The AURORA Evacuator consists of a stainless-steel suction wand connected to a plastic handle with an exiting vacuum tube. A barb connector at the end of the exiting vacuum tube is intended to connect to operating room suction.
The suction wand is configured with an aspiration window near its distal end for fluid and soft tissue to enter. Internal to the suction wand is a rotatable shaft with wire loops. The wire loops are aligned with the aspiration window and are intended to physically break-up soft tissue. The rotatable shaft is connected to a battery powered motor.
The motor and batteries are housed in the ergonomically designed handle. A green power indicator light is located on the top portion of the handle to show that power is available to the device. The handle also contains a suction control vent for user vacuum control and a motor control button for user activation of the rotating wire loops.
The AURORA Evacuator is shipped in a non-functional state that requires removal of a pull tab from the handle to initiate power to the device.
# 6) STATEMENT OF INTENDED USE
The AURORA Evacuator is a powered instrument with a handpiece intended for removal of soft tissue and fluids under direct visualization. Types of direct visualization may include laparoscopic, pelviscopic, endoscopic, percutaneous, and open. Applications include those when access to the surgical site is limited, such as Neurosurgical/Spinal and ENT/Otolaryngological.
# 7) SUMMARY OF TESTING
The following testing was conducted or adopted to demonstrate the safe and effective use of the modified AURORA Evacuators and substantial equivalence to the predicate.
- Biocompatibility Testing per ISO 10993-1, including Cytotoxicity (MEM Elution), Sensitization . (Kligman Maximization), Irritation (Intracutaneous Injection), Systemic Toxicity (Systemic Injection), Material Mediated Pyrogenicity was adopted from the predicate since materials of construction are identical.
{5}------------------------------------------------
- Electrical Safety and Enclosure Protection per IEC 60601-1 and IEC 60529-1 was adopted from . the predicate since the electrical configuration and design has not changed.
- Emissions and Immunity per IEC 60601-1-2 was adopted from the predicate since the electrical . configuration has not changed
- Particulate Testing per USP 36 <788> was adopted from the predicate since there is no change to configuration accept for wand diameter.
- Sterilization per ISO 11135-1 to validate a SAL of 106 was adopted using the predicate data. ●
- Packaging and Shelf-life per ISTA 2A and ASTM F1980 were adopted from the predicate since there are no changes to packaging configuration or package design.
- . Verification of product specifications including materials, multiple bonds evaluations, physical characteristics, audible characteristics, rotational speeds, duration of use and operational temperatures.
- Validation of product performance using surrogate soft tissue materials and surrogate fluid material.
# 8) SUMMARY TECHNOLOGICAL CHARACTERISTICS
The AURORA Evacuator has two (2) modes of operation and one (1) for aspiration and one (1) for breaking up tissue. For aspiration, the AURORA Evacuator uses an external vacuum source present in the operating room. To aspirate, the surgeon covers the suction control vent and positions the aspiration window so that vacuum draws fluid and smaller pieces of soft tissue through the window, into the wand, through the exiting vacuum tube and into a collection receptacle.
For breaking-up soft tissue the AURORA Evacuator uses rotating wire loops that are located inside the wand. Batteries in the handle of the device power a micro gear motor that causes the wire loops to rotate when the motor control button is pushed. The surgeon positions the aspiration window against the tissue to be removed, covers the suction control vent (completely or partially) to draw the tissue into the aspiration window, then depresses the motor control button. The tissue is mechanically broken-up and evacuated via vacuum into the collection receptacle. The wire loops stop rotating when the motor control button is released. The mechanical break-up of tissue can only occur inside of the wand.
# 9) COMPARISON TO PREDICATE
Comparing the modified devices to the predicate devices shows that the primary difference is the rate of aspiration. The modified devices have aspiration rates that are faster than those of the predicate. This is due to the larger diameter wands that are less restrictive to flow. Since lower aspirations rates represent worse case performance, no new questions of safety and effectiveness are introduced by the modified devices.
## 10) CONCLUSION
Upon reviewing the performance data provided in this submission and comparing indicated use, design, materials, principle of operation and overall technological characteristics, the modified AURORA Evacuators have been determined, by Rebound Therapeutics Corporation, to be substantially equivalent to the primary predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.